Published in:
01-06-2018 | Original Article
Validation of cadmium–zinc–telluride camera for measurement of left ventricular systolic performance
Authors:
Elisabeth Coupez, MD, Charles Merlin, MD, Viateur Tuyisenge, PhD, Laurent Sarry, PhD, Bruno Pereira, PhD, Jean René Lusson, MD, PhD, Louis Boyer, MD, PhD, Lucie Cassagnes, MD, PhD
Published in:
Journal of Nuclear Cardiology
|
Issue 3/2018
Login to get access
Abstract
Background
There are paucity of data comparing measurements of left ventricular systolic performance using cadmium–zinc–telluride (CZT) semiconductor cameras with other imaging modalities. This study compared the new system with echocardiography (echo) and cardiac magnetic resonance (CMR) imaging.
Methods
60 Patients presenting with ST-elevated myocardial infarction (MI) were included. Each patient underwent echo, myocardial perfusion imaging using Spectrum Dynamics D-SPECT(r) (CZT-SPECT), and CMR 6 weeks after MI. The primary endpoint was the agreement between CZT-SPECT and CMR for left ventricular ejection fraction (LVEF) measurement.
Results
48 of the 60 patients underwent all 3 studies (echo, CMR, and CZT-SPECT) 40 days after admission. CZT-SPECT and CMR LVEF were well correlated (r = .79, P < .0001), as well as CZT-SPECT vs echo and CMR vs echo (r = .79 and .84, respectively, P < .0001). The segmental LV wall thickening and wall motion also showed good concordance between three techniques.
Conclusions
CZT-SPECT is reliable for LVEF measurement.