How Have Alternative Bearings and Modularity Affected Revision Rates in Total Hip Arthroplasty?

Authors: William M. Mihalko, MD, PhD, Markus A. Wimmer, PhD, Carol A. Pacione, BS, Michel P. Laurent, PhD, Robert F. Murphy, MD, Carson Rider, BS

Published in: Clinical Orthopaedics and Related Research® | Issue 12/2014

Login to get access

Abstract

Background

Total hip arthroplasty (THA) continues to be one of the most successful surgical procedures in the medical field. However, over the last two decades, the use of modularity and alternative bearings in THA has become routine. Given the known problems associated with hard-on-hard bearing couples, including taper failures with more modular stem designs, local and systemic effects from metal-on-metal bearings, and fractures with ceramic-on-ceramic bearings, it is not known whether in aggregate the survivorship of these implants is better or worse than the metal-on-polyethylene bearings that they sought to replace.

Questions/purposes

Have alternative bearings (metal-on-metal and ceramic-on-ceramic) and implant modularity decreased revision rates of primary THAs?

Methods

In this systematic review of MEDLINE and EMBASE, we used several Boolean search strings for each topic and surveyed national registry data from English-speaking countries. Clinical research (Level IV or higher) with ≥ 5 years of followup was included; retrieval studies and case reports were excluded. We included registry data at ≥ 7 years followup. A total of 32 studies (and five registry reports) on metal-on-metal, 19 studies (and five registry reports) on ceramic-on-ceramic, and 20 studies (and one registry report) on modular stem designs met inclusion criteria and were evaluated in detail. Insufficient data were available on metal-on-ceramic and ceramic-on-metal implants, and monoblock acetabular designs were evaluated in another recent systematic review so these were not evaluated here.

Results

There was no evidence in the literature that alternative bearings (either metal-on-metal or ceramic-on-ceramic) in THA have decreased revision rates. Registry data, however, showed that large head metal-on-metal implants have lower 7- to 10-year survivorship than do standard bearings. In THA, modular exchangeable femoral neck implants had a lower 10-year survival rate in both literature reviews and in registry data compared with combined registry primary THA implant survivorship.

Conclusions

Despite improvements in implant technology, there is no evidence that alternative bearings or modularity have resulted in decreased THA revision rates after 5 years. In fact, both large head metal-on-metal THA and added modularity may well lower survivorship and should only be used in select cases in which the mission cannot be achieved without it. Based on this experience, followup and/or postmarket surveillance studies should have a duration of at least 5 years before introducing new alternative bearings or modularity on a widespread scale.
Literature
3.
go back to reference Beldame J, Carreras F, Oger P, Beaufils P. Cementless cups do not increase osteolysis risk in metal-on-metal total hip arthroplasty. Apropos of 106 cases. Orthop Traumatol Surg Res. 2009;95:478–490.PubMedCrossRef Beldame J, Carreras F, Oger P, Beaufils P. Cementless cups do not increase osteolysis risk in metal-on-metal total hip arthroplasty. Apropos of 106 cases. Orthop Traumatol Surg Res. 2009;95:478–490.PubMedCrossRef
4.
go back to reference Benazzo F, Rossi SM, Cecconi D, Piovani L, Ravasi F. Mid-term results of an uncemented femoral stem with modular neck options. Hip Int. 2010;20:427–433.PubMed Benazzo F, Rossi SM, Cecconi D, Piovani L, Ravasi F. Mid-term results of an uncemented femoral stem with modular neck options. Hip Int. 2010;20:427–433.PubMed
5.
go back to reference Bernstein M, Desy NM, Petit A, Zukor DJ, Huk OL, Antoniou J. Long-term follow up and metal ion trend of patients with metal-on-metal total hip arthroplasty. Int Orthop. 2012;36:1807–1812.PubMedCentralPubMedCrossRef Bernstein M, Desy NM, Petit A, Zukor DJ, Huk OL, Antoniou J. Long-term follow up and metal ion trend of patients with metal-on-metal total hip arthroplasty. Int Orthop. 2012;36:1807–1812.PubMedCentralPubMedCrossRef
6.
go back to reference Biant LC, Bruce WJ, Assini JB, Walker PM, Walsh WR. The anatomically difficult primary total hip replacement: medium- to long-term results using a cementless modular stem. J Bone Joint Surg Br. 2008;90:430–435.PubMedCrossRef Biant LC, Bruce WJ, Assini JB, Walker PM, Walsh WR. The anatomically difficult primary total hip replacement: medium- to long-term results using a cementless modular stem. J Bone Joint Surg Br. 2008;90:430–435.PubMedCrossRef
7.
go back to reference Bizot P, Banallec L, Sedel L, Nizard R. Alumina-on-alumina total hip prostheses in patients 40 years of age or younger. Clin Orthop Relat Res. 2000;379:68–76.PubMedCrossRef Bizot P, Banallec L, Sedel L, Nizard R. Alumina-on-alumina total hip prostheses in patients 40 years of age or younger. Clin Orthop Relat Res. 2000;379:68–76.PubMedCrossRef
8.
go back to reference Blakey CM, Eswaramoorthy VK, Hamilton LC, Biant LC, Field RE. Mid-term results of the modular ANCA-Fit femoral component in total hip replacement. J Bone Joint Surg Br. 2009;91:1561–1565.PubMedCrossRef Blakey CM, Eswaramoorthy VK, Hamilton LC, Biant LC, Field RE. Mid-term results of the modular ANCA-Fit femoral component in total hip replacement. J Bone Joint Surg Br. 2009;91:1561–1565.PubMedCrossRef
9.
go back to reference Bolland BJ, Culliford DJ, Langton DJ, Millington JP, Arden NK, Latham JM. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br. 2011;93:608–615.PubMedCrossRef Bolland BJ, Culliford DJ, Langton DJ, Millington JP, Arden NK, Latham JM. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br. 2011;93:608–615.PubMedCrossRef
10.
go back to reference Boyer P, Huten D, Loriaut P, Lestrat V, Jeanrot C, Massin P. Is alumina-on-alumina ceramic bearings total hip replacement the right choice in patients younger than 50 years of age? A 7- to 15-year follow-up study. Orthop Traumatol Surg Res. 2010;96:616–622.PubMedCrossRef Boyer P, Huten D, Loriaut P, Lestrat V, Jeanrot C, Massin P. Is alumina-on-alumina ceramic bearings total hip replacement the right choice in patients younger than 50 years of age? A 7- to 15-year follow-up study. Orthop Traumatol Surg Res. 2010;96:616–622.PubMedCrossRef
11.
go back to reference Brown SR, Davies WA, DeHeer DH, Swanson AB. Long-term survival of McKee-Farrar total hip prostheses. Clin Orthop Relat Res. 2002;402:157–163.PubMedCrossRef Brown SR, Davies WA, DeHeer DH, Swanson AB. Long-term survival of McKee-Farrar total hip prostheses. Clin Orthop Relat Res. 2002;402:157–163.PubMedCrossRef
12.
go back to reference Cameron HU, Keppler L, McTighe T. The role of modularity in primary total hip arthroplasty. J Arthroplasty. 2006;21(Suppl 1):89–92.PubMedCrossRef Cameron HU, Keppler L, McTighe T. The role of modularity in primary total hip arthroplasty. J Arthroplasty. 2006;21(Suppl 1):89–92.PubMedCrossRef
13.
go back to reference Chang JS, Han DJ, Park SK, Sung JH, Ha YC. Cementless total hip arthroplasty in patients with osteonecrosis after kidney transplantation. J Arthroplasty. 2013;28:824–827.PubMedCrossRef Chang JS, Han DJ, Park SK, Sung JH, Ha YC. Cementless total hip arthroplasty in patients with osteonecrosis after kidney transplantation. J Arthroplasty. 2013;28:824–827.PubMedCrossRef
14.
go back to reference Clement ND, Biant LC, Breusch SJ. Total hip arthroplasty: to cement or not to cement the acetabular socket? A critical review of the literature. Arch Orthop Trauma Surg. 2012;132:411–427.PubMedCrossRef Clement ND, Biant LC, Breusch SJ. Total hip arthroplasty: to cement or not to cement the acetabular socket? A critical review of the literature. Arch Orthop Trauma Surg. 2012;132:411–427.PubMedCrossRef
15.
go back to reference Cooper HJ, Urban RM, Wixson RL, Meneghini RM, Jacobs JJ. Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck. J Bone Joint Surg Am. 2013;95:865–872.PubMedCentralPubMedCrossRef Cooper HJ, Urban RM, Wixson RL, Meneghini RM, Jacobs JJ. Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck. J Bone Joint Surg Am. 2013;95:865–872.PubMedCentralPubMedCrossRef
16.
go back to reference D’Antonio JA, Capello WN, Naughton M. Ceramic bearings for total hip arthroplasty have high survivorship at 10 years. Clin Orthop Relat Res. 2012;470:373–381.PubMedCentralPubMedCrossRef D’Antonio JA, Capello WN, Naughton M. Ceramic bearings for total hip arthroplasty have high survivorship at 10 years. Clin Orthop Relat Res. 2012;470:373–381.PubMedCentralPubMedCrossRef
17.
go back to reference Dagnino A, Grappiolo G, Benazzo FM, Learmonth ID, Spotorno L, Portinaro N. Medium-term outcome in patients treated with total hip arthroplasty using a modular femoral stem. Hip Int. 2012;22:274–279.PubMedCrossRef Dagnino A, Grappiolo G, Benazzo FM, Learmonth ID, Spotorno L, Portinaro N. Medium-term outcome in patients treated with total hip arthroplasty using a modular femoral stem. Hip Int. 2012;22:274–279.PubMedCrossRef
18.
go back to reference Dastane MR, Long WT, Wan Z, Chao L, Dorr LD. Metal-on-metal hip arthroplasty does equally well in osteonecrosis and osteoarthritis. Clin Orthop Relat Res. 2008;466:1148–1153.PubMedCentralPubMedCrossRef Dastane MR, Long WT, Wan Z, Chao L, Dorr LD. Metal-on-metal hip arthroplasty does equally well in osteonecrosis and osteoarthritis. Clin Orthop Relat Res. 2008;466:1148–1153.PubMedCentralPubMedCrossRef
19.
go back to reference Daurka JS, Malik AK, Robin DA, Witt JD. The results of uncemented total hip replacement in children with juvenile idiopathic arthritis at ten years. J Bone Joint Surg Br. 2012;94:1618–1624.PubMedCrossRef Daurka JS, Malik AK, Robin DA, Witt JD. The results of uncemented total hip replacement in children with juvenile idiopathic arthritis at ten years. J Bone Joint Surg Br. 2012;94:1618–1624.PubMedCrossRef
20.
go back to reference De la Torre BJ, Chaparro M, Romanillos JO, Zarzoso S, Mosquera M, Rodriguez G. 10 years results of an uncemented metaphyseal fit modular stem in elderly patients. Indian J Orthop. 2011;45:351–358.PubMedCentralPubMedCrossRef De la Torre BJ, Chaparro M, Romanillos JO, Zarzoso S, Mosquera M, Rodriguez G. 10 years results of an uncemented metaphyseal fit modular stem in elderly patients. Indian J Orthop. 2011;45:351–358.PubMedCentralPubMedCrossRef
21.
go back to reference Delaunay CP, Bonnomet F, Clavert P, Laffargue P, Migaud H. THA using metal-on-metal articulation in active patients younger than 50 years. Clin Orthop Relat Res. 2008;466:340–346.PubMedCentralPubMedCrossRef Delaunay CP, Bonnomet F, Clavert P, Laffargue P, Migaud H. THA using metal-on-metal articulation in active patients younger than 50 years. Clin Orthop Relat Res. 2008;466:340–346.PubMedCentralPubMedCrossRef
22.
go back to reference Descamps S, Bouillet B, Boisgard S, Levai JP. High incidence of loosening at 5-year follow-up of a cemented metal-on-metal acetabular component in THR. Eur J Orthop Surg Traumatol. 2009;19:559–563.CrossRef Descamps S, Bouillet B, Boisgard S, Levai JP. High incidence of loosening at 5-year follow-up of a cemented metal-on-metal acetabular component in THR. Eur J Orthop Surg Traumatol. 2009;19:559–563.CrossRef
23.
go back to reference Dorr LD, Wan Z, Longjohn DB, Dubois B, Murken R. Total hip arthroplasty with use of the Metasul metal-on-metal articulation. Four to seven-year results. J Bone Joint Surg Am. 2000;82:789–798.PubMed Dorr LD, Wan Z, Longjohn DB, Dubois B, Murken R. Total hip arthroplasty with use of the Metasul metal-on-metal articulation. Four to seven-year results. J Bone Joint Surg Am. 2000;82:789–798.PubMed
24.
go back to reference Drexler M, Dwyer T, Marmor M, Abolghasemian M, Chakravertty R, Chechik O, Cameron HU. Nineteen-year results of THA using modular 9 mm S-ROM femoral component in patients with small femoral canals. J Arthroplasty. 2013;28:1667–1670.PubMedCrossRef Drexler M, Dwyer T, Marmor M, Abolghasemian M, Chakravertty R, Chechik O, Cameron HU. Nineteen-year results of THA using modular 9 mm S-ROM femoral component in patients with small femoral canals. J Arthroplasty. 2013;28:1667–1670.PubMedCrossRef
25.
go back to reference Engh CA Jr, Ho H, Engh CA, Hamilton WG, Fricka KB. Metal-on-metal total hip arthroplasty adverse local tissue reaction. Semin Arthroplasty. 2010;21:19–23.CrossRef Engh CA Jr, Ho H, Engh CA, Hamilton WG, Fricka KB. Metal-on-metal total hip arthroplasty adverse local tissue reaction. Semin Arthroplasty. 2010;21:19–23.CrossRef
26.
go back to reference Eswaramoorthy V, Moonot P, Kalairajah Y, Biant LC, Field RE. The Metasul metal-on-metal articulation in primary total hip replacement: clinical and radiological results at ten years. J Bone Joint Surg Br. 2008;90:1278–1283.PubMedCrossRef Eswaramoorthy V, Moonot P, Kalairajah Y, Biant LC, Field RE. The Metasul metal-on-metal articulation in primary total hip replacement: clinical and radiological results at ten years. J Bone Joint Surg Br. 2008;90:1278–1283.PubMedCrossRef
27.
go back to reference Gill IP, Webb J, Sloan K, Beaver RJ. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Joint Surg Br. 2012;94:895–900.PubMedCrossRef Gill IP, Webb J, Sloan K, Beaver RJ. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Joint Surg Br. 2012;94:895–900.PubMedCrossRef
28.
go back to reference Girard J, Bocquet D, Autissier G, Fouilleron N, Fron D, Migaud H. Metal-on-metal hip arthroplasty in patients thirty years of age or younger. J Bone Joint Surg Am. 2010;92:2419–2426.PubMedCrossRef Girard J, Bocquet D, Autissier G, Fouilleron N, Fron D, Migaud H. Metal-on-metal hip arthroplasty in patients thirty years of age or younger. J Bone Joint Surg Am. 2010;92:2419–2426.PubMedCrossRef
29.
go back to reference Girard J, Combes A, Herent S, Migaud H. Metal-on-metal cups cemented into reinforcement rings: a possible new acetabular reconstruction procedure for young and active patients. J Arthroplasty. 2011;26:103–109.PubMedCrossRef Girard J, Combes A, Herent S, Migaud H. Metal-on-metal cups cemented into reinforcement rings: a possible new acetabular reconstruction procedure for young and active patients. J Arthroplasty. 2011;26:103–109.PubMedCrossRef
30.
go back to reference Grübl A, Marker M, Brodner W, Giurea A, Heinze G, Meisinger V, Zehetgruber H, Kotz R. Long-term follow-up of metal-on-metal total hip replacement. J Orthop Res. 2007;25:841–848.PubMedCrossRef Grübl A, Marker M, Brodner W, Giurea A, Heinze G, Meisinger V, Zehetgruber H, Kotz R. Long-term follow-up of metal-on-metal total hip replacement. J Orthop Res. 2007;25:841–848.PubMedCrossRef
31.
go back to reference Halma JJ, Vogely HC, Dhert WJ, Van Gaalen SM, de Gast A. Do monoblock cups improve survivorship, decrease wear, or reduce osteolysis in uncemented total hip arthroplasty? Clin Orthop Relat Res. 2013;471:3572–3580.PubMedCrossRef Halma JJ, Vogely HC, Dhert WJ, Van Gaalen SM, de Gast A. Do monoblock cups improve survivorship, decrease wear, or reduce osteolysis in uncemented total hip arthroplasty? Clin Orthop Relat Res. 2013;471:3572–3580.PubMedCrossRef
32.
go back to reference Hasegawa M, Sudo A, Uchida A. Alumina ceramic-on-ceramic total hip replacement with a layered acetabular component. J Bone Joint Surg Br. 2006;88:877–882.PubMedCrossRef Hasegawa M, Sudo A, Uchida A. Alumina ceramic-on-ceramic total hip replacement with a layered acetabular component. J Bone Joint Surg Br. 2006;88:877–882.PubMedCrossRef
33.
go back to reference Hsu JE, Kinsella SD, Garino JP, Lee GC. Ten-year follow-up of patients younger than 50 years with modern ceramic-on-ceramic total hip arthroplasty. Semin Arthroplasty. 2011;22:229–233.CrossRef Hsu JE, Kinsella SD, Garino JP, Lee GC. Ten-year follow-up of patients younger than 50 years with modern ceramic-on-ceramic total hip arthroplasty. Semin Arthroplasty. 2011;22:229–233.CrossRef
34.
go back to reference Hwang KT, Kim YH, Kim YS, Choi IY. Cementless total hip arthroplasty with a metal-on-metal bearing in patients younger than 50 years. J Arthroplasty. 2011;26:1481–1487.PubMedCrossRef Hwang KT, Kim YH, Kim YS, Choi IY. Cementless total hip arthroplasty with a metal-on-metal bearing in patients younger than 50 years. J Arthroplasty. 2011;26:1481–1487.PubMedCrossRef
35.
go back to reference Ito H, Matsuno T, Aok Y, Minami A. Total hip arthroplasty using an Omniflex modular system: 5 to 12 years followup. Clin Orthop Relat Res. 2004;419:98–106.PubMedCrossRef Ito H, Matsuno T, Aok Y, Minami A. Total hip arthroplasty using an Omniflex modular system: 5 to 12 years followup. Clin Orthop Relat Res. 2004;419:98–106.PubMedCrossRef
36.
go back to reference Jack CM, Molloy DO, Walter WL, Zicat BA, Walter WK. The use of ceramic-on-ceramic bearings in isolated revision of the acetabular component. Bone Joint J. 2013;95:333–338.PubMedCrossRef Jack CM, Molloy DO, Walter WL, Zicat BA, Walter WK. The use of ceramic-on-ceramic bearings in isolated revision of the acetabular component. Bone Joint J. 2013;95:333–338.PubMedCrossRef
37.
go back to reference Jameson SS, Mason JM, Baker PN, Jettoo P, Deehan DJ, Reed MR. Factors influencing revision risk following 15,740 single-brand hybrid hip arthroplasties: a cohort study from a National Joint Registry. J Arthroplasty. 2013;28:1152–1159.PubMedCrossRef Jameson SS, Mason JM, Baker PN, Jettoo P, Deehan DJ, Reed MR. Factors influencing revision risk following 15,740 single-brand hybrid hip arthroplasties: a cohort study from a National Joint Registry. J Arthroplasty. 2013;28:1152–1159.PubMedCrossRef
38.
go back to reference Kamath AF, Prieto H, Lewallen DG. Alternative bearings in total hip arthroplasty in the young patient. Orthop Clin North Am. 2013;44:451–462.PubMedCrossRef Kamath AF, Prieto H, Lewallen DG. Alternative bearings in total hip arthroplasty in the young patient. Orthop Clin North Am. 2013;44:451–462.PubMedCrossRef
39.
go back to reference Kawano S, Sonohata M, Shimazaki T, Kitajima M, Mawatari M, Hotokebuchi T. Failure analysis of alumina on alumina total hip arthroplasty with a layered acetabular component: minimum ten-year follow-up study. J Arthroplasty. 2013;28:1822–1827.PubMedCrossRef Kawano S, Sonohata M, Shimazaki T, Kitajima M, Mawatari M, Hotokebuchi T. Failure analysis of alumina on alumina total hip arthroplasty with a layered acetabular component: minimum ten-year follow-up study. J Arthroplasty. 2013;28:1822–1827.PubMedCrossRef
40.
go back to reference Kim SM, Lim SJ, Moon YW, Kim YT, Ko KR, Park YS. Cementless modular total hip arthroplasty in patients younger than fifty with femoral head osteonecrosis: minimum fifteen-year follow-up. J Arthroplasty. 2013;28:504–509.PubMedCrossRef Kim SM, Lim SJ, Moon YW, Kim YT, Ko KR, Park YS. Cementless modular total hip arthroplasty in patients younger than fifty with femoral head osteonecrosis: minimum fifteen-year follow-up. J Arthroplasty. 2013;28:504–509.PubMedCrossRef
41.
go back to reference Kim YH, Choi Y, Kim JS. Cementless total hip arthroplasty with ceramic-on-ceramic bearing in patients younger than 45 years with femoral-head osteonecrosis. Int Orthop. 2010;34:1123–1127.PubMedCentralPubMedCrossRef Kim YH, Choi Y, Kim JS. Cementless total hip arthroplasty with ceramic-on-ceramic bearing in patients younger than 45 years with femoral-head osteonecrosis. Int Orthop. 2010;34:1123–1127.PubMedCentralPubMedCrossRef
42.
go back to reference Kindsfater KA, SychterzTerefenko CJ, Gruen TA, Sherman CM. Minimum 5-year results of modular metal-on-metal total hip arthroplasty. J Arthroplasty. 2012;27:545–550.PubMedCrossRef Kindsfater KA, SychterzTerefenko CJ, Gruen TA, Sherman CM. Minimum 5-year results of modular metal-on-metal total hip arthroplasty. J Arthroplasty. 2012;27:545–550.PubMedCrossRef
43.
go back to reference Korovessis P, Petsinis G, Repanti M, Repantis T. Metallosis after contemporary metal-on-metal total hip arthroplasty. Five to nine-year follow-up. J Bone Joint Surg Am. 2006;88:1183–1191.PubMedCrossRef Korovessis P, Petsinis G, Repanti M, Repantis T. Metallosis after contemporary metal-on-metal total hip arthroplasty. Five to nine-year follow-up. J Bone Joint Surg Am. 2006;88:1183–1191.PubMedCrossRef
44.
go back to reference Krishnan H, Krishnan SP, Blunn G, Skinner JA, Hart AJ. Modular neck femoral stems. Bone Joint J. 2013;95:1011–1021.PubMedCrossRef Krishnan H, Krishnan SP, Blunn G, Skinner JA, Hart AJ. Modular neck femoral stems. Bone Joint J. 2013;95:1011–1021.PubMedCrossRef
45.
go back to reference Kusaba A, Sunami H, Kondo S, Kuroki Y. Uncemented ceramic-on-ceramic bearing couple for dysplastic osteoarthritis: a 5- to 11-year follow-up study. Semin Arthroplasty. 2011;22:240–247.CrossRef Kusaba A, Sunami H, Kondo S, Kuroki Y. Uncemented ceramic-on-ceramic bearing couple for dysplastic osteoarthritis: a 5- to 11-year follow-up study. Semin Arthroplasty. 2011;22:240–247.CrossRef
46.
go back to reference Langton DJ, Joyce TJ, Jameson SS, Lord J, Van Orsouw M, Holland JP, Nargol AV, DeSmet KA. Adverse reaction to metal debris following hip resurfacing: the influence of component type, orientation and volumetric wear. J Bone Joint Surg Br. 2011;93:164–171.PubMedCrossRef Langton DJ, Joyce TJ, Jameson SS, Lord J, Van Orsouw M, Holland JP, Nargol AV, DeSmet KA. Adverse reaction to metal debris following hip resurfacing: the influence of component type, orientation and volumetric wear. J Bone Joint Surg Br. 2011;93:164–171.PubMedCrossRef
47.
go back to reference Lazennec JY, Boyer P, Poupon J, Rousseau MA, Roy C, Ravaud P, Catonné Y. Outcome and serum ion determination up to 11 years after implantation of a cemented metal-on-metal hip prosthesis. Acta Orthop. 2009;80:168–173.PubMedCentralPubMedCrossRef Lazennec JY, Boyer P, Poupon J, Rousseau MA, Roy C, Ravaud P, Catonné Y. Outcome and serum ion determination up to 11 years after implantation of a cemented metal-on-metal hip prosthesis. Acta Orthop. 2009;80:168–173.PubMedCentralPubMedCrossRef
48.
go back to reference Le D, Smith K, Tanzer D, Tanzer M. Modular femoral sleeve and stem implant provides long-term total hip survivorship. Clin Orthop Relat Res. 2011;469:508–513.PubMedCentralPubMedCrossRef Le D, Smith K, Tanzer D, Tanzer M. Modular femoral sleeve and stem implant provides long-term total hip survivorship. Clin Orthop Relat Res. 2011;469:508–513.PubMedCentralPubMedCrossRef
49.
go back to reference Liudahl AA, Liu SS, Goetz DD, Mahoney CR, Callaghan JJ. Metal on metal total hip arthroplasty using modular acetabular shells. J Arthroplasty. 2013;28:867–871.PubMedCrossRef Liudahl AA, Liu SS, Goetz DD, Mahoney CR, Callaghan JJ. Metal on metal total hip arthroplasty using modular acetabular shells. J Arthroplasty. 2013;28:867–871.PubMedCrossRef
50.
go back to reference Lombardi AV Jr, Berend KR, Seng BE, Clarke IC, Adams JB. Delta ceramic-on-alumina ceramic articulation in primary THA: prospective, randomized FDA-IDE study and retrieval analysis. Clin Orthop Relat Res. 2010;468:367–374.PubMedCentralPubMedCrossRef Lombardi AV Jr, Berend KR, Seng BE, Clarke IC, Adams JB. Delta ceramic-on-alumina ceramic articulation in primary THA: prospective, randomized FDA-IDE study and retrieval analysis. Clin Orthop Relat Res. 2010;468:367–374.PubMedCentralPubMedCrossRef
51.
go back to reference Long WT, Dorr LD, Gendelman V. An American experience with metal-on-metal total hip arthroplasties: a 7-year follow-up study. J Arthroplasty. 2004;19(Suppl 3):29–34.PubMed Long WT, Dorr LD, Gendelman V. An American experience with metal-on-metal total hip arthroplasties: a 7-year follow-up study. J Arthroplasty. 2004;19(Suppl 3):29–34.PubMed
52.
go back to reference Malem D, Nagy MT, Ghosh S, Shah B. Catastrophic failure of ceramic-on-ceramic total hip arthroplasty presenting as squeaking hip. BMJ Case Rep. 2013;2013. pii: bcr-2013-008614. doi: 10.1136/bcr-2013-008614. Malem D, Nagy MT, Ghosh S, Shah B. Catastrophic failure of ceramic-on-ceramic total hip arthroplasty presenting as squeaking hip. BMJ Case Rep. 2013;2013. pii: bcr-2013-008614. doi: 10.​1136/​bcr-2013-008614.
53.
go back to reference Mesko JW, D’Antonio JA, Capello WN, Bierbaum BE, Naughton M. Ceramic-on-ceramic hip outcome at a 5- to 10-year interval: has it lived up to its expectations? J Arthroplasty. 2011;26:172–177.PubMedCrossRef Mesko JW, D’Antonio JA, Capello WN, Bierbaum BE, Naughton M. Ceramic-on-ceramic hip outcome at a 5- to 10-year interval: has it lived up to its expectations? J Arthroplasty. 2011;26:172–177.PubMedCrossRef
54.
go back to reference Migaud H, Jobin A, Chantelot C, Giraud F, Laffargue P, Duquennoy A. Cementless metal-on-metal hip arthroplasty in patients less than 50 years of age: comparison with a matched control group using ceramic-on-polyethylene after a minimum 5-year follow-up. J Arthroplasty. 2004;19(Suppl 3):23–28.PubMed Migaud H, Jobin A, Chantelot C, Giraud F, Laffargue P, Duquennoy A. Cementless metal-on-metal hip arthroplasty in patients less than 50 years of age: comparison with a matched control group using ceramic-on-polyethylene after a minimum 5-year follow-up. J Arthroplasty. 2004;19(Suppl 3):23–28.PubMed
55.
go back to reference Milosev I, Trebse R, Kovac S, Cör A, Pisot V. Survivorship and retrieval analysis of Sikomet metal-on-metal total hip replacements at a mean of seven years. J Bone Joint Surg Am. 2006;88:1173–1182.PubMedCrossRef Milosev I, Trebse R, Kovac S, Cör A, Pisot V. Survivorship and retrieval analysis of Sikomet metal-on-metal total hip replacements at a mean of seven years. J Bone Joint Surg Am. 2006;88:1173–1182.PubMedCrossRef
57.
go back to reference Neuerburg C, Impellizzeri F, Goldhahn J, Frey P, Naal FD, von Knoch M, Leunig M, von Knoch F. Survivorship of second-generation metal-on-metal primary total hip replacement. Arch Orthop Trauma Surg. 2012;132:527–353. Neuerburg C, Impellizzeri F, Goldhahn J, Frey P, Naal FD, von Knoch M, Leunig M, von Knoch F. Survivorship of second-generation metal-on-metal primary total hip replacement. Arch Orthop Trauma Surg. 2012;132:527–353.
58.
go back to reference Neumann DR, Thaler C, Hitzl W, Huber M, Hofstädter T, Dorn U. Long-term results of a contemporary metal-on-metal total hip arthroplasty: a 10-year follow-up study. J Arthroplasty. 2010;25:700–708.PubMedCrossRef Neumann DR, Thaler C, Hitzl W, Huber M, Hofstädter T, Dorn U. Long-term results of a contemporary metal-on-metal total hip arthroplasty: a 10-year follow-up study. J Arthroplasty. 2010;25:700–708.PubMedCrossRef
60.
go back to reference Nikolaou VS, Edwards MR, Bogoch E, Schemitsch EH, Waddell JP. A prospective randomised controlled trial comparing three alternative bearing surfaces in primary total hip replacement. J Bone Joint Surg Br. 2012;94:459–465.PubMedCrossRef Nikolaou VS, Edwards MR, Bogoch E, Schemitsch EH, Waddell JP. A prospective randomised controlled trial comparing three alternative bearing surfaces in primary total hip replacement. J Bone Joint Surg Br. 2012;94:459–465.PubMedCrossRef
61.
go back to reference Nikolaou VS, Petit A, Debiparshad K, Huk OL, Zukor DJ, Antoniou J. Metal-on-metal total hip arthroplasty–five- to 11-year follow-up. Bull NYU Hosp Jt Dis. 2011;69(Suppl 1):S77–83.PubMed Nikolaou VS, Petit A, Debiparshad K, Huk OL, Zukor DJ, Antoniou J. Metal-on-metal total hip arthroplasty–five- to 11-year follow-up. Bull NYU Hosp Jt Dis. 2011;69(Suppl 1):S77–83.PubMed
63.
go back to reference Petsatodis GE, Papadopoulos PP, Papavasiliou KA, Hatzokos IG, Agathangelidis FG, Christodoulou AG. Primary cementless total hip arthroplasty with an alumina ceramic-on-ceramic bearing: results after a minimum of twenty years of follow-up. J Bone Joint Surg Am. 2010;92:639–644.PubMedCrossRef Petsatodis GE, Papadopoulos PP, Papavasiliou KA, Hatzokos IG, Agathangelidis FG, Christodoulou AG. Primary cementless total hip arthroplasty with an alumina ceramic-on-ceramic bearing: results after a minimum of twenty years of follow-up. J Bone Joint Surg Am. 2010;92:639–644.PubMedCrossRef
64.
go back to reference Randelli F, Banci L, D’Anna A, Visentin O, Randelli G. Cementless Metasul metal-on-metal total hip arthroplasties at 13 years. J Arthroplasty. 2012;27:186–192.PubMedCrossRef Randelli F, Banci L, D’Anna A, Visentin O, Randelli G. Cementless Metasul metal-on-metal total hip arthroplasties at 13 years. J Arthroplasty. 2012;27:186–192.PubMedCrossRef
65.
go back to reference Reito A, Eskelinen A, Puolakka T, Pajamäki J. Results of metal-on-metal hip resurfacing in patients 40 years old and younger. Arch Orthop Trauma Surg. 2013;133:267–273.PubMedCrossRef Reito A, Eskelinen A, Puolakka T, Pajamäki J. Results of metal-on-metal hip resurfacing in patients 40 years old and younger. Arch Orthop Trauma Surg. 2013;133:267–273.PubMedCrossRef
66.
go back to reference Repantis T, Vitsas V, Korovessis P. Poor mid-term survival of the low-carbide metal-on-metal Zweymüller-plus total hip arthroplasty system: a concise follow-up, at a minimum of ten years, of a previous report. J Bone Joint Surg Am. 2013;95:e331–334.PubMedCrossRef Repantis T, Vitsas V, Korovessis P. Poor mid-term survival of the low-carbide metal-on-metal Zweymüller-plus total hip arthroplasty system: a concise follow-up, at a minimum of ten years, of a previous report. J Bone Joint Surg Am. 2013;95:e331–334.PubMedCrossRef
67.
go back to reference Saito S, Ryu J, Watanabe M, Ishii T, Saigo K. Midterm results of Metasul metal-on-metal total hip arthroplasty. J Arthroplasty. 2006;21:1105–1110.PubMedCrossRef Saito S, Ryu J, Watanabe M, Ishii T, Saigo K. Midterm results of Metasul metal-on-metal total hip arthroplasty. J Arthroplasty. 2006;21:1105–1110.PubMedCrossRef
68.
go back to reference Sakai T, Ohzono K, Nishii T, Miki H, Takao M, Sugano N. A modular femoral neck and head system works well in cementless total hip replacement for patients with developmental dysplasia of the hip. J Bone Joint Surg Br. 2010;92:770–776.PubMedCrossRef Sakai T, Ohzono K, Nishii T, Miki H, Takao M, Sugano N. A modular femoral neck and head system works well in cementless total hip replacement for patients with developmental dysplasia of the hip. J Bone Joint Surg Br. 2010;92:770–776.PubMedCrossRef
69.
go back to reference Seyler TM, Bonutti PM, Shen J, Naughton M, Kester M. Use of an alumina-on-alumina bearing system in total hip arthroplasty for osteonecrosis of the hip. J Bone Joint Surg Am. 2006;88(Suppl 3):116–125.PubMedCrossRef Seyler TM, Bonutti PM, Shen J, Naughton M, Kester M. Use of an alumina-on-alumina bearing system in total hip arthroplasty for osteonecrosis of the hip. J Bone Joint Surg Am. 2006;88(Suppl 3):116–125.PubMedCrossRef
70.
go back to reference Shiga T, Mori M, Hayashida T, Fujiwara Y, Ogura T. Disassembly of a modular femoral component after femoral head prosthetic replacement. J Arthroplasty. 2010;25:659.e17–19. Shiga T, Mori M, Hayashida T, Fujiwara Y, Ogura T. Disassembly of a modular femoral component after femoral head prosthetic replacement. J Arthroplasty. 2010;25:659.e17–19.
71.
go back to reference Sporer SM, Obar RJ, Bernini PM. Primary total hip arthroplasty using a modular proximally coated prosthesis in patients older than 70: two to eight year results. J Arthroplasty. 2004;19:197–203.PubMedCrossRef Sporer SM, Obar RJ, Bernini PM. Primary total hip arthroplasty using a modular proximally coated prosthesis in patients older than 70: two to eight year results. J Arthroplasty. 2004;19:197–203.PubMedCrossRef
72.
go back to reference Steppacher SD, Ecker TM, Tannast M, Murphy SB. Absence of osteolysis in uncemented alumina ceramic-on-ceramic THA in patients younger than 50 years after two to 14 years. Semin Arthroplasty. 2011;22:248–253.CrossRef Steppacher SD, Ecker TM, Tannast M, Murphy SB. Absence of osteolysis in uncemented alumina ceramic-on-ceramic THA in patients younger than 50 years after two to 14 years. Semin Arthroplasty. 2011;22:248–253.CrossRef
73.
go back to reference Suehara H, Fujioka M, Inoue S, Takahashi K, Ueshima K, Kubo T. Clinical and radiographic results for the Richards Modular Hip System prosthesis in total hip arthroplasty: average 10-year follow-up. J Arthroplasty. 2010;25:369–374.PubMedCrossRef Suehara H, Fujioka M, Inoue S, Takahashi K, Ueshima K, Kubo T. Clinical and radiographic results for the Richards Modular Hip System prosthesis in total hip arthroplasty: average 10-year follow-up. J Arthroplasty. 2010;25:369–374.PubMedCrossRef
74.
go back to reference Sugano N, Takao M, Sakai T, Nishii T, Miki H, Ohzono K. Eleven- to 14-year follow-up results of cementless total hip arthroplasty using a third-generation alumina ceramic-on-ceramic bearing. J Arthroplasty. 2012;27:736–741.PubMedCrossRef Sugano N, Takao M, Sakai T, Nishii T, Miki H, Ohzono K. Eleven- to 14-year follow-up results of cementless total hip arthroplasty using a third-generation alumina ceramic-on-ceramic bearing. J Arthroplasty. 2012;27:736–741.PubMedCrossRef
75.
go back to reference Suzuki K, Kawachi S, Matsubara M, Morita S, Jinno T, Shinomiya K. Cementless total hip replacement after previous intertrochanteric valgus osteotomy for advanced osteoarthritis. J Bone Joint Surg Br. 2007;89:1155–1157.PubMedCrossRef Suzuki K, Kawachi S, Matsubara M, Morita S, Jinno T, Shinomiya K. Cementless total hip replacement after previous intertrochanteric valgus osteotomy for advanced osteoarthritis. J Bone Joint Surg Br. 2007;89:1155–1157.PubMedCrossRef
77.
go back to reference Synder M, Drobniewski M, Sibiński M. Long-term results of cementless hip arthroplasty with ceramic-on-ceramic articulation. Int Orthop. 2012;36:2225–2229.PubMedCentralPubMedCrossRef Synder M, Drobniewski M, Sibiński M. Long-term results of cementless hip arthroplasty with ceramic-on-ceramic articulation. Int Orthop. 2012;36:2225–2229.PubMedCentralPubMedCrossRef
78.
go back to reference Takao M, Ohzono K, Nishii T, Miki H, Nakamura N, Sugano N. Cementless modular total hip arthroplasty with subtrochanteric shortening osteotomy for hips with developmental dysplasia. J Bone Joint Surg Am. 2011;93:548–555.PubMedCrossRef Takao M, Ohzono K, Nishii T, Miki H, Nakamura N, Sugano N. Cementless modular total hip arthroplasty with subtrochanteric shortening osteotomy for hips with developmental dysplasia. J Bone Joint Surg Am. 2011;93:548–555.PubMedCrossRef
79.
go back to reference Traina F, De Fine M, Abati CN, Bordini B, Toni A. Outcomes of total hip replacement in patients with slipped capital femoral epiphysis. Arch Orthop Trauma Surg. 2012;132:1133–1139.PubMedCrossRef Traina F, De Fine M, Abati CN, Bordini B, Toni A. Outcomes of total hip replacement in patients with slipped capital femoral epiphysis. Arch Orthop Trauma Surg. 2012;132:1133–1139.PubMedCrossRef
80.
go back to reference Traina F, De Fine M, Biondi F, Tassinari E, Galvani A, Toni A. The influence of the centre of rotation on implant survival using a modular stem hip prosthesis. Int Orthop. 2009;33:1513–1518.PubMedCentralPubMedCrossRef Traina F, De Fine M, Biondi F, Tassinari E, Galvani A, Toni A. The influence of the centre of rotation on implant survival using a modular stem hip prosthesis. Int Orthop. 2009;33:1513–1518.PubMedCentralPubMedCrossRef
81.
go back to reference Traina F, De Fine M, Tassinari E, Sudanese A, Calderoni PP, Toni A. Modular neck prostheses in DDH patients: 11-year results. J Orthop Sci. 2011;16:14–20.PubMedCrossRef Traina F, De Fine M, Tassinari E, Sudanese A, Calderoni PP, Toni A. Modular neck prostheses in DDH patients: 11-year results. J Orthop Sci. 2011;16:14–20.PubMedCrossRef
82.
go back to reference Vassan UT, Sharma S, Chowdary KP, Bhamra MS. Uncemented metal-on-metal acetabular component: follow-up of 112 hips for a minimum of 5 years. Acta Orthop. 2007;78:470–478.PubMedCrossRef Vassan UT, Sharma S, Chowdary KP, Bhamra MS. Uncemented metal-on-metal acetabular component: follow-up of 112 hips for a minimum of 5 years. Acta Orthop. 2007;78:470–478.PubMedCrossRef
83.
go back to reference Weiss RJ, Hailer NP, Stark A, Kärrholm J. Survival of uncemented acetabular monoblock cups: evaluation of 210 hips in the Swedish Hip Arthroplasty Register. Acta Orthop. 2012;83:214–219.PubMedCentralPubMedCrossRef Weiss RJ, Hailer NP, Stark A, Kärrholm J. Survival of uncemented acetabular monoblock cups: evaluation of 210 hips in the Swedish Hip Arthroplasty Register. Acta Orthop. 2012;83:214–219.PubMedCentralPubMedCrossRef
84.
go back to reference Yeung E, Bott PT, Chana R, Jackson MP, Holloway I, Walter WL, Zicat BA, Walter WK. Mid-term results of third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty: a ten-year minimum follow-up. J Bone Joint Surg Am. 2012;94:138–144.PubMedCrossRef Yeung E, Bott PT, Chana R, Jackson MP, Holloway I, Walter WL, Zicat BA, Walter WK. Mid-term results of third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty: a ten-year minimum follow-up. J Bone Joint Surg Am. 2012;94:138–144.PubMedCrossRef
85.
go back to reference Yoon JP, Le Duff MJ, Takamura KM, Hodge S, Amstutz HC. Mid-to-long term follow-up of Transcend metal-on-metal versus Interseal metal-on-polyethylene bearings in total hip arthroplasty. Hip Int. 2011;21:571–576.PubMedCrossRef Yoon JP, Le Duff MJ, Takamura KM, Hodge S, Amstutz HC. Mid-to-long term follow-up of Transcend metal-on-metal versus Interseal metal-on-polyethylene bearings in total hip arthroplasty. Hip Int. 2011;21:571–576.PubMedCrossRef
86.
go back to reference Zijlstra WP, Cheung J, Sietsma MS, van Raay JJ, Deutman R. No superiority of cemented metal-on-metal vs metal-on-polyethylene THA at 5-year follow-up. Orthopedics. 2009;32:479.PubMedCrossRef Zijlstra WP, Cheung J, Sietsma MS, van Raay JJ, Deutman R. No superiority of cemented metal-on-metal vs metal-on-polyethylene THA at 5-year follow-up. Orthopedics. 2009;32:479.PubMedCrossRef
87.
go back to reference Zijlstra WP, van Raay JJ, Bulstra SK, Deutman R. No superiority of cemented metal-on-metal over metal-on-polyethylene THA in a randomized controlled trial at 10-year follow-up. Orthopedics. 2010;33:154. Zijlstra WP, van Raay JJ, Bulstra SK, Deutman R. No superiority of cemented metal-on-metal over metal-on-polyethylene THA in a randomized controlled trial at 10-year follow-up. Orthopedics. 2010;33:154.
Metadata
Title
How Have Alternative Bearings and Modularity Affected Revision Rates in Total Hip Arthroplasty?
Authors
William M. Mihalko, MD, PhD
Markus A. Wimmer, PhD
Carol A. Pacione, BS
Michel P. Laurent, PhD
Robert F. Murphy, MD
Carson Rider, BS
Publication date
01-12-2014
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 12/2014
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-014-3816-2

Other articles of this Issue 12/2014

Clinical Orthopaedics and Related Research® 12/2014 Go to the issue