Published in: Clinical Orthopaedics and Related Research® 8/2011

01-08-2011 | Symposium: UHMWPE for Arthroplasty: From Powder to Debris

History and Systematic Review of Wear and Osteolysis Outcomes for First-generation Highly Crosslinked Polyethylene

Authors: Steven M. Kurtz, PhD, Heather A. Gawel, MS, Jasmine D. Patel, PhD

Published in: Clinical Orthopaedics and Related Research® | Issue 8/2011

Login to get access

Abstract

Background

Highly crosslinked polyethylene (HXLPE) was introduced to reduce wear and osteolysis in total joint arthroplasty. While many studies report wear and osteolysis associated with HXLPE, analytical techniques, clinical study design and followup, HXLPE formulation and implant design characteristics, and patient populations differ substantially among investigations, complicating a unified perspective.

Questions/purposes

Literature on first-generation HXLPE was summarized. We systematically reviewed the radiographic wear data and incidence of osteolysis for HXLPE in hip and knee arthroplasty.

Methods

PubMed identified 391 studies; 28 met inclusion criteria for a weighted-averages analysis of two-dimensional femoral head penetration rates. To determine the incidence of osteolysis, we estimated a pooled odds ratio using a random-effects model.

Results

Weighted-averages analyses of femoral head penetration rates in HXLPE liners and conventional UHMWPE liners resulted, respectively, in a mean two-dimensional linear penetration rate of 0.042 mm/year based on 28 studies (n = 1503 hips) and 0.137 mm/year based on 18 studies (n = 695 hips). The pooled odds ratio for the risk of osteolysis in HXLPE versus conventional liners was 0.13 (95% confidence interval, 0.06–0.27) among studies with minimum 5-year followup. We identified two clinical studies of HXLPE in TKA, preventing systematic analysis of outcomes.

Conclusions

HXLPE liner studies consistently report lower femoral head penetration and an 87% lower risk of osteolysis. Reduction in femoral head penetration or osteolysis risk is not established for large-diameter (> 32 mm) metallic femoral heads or ceramic femoral heads of any size. Few studies document the clinical performance of HXLPE in knees.
Appendix
Available only for authorised users
Literature
1.
go back to reference Ainsworth R, Farling G, Bardos D. An improved bearing material for joint replacement prostheses: carbon fiber-reinforced UHMW polyethylene. In: Transactions of the 3rd Annual Meeting of the Society for Biomaterials. Minneapolis, MN: Society for Biomaterials; 1977:119. Ainsworth R, Farling G, Bardos D. An improved bearing material for joint replacement prostheses: carbon fiber-reinforced UHMW polyethylene. In: Transactions of the 3rd Annual Meeting of the Society for Biomaterials. Minneapolis, MN: Society for Biomaterials; 1977:119.
2.
go back to reference American Society for Testing and Materials. ASTM F2565-06. Standard Guide for Extensively Irradiation-Crosslinked Ultra-High Molecular Weight Polyethylene Fabricated Forms for Surgical Implant Applications. West Conshohocken, PA: American Society for Testing and Materials; 2006. American Society for Testing and Materials. ASTM F2565-06. Standard Guide for Extensively Irradiation-Crosslinked Ultra-High Molecular Weight Polyethylene Fabricated Forms for Surgical Implant Applications. West Conshohocken, PA: American Society for Testing and Materials; 2006.
3.
go back to reference Amstutz HC, Campbell P, McKellop H, Schmalzreid TP, Gillespie WJ, Howie D, Jacobs J, Medley J, Merritt K. Metal on metal total hip replacement workshop consensus document. Clin Orthop Relat Res. 1996;329(Suppl):S297–S303.PubMed Amstutz HC, Campbell P, McKellop H, Schmalzreid TP, Gillespie WJ, Howie D, Jacobs J, Medley J, Merritt K. Metal on metal total hip replacement workshop consensus document. Clin Orthop Relat Res. 1996;329(Suppl):S297–S303.PubMed
4.
go back to reference August AC, Aldam CH, Pynsent PB. The McKee-Farrar hip arthroplasty: a long-term study. J Bone Joint Surg Br. 1986;68:520–527.PubMed August AC, Aldam CH, Pynsent PB. The McKee-Farrar hip arthroplasty: a long-term study. J Bone Joint Surg Br. 1986;68:520–527.PubMed
5.
go back to reference Ayers DC, Hays PL, Drew JM, Eskander MS, Osuch D, Bragdon CR. Two-year radiostereometric analysis evaluation of femoral head penetration in a challenging population of young total hip arthroplasty patients. J Arthroplasty. 2009;24:9–14.PubMed Ayers DC, Hays PL, Drew JM, Eskander MS, Osuch D, Bragdon CR. Two-year radiostereometric analysis evaluation of femoral head penetration in a challenging population of young total hip arthroplasty patients. J Arthroplasty. 2009;24:9–14.PubMed
6.
go back to reference Baker DA, Bellare A, Pruitt L. The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic-grade polyethylene. J Biomed Mater Res A. 2003;66:146–154.PubMed Baker DA, Bellare A, Pruitt L. The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic-grade polyethylene. J Biomed Mater Res A. 2003;66:146–154.PubMed
7.
go back to reference Baker DA, Hastings RS, Pruitt L. Study of fatigue resistance of chemical and radiation crosslinked medical grade ultrahigh molecular weight polyethylene. J Biomed Mater Res. 1999;46:573–581.PubMed Baker DA, Hastings RS, Pruitt L. Study of fatigue resistance of chemical and radiation crosslinked medical grade ultrahigh molecular weight polyethylene. J Biomed Mater Res. 1999;46:573–581.PubMed
8.
go back to reference Bartel DL, Bicknell VL, Wright TM. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am. 1986;68:1041–1051.PubMed Bartel DL, Bicknell VL, Wright TM. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am. 1986;68:1041–1051.PubMed
9.
go back to reference Bartel DL, Rawlinson JJ, Burstein AH, Ranawat CS, Flynn WF Jr. Stresses in polyethylene components of contemporary total knee replacements. Clin Orthop Relat Res. 1995;317:76–82.PubMed Bartel DL, Rawlinson JJ, Burstein AH, Ranawat CS, Flynn WF Jr. Stresses in polyethylene components of contemporary total knee replacements. Clin Orthop Relat Res. 1995;317:76–82.PubMed
10.
go back to reference Beksac B, Salas A, Gonzalez Della Valle A, Salvati EA. Wear is reduced in THA performed with highly cross-linked polyethylene. Clin Orthop Relat Res. 2009;467:1765–1772.PubMed Beksac B, Salas A, Gonzalez Della Valle A, Salvati EA. Wear is reduced in THA performed with highly cross-linked polyethylene. Clin Orthop Relat Res. 2009;467:1765–1772.PubMed
11.
go back to reference Bell CJ, Walker PS, Abeysundera MR, Simmons JM, King PM, Blunn GW. Effect of oxidation on delamination of ultrahigh-molecular-weight polyethylene tibial components. J Arthroplasty. 1998;13:280–290.PubMed Bell CJ, Walker PS, Abeysundera MR, Simmons JM, King PM, Blunn GW. Effect of oxidation on delamination of ultrahigh-molecular-weight polyethylene tibial components. J Arthroplasty. 1998;13:280–290.PubMed
12.
go back to reference Bhambri SK, Gsell R, Kirkpatrick L, Swarts D, Blanchard CR, Crowninshield RD. The effect of aging on mechanical properties of melt-annealed highly crosslinked UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Bhambri SK, Gsell R, Kirkpatrick L, Swarts D, Blanchard CR, Crowninshield RD. The effect of aging on mechanical properties of melt-annealed highly crosslinked UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
13.
go back to reference Bitsch RG, Loidolt T, Heisel C, Ball S, Schmalzried TP. Reduction of osteolysis with use of Marathon cross-linked polyethylene: a concise follow-up, at a minimum of five years, of a previous report. J Bone Joint Surg Am. 2008;90:1487–1491.PubMed Bitsch RG, Loidolt T, Heisel C, Ball S, Schmalzried TP. Reduction of osteolysis with use of Marathon cross-linked polyethylene: a concise follow-up, at a minimum of five years, of a previous report. J Bone Joint Surg Am. 2008;90:1487–1491.PubMed
14.
go back to reference Bizot P, Nizard R, Lerouge S, Prudhommeaux F, Sedel L. Ceramic/ceramic total hip arthroplasty. J Orthop Sci. 2000;5:622–627.PubMed Bizot P, Nizard R, Lerouge S, Prudhommeaux F, Sedel L. Ceramic/ceramic total hip arthroplasty. J Orthop Sci. 2000;5:622–627.PubMed
15.
go back to reference Blunn GW, Joshi AB, Minns RJ, Lidgren L, Lilley P, Ryd L, Engelbrecht E, Walker PS. Wear in retrieved condylar knee arthroplasties: a comparison of wear in different designs of 280 retrieved condylar knee prostheses. J Arthroplasty. 1997;12:281–290.PubMed Blunn GW, Joshi AB, Minns RJ, Lidgren L, Lilley P, Ryd L, Engelbrecht E, Walker PS. Wear in retrieved condylar knee arthroplasties: a comparison of wear in different designs of 280 retrieved condylar knee prostheses. J Arthroplasty. 1997;12:281–290.PubMed
16.
go back to reference Bostrom MP, Bennett AP, Rimnac CM, Wright TM. The natural history of ultra high molecular weight polyethylene. Clin Orthop Relat Res. 1994;309:20–28.PubMed Bostrom MP, Bennett AP, Rimnac CM, Wright TM. The natural history of ultra high molecular weight polyethylene. Clin Orthop Relat Res. 1994;309:20–28.PubMed
17.
go back to reference Boutin P, Christel P, Dorlot JM, Meunier A, de Roquancourt A, Blanquaert D, Herman S, Sedel L, Witvoet J. The use of dense alumina-alumina ceramic combination in total hip replacement. J Biomed Mater Res. 1988;22:1203–1232.PubMed Boutin P, Christel P, Dorlot JM, Meunier A, de Roquancourt A, Blanquaert D, Herman S, Sedel L, Witvoet J. The use of dense alumina-alumina ceramic combination in total hip replacement. J Biomed Mater Res. 1988;22:1203–1232.PubMed
18.
go back to reference Bozic KJ, Kurtz S, Lau E, Ong K, Chiu V, Vail TP, Rubash HE, Berry DJ. The epidemiology of bearing surface usage in total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:1614–1620.PubMed Bozic KJ, Kurtz S, Lau E, Ong K, Chiu V, Vail TP, Rubash HE, Berry DJ. The epidemiology of bearing surface usage in total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:1614–1620.PubMed
19.
go back to reference Bragdon CR, Barrett S, Martell JM, Greene ME, Malchau H, Harris WH. Steady-state penetration rates of electron beam-irradiated, highly cross-linked polyethylene at an average 45-month follow-up. J Arthroplasty. 2006;21:935–943.PubMed Bragdon CR, Barrett S, Martell JM, Greene ME, Malchau H, Harris WH. Steady-state penetration rates of electron beam-irradiated, highly cross-linked polyethylene at an average 45-month follow-up. J Arthroplasty. 2006;21:935–943.PubMed
20.
go back to reference Bragdon CR, Greene ME, Freiberg AA, Harris WH, Malchau H. Radiostereometric analysis comparison of wear of highly cross-linked polyethylene against 36- vs 28-mm femoral heads. J Arthroplasty. 2007;22:125–129.PubMed Bragdon CR, Greene ME, Freiberg AA, Harris WH, Malchau H. Radiostereometric analysis comparison of wear of highly cross-linked polyethylene against 36- vs 28-mm femoral heads. J Arthroplasty. 2007;22:125–129.PubMed
21.
go back to reference Bragdon CR, Kwon YM, Geller JA, Greene ME, Freiberg AA, Harris WH, Malchau H. Minimum 6-year followup of highly cross-linked polyethylene in THA. Clin Orthop Relat Res. 2007;465:122–127.PubMed Bragdon CR, Kwon YM, Geller JA, Greene ME, Freiberg AA, Harris WH, Malchau H. Minimum 6-year followup of highly cross-linked polyethylene in THA. Clin Orthop Relat Res. 2007;465:122–127.PubMed
22.
go back to reference Calvert GT, Devane PA, Fielden J, Adams K, Horne JG. A double-blind, prospective, randomized controlled trial comparing highly cross-linked and conventional polyethylene in primary total hip arthroplasty. J Arthroplasty. 2009;24:505–510.PubMed Calvert GT, Devane PA, Fielden J, Adams K, Horne JG. A double-blind, prospective, randomized controlled trial comparing highly cross-linked and conventional polyethylene in primary total hip arthroplasty. J Arthroplasty. 2009;24:505–510.PubMed
23.
go back to reference Champion AR, Li S, Saum K, Howard E, Simmons W. The effect of crystallinity on the physical properties of UHMWPE. Trans Orthop Res Soc. 1994;19:585. Champion AR, Li S, Saum K, Howard E, Simmons W. The effect of crystallinity on the physical properties of UHMWPE. Trans Orthop Res Soc. 1994;19:585.
24.
go back to reference Chan FW, Bobyn JD, Medley JB, Krygier JJ, Yue S, Tanzer M. Engineering issues and wear performance of metal on metal hip implants. Clin Orthop Relat Res. 1996;333:96–107.PubMed Chan FW, Bobyn JD, Medley JB, Krygier JJ, Yue S, Tanzer M. Engineering issues and wear performance of metal on metal hip implants. Clin Orthop Relat Res. 1996;333:96–107.PubMed
25.
go back to reference Charnley J. Arthroplasty of the hip: a new operation. Lancet 1961;I:1129–1132. Charnley J. Arthroplasty of the hip: a new operation. Lancet 1961;I:1129–1132.
26.
go back to reference Charnley J. Tissue reaction to the polytetraflouroethyene. Lancet. 1963;II:1379. Charnley J. Tissue reaction to the polytetraflouroethyene. Lancet. 1963;II:1379.
27.
go back to reference Charnley J. Low friction principle. In: Low Friction Arthroplasty of the Hip: Theory and Practice. Berlin, Germany: Springer-Verlag; 1979:3–16. Charnley J. Low friction principle. In: Low Friction Arthroplasty of the Hip: Theory and Practice. Berlin, Germany: Springer-Verlag; 1979:3–16.
28.
go back to reference Charnley J. Personal communication, 1971: Sterilization of Charnley total hip sockets of high density polyethylene. In: Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009:39–40. Charnley J. Personal communication, 1971: Sterilization of Charnley total hip sockets of high density polyethylene. In: Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009:39–40.
29.
go back to reference Charnley J, Halley DK. Rate of wear in total hip replacement. Clin Orthop Relat Res. 1975;112:170–179.PubMed Charnley J, Halley DK. Rate of wear in total hip replacement. Clin Orthop Relat Res. 1975;112:170–179.PubMed
30.
go back to reference Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty. 1996;11:377–389.PubMed Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty. 1996;11:377–389.PubMed
31.
go back to reference Collier JP, Sutula LC, Currier BH, Currier JH, Wooding RE, Williams IR, Farber KB, Mayor MB. Overview of polyethylene as a bearing material: comparison of sterilization methods. Clin Orthop Relat Res. 1996;333:76–86.PubMed Collier JP, Sutula LC, Currier BH, Currier JH, Wooding RE, Williams IR, Farber KB, Mayor MB. Overview of polyethylene as a bearing material: comparison of sterilization methods. Clin Orthop Relat Res. 1996;333:76–86.PubMed
32.
go back to reference Collier MB, Engh CA Jr, Engh GA. Shelf age of the polyethylene tibial component and outcome of unicondylar knee arthroplasty. J Bone Joint Surg Am. 2004;86:763–769.PubMed Collier MB, Engh CA Jr, Engh GA. Shelf age of the polyethylene tibial component and outcome of unicondylar knee arthroplasty. J Bone Joint Surg Am. 2004;86:763–769.PubMed
33.
go back to reference Connelly GM, Rimnac CM, Wright TM, Hertzberg RW, Manson JA. Fatigue crack propagation behavior of ultrahigh molecular weight polyethylene. J Orthop Res. 1984;2:119–125.PubMed Connelly GM, Rimnac CM, Wright TM, Hertzberg RW, Manson JA. Fatigue crack propagation behavior of ultrahigh molecular weight polyethylene. J Orthop Res. 1984;2:119–125.PubMed
34.
go back to reference Costa L, Brach de Prever EM. UHMWPE for Arthroplasty. Turin, Italy: Edizioni Minerva Medica; 2000. Costa L, Brach de Prever EM. UHMWPE for Arthroplasty. Turin, Italy: Edizioni Minerva Medica; 2000.
35.
go back to reference Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg Am. 2007;89:2023–2029.PubMed Currier BH, Currier JH, Mayor MB, Lyford KA, Collier JP, Van Citters DW. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups. J Bone Joint Surg Am. 2007;89:2023–2029.PubMed
36.
go back to reference Currier BH, Currier JH, Mayor MB, Lyford KA, Van Citters DW, Collier JP. In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings. J Arthroplasty. 2007;22:721–731.PubMed Currier BH, Currier JH, Mayor MB, Lyford KA, Van Citters DW, Collier JP. In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings. J Arthroplasty. 2007;22:721–731.PubMed
37.
go back to reference D’Antonio JA, Manley MT, Capello WN, Bierbaum BE, Ramakrishnan R, Naughton M, Sutton K. Five-year experience with Crossfire highly cross-linked polyethylene. Clin Orthop Relat Res. 2005;441:143–150.PubMed D’Antonio JA, Manley MT, Capello WN, Bierbaum BE, Ramakrishnan R, Naughton M, Sutton K. Five-year experience with Crossfire highly cross-linked polyethylene. Clin Orthop Relat Res. 2005;441:143–150.PubMed
38.
go back to reference Digas G, Karrholm J, Thanner J, Herberts P. 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis. Acta Orthop. 2007;78:746–754.PubMed Digas G, Karrholm J, Thanner J, Herberts P. 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis. Acta Orthop. 2007;78:746–754.PubMed
39.
go back to reference Digas G, Karrholm J, Thanner J, Malchau H, Herberts P. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis. Clin Orthop Relat Res. 2004;429:6–16.PubMed Digas G, Karrholm J, Thanner J, Malchau H, Herberts P. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis. Clin Orthop Relat Res. 2004;429:6–16.PubMed
40.
go back to reference Dorr LD, Hilton KR, Wan Z, Markovich GD, Bloebaum R. Modern metal on metal articulation for total hip replacements. Clin Orthop Relat Res. 1996;333:108–117.PubMed Dorr LD, Hilton KR, Wan Z, Markovich GD, Bloebaum R. Modern metal on metal articulation for total hip replacements. Clin Orthop Relat Res. 1996;333:108–117.PubMed
41.
go back to reference Dorr LD, Wan Z, Shahrdar C, Sirianni L, Boutary M, Yun A. Clinical performance of a Durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years. J Bone Joint Surg Am. 2005;87:1816–1821.PubMed Dorr LD, Wan Z, Shahrdar C, Sirianni L, Boutary M, Yun A. Clinical performance of a Durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years. J Bone Joint Surg Am. 2005;87:1816–1821.PubMed
42.
go back to reference Duffy GP, Wannomae KK, Rowell SL, Muratoglu OK. Fracture of a cross-linked polyethylene liner due to impingement. J Arthroplasty. 2009;24:158.e15–19. Duffy GP, Wannomae KK, Rowell SL, Muratoglu OK. Fracture of a cross-linked polyethylene liner due to impingement. J Arthroplasty. 2009;24:158.e15–19.
43.
go back to reference Dumbleton JH, D’Antonio JA, Manley MT, Capello WN, Wang A. The basis for a second-generation highly cross-linked UHMWPE. Clin Orthop Relat Res. 2006;453:265–271.PubMed Dumbleton JH, D’Antonio JA, Manley MT, Capello WN, Wang A. The basis for a second-generation highly cross-linked UHMWPE. Clin Orthop Relat Res. 2006;453:265–271.PubMed
44.
go back to reference Engh CA Jr, Stepniewski AS, Ginn SD, Beykirch SE, Sychterz-Terefenko CJ, Hopper RH Jr, Engh CA. A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners. J Arthroplasty. 2006;21:17–25.PubMed Engh CA Jr, Stepniewski AS, Ginn SD, Beykirch SE, Sychterz-Terefenko CJ, Hopper RH Jr, Engh CA. A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners. J Arthroplasty. 2006;21:17–25.PubMed
45.
go back to reference Fukui K, Kaneuji A, Sugimori T, Ichiseki T, Kitamura K, Matsumoto T. Wear comparison between a highly cross-linked polyethylene and conventional polyethylene against a zirconia femoral head: minimum 5-year follow-up. J Arthroplasty. 2011;26:45–49.PubMed Fukui K, Kaneuji A, Sugimori T, Ichiseki T, Kitamura K, Matsumoto T. Wear comparison between a highly cross-linked polyethylene and conventional polyethylene against a zirconia femoral head: minimum 5-year follow-up. J Arthroplasty. 2011;26:45–49.PubMed
46.
go back to reference Garcia-Rey E, Garcia-Cimbrelo E, Cruz-Pardos A, Ortega-Chamarro J. New polyethylenes in total hip replacement: a prospective, comparative clinical study of two types of liner. J Bone Joint Surg Br. 2008;90:149–153.PubMed Garcia-Rey E, Garcia-Cimbrelo E, Cruz-Pardos A, Ortega-Chamarro J. New polyethylenes in total hip replacement: a prospective, comparative clinical study of two types of liner. J Bone Joint Surg Br. 2008;90:149–153.PubMed
47.
go back to reference Garino JP. Modern ceramic-on-ceramic total hip systems in the United States: early results. Clin Orthop Relat Res. 2000;379:41–47.PubMed Garino JP. Modern ceramic-on-ceramic total hip systems in the United States: early results. Clin Orthop Relat Res. 2000;379:41–47.PubMed
48.
go back to reference Garvin KL, Hartman CW, Mangla J, Murdoch N, Martell JM. Wear analysis in THA utilizing oxidized zirconium and crosslinked polyethylene. Clin Orthop Relat Res. 2009;467:141–145.PubMed Garvin KL, Hartman CW, Mangla J, Murdoch N, Martell JM. Wear analysis in THA utilizing oxidized zirconium and crosslinked polyethylene. Clin Orthop Relat Res. 2009;467:141–145.PubMed
49.
go back to reference Geller JA, Malchau H, Bragdon C, Greene M, Harris WH, Freiberg AA. Large diameter femoral heads on highly cross-linked polyethylene: minimum 3-year results. Clin Orthop Relat Res. 2006;447:53–59.PubMed Geller JA, Malchau H, Bragdon C, Greene M, Harris WH, Freiberg AA. Large diameter femoral heads on highly cross-linked polyethylene: minimum 3-year results. Clin Orthop Relat Res. 2006;447:53–59.PubMed
50.
go back to reference Gioe TJ, Sharma A, Tatman P, Mehle S. Do “premium” joint implants add value? Analysis of high cost joint implants in a community registry. Clin Orthop Relat Res. 2011;469:48–54.PubMed Gioe TJ, Sharma A, Tatman P, Mehle S. Do “premium” joint implants add value? Analysis of high cost joint implants in a community registry. Clin Orthop Relat Res. 2011;469:48–54.PubMed
51.
go back to reference Glyn-Jones S, Isaac S, Hauptfleisch J, McLardy-Smith P, Murray DW, Gill HS. Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty? A double-blind, randomized, and controlled trial using roentgen stereophotogrammetric analysis. J Arthroplasty. 2008;23:337–343.PubMed Glyn-Jones S, Isaac S, Hauptfleisch J, McLardy-Smith P, Murray DW, Gill HS. Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty? A double-blind, randomized, and controlled trial using roentgen stereophotogrammetric analysis. J Arthroplasty. 2008;23:337–343.PubMed
52.
go back to reference Glyn-Jones S, McLardy-Smith P, Gill HS, Murray DW. The creep and wear of highly cross-linked polyethylene: a three-year randomised, controlled trial using radiostereometric analysis. J Bone Joint Surg Br. 2008;90:556–561.PubMed Glyn-Jones S, McLardy-Smith P, Gill HS, Murray DW. The creep and wear of highly cross-linked polyethylene: a three-year randomised, controlled trial using radiostereometric analysis. J Bone Joint Surg Br. 2008;90:556–561.PubMed
53.
go back to reference Good V, Widding K, Scott M, Jani S. The sensitivity of crosslinked UHMWPE to abrasive wear: hips vs. knees. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Good V, Widding K, Scott M, Jani S. The sensitivity of crosslinked UHMWPE to abrasive wear: hips vs. knees. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
54.
go back to reference Greer K, King R, Chan FW. Effects of raw material, irradiation dose, and irradiation source on crosslinking of UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Greer K, King R, Chan FW. Effects of raw material, irradiation dose, and irradiation source on crosslinking of UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
55.
go back to reference Griffith MJ, Seidenstein MK, Williams D, Charnley J. Socket wear in Charnley low friction arthroplasty of the hip. Clin Orthop Relat Res. 1978;137:37–47.PubMed Griffith MJ, Seidenstein MK, Williams D, Charnley J. Socket wear in Charnley low friction arthroplasty of the hip. Clin Orthop Relat Res. 1978;137:37–47.PubMed
56.
go back to reference Grobbelaar CJ, Du Plessis TA, Marais F. The radiation improvment of polyethylene prostheses: a preliminary study. J Bone Joint Surg Br. 1978;60:370–374.PubMed Grobbelaar CJ, Du Plessis TA, Marais F. The radiation improvment of polyethylene prostheses: a preliminary study. J Bone Joint Surg Br. 1978;60:370–374.PubMed
57.
go back to reference Halley D, Glassman A, Crowninshield RD. Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head: a case report. J Bone Joint Surg Am. 2004;86:827–830.PubMed Halley D, Glassman A, Crowninshield RD. Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head: a case report. J Bone Joint Surg Am. 2004;86:827–830.PubMed
58.
go back to reference Hammerberg EM, Zhinian W, Manish D, Lawrence DD. Wear and range of motion of different femoral head sizes. J Arthroplasty. 2010;25:839–843.PubMed Hammerberg EM, Zhinian W, Manish D, Lawrence DD. Wear and range of motion of different femoral head sizes. J Arthroplasty. 2010;25:839–843.PubMed
59.
go back to reference Harris WH. The problem is osteolysis. Clin Orthop Relat Res. 1995;311:46–53.PubMed Harris WH. The problem is osteolysis. Clin Orthop Relat Res. 1995;311:46–53.PubMed
60.
go back to reference Heisel C, Silva M, dela Rosa MA, Schmalzried TP. Short-term in vivo wear of cross-linked polyethylene. J Bone Joint Surg Am. 2004;86:748–751.PubMed Heisel C, Silva M, dela Rosa MA, Schmalzried TP. Short-term in vivo wear of cross-linked polyethylene. J Bone Joint Surg Am. 2004;86:748–751.PubMed
61.
go back to reference Heisel C, Silva M, Schmalzried TP. In vivo wear of bilateral total hip replacements: conventional versus crosslinked polyethylene. Arch Orthop Trauma Surg. 2005;125:555–557.PubMed Heisel C, Silva M, Schmalzried TP. In vivo wear of bilateral total hip replacements: conventional versus crosslinked polyethylene. Arch Orthop Trauma Surg. 2005;125:555–557.PubMed
62.
go back to reference Hodrick JT, Severson EP, McAlister DS, Dahl B, Hofmann AA. Highly crosslinked polyethylene is safe for use in total knee arthroplasty. Clin Orthop Relat Res. 2008;466:2806–2812.PubMed Hodrick JT, Severson EP, McAlister DS, Dahl B, Hofmann AA. Highly crosslinked polyethylene is safe for use in total knee arthroplasty. Clin Orthop Relat Res. 2008;466:2806–2812.PubMed
63.
go back to reference Hopper RH Jr, Young AM, Orishimo KF, Engh CA Jr. Effect of terminal sterilization with gas plasma or gamma radiation on wear of polyethylene liners. J Bone Joint Surg Am. 2003;85:464–468.PubMed Hopper RH Jr, Young AM, Orishimo KF, Engh CA Jr. Effect of terminal sterilization with gas plasma or gamma radiation on wear of polyethylene liners. J Bone Joint Surg Am. 2003;85:464–468.PubMed
64.
go back to reference Hopper RH Jr, Young AM, Orishimo KF, McAuley JP. Correlation between early and late wear rates in total hip arthroplasty with application to the performance of marathon cross-linked polyethylene liners. J Arthroplasty. 2003;18:60–67.PubMed Hopper RH Jr, Young AM, Orishimo KF, McAuley JP. Correlation between early and late wear rates in total hip arthroplasty with application to the performance of marathon cross-linked polyethylene liners. J Arthroplasty. 2003;18:60–67.PubMed
65.
go back to reference Ianuzzi A, Kurtz SM. Sterilization of joint replacement materials. In: Revell PA, ed. Joint Replacement Technology. Cambridge, UK: Woodhead Publishing Ltd; 2008:407–430. Ianuzzi A, Kurtz SM. Sterilization of joint replacement materials. In: Revell PA, ed. Joint Replacement Technology. Cambridge, UK: Woodhead Publishing Ltd; 2008:407–430.
66.
go back to reference Ise K, Kawanabe K, Tamura J, Akiyama H, Goto K, Nakamura T. Clinical results of the wear performance of cross-linked polyethylene in total hip arthroplasty: prospective randomized trial. J Arthroplasty. 2009;24:1216–1220.PubMed Ise K, Kawanabe K, Tamura J, Akiyama H, Goto K, Nakamura T. Clinical results of the wear performance of cross-linked polyethylene in total hip arthroplasty: prospective randomized trial. J Arthroplasty. 2009;24:1216–1220.PubMed
67.
go back to reference Jacobs JJ, Shanbhag A, Glant TT, Black J, Galante JO. Wear debris in total joint replacements. J Am Acad Orthop Surg. 1994;2:212–220.PubMed Jacobs JJ, Shanbhag A, Glant TT, Black J, Galante JO. Wear debris in total joint replacements. J Am Acad Orthop Surg. 1994;2:212–220.PubMed
68.
go back to reference Jani S, Scott M. Wear debris generation in joint simulator testing of crosslinked UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Jani S, Scott M. Wear debris generation in joint simulator testing of crosslinked UHMWPE. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
69.
go back to reference Kawate K, Ohmura T, Kawahara I, Tamai K, Ueha T, Takemura K. Differences in highly cross-linked polyethylene wear between zirconia and cobalt-chromium femoral heads in Japanese patients: a prospective, randomized study. J Arthroplasty. 2009;24:1221–1224.PubMed Kawate K, Ohmura T, Kawahara I, Tamai K, Ueha T, Takemura K. Differences in highly cross-linked polyethylene wear between zirconia and cobalt-chromium femoral heads in Japanese patients: a prospective, randomized study. J Arthroplasty. 2009;24:1221–1224.PubMed
70.
go back to reference Kiely PD, Harty JA, McElwain JP. Hylamer wear rates and shelf life: a clinical correlation. Acta Orthop Belg. 2005;71:429–434.PubMed Kiely PD, Harty JA, McElwain JP. Hylamer wear rates and shelf life: a clinical correlation. Acta Orthop Belg. 2005;71:429–434.PubMed
71.
go back to reference Kim YH, Kim JS, Choi YW, Kwon OR. Intermediate results of simultaneous alumina-on-alumina bearing and alumina-on-highly cross-linked polyethylene bearing total hip arthroplasties. J Arthroplasty. 2009;24:885–891.PubMed Kim YH, Kim JS, Choi YW, Kwon OR. Intermediate results of simultaneous alumina-on-alumina bearing and alumina-on-highly cross-linked polyethylene bearing total hip arthroplasties. J Arthroplasty. 2009;24:885–891.PubMed
72.
go back to reference Kothari M, Bartel DL, Booker JF. Surface geometry of retrieved McKee-Farrar total hip replacements. Clin Orthop Relat Res. 1996;329(Suppl):S141–S147.PubMed Kothari M, Bartel DL, Booker JF. Surface geometry of retrieved McKee-Farrar total hip replacements. Clin Orthop Relat Res. 1996;329(Suppl):S141–S147.PubMed
73.
go back to reference Krushell RJ, Fingeroth RH, Cushing MC. Early femoral head penetration of a highly crosslinked polyethylene liner vs a conventional polyethlene liner. J Arthroplasty. 2005;20:73–76.PubMed Krushell RJ, Fingeroth RH, Cushing MC. Early femoral head penetration of a highly crosslinked polyethylene liner vs a conventional polyethlene liner. J Arthroplasty. 2005;20:73–76.PubMed
74.
go back to reference Kurtz SM. The UHMWPE Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement. New York, NY: Academic Press; 2004. Kurtz SM. The UHMWPE Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement. New York, NY: Academic Press; 2004.
75.
go back to reference Kurtz SM. Packaging and sterilization of UHMWPE. In: Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009:21–30. Kurtz SM. Packaging and sterilization of UHMWPE. In: Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009:21–30.
76.
go back to reference Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009. Kurtz SM, ed. The UHMWPE Biomaterials Handbook: Ultra-High Molecular Weight Polyethylene in Total Joint Replacement and Medical Devices. 2nd ed. Burlington, MA: Academic Press; 2009.
77.
go back to reference Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
78.
go back to reference Kurtz SM, Hozack WJ, Purtill JJ, Marcolongo M, Kraay MJ, Goldberg VM, Sharkey PF, Parvizi J, Rimnac CM, Edidin AA. Significance of in vivo degradation for polyethylene in total hip arthroplasty. Clin Orthop Relat Res. 2006;453:47–57.PubMed Kurtz SM, Hozack WJ, Purtill JJ, Marcolongo M, Kraay MJ, Goldberg VM, Sharkey PF, Parvizi J, Rimnac CM, Edidin AA. Significance of in vivo degradation for polyethylene in total hip arthroplasty. Clin Orthop Relat Res. 2006;453:47–57.PubMed
79.
go back to reference Kurtz SM, Manley M, Wang A, Taylor S, Dumbleton J. Comparison of the properties of annealed crosslinked (Crossfire) and conventional polyethylene as hip bearing materials. Bull Hosp Jt Dis. 2002-2003;61:17–26. Kurtz SM, Manley M, Wang A, Taylor S, Dumbleton J. Comparison of the properties of annealed crosslinked (Crossfire) and conventional polyethylene as hip bearing materials. Bull Hosp Jt Dis. 2002-2003;61:17–26.
80.
go back to reference Kurtz SM, Mazzucco D, Rimnac CM, Schroeder D. Anisotropy and oxidative resistance of highly crosslinked UHMWPE after deformation processing by solid-state ram extrusion. Biomaterials. 2006;27:24–34.PubMed Kurtz SM, Mazzucco D, Rimnac CM, Schroeder D. Anisotropy and oxidative resistance of highly crosslinked UHMWPE after deformation processing by solid-state ram extrusion. Biomaterials. 2006;27:24–34.PubMed
81.
go back to reference Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and crosslinking of ultra- high molecular weight polyethylene for total joint arthroplasty. Biomaterials. 1999;20:1659–1688.PubMed Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and crosslinking of ultra- high molecular weight polyethylene for total joint arthroplasty. Biomaterials. 1999;20:1659–1688.PubMed
82.
go back to reference Kurtz SM, Rimnac CM, Hozack WJ, Turner J, Marcolongo M, Goldberg VM, Kraay MJ, Edidin AA. In vivo degradation of polyethylene liners after gamma sterilization in air. J Bone Joint Surg Am. 2005;87:815–823.PubMed Kurtz SM, Rimnac CM, Hozack WJ, Turner J, Marcolongo M, Goldberg VM, Kraay MJ, Edidin AA. In vivo degradation of polyethylene liners after gamma sterilization in air. J Bone Joint Surg Am. 2005;87:815–823.PubMed
83.
go back to reference Lachiewicz PF, Heckman DS, Soileau ES, Mangla J, Martell JM. Femoral head size and wear of highly cross-linked polyethylene at 5 to 8 years. Clin Orthop Relat Res. 2009;467:3290–3296.PubMed Lachiewicz PF, Heckman DS, Soileau ES, Mangla J, Martell JM. Femoral head size and wear of highly cross-linked polyethylene at 5 to 8 years. Clin Orthop Relat Res. 2009;467:3290–3296.PubMed
84.
go back to reference Laurent MP, Blanchard CR, Yao JQ, Johnson TS, Gilbertson LN, Swarts D, Crowninshield RD. The wear of highly crosslinked UHMWPE in the presence of abrasive particles: hip and knee simulator studies. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Laurent MP, Blanchard CR, Yao JQ, Johnson TS, Gilbertson LN, Swarts D, Crowninshield RD. The wear of highly crosslinked UHMWPE in the presence of abrasive particles: hip and knee simulator studies. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
85.
go back to reference Leung SB, Egawa H, Stepniewski A, Beykirch S, Engh CA Jr, Engh CA Sr. Incidence and volume of pelvic osteolysis at early follow-up with highly cross-linked and noncross-linked polyethylene. J Arthroplasty. 2007;22:134–139.PubMed Leung SB, Egawa H, Stepniewski A, Beykirch S, Engh CA Jr, Engh CA Sr. Incidence and volume of pelvic osteolysis at early follow-up with highly cross-linked and noncross-linked polyethylene. J Arthroplasty. 2007;22:134–139.PubMed
86.
go back to reference Li S, Burstein AH. Ultra-high molecular weight polyethylene: the material and its use in total joint implants. J Bone Joint Surg Am. 1994;76:1080–1090.PubMed Li S, Burstein AH. Ultra-high molecular weight polyethylene: the material and its use in total joint implants. J Bone Joint Surg Am. 1994;76:1080–1090.PubMed
87.
go back to reference Li S, Chang JD, Barrena EG, Furman BD, Wright TM, Salvati E. Nonconsolidated polyethylene particles and oxidation in Charnley acetabular cups. Clin Orthop Relat Res. 1995;319:54–63.PubMed Li S, Chang JD, Barrena EG, Furman BD, Wright TM, Salvati E. Nonconsolidated polyethylene particles and oxidation in Charnley acetabular cups. Clin Orthop Relat Res. 1995;319:54–63.PubMed
88.
go back to reference MacDonald D, Kurtz SM, Sakona A, Rimnac C. In vivo oxidation, oxidation potential, and mechanical properties of first generation highly crosslinked polyethylenes. Clin Orthop Relat Res. 2010 December 16 [Epub ahead of print]. MacDonald D, Kurtz SM, Sakona A, Rimnac C. In vivo oxidation, oxidation potential, and mechanical properties of first generation highly crosslinked polyethylenes. Clin Orthop Relat Res. 2010 December 16 [Epub ahead of print].
89.
go back to reference Mall NA, Nunley RM, Zhu JJ, Maloney WJ, Barrack RL, Clohisy JC. The incidence of acetabular osteolysis in young patients with conventional versus highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:372–381.PubMed Mall NA, Nunley RM, Zhu JJ, Maloney WJ, Barrack RL, Clohisy JC. The incidence of acetabular osteolysis in young patients with conventional versus highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:372–381.PubMed
90.
go back to reference Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH. In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J Arthroplasty. 2005;20:880–886.PubMed Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH. In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J Arthroplasty. 2005;20:880–886.PubMed
91.
go back to reference Martell JM, Verner JJ, Incavo SJ. Clinical performance of a highly cross-linked polyethylene at two years in total hip arthroplasty: a randomized prospective trial. J Arthroplasty. 2003;18:55–59.PubMed Martell JM, Verner JJ, Incavo SJ. Clinical performance of a highly cross-linked polyethylene at two years in total hip arthroplasty: a randomized prospective trial. J Arthroplasty. 2003;18:55–59.PubMed
92.
go back to reference McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.PubMed McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.PubMed
93.
go back to reference McGovern TF, Ammeen DJ, Collier JP, Currier BH, Engh GA. Rapid polyethylene failure of unicondylar tibial components sterilized with gamma irradiation in air and implanted after a long shelf life. J Bone Joint Surg Am. 2002;84:901–906.PubMed McGovern TF, Ammeen DJ, Collier JP, Currier BH, Engh GA. Rapid polyethylene failure of unicondylar tibial components sterilized with gamma irradiation in air and implanted after a long shelf life. J Bone Joint Surg Am. 2002;84:901–906.PubMed
94.
go back to reference McKee GK, Watson-Farrar J. Replacement of arthritic hips by the McKee-Farrar prosthesis. J Bone Joint Surg Br. 1966;48:245–259.PubMed McKee GK, Watson-Farrar J. Replacement of arthritic hips by the McKee-Farrar prosthesis. J Bone Joint Surg Br. 1966;48:245–259.PubMed
95.
go back to reference McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–167.PubMed McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–167.PubMed
96.
go back to reference McKellop HA, Shen FW, Campbell P, Ota T. Effect of molecular weight, calcium stearate, and sterilization methods on the wear of ultra-high molecular weight polyethylene acetabular cups in a hip simulator. J Orthop Res. 1999;17:329–339.PubMed McKellop HA, Shen FW, Campbell P, Ota T. Effect of molecular weight, calcium stearate, and sterilization methods on the wear of ultra-high molecular weight polyethylene acetabular cups in a hip simulator. J Orthop Res. 1999;17:329–339.PubMed
97.
go back to reference McKnulty D, Swope S. The effect of crosslinking UHMWPE on in-vitro wear rates of fixed and mobile bearing knees. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. McKnulty D, Swope S. The effect of crosslinking UHMWPE on in-vitro wear rates of fixed and mobile bearing knees. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
98.
go back to reference Medel F, Kurtz SM, Klein G, Levine H, Sharkey P, Austin M, Kraay M, Rimnac CM. Clinical, surface damage and oxidative performance of Poly II tibial inserts after long-term implantation. J Long Term Eff Med Implants. 2008;18:151–165.PubMed Medel F, Kurtz SM, Klein G, Levine H, Sharkey P, Austin M, Kraay M, Rimnac CM. Clinical, surface damage and oxidative performance of Poly II tibial inserts after long-term implantation. J Long Term Eff Med Implants. 2008;18:151–165.PubMed
99.
go back to reference Minoda Y, Aihara M, Sakawa A, Fukuoka S, Hayakawa K, Tomita M, Umeda N, Ohzono K. Comparison between highly cross-linked and conventional polyethylene in total knee arthroplasty. Knee. 2009;16:348–351.PubMed Minoda Y, Aihara M, Sakawa A, Fukuoka S, Hayakawa K, Tomita M, Umeda N, Ohzono K. Comparison between highly cross-linked and conventional polyethylene in total knee arthroplasty. Knee. 2009;16:348–351.PubMed
100.
go back to reference Miyanishi K, Hara T, Kaminomachi S, Maekawa M, Iwamoto M, Torisu T. Short-term wear of Japanese highly cross-linked polyethylene in cementless THA. Arch Orthop Trauma Surg. 2008;128:995–1000.PubMed Miyanishi K, Hara T, Kaminomachi S, Maekawa M, Iwamoto M, Torisu T. Short-term wear of Japanese highly cross-linked polyethylene in cementless THA. Arch Orthop Trauma Surg. 2008;128:995–1000.PubMed
101.
go back to reference Moore KD, Beck PR, Petersen DW, Cuckler JM, Lemons JE, Eberhardt AW. Early failure of a cross-linked polyethylene acetabular liner: a case report. J Bone Joint Surg Am. 2008;90:2499–2504.PubMed Moore KD, Beck PR, Petersen DW, Cuckler JM, Lemons JE, Eberhardt AW. Early failure of a cross-linked polyethylene acetabular liner: a case report. J Bone Joint Surg Am. 2008;90:2499–2504.PubMed
102.
go back to reference Mu Z, Tian J, Wu T, Yang J, Pei F. A systematic review of radiological outcomes of highly cross-linked polyethylene versus conventional polyethylene in total hip arthroplasty. Int Orthop. 2009;33:599–604.PubMed Mu Z, Tian J, Wu T, Yang J, Pei F. A systematic review of radiological outcomes of highly cross-linked polyethylene versus conventional polyethylene in total hip arthroplasty. Int Orthop. 2009;33:599–604.PubMed
103.
go back to reference Muratoglu OK, Bragdon CR, O’Connor DO, Jasty M, Harris WH. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP Paul Award. J Arthroplasty. 2001;16:149–160.PubMed Muratoglu OK, Bragdon CR, O’Connor DO, Jasty M, Harris WH. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP Paul Award. J Arthroplasty. 2001;16:149–160.PubMed
104.
go back to reference Mutimer J, Devane PA, Adams K, Horne JG. Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years. Clin Orthop Relat Res. 2010;468:3228–3233.PubMed Mutimer J, Devane PA, Adams K, Horne JG. Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years. Clin Orthop Relat Res. 2010;468:3228–3233.PubMed
105.
go back to reference Nair PD. Currently practised sterilization methods—some inadvertent consequences. J Biomater Appl. 1995;10:121–135.PubMed Nair PD. Currently practised sterilization methods—some inadvertent consequences. J Biomater Appl. 1995;10:121–135.PubMed
106.
go back to reference Nakahara I, Nakamura N, Nishii T, Miki H, Sakai T, Sugano N. Minimum five-year follow-up wear measurement of longevity highly cross-linked polyethylene cup against cobalt-chromium or zirconia heads. J Arthroplasty. 2010;25:1182–1187.PubMed Nakahara I, Nakamura N, Nishii T, Miki H, Sakai T, Sugano N. Minimum five-year follow-up wear measurement of longevity highly cross-linked polyethylene cup against cobalt-chromium or zirconia heads. J Arthroplasty. 2010;25:1182–1187.PubMed
107.
go back to reference Nizard RS, Sedel L, Christel P, Meunier A, Soudry M, Witvoet J. Ten-year survivorship of cemented ceramic-ceramic total hip prosthesis. Clin Orthop Relat Res. 1992;282:53–63.PubMed Nizard RS, Sedel L, Christel P, Meunier A, Soudry M, Witvoet J. Ten-year survivorship of cemented ceramic-ceramic total hip prosthesis. Clin Orthop Relat Res. 1992;282:53–63.PubMed
108.
go back to reference Olyslaegers C, Defoort K, Simon JP, Vandenberghe L. Wear in conventional and highly cross-linked polyethylene cups: a 5-year follow-up study. J Arthroplasty. 2008;23:489–494.PubMed Olyslaegers C, Defoort K, Simon JP, Vandenberghe L. Wear in conventional and highly cross-linked polyethylene cups: a 5-year follow-up study. J Arthroplasty. 2008;23:489–494.PubMed
109.
go back to reference Oonishi H, Clarke IC, Masuda S, Amino H. Study of retrieved acetabular sockets made from high-dose, cross-linked polyethylene. J Arthroplasty. 2001;16:129–133.PubMed Oonishi H, Clarke IC, Masuda S, Amino H. Study of retrieved acetabular sockets made from high-dose, cross-linked polyethylene. J Arthroplasty. 2001;16:129–133.PubMed
110.
go back to reference Oonishi H, Kim SC, Takao Y, Kyomoto M, Iwamoto M, Ueno M. Wear of highly cross-linked polyethylene acetabular cup in Japan. J Arthroplasty. 2006;21:944–949.PubMed Oonishi H, Kim SC, Takao Y, Kyomoto M, Iwamoto M, Ueno M. Wear of highly cross-linked polyethylene acetabular cup in Japan. J Arthroplasty. 2006;21:944–949.PubMed
111.
go back to reference Oonishi H, Takayama Y, Tsuji E. Improvement of polyethylene by irradiation in artificial joints. Rad Phys Chem. 1992;39:495–504. Oonishi H, Takayama Y, Tsuji E. Improvement of polyethylene by irradiation in artificial joints. Rad Phys Chem. 1992;39:495–504.
112.
go back to reference Oral E, Wannomae KK, Hawkins N, Harris WH, Muratoglu OK. Alpha-tocopherol-doped irradiated UHMWPE for high fatigue resistance and low wear. Biomaterials. 2004;25:5515–5522.PubMed Oral E, Wannomae KK, Hawkins N, Harris WH, Muratoglu OK. Alpha-tocopherol-doped irradiated UHMWPE for high fatigue resistance and low wear. Biomaterials. 2004;25:5515–5522.PubMed
113.
go back to reference Premnath V, Harris WH, Jasty M, Merrill EW. Gamma sterilization of UHMWPE articular implants: an analysis of the oxidation problem. Biomaterials. 1996;17:1741–1753.PubMed Premnath V, Harris WH, Jasty M, Merrill EW. Gamma sterilization of UHMWPE articular implants: an analysis of the oxidation problem. Biomaterials. 1996;17:1741–1753.PubMed
114.
go back to reference Rajadhyaksha AD, Brotea C, Cheung Y, Kuhn C, Ramakrishnan R, Zelicof SB. Five-year comparative study of highly cross-linked (Crossfire) and traditional polyethylene. J Arthroplasty. 2009;24:161–167.PubMed Rajadhyaksha AD, Brotea C, Cheung Y, Kuhn C, Ramakrishnan R, Zelicof SB. Five-year comparative study of highly cross-linked (Crossfire) and traditional polyethylene. J Arthroplasty. 2009;24:161–167.PubMed
115.
go back to reference Ramani K, Parasnis NC. Process-induced effects in compression molding of ultra-high molecular weight polyethylene (UHMWPE). In: Gsell RA, Stein HL, Ploskonka JJ, eds. Characterization and Properties of Ultra-High Molecular Weight Polyethylene. West Conshohocken, PA: American Society for Testing and Materials; 1998:5–23. Ramani K, Parasnis NC. Process-induced effects in compression molding of ultra-high molecular weight polyethylene (UHMWPE). In: Gsell RA, Stein HL, Ploskonka JJ, eds. Characterization and Properties of Ultra-High Molecular Weight Polyethylene. West Conshohocken, PA: American Society for Testing and Materials; 1998:5–23.
116.
go back to reference Reeves EA, Barton DC, FitzPatrick DP, Fisher J. Comparison of gas plasma and gamma irradiation in air sterilization on the delamination wear of the ultra-high molecular weight polyethylene used in knee replacements. Proc Inst Mech Eng H. 2000;214:249–255.PubMed Reeves EA, Barton DC, FitzPatrick DP, Fisher J. Comparison of gas plasma and gamma irradiation in air sterilization on the delamination wear of the ultra-high molecular weight polyethylene used in knee replacements. Proc Inst Mech Eng H. 2000;214:249–255.PubMed
117.
go back to reference Ries MD, Weaver K, Beals N. Safety and efficacy of ethylene oxide sterilized polyethylene in total knee arthroplasty. Clin Orthop Relat Res. 1996;331:159–163.PubMed Ries MD, Weaver K, Beals N. Safety and efficacy of ethylene oxide sterilized polyethylene in total knee arthroplasty. Clin Orthop Relat Res. 1996;331:159–163.PubMed
118.
go back to reference Ries MD, Weaver K, Rose RM, Gunther J, Sauer W, Beals N. Fatigue strength of polyethylene after sterilization by gamma irradiation or ethylene oxide. Clin Orthop Relat Res. 1996;333:87–95.PubMed Ries MD, Weaver K, Rose RM, Gunther J, Sauer W, Beals N. Fatigue strength of polyethylene after sterilization by gamma irradiation or ethylene oxide. Clin Orthop Relat Res. 1996;333:87–95.PubMed
119.
go back to reference Rimnac CM, Klein RW, Betts F, Wright TM. Post-irradiation aging of ultra-high molecular weight polyethylene. J Bone Joint Surg Am. 1994;76:1052–1056.PubMed Rimnac CM, Klein RW, Betts F, Wright TM. Post-irradiation aging of ultra-high molecular weight polyethylene. J Bone Joint Surg Am. 1994;76:1052–1056.PubMed
120.
go back to reference Ring PA. Complete replacement arthroplasty of the hip by the ring prosthesis. J Bone Joint Surg Br. 1968;50:720–731.PubMed Ring PA. Complete replacement arthroplasty of the hip by the ring prosthesis. J Bone Joint Surg Br. 1968;50:720–731.PubMed
121.
go back to reference Rohrl S, Nivbrant B, Mingguo L, Hewitt B. In vivo wear and migration of highly cross-linked polyethylene cups a radiostereometry analysis study. J Arthroplasty. 2005;20:409–413.PubMed Rohrl S, Nivbrant B, Mingguo L, Hewitt B. In vivo wear and migration of highly cross-linked polyethylene cups a radiostereometry analysis study. J Arthroplasty. 2005;20:409–413.PubMed
122.
go back to reference Rohrl SM, Li MG, Nilsson KG, Nivbrant B. Very low wear of non-remelted highly cross-linked polyethylene cups: an RSA study lasting up to 6 years. Acta Orthop. 2007;78:739–745.PubMed Rohrl SM, Li MG, Nilsson KG, Nivbrant B. Very low wear of non-remelted highly cross-linked polyethylene cups: an RSA study lasting up to 6 years. Acta Orthop. 2007;78:739–745.PubMed
123.
go back to reference Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: the concept of effective joint space. J Bone Joint Surg Am. 1992;74:849–863.PubMed Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: the concept of effective joint space. J Bone Joint Surg Am. 1992;74:849–863.PubMed
124.
go back to reference Schmidt MB, Hamilton JV. The effects of calcium stearate on the properties of UHMWPE. Trans Orthop Res Soc. 1996;21:22. Schmidt MB, Hamilton JV. The effects of calcium stearate on the properties of UHMWPE. Trans Orthop Res Soc. 1996;21:22.
125.
go back to reference Scott DL, Campbell PA, McClung CD, Schmalzried TP. Factors contributing to rapid wear and osteolysis in hips with modular acetabular bearings made of hylamer. J Arthroplasty. 2000;15:35–46.PubMed Scott DL, Campbell PA, McClung CD, Schmalzried TP. Factors contributing to rapid wear and osteolysis in hips with modular acetabular bearings made of hylamer. J Arthroplasty. 2000;15:35–46.PubMed
126.
go back to reference Sedel L, Kerboull L, Christel P, Meunier A, Witvoet J. Alumina-on-alumina hip replacement: results and survivorship in young patients. J Bone Joint Surg Br. 1990;72:658–663.PubMed Sedel L, Kerboull L, Christel P, Meunier A, Witvoet J. Alumina-on-alumina hip replacement: results and survivorship in young patients. J Bone Joint Surg Br. 1990;72:658–663.PubMed
127.
go back to reference Shia DS, Clohisy JC, Schinsky MF, Martell JM, Maloney WJ. THA with highly cross-linked polyethylene in patients 50 years or younger. Clin Orthop Relat Res. 2009;467:2059–2065.PubMed Shia DS, Clohisy JC, Schinsky MF, Martell JM, Maloney WJ. THA with highly cross-linked polyethylene in patients 50 years or younger. Clin Orthop Relat Res. 2009;467:2059–2065.PubMed
128.
go back to reference Streicher RM. Ionizing irradiation for sterilization and modification of high molecular weight polyethylenes. Plast Rubber Proc Appl. 1988;10:221–229. Streicher RM. Ionizing irradiation for sterilization and modification of high molecular weight polyethylenes. Plast Rubber Proc Appl. 1988;10:221–229.
129.
go back to reference Sutula LC, Collier JP, Saum KA, Currier BH, Currier JH, Sanford WM, Mayor MB, Wooding RE, Sperling DK, Williams IR, Kasprzack DJ, Suprenant VA. Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin Orthop Relat Res. 1995;319:28–40.PubMed Sutula LC, Collier JP, Saum KA, Currier BH, Currier JH, Sanford WM, Mayor MB, Wooding RE, Sperling DK, Williams IR, Kasprzack DJ, Suprenant VA. Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin Orthop Relat Res. 1995;319:28–40.PubMed
130.
go back to reference Swarts D, Gsell R, King R, Devanathan D, Wallace S, Lin S, Rohr W. Aging of calcium stearate-free polyethylene. In: Transactions of the 5th World Biomaterials Congress. Mount Laurel, NJ: Society for Biomaterials; 1996:196. Swarts D, Gsell R, King R, Devanathan D, Wallace S, Lin S, Rohr W. Aging of calcium stearate-free polyethylene. In: Transactions of the 5th World Biomaterials Congress. Mount Laurel, NJ: Society for Biomaterials; 1996:196.
131.
go back to reference Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2212–2217.PubMed Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2212–2217.PubMed
132.
go back to reference Triclot P, Grosjean G, El Masri F, Courpied JP, Hamadouche M. A comparison of the penetration rate of two polyethylene acetabular liners of different levels of cross-linking: a prospective randomised trial. J Bone Joint Surg Br. 2007;89:1439–1445.PubMed Triclot P, Grosjean G, El Masri F, Courpied JP, Hamadouche M. A comparison of the penetration rate of two polyethylene acetabular liners of different levels of cross-linking: a prospective randomised trial. J Bone Joint Surg Br. 2007;89:1439–1445.PubMed
133.
go back to reference Wang A, Essner A, Polineni VK, Stark C, Dumbleton JH. Lubrication and wear of ultra-high molecular weight polyethylene in total joint replacements. Tribol Int. 1998;31:17–33. Wang A, Essner A, Polineni VK, Stark C, Dumbleton JH. Lubrication and wear of ultra-high molecular weight polyethylene in total joint replacements. Tribol Int. 1998;31:17–33.
134.
go back to reference Wang A, Manley M, Serekian P. Wear and structural fatigue simulation of crosslinked ultra-high molecular weight polyethylene for hip and knee bearing applications. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Wang A, Manley M, Serekian P. Wear and structural fatigue simulation of crosslinked ultra-high molecular weight polyethylene for hip and knee bearing applications. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
135.
go back to reference Waugh W. John Charnley: The Man and the Hip. London, UK: Springer-Verlag; 1990. Waugh W. John Charnley: The Man and the Hip. London, UK: Springer-Verlag; 1990.
136.
go back to reference White SE, Paxson RD, Tanner MG, Whiteside LA. Effects of sterilization on wear in total knee arthroplasty. Clin Orthop Relat Res. 1996;331:164–171.PubMed White SE, Paxson RD, Tanner MG, Whiteside LA. Effects of sterilization on wear in total knee arthroplasty. Clin Orthop Relat Res. 1996;331:164–171.PubMed
137.
go back to reference Whittaker JP, Charron KD, McCalden RW, Macdonald SJ, Bourne RB. Comparison of steady state femoral head penetration rates between two highly cross-linked polyethylenes in total hip arthroplasty. J Arthroplasty. 2010;25:680–686.PubMed Whittaker JP, Charron KD, McCalden RW, Macdonald SJ, Bourne RB. Comparison of steady state femoral head penetration rates between two highly cross-linked polyethylenes in total hip arthroplasty. J Arthroplasty. 2010;25:680–686.PubMed
138.
go back to reference Willert HG, Bertram H, Buchhorn GH. Osteolysis in alloarthroplasty of the hip: the role of ultra-high molecular weight polyethylene wear particles. Clin Orthop Relat Res. 1990;258:95–107.PubMed Willert HG, Bertram H, Buchhorn GH. Osteolysis in alloarthroplasty of the hip: the role of ultra-high molecular weight polyethylene wear particles. Clin Orthop Relat Res. 1990;258:95–107.PubMed
139.
go back to reference Willert HG, Buchhorn GH, Eyerer P. Ultra-high Molecular Weight Polyethlene as Biomaterial in Orthopedic Surgery. Toronto, Canada: Hogrefe & Huber Publishers; 1991. Willert HG, Buchhorn GH, Eyerer P. Ultra-high Molecular Weight Polyethlene as Biomaterial in Orthopedic Surgery. Toronto, Canada: Hogrefe & Huber Publishers; 1991.
140.
go back to reference Williams IR, Mayor MB, Collier JP. The impact of sterilization method on wear in knee arthroplasty. Clin Orthop Relat Res. 1998;356:170–180.PubMed Williams IR, Mayor MB, Collier JP. The impact of sterilization method on wear in knee arthroplasty. Clin Orthop Relat Res. 1998;356:170–180.PubMed
141.
go back to reference Winter M, Griss P, Scheller G, Moser T. Ten- to 14-year results of a ceramic hip prosthesis. Clin Orthop Relat Res. 1992;282:73–80.PubMed Winter M, Griss P, Scheller G, Moser T. Ten- to 14-year results of a ceramic hip prosthesis. Clin Orthop Relat Res. 1992;282:73–80.PubMed
142.
go back to reference Wright TM, Bartel DL. The problem of surface damage in polyethylene total knee components. Clin Orthop Relat Res. 1986;205:67–74.PubMed Wright TM, Bartel DL. The problem of surface damage in polyethylene total knee components. Clin Orthop Relat Res. 1986;205:67–74.PubMed
143.
go back to reference Wright TM, Fukubayashi T, Burstein AH. The effect of carbon fiber reinforcement on contact area, contact pressure, and time-dependent deformation in polyethylene tibial components. J Biomed Mater Res. 1981;15:719–730.PubMed Wright TM, Fukubayashi T, Burstein AH. The effect of carbon fiber reinforcement on contact area, contact pressure, and time-dependent deformation in polyethylene tibial components. J Biomed Mater Res. 1981;15:719–730.PubMed
144.
go back to reference Wroblewski BM, Siney PD, Dowson D, Collins SN. Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J Bone Joint Surg Br. 1996;78:280–285.PubMed Wroblewski BM, Siney PD, Dowson D, Collins SN. Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J Bone Joint Surg Br. 1996;78:280–285.PubMed
145.
go back to reference Wrona M, Mayor MB, Collier JP, Jensen RE. The correlation between fusion defects and damage in tibial polyethylene bearings. Clin Orthop Relat Res. 1994;299:92–103.PubMed Wrona M, Mayor MB, Collier JP, Jensen RE. The correlation between fusion defects and damage in tibial polyethylene bearings. Clin Orthop Relat Res. 1994;299:92–103.PubMed
146.
go back to reference Yao JQ, Lu MP, Johnson TS, Gilbertson LN, Swarts D, Blanchard CR, Crowninshield RD. Improved resistance to wear, delamination and posterior loading fatigue damage of electron beam irradiated, melt-annealed, highly crosslinked UHMWPE knee inserts. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004. Yao JQ, Lu MP, Johnson TS, Gilbertson LN, Swarts D, Blanchard CR, Crowninshield RD. Improved resistance to wear, delamination and posterior loading fatigue damage of electron beam irradiated, melt-annealed, highly crosslinked UHMWPE knee inserts. In: Kurtz SM, Gsell R, Martell J, eds. ASTM STP 1445. Highly Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene For Joint Replacements. West Conshohocken, PA: American Society for Testing and Materials; 2004.
Metadata
Title
History and Systematic Review of Wear and Osteolysis Outcomes for First-generation Highly Crosslinked Polyethylene
Authors
Steven M. Kurtz, PhD
Heather A. Gawel, MS
Jasmine D. Patel, PhD
Publication date
01-08-2011
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 8/2011
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-011-1872-4

Other articles of this Issue 8/2011

Clinical Orthopaedics and Related Research® 8/2011 Go to the issue

Symposium: Bone Quality: From Bench to Bedside

Biographical Sketch: Fuller Albright, MD 1900–1969

Symposium: Bone Quality: From Bench to Bedside

Bone Quality: From Bench to Bedside: Opening Editorial Comment

Symposium: Bone Quality: From Bench to Bedside

Diseases Affecting Bone Quality: Beyond Osteoporosis

Symposium: Bone Quality: From Bench to Bedside

Surgical Treatment Options in Patients With Impaired Bone Quality

Symposium: Bone Quality: From Bench to Bedside

Raman Assessment of Bone Quality