Published in: Canadian Journal of Anesthesia/Journal canadien d'anesthésie 5/2013

01-05-2013 | Reports of Original Investigations

A novel method to position an endotracheal tube at the correct depth using an infrared sensor stylet

Authors: Yeongtak Song, Msc, Jaehoon Oh, MD, Youngjoon Chee, PhD, Taeho Lim, MD, PhD, Hyunggoo Kang, MD, Youngsuk Cho, MD

Published in: Canadian Journal of Anesthesia/Journal canadien d'anesthésie | Issue 5/2013

Login to get access

Abstract

Purpose

This study is a feasibility assessment to determine the ability of novice users to utilize an infrared (IR) sensor stylet as a guide to position the tip of the endotracheal tube (ETT) 40 mm proximal to the carina in the swine trachea.

Methods

We developed a stylet system using an IR sensor attached to the tip of a stylet to facilitate measuring the distance of the ETT from the carina. The indicator lamp of the IR sensor system turns on through calibration when the ETT tip arrives at a point 20 mm proximal to the carina. In order to place the ETT tip 40 mm (middle of 20-60 mm) from the carina after the indicator lamp turns on, the operator uses the ETT marker to withdraw the ETT 20 mm. For this study, five fresh swine airways were used in random order after drawing lots, and ten novices were recruited to attempt the procedure ten times on each swine’s airway.

Results

Five hundred endotracheal intubations were performed. For the target distance of 40 mm from the ETT tip to the carina, the mean (standard deviation) of the total data set was 37.9 (3.5) mm; all data were within a 20-60 mm range (500/500), and 98.2% (491/500) of the attempts fell within the 30-50 mm range.

Conclusions

The IR sensor stylet system can facilitate correct positioning of the ETT tip at an appropriate depth above the carina in the swine trachea. Evaluation of the IR sensor stylet methodology in human subjects is warranted.
Literature
1.
go back to reference Neumar RW, Otto CW, Link MS, et al. Part 8: Adult advanced cardiovascular life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation 2010; 122: S729-67.PubMedCrossRef Neumar RW, Otto CW, Link MS, et al. Part 8: Adult advanced cardiovascular life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation 2010; 122: S729-67.PubMedCrossRef
2.
go back to reference Salem MR. Verification of endotracheal tube position. Anesthesiol Clin North America 2001; 19: 813-39.PubMedCrossRef Salem MR. Verification of endotracheal tube position. Anesthesiol Clin North America 2001; 19: 813-39.PubMedCrossRef
3.
go back to reference Szekely SM, Webb RK, Williamson JA, Russell WJ. The Australian Incident Monitoring Study. Problems related to the endotracheal tube: an analysis of 2000 incident reports. Anaesth Intensive Care 1993; 21: 611-6. Szekely SM, Webb RK, Williamson JA, Russell WJ. The Australian Incident Monitoring Study. Problems related to the endotracheal tube: an analysis of 2000 incident reports. Anaesth Intensive Care 1993; 21: 611-6.
4.
go back to reference Owen RL, Cheney FW. Endobronchial intubation: A preventable complication. Anesthesiology 1987; 67: 255-7.PubMedCrossRef Owen RL, Cheney FW. Endobronchial intubation: A preventable complication. Anesthesiology 1987; 67: 255-7.PubMedCrossRef
5.
go back to reference Kim JT, Kim HJ, Ahn W, et al. Head rotation, flexion, and extension alter endotracheal tube position in adults and children. Can J Anesth 2009; 56: 751-6.PubMedCrossRef Kim JT, Kim HJ, Ahn W, et al. Head rotation, flexion, and extension alter endotracheal tube position in adults and children. Can J Anesth 2009; 56: 751-6.PubMedCrossRef
6.
go back to reference Morray JP, Geiduschek JM, Caplan RA, Posner KL, Gild WM, Cheney FW. A comparison of pediatric and adult anesthesia closed malpractice claims. Anesthesiology 1993; 78: 461-7.PubMedCrossRef Morray JP, Geiduschek JM, Caplan RA, Posner KL, Gild WM, Cheney FW. A comparison of pediatric and adult anesthesia closed malpractice claims. Anesthesiology 1993; 78: 461-7.PubMedCrossRef
7.
go back to reference Brunel W, Coleman DL, Schwartz DE, Peper E, Cohen NH. Assessment of routine chest roentgenograms and the physical examination to confirm endotracheal tube position. Chest 1989; 96: 1043-5.PubMedCrossRef Brunel W, Coleman DL, Schwartz DE, Peper E, Cohen NH. Assessment of routine chest roentgenograms and the physical examination to confirm endotracheal tube position. Chest 1989; 96: 1043-5.PubMedCrossRef
8.
go back to reference Rigini N, Boaz M, Ezri T, et al. Prompt correction of endotracheal tube positioning after intubation prevents further inappropriate positions. J Clin Anesth 2011; 23: 367-71.PubMedCrossRef Rigini N, Boaz M, Ezri T, et al. Prompt correction of endotracheal tube positioning after intubation prevents further inappropriate positions. J Clin Anesth 2011; 23: 367-71.PubMedCrossRef
9.
go back to reference Falzone E, Hoffmann C, Pasquier P, Boutonnet M, Tourtier JP. Proper depth placement of oral endotracheal tubes in adults: 21/23 cm or 20/22 cm? J Clin Anesth 2012; 24: 165-6.PubMedCrossRef Falzone E, Hoffmann C, Pasquier P, Boutonnet M, Tourtier JP. Proper depth placement of oral endotracheal tubes in adults: 21/23 cm or 20/22 cm? J Clin Anesth 2012; 24: 165-6.PubMedCrossRef
10.
go back to reference Knapp S, Kofler J, Stoiser B, et al. The assessment of four different methods to verify tracheal tube placement in the critical care setting. Anesth Analg 1999; 88: 766-70.PubMed Knapp S, Kofler J, Stoiser B, et al. The assessment of four different methods to verify tracheal tube placement in the critical care setting. Anesth Analg 1999; 88: 766-70.PubMed
11.
go back to reference Grmec S. Comparison of three different methods to confirm tracheal tube placement in emergency intubation. Intensive Care Med 2002; 28: 701-4.PubMedCrossRef Grmec S. Comparison of three different methods to confirm tracheal tube placement in emergency intubation. Intensive Care Med 2002; 28: 701-4.PubMedCrossRef
12.
go back to reference Dietrich KA, Conrad SA, Romero MD, Lloyd JL, Fontenot EE. Accuracy of flexible fiberoptic endoscopy in identifying abnormal endotracheal tube positions. Pediatr Emerg Care 1990; 6: 257-9.PubMedCrossRef Dietrich KA, Conrad SA, Romero MD, Lloyd JL, Fontenot EE. Accuracy of flexible fiberoptic endoscopy in identifying abnormal endotracheal tube positions. Pediatr Emerg Care 1990; 6: 257-9.PubMedCrossRef
13.
go back to reference Stone DJ. Airway management. In: Miller R (Ed.). Anesthesia, 5th ed. Churchill Livingstone; 2000: 1431-2. Stone DJ. Airway management. In: Miller R (Ed.). Anesthesia, 5th ed. Churchill Livingstone; 2000: 1431-2.
14.
go back to reference Sitzwohl C, Langheinrich A, Schober A, et al. Endobronchial intubation detected by insertion depth of endotracheal tube, bilateral auscultation, or observation of chest movements: randomised trial. BMJ 2010; 341: c5943.PubMedCrossRef Sitzwohl C, Langheinrich A, Schober A, et al. Endobronchial intubation detected by insertion depth of endotracheal tube, bilateral auscultation, or observation of chest movements: randomised trial. BMJ 2010; 341: c5943.PubMedCrossRef
15.
go back to reference Chun R, Kirkpatrick AW, Sirois M, et al. Where’s the tube? Evaluation of hand-held ultrasound in confirming endotracheal tube placement. Prehosp Disaster Med 2004; 19: 366-9.PubMed Chun R, Kirkpatrick AW, Sirois M, et al. Where’s the tube? Evaluation of hand-held ultrasound in confirming endotracheal tube placement. Prehosp Disaster Med 2004; 19: 366-9.PubMed
16.
go back to reference Lederman D, Lampotang S, Shamir MY. Automatic endotracheal tube position confirmation system based on image classification - a preliminary assessment. Med Eng Phys 2011; 33: 1017-26.PubMedCrossRef Lederman D, Lampotang S, Shamir MY. Automatic endotracheal tube position confirmation system based on image classification - a preliminary assessment. Med Eng Phys 2011; 33: 1017-26.PubMedCrossRef
17.
go back to reference Shamir MY, Lederman D, Gravenstein D. Automatic computerized endotracheal tube position verification: an animal model evaluation. Anesth Analg 2011; 113: 1411-5.PubMedCrossRef Shamir MY, Lederman D, Gravenstein D. Automatic computerized endotracheal tube position verification: an animal model evaluation. Anesth Analg 2011; 113: 1411-5.PubMedCrossRef
18.
go back to reference Reicher J, Reicher D, Reicher M. Use of radio frequency identification (RFID) tags in bedside monitoring of endotracheal tube position. J Clin Monit Comput 2007; 21: 155-8.PubMedCrossRef Reicher J, Reicher D, Reicher M. Use of radio frequency identification (RFID) tags in bedside monitoring of endotracheal tube position. J Clin Monit Comput 2007; 21: 155-8.PubMedCrossRef
19.
go back to reference Mohammad T. Using ultrasonic and infrared sensors for distance measurement. World Academy of Science, Engineering and Technology 2009; 27: 293-8. Mohammad T. Using ultrasonic and infrared sensors for distance measurement. World Academy of Science, Engineering and Technology 2009; 27: 293-8.
20.
go back to reference Schwartz DE, Matthay MA, Cohen NH. Death and other complications of emergency airway management in critically ill adults. A prospective investigation of 297 tracheal intubations. Anesthesiology 1995; 82: 367-76.PubMedCrossRef Schwartz DE, Matthay MA, Cohen NH. Death and other complications of emergency airway management in critically ill adults. A prospective investigation of 297 tracheal intubations. Anesthesiology 1995; 82: 367-76.PubMedCrossRef
21.
go back to reference Hagberg CA. Benumof’s airway management. Principles and Practice, 2nd ed. Mosby: Elsevier; 2007: 1061-3. Hagberg CA. Benumof’s airway management. Principles and Practice, 2nd ed. Mosby: Elsevier; 2007: 1061-3.
22.
go back to reference Gilbert D, Benumof JL. Biphasic carbon dioxide elimination waveform with right mainstem bronchial intubation. Anesth Analg 1989; 69: 829-32.PubMedCrossRef Gilbert D, Benumof JL. Biphasic carbon dioxide elimination waveform with right mainstem bronchial intubation. Anesth Analg 1989; 69: 829-32.PubMedCrossRef
23.
go back to reference Stewart RD, LaRosee A, Stoy WA, Heller MB. Use of a lighted stylet to confirm correct endotracheal tube placement. Chest 1987; 92: 900-3.PubMedCrossRef Stewart RD, LaRosee A, Stoy WA, Heller MB. Use of a lighted stylet to confirm correct endotracheal tube placement. Chest 1987; 92: 900-3.PubMedCrossRef
Metadata
Title
A novel method to position an endotracheal tube at the correct depth using an infrared sensor stylet
Authors
Yeongtak Song, Msc
Jaehoon Oh, MD
Youngjoon Chee, PhD
Taeho Lim, MD, PhD
Hyunggoo Kang, MD
Youngsuk Cho, MD
Publication date
01-05-2013
Publisher
Springer-Verlag
Published in
Canadian Journal of Anesthesia/Journal canadien d'anesthésie / Issue 5/2013
Print ISSN: 0832-610X
Electronic ISSN: 1496-8975
DOI
https://doi.org/10.1007/s12630-013-9898-6

Other articles of this Issue 5/2013

Canadian Journal of Anesthesia/Journal canadien d'anesthésie 5/2013 Go to the issue

Book and New Media Reviews

The 5-Minute ICU Consult