01-09-2017 | Original Article

Mechanical load exerted by PET-G aligners during mesial and distal derotation of a mandibular canine

An in vitro study

Authors: F. Elkholy, B. Mikhaiel, F. Schmidt, B. G. Lapatki

Published in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie | Issue 5/2017

Login to get access

Abstract

Introduction

The six force–moment (F/M) components exerted by aligners of different thickness during simulated mesiorotation and distorotation of a mandibular canine were studied.

Materials and methods

An acrylic mandibular model with a separated right canine mounted on a hexapod via a 3D F/M sensor was used. Duran+® aligners (Scheu Dental, Germany) of thickness 0.5, 0.625, and 0.75 mm were fabricated on plaster models with the measurement tooth in its neutral position. The F/M values were recorded during progressive mesiorotation or distorotation of tooth 43 in 1° steps up to ±15°, corresponding to 0.5 mm displacements of the tooth’s interdental contacts. Each rotation step included renewed seating of the aligner on the acrylic model. Three aligners were tested three times each for each thickness and direction of rotation.

Results

The median rotational moments for the 0.5 mm aligner and 15° distorotation of tooth 43 was 27.49 Nmm (interquartile range, IQR 1.45 Nmm). The corresponding values for the 0.625 and 0.75 mm aligners were 41.04 Nmm (IQR 5.62 Nmm) and 42.48 Nmm (IQR 2.17 Nmm), respectively. The average rotational moments for distorotation were 15% higher than for mesiorotation (p = 0.01). Relatively high collateral F/M components, specifically an intrusive force and labiolingual and mesiodistal tipping moments, were observed.

Conclusion

To avoid overloading of periodontal structures, derotation of lower canines should be limited to 10° per setup step, leading to rotational moments of about 15 Nmm. The mechanical behavior of the 0.625 and 0.75 mm aligners were similar; thus, it may be omitted from the aligner sequence. Further studies are required to investigate specific aligner modifications or attachments for minimizing collateral F/M components or unwanted movements, respectively, during canine derotation.
Literature
2.
go back to reference Elkholy F, Panchaphongsaphak T, Kilic F et al (2015) Forces and moments delivered by PET-G aligners to an upper central incisor for labial and palatal translation. J Orofac Orthop 76:460–475CrossRefPubMed Elkholy F, Panchaphongsaphak T, Kilic F et al (2015) Forces and moments delivered by PET-G aligners to an upper central incisor for labial and palatal translation. J Orofac Orthop 76:460–475CrossRefPubMed
3.
go back to reference Elkholy F, Schmidt F, Jäger R et al (2016) Forces and moments delivered by novel, thinner PET-G aligners during labiopalatal bodily movement of a maxillary central incisor: an in vitro study. Angle Orthod 86:883–890CrossRefPubMed Elkholy F, Schmidt F, Jäger R et al (2016) Forces and moments delivered by novel, thinner PET-G aligners during labiopalatal bodily movement of a maxillary central incisor: an in vitro study. Angle Orthod 86:883–890CrossRefPubMed
4.
go back to reference Elkholy F, Schmidt F, Jäger R et al (2017) Forces and moments applied during derotation of a maxillary central incisor with thinner aligners: an in vitro study. Am J Orthod Dentofac Orthop 151:407–415CrossRef Elkholy F, Schmidt F, Jäger R et al (2017) Forces and moments applied during derotation of a maxillary central incisor with thinner aligners: an in vitro study. Am J Orthod Dentofac Orthop 151:407–415CrossRef
6.
go back to reference Göz G, Diedrich P, Bourauel C et al (2004) Stellungnahme zur Invisalign-Methode. Zahnärztl Mitt 94:2632–2633 Göz G, Diedrich P, Bourauel C et al (2004) Stellungnahme zur Invisalign-Methode. Zahnärztl Mitt 94:2632–2633
7.
go back to reference Hahn W, Engelke B, Jung K et al (2010) Initial forces and moments delivered by removable thermoplastic appliances during rotation of an upper central incisor. Angle Orthod 80:239–246CrossRefPubMed Hahn W, Engelke B, Jung K et al (2010) Initial forces and moments delivered by removable thermoplastic appliances during rotation of an upper central incisor. Angle Orthod 80:239–246CrossRefPubMed
8.
go back to reference Hahn W, Engelke B, Jung K et al (2011) The influence of occlusal forces on force delivery properties of aligners during rotation of an upper central incisor. Angle Orthod 81:1057–1063CrossRefPubMed Hahn W, Engelke B, Jung K et al (2011) The influence of occlusal forces on force delivery properties of aligners during rotation of an upper central incisor. Angle Orthod 81:1057–1063CrossRefPubMed
9.
go back to reference Hahn W, Zapf A, Dathe H et al (2010) Torquing an upper central incisor with aligners–acting forces and biomechanical principles. Eur J Orthod 32:607–613CrossRefPubMed Hahn W, Zapf A, Dathe H et al (2010) Torquing an upper central incisor with aligners–acting forces and biomechanical principles. Eur J Orthod 32:607–613CrossRefPubMed
10.
go back to reference Kravitz ND, Kusnoto B, Agran B et al (2008) Influence of attachments and interproximal reduction on the accuracy of canine rotation with Invisalign. A prospective clinical study. Angle Orthod 78:682–687CrossRefPubMed Kravitz ND, Kusnoto B, Agran B et al (2008) Influence of attachments and interproximal reduction on the accuracy of canine rotation with Invisalign. A prospective clinical study. Angle Orthod 78:682–687CrossRefPubMed
11.
go back to reference Li X, Ren C, Wang Z et al (2016) Changes in force associated with the amount of aligner activation and lingual bodily movement of the maxillary central incisor. Korean J Orthod 46:65–72CrossRefPubMedPubMedCentral Li X, Ren C, Wang Z et al (2016) Changes in force associated with the amount of aligner activation and lingual bodily movement of the maxillary central incisor. Korean J Orthod 46:65–72CrossRefPubMedPubMedCentral
12.
go back to reference Lombardo L, Martines E, Mazzanti V et al (2017) Stress relaxation properties of four orthodontic aligner materials: a 24-hour in vitro study. Angle Orthod 87:11–18CrossRefPubMed Lombardo L, Martines E, Mazzanti V et al (2017) Stress relaxation properties of four orthodontic aligner materials: a 24-hour in vitro study. Angle Orthod 87:11–18CrossRefPubMed
13.
go back to reference Proffit WR (2007) The biologic basis of orthodontic therapy. In: Proffit WR, Fields HW, Sarver DM (eds) Contemporary orthodontics. Mosby Elsevier, St. Louis Mo, pp 331–358 Proffit WR (2007) The biologic basis of orthodontic therapy. In: Proffit WR, Fields HW, Sarver DM (eds) Contemporary orthodontics. Mosby Elsevier, St. Louis Mo, pp 331–358
14.
go back to reference Rossini G, Parrini S, Castroflorio T et al (2015) Efficacy of clear aligners in controlling orthodontic tooth movement: a systematic review. Angle Orthod 85:881–889CrossRefPubMed Rossini G, Parrini S, Castroflorio T et al (2015) Efficacy of clear aligners in controlling orthodontic tooth movement: a systematic review. Angle Orthod 85:881–889CrossRefPubMed
15.
go back to reference Simon M, Keilig L, Schwarze J et al (2014) Forces and moments generated by removable thermoplastic aligners: incisor torque, premolar derotation, and molar distalization. Am J Orthod Dentofac Orthop 145:728–736CrossRef Simon M, Keilig L, Schwarze J et al (2014) Forces and moments generated by removable thermoplastic aligners: incisor torque, premolar derotation, and molar distalization. Am J Orthod Dentofac Orthop 145:728–736CrossRef
Metadata
Title
Mechanical load exerted by PET-G aligners during mesial and distal derotation of a mandibular canine
An in vitro study
Authors
F. Elkholy
B. Mikhaiel
F. Schmidt
B. G. Lapatki
Publication date
01-09-2017
Publisher
Springer Medizin
Published in
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie / Issue 5/2017
Print ISSN: 1434-5293
Electronic ISSN: 1615-6714
DOI
https://doi.org/10.1007/s00056-017-0090-4

Other articles of this Issue 5/2017

Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 5/2017 Go to the issue