Published in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 2/2015

01-03-2015 | Original article

Artifacts in orthodontic bracket systems in cone-beam computed tomography and multislice computed tomography

Authors: V. Hirschinger, S. Hanke, U. Hirschfelder, Dr. E. Hofmann

Published in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie | Issue 2/2015

Login to get access

Abstract

Objectives

The purpose of this study was to quantify artifacts caused by different bracket systems in cone-beam computed tomography (CBCT) and multislice computed tomography (MSCT) scans.

Methods

Orthodontic brackets of four different systems were consecutively bonded to the surface of a residual molar on a human cadaveric mandible. One MSCT system and three CBCT units were used to scan each of the four bonded brackets, in addition to obtaining a blank reference scan of the tooth surface. All datasets were registered to the reference dataset using visualization software (Analyze 11.0® by AnalyzeDirect). Artifact-related reductions in image quality were expressed in percent of theoretical maximum standard deviations (SD) obtained for the gray values of the adjacent voxels, with higher percentages correlating more pronounced artifacts.

Results

Both the SD percentages for three defined line profiles and their mean values were almost invariably higher with the MSCT system than with the CBCT units. Looking into the individual SD percentages, two of the CBCT units (Pax Zenith 3D® and Picasso Trio®; both Vatech) produced higher values than the MSCT system (SOMATOM Definition AS+®; Siemens) in some line profiles. The titanium bracket, in particular, was associated with marked differences between the two scanner technologies, as the mean artifact intensities from this bracket were particularly high with the MSCT unit and relatively low with the CBCT units. The artifact intensities observed with the other three bracket systems varied widely depending on which scanner was used.

Conclusion

Different artifact intensities were noted depending on the composition of the bracket system and on the scanner technology (MSCT/CBCT). While the artifacts manifested themselves differently with different scanners, their adverse effects were comparable. However, given the variable severity of the artifacts observed depending on the materials scanned and the scanners used, a blanket recommendation for or against MSCT or CBCT units cannot be given on the basis of this study.
Literature
1.
go back to reference Bal M, Spies L (2006) Metal artifact reduction in CT using tissue-class modeling and adaptive prefiltering. Med Phys 33:2852–2859CrossRefPubMed Bal M, Spies L (2006) Metal artifact reduction in CT using tissue-class modeling and adaptive prefiltering. Med Phys 33:2852–2859CrossRefPubMed
2.
3.
go back to reference Berlemann U, Heini P, Müller U et al (1997) Reliability of pedicle screw assessment utilizing plain radiographs versus CT reconstruction. Eur Spine J 6:406–411CrossRefPubMedCentralPubMed Berlemann U, Heini P, Müller U et al (1997) Reliability of pedicle screw assessment utilizing plain radiographs versus CT reconstruction. Eur Spine J 6:406–411CrossRefPubMedCentralPubMed
4.
go back to reference Carrafiello G, Dizonno M, Colli V et al (2010) Comparative study of jaws with multislice computed tomography and cone-beam computed tomography. Radiol Med 115:600–611CrossRefPubMed Carrafiello G, Dizonno M, Colli V et al (2010) Comparative study of jaws with multislice computed tomography and cone-beam computed tomography. Radiol Med 115:600–611CrossRefPubMed
5.
go back to reference Coppenrath E, Draenert F, Lechel U et al (2008) Schnittbildverfahren zur dentomaxillofacialen Diagnostik: Dosisvergleich von Dental-MSCT und New Tom 9000 DVT. Fortschr Röntgenstr 180:396–401CrossRef Coppenrath E, Draenert F, Lechel U et al (2008) Schnittbildverfahren zur dentomaxillofacialen Diagnostik: Dosisvergleich von Dental-MSCT und New Tom 9000 DVT. Fortschr Röntgenstr 180:396–401CrossRef
6.
go back to reference Damstra J, Fourie Z, Huddleston Slater JJ et al (2010) Accuracy of linear measurements from cone-beam computed tomography-derived surface models of different voxel sizes. Am J Orthod Dentofacial Orthop 137(1):16.e1–16.e6 (discussion 16–17)CrossRefPubMed Damstra J, Fourie Z, Huddleston Slater JJ et al (2010) Accuracy of linear measurements from cone-beam computed tomography-derived surface models of different voxel sizes. Am J Orthod Dentofacial Orthop 137(1):16.e1–16.e6 (discussion 16–17)CrossRefPubMed
7.
go back to reference De Man B (2001) Iterative reconstruction for reduction of metal artifacts in computed tomography. Dissertation, University of Leuven, Belgium De Man B (2001) Iterative reconstruction for reduction of metal artifacts in computed tomography. Dissertation, University of Leuven, Belgium
8.
go back to reference Draenert FG, Coppenrath E, Herzog P et al (2007) Beam hardening artefacts occur in dental implant scans with the NewTom cone beam CT but not with the dental 4-row multidetector CT. Dentomaxillofac Radiol 36:198–203CrossRefPubMed Draenert FG, Coppenrath E, Herzog P et al (2007) Beam hardening artefacts occur in dental implant scans with the NewTom cone beam CT but not with the dental 4-row multidetector CT. Dentomaxillofac Radiol 36:198–203CrossRefPubMed
9.
go back to reference Farman AG (2005) ALARA still aplies. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 100(4):395–397CrossRefPubMed Farman AG (2005) ALARA still aplies. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 100(4):395–397CrossRefPubMed
10.
go back to reference Gayer A (2009) Quantifizierung von Metallartefakten in der Dental-Computertomographie. Diplomarbeit, Universitätsklinik für Radiodiagnostik Medizinische Universität Wien. http://ub.meduniwien.ac.at/edocmed/?f_file=1&f_ac=AC07452305 Gayer A (2009) Quantifizierung von Metallartefakten in der Dental-Computertomographie. Diplomarbeit, Universitätsklinik für Radiodiagnostik Medizinische Universität Wien. http://​ub.​meduniwien.​ac.​at/​edocmed/​?​f_​file=​1&​f_​ac=​AC07452305
11.
go back to reference Goldman LW (2007) Principles of CT: radiation dose and image quality. J Nucl Med Technol 35:213–225CrossRefPubMed Goldman LW (2007) Principles of CT: radiation dose and image quality. J Nucl Med Technol 35:213–225CrossRefPubMed
12.
go back to reference Holberg C, Steinhäuser S, Geis P, Rudzki-Janson I (2005) Cone-beam computed tomography in orthodontics: benefits and limitations. J Orofac Orthop 66:434–444CrossRefPubMed Holberg C, Steinhäuser S, Geis P, Rudzki-Janson I (2005) Cone-beam computed tomography in orthodontics: benefits and limitations. J Orofac Orthop 66:434–444CrossRefPubMed
13.
go back to reference Horner K, Islam M, Flygare L et al (2009) Basic principles for use of dental cone beam CT: consensus guidelines of the European Academy of Dental and Maxillofacial Radiology. Dentomaxillofac Radiol 38:187–195CrossRefPubMed Horner K, Islam M, Flygare L et al (2009) Basic principles for use of dental cone beam CT: consensus guidelines of the European Academy of Dental and Maxillofacial Radiology. Dentomaxillofac Radiol 38:187–195CrossRefPubMed
14.
go back to reference Imai K (2007) Analysis of streak artefacts on CT images using statistics of extremes. Br J Radiol 80:911–918CrossRefPubMed Imai K (2007) Analysis of streak artefacts on CT images using statistics of extremes. Br J Radiol 80:911–918CrossRefPubMed
15.
go back to reference Imai K, Ikeda M, Enchi Y, Niimi T (2009) Statistical characteristics of streak artifacts on CT images: relationship between streak artifacts and mA s values. Med Phys 36:492–499CrossRefPubMed Imai K, Ikeda M, Enchi Y, Niimi T (2009) Statistical characteristics of streak artifacts on CT images: relationship between streak artifacts and mA s values. Med Phys 36:492–499CrossRefPubMed
16.
go back to reference Kalender WA, Hebel R, Ebersberger J (1987) Reduction of CT artifacts caused by metallic implants. Radiology 164(2):576–577CrossRefPubMed Kalender WA, Hebel R, Ebersberger J (1987) Reduction of CT artifacts caused by metallic implants. Radiology 164(2):576–577CrossRefPubMed
17.
go back to reference Klingenbeck-Regn K, Schaller S, Flohr T et al (1999) Subsecond multi-slice computed tomography: basics and applications. Eur J Radiol 31:110–124CrossRefPubMed Klingenbeck-Regn K, Schaller S, Flohr T et al (1999) Subsecond multi-slice computed tomography: basics and applications. Eur J Radiol 31:110–124CrossRefPubMed
18.
go back to reference Kyriakou Y, Kolditz D, Langner O et al (2011) Digital volume tomography (DVT) and multislice spiral CT (MSCT): an objective examination of dose and image quality. Rofo 183(2):144–153CrossRefPubMed Kyriakou Y, Kolditz D, Langner O et al (2011) Digital volume tomography (DVT) and multislice spiral CT (MSCT): an objective examination of dose and image quality. Rofo 183(2):144–153CrossRefPubMed
19.
go back to reference Lee R, Azevedo B, Shintaku W et al (2008) Patient movement in three different CBCT units. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 105:e55CrossRef Lee R, Azevedo B, Shintaku W et al (2008) Patient movement in three different CBCT units. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 105:e55CrossRef
20.
go back to reference Naitoh M, Saburi K, Gotoh K et al (2013) Metal artifacts from posterior mandibular implants as seen in CBCT. Implant Dent 22(2):151–154CrossRefPubMed Naitoh M, Saburi K, Gotoh K et al (2013) Metal artifacts from posterior mandibular implants as seen in CBCT. Implant Dent 22(2):151–154CrossRefPubMed
21.
go back to reference Pauwels R, Stamatakis H, Bosmans H et al (2013) Quantification of metal artifacts on cone beam computed tomography images. Clin Oral Implants Res 24:94–99CrossRefPubMed Pauwels R, Stamatakis H, Bosmans H et al (2013) Quantification of metal artifacts on cone beam computed tomography images. Clin Oral Implants Res 24:94–99CrossRefPubMed
22.
go back to reference Perrella A, Lopes PM, Rocha RG et al (2010) Influence of dental metallic artifact from multislice CT in the assessment of simulated mandibular lesions. J Appl Oral Sci 18:149–154CrossRefPubMed Perrella A, Lopes PM, Rocha RG et al (2010) Influence of dental metallic artifact from multislice CT in the assessment of simulated mandibular lesions. J Appl Oral Sci 18:149–154CrossRefPubMed
23.
go back to reference Prell D, KyriakouY, Beister M, Kalender W (2009) A novel forward projection-based metal artifact reduction method for flat-detector computed tomography. Phys Med Biol 54(21):6575–6591CrossRefPubMed Prell D, KyriakouY, Beister M, Kalender W (2009) A novel forward projection-based metal artifact reduction method for flat-detector computed tomography. Phys Med Biol 54(21):6575–6591CrossRefPubMed
24.
go back to reference Razavi T, Palmer RM, Davies J et al (2010) Accuracy of measuring the cortical bone thickness adjacent to dental implants using cone beam computed tomography. Clin Oral Implants Res 21:718–725CrossRefPubMed Razavi T, Palmer RM, Davies J et al (2010) Accuracy of measuring the cortical bone thickness adjacent to dental implants using cone beam computed tomography. Clin Oral Implants Res 21:718–725CrossRefPubMed
25.
go back to reference Sanders MA, Hoyjberg C, Chu CB et al (2007) Common orthodontic appliances cause artifacts that degrade the diagnostic quality of CBCT images. J Calif Dent Assoc 35:850–857PubMed Sanders MA, Hoyjberg C, Chu CB et al (2007) Common orthodontic appliances cause artifacts that degrade the diagnostic quality of CBCT images. J Calif Dent Assoc 35:850–857PubMed
26.
go back to reference Schulze RKW, Berndt D, d’Hoedt B (2010) On cone-beam computed tomography artifacts induced by titanium implants. Clin Oral Implants Res 21:100–107CrossRefPubMed Schulze RKW, Berndt D, d’Hoedt B (2010) On cone-beam computed tomography artifacts induced by titanium implants. Clin Oral Implants Res 21:100–107CrossRefPubMed
28.
go back to reference Schulze R (2013) s2k-Leitlinie: Dentale digitale Volumentomographie. http://www.awmf.org/uploads/tx_szleitlinien/083-005l_S2k_Dentale_Volumentomographie_2013-10.pdf. Zugegriffen 25. November 2014 Schulze R (2013) s2k-Leitlinie: Dentale digitale Volumentomographie. http://​www.​awmf.​org/​uploads/​tx_​szleitlinien/​083-005l_​S2k_​Dentale_​Volumentomograph​ie_​2013-10.​pdf.​ Zugegriffen 25. November 2014
29.
go back to reference Williamson JF, Whiting BR, Benac J et al (2002) Prospects for quantitative computed tomography imaging in the presence of foreign metal bodies using statistical image reconstruction. Med Phys 29:2404–2418CrossRefPubMed Williamson JF, Whiting BR, Benac J et al (2002) Prospects for quantitative computed tomography imaging in the presence of foreign metal bodies using statistical image reconstruction. Med Phys 29:2404–2418CrossRefPubMed
30.
go back to reference Zhang Y, Zhang L, Zhu XR et al (2007) Reducing metal artifacts in cone-beam CT images by preprocessing projection data. Int J Radiat Oncol Biol Phys 67:924–932CrossRefPubMed Zhang Y, Zhang L, Zhu XR et al (2007) Reducing metal artifacts in cone-beam CT images by preprocessing projection data. Int J Radiat Oncol Biol Phys 67:924–932CrossRefPubMed
Metadata
Title
Artifacts in orthodontic bracket systems in cone-beam computed tomography and multislice computed tomography
Authors
V. Hirschinger
S. Hanke
U. Hirschfelder
Dr. E. Hofmann
Publication date
01-03-2015
Publisher
Springer Berlin Heidelberg
Published in
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie / Issue 2/2015
Print ISSN: 1434-5293
Electronic ISSN: 1615-6714
DOI
https://doi.org/10.1007/s00056-014-0278-9

Other articles of this Issue 2/2015

Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 2/2015 Go to the issue

Informationen

Mitteilungsseiten