Published in: Clinical Orthopaedics and Related Research® 12/2010

01-12-2010 | Basic Research

Bioabsorbable Tricalcium Phosphate Bone Cement Strengthens Fixation of Suture Anchors

Authors: Rayshad Oshtory, MD MBA, Derek P. Lindsey, MS, Nicholas J. Giori, MD, PhD, Faisal M. Mirza, MD, FRCSC

Published in: Clinical Orthopaedics and Related Research® | Issue 12/2010

Login to get access

Abstract

Background

Failure of suture anchor fixation in rotator cuff repair can occur at different interfaces. Prior studies show fixation at the bone-anchor interface can be augmented using polymethylmethacrylate (PMMA) cement, and screw fixation into bone can be strengthened using bioabsorbable tricalcium phosphate cement.

Questions/purposes

We wished to determine whether augmentation of suture anchor fixation using bioabsorbable tricalcium phosphate cement would increase pullout strength of suture anchors from bone and the number of cycles to failure, to determine the mode of failure after cement augmentation, and to compare strength and mode of failure with those after augmentation with PMMA.

Methods

We used 10 matched pairs of cadaveric proximal humeri and implanted a metal screw-type suture anchor in one side and on the other side injected tricalcium phosphate cement into the anchor holes before anchor placement. We tested all specimens to failure using a ramped cyclic loading protocol.

Results

Tricalcium phosphate cement augmentation increased the final load to failure by 29% and the number of cycles to failure by 20%. Visual inspection confirmed that failure occurred at the cement-bone interface.

Conclusions

Tricalcium phosphate cement appears to augment suture anchor fixation into bone, reducing the risk of anchor pullout and failure.

Clinical Relevance

When relying on suture anchor fixation in bone of questionable quality, we suggest considering augmentation of suture anchor fixation with bioabsorbable cement. This method also provides potential for bioabsorbability and may be more amenable to arthroscopic application.
Literature
1.
go back to reference Bishop J, Klepps S, Lo IK, Bird J, Gladstone JN, Flatow EL. Cuff integrity after arthroscopic versus open rotator cuff repair: a prospective study. J Shoulder Elbow Surg. 2006;15:290–299.CrossRefPubMed Bishop J, Klepps S, Lo IK, Bird J, Gladstone JN, Flatow EL. Cuff integrity after arthroscopic versus open rotator cuff repair: a prospective study. J Shoulder Elbow Surg. 2006;15:290–299.CrossRefPubMed
2.
go back to reference Calvert PT, Packer NP, Stoker DJ, Bayley JI, Kessel L. Arthrography of the shoulder after operative repair of the torn rotator cuff. J Bone Joint Surg Br. 1986;68:147–150.PubMed Calvert PT, Packer NP, Stoker DJ, Bayley JI, Kessel L. Arthrography of the shoulder after operative repair of the torn rotator cuff. J Bone Joint Surg Br. 1986;68:147–150.PubMed
3.
go back to reference Collinge C, Merk B, Lautenschlager EP. Mechanical evaluation of fracture fixation augmented with tricalcium phosphate bone cement in a porous osteoporotic cancellous bone model. J Orthop Trauma. 2007;21:124–128.CrossRefPubMed Collinge C, Merk B, Lautenschlager EP. Mechanical evaluation of fracture fixation augmented with tricalcium phosphate bone cement in a porous osteoporotic cancellous bone model. J Orthop Trauma. 2007;21:124–128.CrossRefPubMed
4.
go back to reference Cook SD, Salkeld SL, Stanley T, Faciane A, Miller SD. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine J. 2004;4:402–408.CrossRefPubMed Cook SD, Salkeld SL, Stanley T, Faciane A, Miller SD. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine J. 2004;4:402–408.CrossRefPubMed
5.
go back to reference Cummins CA, Murrell GA. Mode of failure for rotator cuff repair with suture anchors identified at revision surgery. J Shoulder Elbow Surg. 2003;12:128–133.CrossRefPubMed Cummins CA, Murrell GA. Mode of failure for rotator cuff repair with suture anchors identified at revision surgery. J Shoulder Elbow Surg. 2003;12:128–133.CrossRefPubMed
6.
go back to reference Djurasovic M, Marra G, Arroyo JS, Pollock RG, Flatow EL, Bigliani LU. Revision rotator cuff repair: factors influencing results. J Bone Joint Surg Am. 2001;83:1849–1855.PubMed Djurasovic M, Marra G, Arroyo JS, Pollock RG, Flatow EL, Bigliani LU. Revision rotator cuff repair: factors influencing results. J Bone Joint Surg Am. 2001;83:1849–1855.PubMed
7.
go back to reference Elder S, Frankenburg E, Goulet J, Yetkinler D, Poser R, Goldstein S. Biomechanical evaluation of calcium phosphate cement-augmented fixation of unstable intertrochanteric fractures. J Orthop Trauma. 2000;14:386–393.CrossRefPubMed Elder S, Frankenburg E, Goulet J, Yetkinler D, Poser R, Goldstein S. Biomechanical evaluation of calcium phosphate cement-augmented fixation of unstable intertrochanteric fractures. J Orthop Trauma. 2000;14:386–393.CrossRefPubMed
8.
go back to reference Favus MJ, ed. Primer on the Metabolic Bone Diseases and Disorders of Mineral Metabolism. Ed 4. Philadelphia, PA: Lippincott, Williams & Wilkins; 1999. Favus MJ, ed. Primer on the Metabolic Bone Diseases and Disorders of Mineral Metabolism. Ed 4. Philadelphia, PA: Lippincott, Williams & Wilkins; 1999.
9.
go back to reference Giori NJ, Sohn DH, Mirza FM, Lindsey DP, Lee AT. Bone cement improves suture anchor fixation. Clin Orthop Relat Res. 2006;451:236–241.CrossRefPubMed Giori NJ, Sohn DH, Mirza FM, Lindsey DP, Lee AT. Bone cement improves suture anchor fixation. Clin Orthop Relat Res. 2006;451:236–241.CrossRefPubMed
10.
go back to reference Jang JS, Lee SH, Rhee CH, Lee SH. Polymethylmethacrylate-augmented screw fixation for stabilization in metastatic spinal tumors: technical note. J Neurosurg. 2002;96(1 suppl):131–134.PubMed Jang JS, Lee SH, Rhee CH, Lee SH. Polymethylmethacrylate-augmented screw fixation for stabilization in metastatic spinal tumors: technical note. J Neurosurg. 2002;96(1 suppl):131–134.PubMed
11.
go back to reference Klepps S, Bishop J, Lin J, Cahlon O, Strauss A, Hayes P, Flatow EL. Prospective evaluation of the effect of rotator cuff integrity on the outcome of open rotator cuff repairs. Am J Sports Med. 2004;32:1716–1722.CrossRefPubMed Klepps S, Bishop J, Lin J, Cahlon O, Strauss A, Hayes P, Flatow EL. Prospective evaluation of the effect of rotator cuff integrity on the outcome of open rotator cuff repairs. Am J Sports Med. 2004;32:1716–1722.CrossRefPubMed
12.
go back to reference Magee T, Shapiro M, Hewell G, Williams D. Complications of rotator cuff surgery in which bioabsorbable anchors are used. AJR Am J Roentgenol. 2003;181:1227–1231.PubMed Magee T, Shapiro M, Hewell G, Williams D. Complications of rotator cuff surgery in which bioabsorbable anchors are used. AJR Am J Roentgenol. 2003;181:1227–1231.PubMed
13.
go back to reference Meyer DC, Fucentese SF, Koller B, Gerber C. Association of osteopenia of the humeral head with full-thickness rotator cuff tears. J Shoulder Elbow Surg. 2004;13:333–337.CrossRefPubMed Meyer DC, Fucentese SF, Koller B, Gerber C. Association of osteopenia of the humeral head with full-thickness rotator cuff tears. J Shoulder Elbow Surg. 2004;13:333–337.CrossRefPubMed
14.
go back to reference Meyer DC, Jacob HA, Pistoia W, von Roll A, Gerber C. The use of acrylic bone cement for suture anchoring. Clin Orthop Relat Res. 2003;410:295–302.CrossRefPubMed Meyer DC, Jacob HA, Pistoia W, von Roll A, Gerber C. The use of acrylic bone cement for suture anchoring. Clin Orthop Relat Res. 2003;410:295–302.CrossRefPubMed
15.
go back to reference Moore DC, Frankenburg EP, Goulet JA, Goldstein SA. Hip screw augmentation with an in situ-setting calcium phosphate cement: an in vitro biomechanical analysis. J Orthop Trauma. 1997;11:577–583.CrossRefPubMed Moore DC, Frankenburg EP, Goulet JA, Goldstein SA. Hip screw augmentation with an in situ-setting calcium phosphate cement: an in vitro biomechanical analysis. J Orthop Trauma. 1997;11:577–583.CrossRefPubMed
16.
go back to reference Moore DC, Maitra RS, Farjo LA, Graziano GP, Goldstein SA. Restoration of pedicle screw fixation with an in situ setting calcium phosphate cement. Spine (Phila PA 1976). 1997;22:1696–1705. Moore DC, Maitra RS, Farjo LA, Graziano GP, Goldstein SA. Restoration of pedicle screw fixation with an in situ setting calcium phosphate cement. Spine (Phila PA 1976). 1997;22:1696–1705.
17.
go back to reference Motzkin NE, Chao EY, An KN, Wikenheiser MA, Lewallen DG. Pull-out strength of screws from polymethylmethacrylate cement. J Bone Joint Surg Br. 1994;76:320–323.PubMed Motzkin NE, Chao EY, An KN, Wikenheiser MA, Lewallen DG. Pull-out strength of screws from polymethylmethacrylate cement. J Bone Joint Surg Br. 1994;76:320–323.PubMed
18.
go back to reference Renner SM, Lim TH, Kim WJ, Katolik L, An HS, Andersson GB. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method. Spine (Phila PA 1976). 2004;29:E212–E216. Renner SM, Lim TH, Kim WJ, Katolik L, An HS, Andersson GB. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method. Spine (Phila PA 1976). 2004;29:E212–E216.
19.
go back to reference Rohmiller MT, Schwalm D, Glattes RC, Elalayli TG, Spengler DM. Evaluation of calcium sulfate paste for augmentation of lumbar pedicle screw pullout strength. Spine J. 2002;2:255–260.CrossRefPubMed Rohmiller MT, Schwalm D, Glattes RC, Elalayli TG, Spengler DM. Evaluation of calcium sulfate paste for augmentation of lumbar pedicle screw pullout strength. Spine J. 2002;2:255–260.CrossRefPubMed
20.
go back to reference Sugaya H, Maeda K, Matsuki K, Moriishi J. Repair integrity and functional outcome after arthroscopic double-row rotator cuff repair: a prospective outcome study. J Bone Joint Surg Am. 2007;89:953–960.CrossRefPubMed Sugaya H, Maeda K, Matsuki K, Moriishi J. Repair integrity and functional outcome after arthroscopic double-row rotator cuff repair: a prospective outcome study. J Bone Joint Surg Am. 2007;89:953–960.CrossRefPubMed
21.
go back to reference Tingart MJ, Apreleva M, Zurakowski D, Warner JJ. Pullout strength of suture anchors used in rotator cuff repair. J Bone Joint Surg Am. 2003;85:2190–2198.CrossRefPubMed Tingart MJ, Apreleva M, Zurakowski D, Warner JJ. Pullout strength of suture anchors used in rotator cuff repair. J Bone Joint Surg Am. 2003;85:2190–2198.CrossRefPubMed
22.
go back to reference Wittenberg RH, Lee KS, Shea M, White AA III, Hayes WC. Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clin Orthop Relat Res. 1993;296:278–287.PubMed Wittenberg RH, Lee KS, Shea M, White AA III, Hayes WC. Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clin Orthop Relat Res. 1993;296:278–287.PubMed
Metadata
Title
Bioabsorbable Tricalcium Phosphate Bone Cement Strengthens Fixation of Suture Anchors
Authors
Rayshad Oshtory, MD MBA
Derek P. Lindsey, MS
Nicholas J. Giori, MD, PhD
Faisal M. Mirza, MD, FRCSC
Publication date
01-12-2010
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 12/2010
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-010-1412-7

Other articles of this Issue 12/2010

Clinical Orthopaedics and Related Research® 12/2010 Go to the issue

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

The Repair Process of Osteonecrosis After a Transtrochanteric Rotational Osteotomy

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

The Use of a Cemented Dual Mobility Socket to Treat Recurrent Dislocation

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Is There a Risk in Placing a Ceramic Head on a Previously Implanted Trunion?

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

What is the Midterm Survivorship and Function After Hip Resurfacing?

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Biographical Sketch: Sir John Charnley MD, 1911–1982

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Function and Fixation of Total Hip Arthroplasty in Patients 25 Years of Age or Younger