Abstract

Background

The local and systemic effects of wear debris and corrosion products remain a concern in arthroplasty and reaction to corrosion or wear products from modular junctions has been reported in primary and revision total joint arthroplasties. These effects have not been reported previously for unipolar hemiarthroplasties where there is no prosthetic bearing surface to contribute to the phenomenon. This may have implications for clinical surveillance and implant design.

Case Description

We report the case of a 72-year-old man who had symptomatic pseudotumor formation, confirmed by pathologic examination of the excised pseudotumor, with a large-head modular hip hemiarthroplasty. Metallosis and corrosion of the modular head/neck taper junction were noted at the time of revision surgery.

Literature Review

To our knowledge, this is the first report of pseudotumor formation where the corrosion or wear products arose from the modular junction of the implant with no bearing couple present to contribute wear debris that may influence the formation of the pseudotumor.

Purposes and Clinical Relevance

Adverse tissue reactions to wear debris generated at prosthetic articulating surfaces and corrosion and wear products from nonarticulating prosthetic junctions have been reported. The problem has been reported to be higher in metal-on-metal bearing couples and in large-diameter hip arthroplasties. Mixed-alloy junctions appear to be more susceptible to corrosion. We believe that corrosion should be considered a possible diagnostic entity when investigating persistent symptoms after hemiarthroplasty and may be avoided with the use of monoblock components.
Literature
1.
go back to reference Brooker AF, Bowerman JW, Robinson BG, Riley LH Jr. Ectopic ossification following total hip replacement: incidence and a method of classification. J Bone Joint Surg Am. 1973;55:1629–1632.PubMed Brooker AF, Bowerman JW, Robinson BG, Riley LH Jr. Ectopic ossification following total hip replacement: incidence and a method of classification. J Bone Joint Surg Am. 1973;55:1629–1632.PubMed
2.
go back to reference Carli A, Reuven A, Zukor DJ, Antoniou J. Adverse soft-tissue reactions around non-metal-on-metal total hip arthroplasty: a systematic review of the literature. Bull NYU Hosp Jt Dis. 2011;69(suppl 1):S47–51.PubMed Carli A, Reuven A, Zukor DJ, Antoniou J. Adverse soft-tissue reactions around non-metal-on-metal total hip arthroplasty: a systematic review of the literature. Bull NYU Hosp Jt Dis. 2011;69(suppl 1):S47–51.PubMed
3.
go back to reference Chmell MJ, Rispler D, Poss R. The impact of modularity in total hip arthroplasty. Clin Orthop Relat Res. 1995;319:77–84.PubMed Chmell MJ, Rispler D, Poss R. The impact of modularity in total hip arthroplasty. Clin Orthop Relat Res. 1995;319:77–84.PubMed
4.
go back to reference Cooper HJ, Valle Della CJ, Berger RA, Tetreault M, Paprosky WG, Sporer SM, Jacobs JJ. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am. 2012;94:1655–1661.PubMedCrossRef Cooper HJ, Valle Della CJ, Berger RA, Tetreault M, Paprosky WG, Sporer SM, Jacobs JJ. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am. 2012;94:1655–1661.PubMedCrossRef
5.
go back to reference Garbuz DS, Tanzer M, Greidanus NV, Masri BA, Duncan CP. The John Charnley Award: Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res. 2009;468:318–325.PubMedCrossRef Garbuz DS, Tanzer M, Greidanus NV, Masri BA, Duncan CP. The John Charnley Award: Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res. 2009;468:318–325.PubMedCrossRef
6.
go back to reference Gilbert JL, Buckley CA, Jacobs JJ. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations: the effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res. 1993;27:1533–1544.PubMedCrossRef Gilbert JL, Buckley CA, Jacobs JJ. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations: the effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res. 1993;27:1533–1544.PubMedCrossRef
7.
go back to reference Gill IP, Webb J, Sloan K, Beaver RJ. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Joint Surg Br. 2012;94:895–900.PubMed Gill IP, Webb J, Sloan K, Beaver RJ. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Joint Surg Br. 2012;94:895–900.PubMed
8.
go back to reference Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S. A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res. 2002;401:149–161.PubMedCrossRef Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S. A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res. 2002;401:149–161.PubMedCrossRef
9.
go back to reference Jacobs JJ, Urban RM, Gilbert JL, Skipor AK, Black J, Jasty M, Galante JO. Local and distant products from modularity. Clin Orthop Relat Res. 1995;319:94–105.PubMed Jacobs JJ, Urban RM, Gilbert JL, Skipor AK, Black J, Jasty M, Galante JO. Local and distant products from modularity. Clin Orthop Relat Res. 1995;319:94–105.PubMed
10.
go back to reference Langton DJ, Jameson SS, Joyce TJ, Gandhi JN, Sidaginamale R, Mereddy P, Lord J, Nargol AV. Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br. 2011;93:1011–1016.PubMed Langton DJ, Jameson SS, Joyce TJ, Gandhi JN, Sidaginamale R, Mereddy P, Lord J, Nargol AV. Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br. 2011;93:1011–1016.PubMed
11.
go back to reference Lindgren JU, Brismar BH, Wikstrom AC. Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis. J Bone Joint Surg Br. 2011;93:1427–1430.PubMed Lindgren JU, Brismar BH, Wikstrom AC. Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis. J Bone Joint Surg Br. 2011;93:1427–1430.PubMed
12.
go back to reference Lucas LC, Buchanan RA, Lemons JE. Investigations on the galvanic corrosion of multialloy total hip prostheses. J Biomed Mater Res. 1981;15:731–747.PubMedCrossRef Lucas LC, Buchanan RA, Lemons JE. Investigations on the galvanic corrosion of multialloy total hip prostheses. J Biomed Mater Res. 1981;15:731–747.PubMedCrossRef
13.
go back to reference Pandit H, Vlychou M, Whitwell D, Crook D, Luqmani R, Ostlere S, Murray DW, Athanasou NA. Necrotic granulomatous pseudotumours in bilateral resurfacing hip arthroplasties: evidence for a type IV immune response. Virchows Arch. 2008;453:529–534.PubMedCrossRef Pandit H, Vlychou M, Whitwell D, Crook D, Luqmani R, Ostlere S, Murray DW, Athanasou NA. Necrotic granulomatous pseudotumours in bilateral resurfacing hip arthroplasties: evidence for a type IV immune response. Virchows Arch. 2008;453:529–534.PubMedCrossRef
14.
go back to reference Parry MC, Bhabra G, Sood A, Machado F, Cartwright L, Saunders M, Ingham E, Newson R, Blom AW, Case CP. Thresholds for indirect DNA damage across cellular barriers for orthopaedic biomaterials. Biomaterials. 2010;31:4477–4483.PubMedCrossRef Parry MC, Bhabra G, Sood A, Machado F, Cartwright L, Saunders M, Ingham E, Newson R, Blom AW, Case CP. Thresholds for indirect DNA damage across cellular barriers for orthopaedic biomaterials. Biomaterials. 2010;31:4477–4483.PubMedCrossRef
15.
go back to reference Polyzois I, Nikolopoulos D, Michos I, Patsouris E, Theocharis S. Local and systemic toxicity of nanoscale debris particles in total hip arthroplasty. J Appl Toxicol. 2012;32:255–269.PubMedCrossRef Polyzois I, Nikolopoulos D, Michos I, Patsouris E, Theocharis S. Local and systemic toxicity of nanoscale debris particles in total hip arthroplasty. J Appl Toxicol. 2012;32:255–269.PubMedCrossRef
16.
go back to reference Restrepo C, Post ZD, Kai B, Hozack WJ. The effect of stem design on the prevalence of squeaking following ceramic-on-ceramic bearing total hip arthroplasty. J Bone Joint Surg Am. 2010;92:550–557.PubMedCrossRef Restrepo C, Post ZD, Kai B, Hozack WJ. The effect of stem design on the prevalence of squeaking following ceramic-on-ceramic bearing total hip arthroplasty. J Bone Joint Surg Am. 2010;92:550–557.PubMedCrossRef
17.
go back to reference Smith AJ, Dieppe PA, Vernon K, Porter M, Blom AW; National Joint Registry of England and Wales. Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales. Lancet. 2012;379:1199–1204.PubMedCrossRef Smith AJ, Dieppe PA, Vernon K, Porter M, Blom AW; National Joint Registry of England and Wales. Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales. Lancet. 2012;379:1199–1204.PubMedCrossRef
18.
go back to reference Svensson O, Mathiesen EB, Reinholt FP, Blomgren G. Formation of a fulminant soft-tissue pseudotumor after uncemented hip arthroplasty: a case report. J Bone Joint Surg Am. 1988;70:1238–1242.PubMed Svensson O, Mathiesen EB, Reinholt FP, Blomgren G. Formation of a fulminant soft-tissue pseudotumor after uncemented hip arthroplasty: a case report. J Bone Joint Surg Am. 1988;70:1238–1242.PubMed
19.
go back to reference Walsh AJ, Nikolaou VS, Antoniou J. Inflammatory pseudotumor complicating metal-on-highly cross-linked polyethylene total hip arthroplasty. J Arthroplasty. 2012;27:324.e5–8. Walsh AJ, Nikolaou VS, Antoniou J. Inflammatory pseudotumor complicating metal-on-highly cross-linked polyethylene total hip arthroplasty. J Arthroplasty. 2012;27:324.e5–8.
20.
go back to reference Williams DH, Greidanus NV, Masri BA, Duncan CP, Garbuz DS. Prevalence of pseudotumor in asymptomatic patients after metal-on-metal hip arthroplasty. J Bone Joint Surg Am. 2011;93:2164–2171.PubMedCrossRef Williams DH, Greidanus NV, Masri BA, Duncan CP, Garbuz DS. Prevalence of pseudotumor in asymptomatic patients after metal-on-metal hip arthroplasty. J Bone Joint Surg Am. 2011;93:2164–2171.PubMedCrossRef
Metadata
Title
Adverse Local Tissue Reaction Associated With a Modular Hip Hemiarthroplasty
Authors
Michael R. Whitehouse, PhD, MSc, FRCS (Tr & Orth)
Makoto Endo, MD, PhD
Bassam A. Masri, MD, FRCS(C)
Publication date
01-12-2013
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 12/2013
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-013-3133-1

Other articles of this Issue 12/2013

Clinical Orthopaedics and Related Research® 12/2013 Go to the issue