Skip to main content
Top
Published in: Clinical Orthopaedics and Related Research® 2/2011

01-02-2011 | Symposium: Papers Presented at the Hip Society Meetings 2010

Hard-on-Hard Total Hip Impingement Causes Extreme Contact Stress Concentrations

Authors: Jacob M. Elkins, MS, Megan K. O’Brien, BS, Nicholas J. Stroud, BSE, Douglas R. Pedersen, PhD, John J. Callaghan, MD, Thomas D. Brown, PhD

Published in: Clinical Orthopaedics and Related Research® | Issue 2/2011

Login to get access

Abstract

Background

Impingement events, in addition to their role immediately proximate to frank dislocation, hold the potential to damage new-generation hard-on-hard bearings as a result of the relatively unforgiving nature of the materials and designs. Because of the higher stiffness and tighter design tolerances of metal-on-metal and ceramic implants, surgical positioning plausibly has become even more important.

Questions/purposes

We asked (1) whether, and under what cup orientation conditions, hard-on-hard impingements might challenge implant material failure strength; and (2) whether particle generation propensity at impingement and egress sites would show similar dependence on cup orientation.

Methods

Realistic computational simulations were enabled by multistage finite element analyses, addressing both global construct motion and loading, and focal stress concentrations at neck impingement and rim egress sites. The global model, validated by a cadaveric simulation in a servohydraulic hip simulator, included both hardware components and advanced anisotropic capsule characterization. Parametric computational runs explored the effect of cup orientation for both ceramic-on-ceramic and metal-on-metal bearing couples for two distinct motion sequences associated with dislocation.

Results

Stress concentrations from impingement increased nearly linearly with increased cup tilt and with cup anteversion. In some situations, peak values of stress approached or exceeded 1 GPa, levels challenging the yield strength of cobalt-chromium implants, and potentially the fracture strength of ceramics. The tendency for impingement events to generate debris, indexed in terms of a new scraping severity metric, showed orientation dependences similar to that for bulk material failure.

Conclusions

Damage propensity arising from impingement events in hard total hip bearings is highly orientation-dependent.
Literature
1.
go back to reference Archard JF. Contact and rubbing of flat surfaces. J Appl Phys. 1953;24:981–988.CrossRef Archard JF. Contact and rubbing of flat surfaces. J Appl Phys. 1953;24:981–988.CrossRef
2.
go back to reference Barrack RL, Burak C, Skinner HB. Concerns about ceramics in THA. Clin Orthop Relat Res. 2004;429:73–79.CrossRefPubMed Barrack RL, Burak C, Skinner HB. Concerns about ceramics in THA. Clin Orthop Relat Res. 2004;429:73–79.CrossRefPubMed
3.
go back to reference Barrack RL, Schmalzried TP. Impingement and rim wear associated with early osteolysis after a total hip replacement—a case report. J Bone Joint Surg Am. 2002;84:1218–1220.PubMed Barrack RL, Schmalzried TP. Impingement and rim wear associated with early osteolysis after a total hip replacement—a case report. J Bone Joint Surg Am. 2002;84:1218–1220.PubMed
4.
go back to reference Brockett C, Williams S, Jin Z, Isaac G, Fisher J. Friction of total hip replacements with different bearings and loading conditions. J Biomed Mater Res Appl Biomater. 2007;81:508–515. Brockett C, Williams S, Jin Z, Isaac G, Fisher J. Friction of total hip replacements with different bearings and loading conditions. J Biomed Mater Res Appl Biomater. 2007;81:508–515.
5.
go back to reference Brown TD, Callaghan JJ. Impingement in total hip replacement: mechanisms and consequences. Curr Orthop. 2008;22:376–391.CrossRefPubMed Brown TD, Callaghan JJ. Impingement in total hip replacement: mechanisms and consequences. Curr Orthop. 2008;22:376–391.CrossRefPubMed
6.
go back to reference Coventry MB. Late dislocations in patients with Charnley total hip arthroplasty. J Bone Joint Surg Am. 1985;67:832–841.PubMed Coventry MB. Late dislocations in patients with Charnley total hip arthroplasty. J Bone Joint Surg Am. 1985;67:832–841.PubMed
7.
go back to reference Holley KG, Furman BD, Babalola OM, Lipman JD, Padgett DE, Wright TM. Impingement of acetabular cups in a hip simulator—comparison of highly cross-linked and conventional polyethylene. J Arthroplasty. 2005;20(Suppl 3):77–86.CrossRefPubMed Holley KG, Furman BD, Babalola OM, Lipman JD, Padgett DE, Wright TM. Impingement of acetabular cups in a hip simulator—comparison of highly cross-linked and conventional polyethylene. J Arthroplasty. 2005;20(Suppl 3):77–86.CrossRefPubMed
8.
go back to reference ISO 5832. Implants for Surgery—Metallic Materials. Geneva: International Organization for Standardization; 1996. ISO 5832. Implants for Surgery—Metallic Materials. Geneva: International Organization for Standardization; 1996.
9.
go back to reference ISO 6474. Implants for Surgery—Ceramic Materials Based on High Purity Alumina. Geneva: International Organization for Standardization; 1994. ISO 6474. Implants for Surgery—Ceramic Materials Based on High Purity Alumina. Geneva: International Organization for Standardization; 1994.
10.
go back to reference Liu F, Leslie I, Williams S, Fisher J, Jin Z. Development of computational wear simulations of metal-on-metal hip resurfacing replacements. J Biomech. 2008;41:686–694.CrossRefPubMed Liu F, Leslie I, Williams S, Fisher J, Jin Z. Development of computational wear simulations of metal-on-metal hip resurfacing replacements. J Biomech. 2008;41:686–694.CrossRefPubMed
11.
go back to reference Lundberg HJ, Liu SS, Callaghan JJ, Pedersen DR, O’Rourke MR, Goetz DD, Vittetoe DA, Clohisy JC, Brown TD. Association of third body embedment with rim damage in retrieved acetabular liners. Clin Orthop Relat Res. 2007;465:133–139.PubMed Lundberg HJ, Liu SS, Callaghan JJ, Pedersen DR, O’Rourke MR, Goetz DD, Vittetoe DA, Clohisy JC, Brown TD. Association of third body embedment with rim damage in retrieved acetabular liners. Clin Orthop Relat Res. 2007;465:133–139.PubMed
12.
go back to reference Nadzadi ME, Pedersen DR, Callaghan JJ, Brown TD. Effects of acetabular component orientation on dislocation propensity for small-head-size total hip arthroplasty. Clin Biomech (Bristol, Avon).2002;17:32–40.CrossRef Nadzadi ME, Pedersen DR, Callaghan JJ, Brown TD. Effects of acetabular component orientation on dislocation propensity for small-head-size total hip arthroplasty. Clin Biomech (Bristol, Avon).2002;17:32–40.CrossRef
13.
go back to reference Nadzadi ME, Pedersen DR, Yack HJ, Callaghan JJ, Brown TD. Kinematics, kinetics, and finite element analysis of commonplace maneuvers at risk for total hip dislocation. J Biomech. 2003;36:577–591.CrossRefPubMed Nadzadi ME, Pedersen DR, Yack HJ, Callaghan JJ, Brown TD. Kinematics, kinetics, and finite element analysis of commonplace maneuvers at risk for total hip dislocation. J Biomech. 2003;36:577–591.CrossRefPubMed
14.
go back to reference Pedersen DR, Callaghan JJ, Brown TD. Activity-dependence of the ‘safe zone’ for impingement versus dislocation avoidance. Med Engr Phys. 2005;27:323–328.CrossRef Pedersen DR, Callaghan JJ, Brown TD. Activity-dependence of the ‘safe zone’ for impingement versus dislocation avoidance. Med Engr Phys. 2005;27:323–328.CrossRef
15.
go back to reference Scifert CF, Noble PC, Brown TD, Bartz RL, Kadakia N, Sugano N, Johnston RC, Pedersen DR, Callaghan JJ. (2001) Experimental and computational simulation of total hip arthroplasty dislocation. Orthop Clin North Am 32:552–567, vii.CrossRef Scifert CF, Noble PC, Brown TD, Bartz RL, Kadakia N, Sugano N, Johnston RC, Pedersen DR, Callaghan JJ. (2001) Experimental and computational simulation of total hip arthroplasty dislocation. Orthop Clin North Am 32:552–567, vii.CrossRef
16.
go back to reference Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA. Impingement in total hip arthroplasty: a study of retrieved acetabular components. J Arthroplasty. 2005;20:427–435.CrossRefPubMed Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA. Impingement in total hip arthroplasty: a study of retrieved acetabular components. J Arthroplasty. 2005;20:427–435.CrossRefPubMed
17.
go back to reference Ugural AC, Fenster SK. Advanced Strength and Applied Elasticity. 4th ed. Upper Saddle River, NJ: Prentice Hall; 2008. Ugural AC, Fenster SK. Advanced Strength and Applied Elasticity. 4th ed. Upper Saddle River, NJ: Prentice Hall; 2008.
18.
go back to reference Urban RM, Tomlinson MJ, Hall DJ, Jacobs JJ. Accumulation in liver and spleen of metal particles generated at nonbearing surfaces in hip arthroplasty. J Arthroplasty. 2004;19(Suppl 3):94–101.CrossRefPubMed Urban RM, Tomlinson MJ, Hall DJ, Jacobs JJ. Accumulation in liver and spleen of metal particles generated at nonbearing surfaces in hip arthroplasty. J Arthroplasty. 2004;19(Suppl 3):94–101.CrossRefPubMed
19.
go back to reference Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, Lohmann CH. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Joint Surg Am. 2005;87:28–36.CrossRefPubMed Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, Lohmann CH. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Joint Surg Am. 2005;87:28–36.CrossRefPubMed
Metadata
Title
Hard-on-Hard Total Hip Impingement Causes Extreme Contact Stress Concentrations
Authors
Jacob M. Elkins, MS
Megan K. O’Brien, BS
Nicholas J. Stroud, BSE
Douglas R. Pedersen, PhD
John J. Callaghan, MD
Thomas D. Brown, PhD
Publication date
01-02-2011
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 2/2011
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-010-1632-x

Other articles of this Issue 2/2011

Clinical Orthopaedics and Related Research® 2/2011 Go to the issue

Symposium: Papers Presented at the Hip Society Meetings 2010

A Comparison of Two Implant Systems in Restoration of Hip Geometry in Arthroplasty

Symposium: Papers Presented at the Hip Society Meetings 2010

Improving Cup Positioning Using a Mechanical Navigation Instrument

Symposium: Papers Presented at the Hip Society Meetings 2010

A Monoblock Porous Tantalum Acetabular Cup Has No Osteolysis on CT at 10 Years