Published in: Clinical Orthopaedics and Related Research® 1/2014

01-01-2014 | Symposium: 2013 Knee Society Proceedings

Inter- and Intraobserver Reliability of Two-dimensional CT Scan for Total Knee Arthroplasty Component Malrotation

Authors: Beau Konigsberg, MD, Ryan Hess, MD, Curtis Hartman, MD, Lynette Smith, MS, Kevin L. Garvin, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 1/2014

Login to get access

Abstract

Background

Rotational malalignment of total knee arthroplasty (TKA) has been correlated with patellofemoral maltracking, knee instability, and stiffness. CT is the most accurate method to assess rotational alignment of prosthetic components after TKA, but inter- and intraobserver reliability of CT scans for this use has not been well documented.

Questions/purposes

The objective of this study was to determine the inter- and intraobserver reliability and the repeatability of the measurement of TKA component rotation using two-dimensional CT scans.

Methods

Fifty-two CT scans of TKAs being evaluated for revision surgery were measured by three different physicians. An orthopaedic resident and attending measured the same scans twice (more than 2 weeks apart) and a musculoskeletal radiologist measured them once. To assess interobserver reliability, intraclass correlation coefficients (ICCs) with two-way mixed-effects analysis of variance models as well as 95% confidence intervals for each were done. The repeatability coefficient was calculated as well, which is defined as the difference in measurements that include 95% of the values. This indicates the magnitude of variability among measurements in the same scale, which in this study is degrees.

Results

The interobserver ICC measurement for the femoral component was 0.386 (poor), and it was 0.670 (good) for the tibial component. The interobserver ICC for the combined rotation measurements was 0.617 (good). The intraobserver ICC for the femoral component was 0.606 (good), and it was 0.809 (very good) for the tibial component. The intraobserver ICC for combined rotation was 0.751 (good). The intraobserver repeatability coefficient for the femoral component was 0.49°, 10.64° for the tibial component, and 12.29° for combined rotation.

Conclusions

In this study, the inter- and intraobserver reliability, and the repeatability, of TKA component rotation were variable. This raises concern about whether CT scan is diagnostic in the assessment of component malrotation after TKA.

Level of Evidence

Level IV, diagnostic study. See Guidelines for Authors for a complete description of levels of evidence.
Literature
1.
go back to reference Bargren JH. Total knee dislocation due to rotatory malalignment of tibial component: A case report. Clin Orthop Relat Res. 1980;147:271–274.PubMed Bargren JH. Total knee dislocation due to rotatory malalignment of tibial component: A case report. Clin Orthop Relat Res. 1980;147:271–274.PubMed
2.
go back to reference Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L. Component rotation and anterior knee pain after total knee arthroplasty. Clin Orthop Relat Res. 2001;392:46–55.PubMedCrossRef Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L. Component rotation and anterior knee pain after total knee arthroplasty. Clin Orthop Relat Res. 2001;392:46–55.PubMedCrossRef
3.
go back to reference Bedard M, Vince KG, Redfern J, Collen SR. Internal rotation of the tibial component is frequent in stiff total knee arthroplasty. Clin Orthop Relat Res. 2011;8:2346–2355.CrossRef Bedard M, Vince KG, Redfern J, Collen SR. Internal rotation of the tibial component is frequent in stiff total knee arthroplasty. Clin Orthop Relat Res. 2011;8:2346–2355.CrossRef
4.
go back to reference Berger RA, Crossett LS, Jacobs JJ, Rubash HE. Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res. 1998;356:144–153.PubMedCrossRef Berger RA, Crossett LS, Jacobs JJ, Rubash HE. Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res. 1998;356:144–153.PubMedCrossRef
5.
go back to reference Berger RA, Rubash HE. Rotational instability and malrotation after total knee arthroplasty. Orthop Clin North Am. 2001;4:639–47, ix. Berger RA, Rubash HE. Rotational instability and malrotation after total knee arthroplasty. Orthop Clin North Am. 2001;4:639–47, ix.
6.
go back to reference Berger RA, Rubash HE, Seel MJ, Thompson WH, Crossett LS. Determining the rotational alignment of the femoral component in total knee arthroplasty using the epicondylar axis. Clin Orthop Relat Res. 1993;286:40–47.PubMed Berger RA, Rubash HE, Seel MJ, Thompson WH, Crossett LS. Determining the rotational alignment of the femoral component in total knee arthroplasty using the epicondylar axis. Clin Orthop Relat Res. 1993;286:40–47.PubMed
7.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;8476:307–310.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;8476:307–310.CrossRef
8.
go back to reference British Standards Institution. Precision of test methods. In: Guide for the Determination and Reproducibility for a Standard Test Method. London, UK: London BSI; 1979. British Standards Institution. Precision of test methods. In: Guide for the Determination and Reproducibility for a Standard Test Method. London, UK: London BSI; 1979.
9.
go back to reference Christensen CP, Crawford JJ, Olin MD, Vail TP. Revision of the stiff total knee arthroplasty. J Arthroplasty. 2002;4:409–415.CrossRef Christensen CP, Crawford JJ, Olin MD, Vail TP. Revision of the stiff total knee arthroplasty. J Arthroplasty. 2002;4:409–415.CrossRef
10.
go back to reference Eckhoff DG, Metzger RG, Vandewalle MV. Malrotation associated with implant alignment technique in total knee arthroplasty. Clin Orthop Relat Res. 1995;321:28–31.PubMed Eckhoff DG, Metzger RG, Vandewalle MV. Malrotation associated with implant alignment technique in total knee arthroplasty. Clin Orthop Relat Res. 1995;321:28–31.PubMed
11.
go back to reference Eckhoff DG, Piatt BE, Gnadinger CA, Blaschke RC. Assessing rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 1995;318:176–181.PubMed Eckhoff DG, Piatt BE, Gnadinger CA, Blaschke RC. Assessing rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 1995;318:176–181.PubMed
12.
go back to reference Hirschmann MT, Konala P, Amsler F, Iranpour F, Friederich NF, Cobb JP. The position and orientation of total knee replacement components: A comparison of conventional radiographs, transverse 2D-CT slices and 3D-CT reconstruction. J Bone Joint Surg Br. 2011;5:629–633.CrossRef Hirschmann MT, Konala P, Amsler F, Iranpour F, Friederich NF, Cobb JP. The position and orientation of total knee replacement components: A comparison of conventional radiographs, transverse 2D-CT slices and 3D-CT reconstruction. J Bone Joint Surg Br. 2011;5:629–633.CrossRef
13.
go back to reference Incavo SJ, Wild JJ, Coughlin KM, Beynnon BD. Early revision for component malrotation in total knee arthroplasty. Clin Orthop Relat Res. 2007:131–136. Incavo SJ, Wild JJ, Coughlin KM, Beynnon BD. Early revision for component malrotation in total knee arthroplasty. Clin Orthop Relat Res. 2007:131–136.
14.
go back to reference Insall JN, Kelly M. The total condylar prosthesis. Clin Orthop Relat Res. 1986;205:43–48.PubMed Insall JN, Kelly M. The total condylar prosthesis. Clin Orthop Relat Res. 1986;205:43–48.PubMed
15.
go back to reference Jazrawi LM, Birdzell L, Kummer FJ, Di Cesare PE. The accuracy of computed tomography for determining femoral and tibial total knee arthroplasty component rotation. J Arthroplasty. 2000;6:761–766.CrossRef Jazrawi LM, Birdzell L, Kummer FJ, Di Cesare PE. The accuracy of computed tomography for determining femoral and tibial total knee arthroplasty component rotation. J Arthroplasty. 2000;6:761–766.CrossRef
16.
go back to reference Kanekasu K, Kondo M, Kadoya Y. Axial radiography of the distal femur to assess rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 2005;434:193–197.PubMedCrossRef Kanekasu K, Kondo M, Kadoya Y. Axial radiography of the distal femur to assess rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 2005;434:193–197.PubMedCrossRef
17.
go back to reference Nagamine R, White SE, McCarthy DS, Whiteside LA. Effect of rotational malposition of the femoral component on knee stability kinematics after total knee arthroplasty. J Arthroplasty. 1995;3:265–270.CrossRef Nagamine R, White SE, McCarthy DS, Whiteside LA. Effect of rotational malposition of the femoral component on knee stability kinematics after total knee arthroplasty. J Arthroplasty. 1995;3:265–270.CrossRef
18.
go back to reference Ranawat CS. The patellofemoral joint in total condylar knee arthroplasty. pros and cons based on five- to ten-year follow-up observations. Clin Orthop Relat Res. 1986;205:93–99.PubMed Ranawat CS. The patellofemoral joint in total condylar knee arthroplasty. pros and cons based on five- to ten-year follow-up observations. Clin Orthop Relat Res. 1986;205:93–99.PubMed
19.
go back to reference Rand JA, Coventry MB. Ten-year evaluation of geometric total knee arthroplasty. Clin Orthop Relat Res. 1988;232:168–173.PubMed Rand JA, Coventry MB. Ten-year evaluation of geometric total knee arthroplasty. Clin Orthop Relat Res. 1988;232:168–173.PubMed
20.
go back to reference Torga-Spak R, Parikh SN, Stuchin SA. Anterior knee pain due to biplanar rotatory malalignment of the femoral component in total knee arthroplasty. case report. J Knee Surg. 2004;2:113–116. Torga-Spak R, Parikh SN, Stuchin SA. Anterior knee pain due to biplanar rotatory malalignment of the femoral component in total knee arthroplasty. case report. J Knee Surg. 2004;2:113–116.
21.
go back to reference Vince KG, Insall JN, Kelly MA. The total condylar prosthesis. 10- to 12-year results of a cemented knee replacement. J Bone Joint Surg Br. 1989;5:793–797. Vince KG, Insall JN, Kelly MA. The total condylar prosthesis. 10- to 12-year results of a cemented knee replacement. J Bone Joint Surg Br. 1989;5:793–797.
Metadata
Title
Inter- and Intraobserver Reliability of Two-dimensional CT Scan for Total Knee Arthroplasty Component Malrotation
Authors
Beau Konigsberg, MD
Ryan Hess, MD
Curtis Hartman, MD
Lynette Smith, MS
Kevin L. Garvin, MD
Publication date
01-01-2014
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 1/2014
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-013-3111-7

Other articles of this Issue 1/2014

Clinical Orthopaedics and Related Research® 1/2014 Go to the issue

Symposium: 2013 Knee Society Proceedings

Do Patients Return to Work After Total Knee Arthroplasty?