Skip to main content
Top
Published in: Clinical Orthopaedics and Related Research® 7/2009

01-07-2009 | Original Article

Computer-navigated Iliosacral Screw Insertion Reduces Malposition Rate and Radiation Exposure

Authors: Jörn Zwingmann, MD, Gerhard Konrad, MD, Elmar Kotter, MD, Norbert P. Südkamp, Prof, Michael Oberst, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 7/2009

Login to get access

Abstract

Insertion of percutaneous iliosacral screws with fluoroscopic guidance is associated with a relatively high screw malposition rate and long radiation exposure. We asked whether radiation exposure was reduced and screw position improved in patients having percutaneous iliosacral screw insertion using computer-assisted navigation compared with patients having conventional fluoroscopic screw placement. We inserted 26 screws in 24 patients using the navigation system and 35 screws in 32 patients using the conventional fluoroscopic technique. Two subgroups were analyzed, one in which only one iliosacral screw was placed and another with additional use of an external fixator. We determined screw positions by computed tomography and compared operation time, radiation exposure, and screw position. We observed no difference in operative times. Radiation exposure was reduced for the patients and operating room personnel with computer assistance. The postoperative computed tomography scan showed better screw position and fewer malpositioned screws in the three-dimensional navigated groups. Computer navigation reduced malposition rate and radiation exposure.
Level of Evidence: Level II, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.
Literature
1.
go back to reference Altman DT, Jones CB, Routt ML Jr. Superior gluteal artery injury during iliosacral screw placement. J Orthop Trauma. 1999;13:220–227.PubMedCrossRef Altman DT, Jones CB, Routt ML Jr. Superior gluteal artery injury during iliosacral screw placement. J Orthop Trauma. 1999;13:220–227.PubMedCrossRef
2.
go back to reference Arand M, Kinzl L, Gebhard F. Computer-guidance in percutaneous screw stabilization of the iliosacral joint. Clin Orthop Relat Res. 2004;422:201–207.PubMedCrossRef Arand M, Kinzl L, Gebhard F. Computer-guidance in percutaneous screw stabilization of the iliosacral joint. Clin Orthop Relat Res. 2004;422:201–207.PubMedCrossRef
3.
go back to reference Baker SP, O’Neill B, Haddon W Jr, Long WB. The Injury Severity Score: a method for describing patients with multiple injuries and evaluating emergency care. J Trauma. 1974;14:187–196.PubMedCrossRef Baker SP, O’Neill B, Haddon W Jr, Long WB. The Injury Severity Score: a method for describing patients with multiple injuries and evaluating emergency care. J Trauma. 1974;14:187–196.PubMedCrossRef
4.
go back to reference Barrick EF, O’Mara JW, Lane HE 3rd. Iliosacral screw insertion using computer-assisted CT image guidance: a laboratory study. Comput Aided Surg. 1998;3:289–296.PubMedCrossRef Barrick EF, O’Mara JW, Lane HE 3rd. Iliosacral screw insertion using computer-assisted CT image guidance: a laboratory study. Comput Aided Surg. 1998;3:289–296.PubMedCrossRef
5.
go back to reference Briem D, Windolf J, Rueger JM. [Percutaneous, 2D-fluoroscopic navigated iliosacral screw placement in the supine position: technique, possibilities, and limits] [in German]. Unfallchirurg. 2007;110:393–401.PubMedCrossRef Briem D, Windolf J, Rueger JM. [Percutaneous, 2D-fluoroscopic navigated iliosacral screw placement in the supine position: technique, possibilities, and limits] [in German]. Unfallchirurg. 2007;110:393–401.PubMedCrossRef
6.
go back to reference Cole JD, Blum DA, Ansel LJ. Outcome after fixation of unstable posterior pelvic ring injuries. Clin Orthop Relat Res. 1996;329:160–179.PubMedCrossRef Cole JD, Blum DA, Ansel LJ. Outcome after fixation of unstable posterior pelvic ring injuries. Clin Orthop Relat Res. 1996;329:160–179.PubMedCrossRef
7.
go back to reference Collinge C, Coons D, Tornetta P, Aschenbrenner J. Standard multiplanar fluoroscopy versus a fluoroscopically based navigation system for the percutaneous insertion of iliosacral screws: a cadaver model. J Orthop Trauma. 2005;19:254–258.PubMedCrossRef Collinge C, Coons D, Tornetta P, Aschenbrenner J. Standard multiplanar fluoroscopy versus a fluoroscopically based navigation system for the percutaneous insertion of iliosacral screws: a cadaver model. J Orthop Trauma. 2005;19:254–258.PubMedCrossRef
8.
go back to reference Day AC, Stott PM, Boden RA. The accuracy of computer-assisted percutaneous iliosacral screw placement. Clin Orthop Relat Res. 2007;463:179–186.PubMed Day AC, Stott PM, Boden RA. The accuracy of computer-assisted percutaneous iliosacral screw placement. Clin Orthop Relat Res. 2007;463:179–186.PubMed
9.
go back to reference Ebraheim NA, Coombs R, Jackson WT, Rusin JJ. Percutaneous computed tomography-guided stabilization of posterior pelvic fractures. Clin Orthop Relat Res. 1994;307:222–228.PubMed Ebraheim NA, Coombs R, Jackson WT, Rusin JJ. Percutaneous computed tomography-guided stabilization of posterior pelvic fractures. Clin Orthop Relat Res. 1994;307:222–228.PubMed
10.
go back to reference Ebraheim NA, Haman SP, Xu R, Stanescu S, Yeasting RA. The lumbosacral nerves in relation to dorsal S1 screw placement and their locations on plain radiographs. Orthopedics. 2000;23:245–247.PubMed Ebraheim NA, Haman SP, Xu R, Stanescu S, Yeasting RA. The lumbosacral nerves in relation to dorsal S1 screw placement and their locations on plain radiographs. Orthopedics. 2000;23:245–247.PubMed
11.
go back to reference Gautier E, Bachler R, Heini PF, Nolte LP. Accuracy of computer-guided screw fixation of the sacroiliac joint. Clin Orthop Relat Res. 2001;393:310–317.PubMedCrossRef Gautier E, Bachler R, Heini PF, Nolte LP. Accuracy of computer-guided screw fixation of the sacroiliac joint. Clin Orthop Relat Res. 2001;393:310–317.PubMedCrossRef
12.
go back to reference Goldberg BA, Lindsey RW, Foglar C, Hedrick TD, Miclau T, Hadad JL. Imaging assessment of sacroiliac screw placement relative to the neuroforamen. Spine. 1998;23:585–589.PubMedCrossRef Goldberg BA, Lindsey RW, Foglar C, Hedrick TD, Miclau T, Hadad JL. Imaging assessment of sacroiliac screw placement relative to the neuroforamen. Spine. 1998;23:585–589.PubMedCrossRef
13.
go back to reference Hinsche AF, Giannoudis PV, Smith RM. Fluoroscopy-based multiplanar image guidance for insertion of sacroiliac screws. Clin Orthop Relat Res. 2002;395:135–144.PubMedCrossRef Hinsche AF, Giannoudis PV, Smith RM. Fluoroscopy-based multiplanar image guidance for insertion of sacroiliac screws. Clin Orthop Relat Res. 2002;395:135–144.PubMedCrossRef
14.
go back to reference Hufner T, Geerling J, Gansslen A, Kendoff D, Citak C, Grutzner P, Krettek C. [Computer-assisted surgery for pelvic injuries] [in German]. Chirurg. 2004;75:961–966.PubMedCrossRef Hufner T, Geerling J, Gansslen A, Kendoff D, Citak C, Grutzner P, Krettek C. [Computer-assisted surgery for pelvic injuries] [in German]. Chirurg. 2004;75:961–966.PubMedCrossRef
15.
go back to reference Keating JF, Werier J, Blachut P, Broekhuyse H, Meek RN, O’Brien PJ. Early fixation of the vertically unstable pelvis: the role of iliosacral screw fixation of the posterior lesion. J Orthop Trauma. 1999;13:107–113.PubMedCrossRef Keating JF, Werier J, Blachut P, Broekhuyse H, Meek RN, O’Brien PJ. Early fixation of the vertically unstable pelvis: the role of iliosacral screw fixation of the posterior lesion. J Orthop Trauma. 1999;13:107–113.PubMedCrossRef
16.
go back to reference Matta JM, Saucedo T. Internal fixation of pelvic ring fractures. Clin Orthop Relat Res. 1989;242:83–97.PubMed Matta JM, Saucedo T. Internal fixation of pelvic ring fractures. Clin Orthop Relat Res. 1989;242:83–97.PubMed
17.
go back to reference Nelson DW, Duwelius PJ. CT-guided fixation of sacral fractures and sacroiliac joint disruptions. Radiology. 1991;180:527–532.PubMed Nelson DW, Duwelius PJ. CT-guided fixation of sacral fractures and sacroiliac joint disruptions. Radiology. 1991;180:527–532.PubMed
18.
go back to reference Rampersaud YR, Foley KT, Shen AC, Williams S, Solomito M. Radiation exposure to the spine surgeon during fluoroscopically assisted pedicle screw insertion. Spine. 2000;25:2637–2645.PubMedCrossRef Rampersaud YR, Foley KT, Shen AC, Williams S, Solomito M. Radiation exposure to the spine surgeon during fluoroscopically assisted pedicle screw insertion. Spine. 2000;25:2637–2645.PubMedCrossRef
19.
go back to reference Routt ML Jr, Kregor PJ, Simonian PT, Mayo KA. Early results of percutaneous iliosacral screws placed with the patient in the supine position. J Orthop Trauma. 1995;9:207–214.PubMedCrossRef Routt ML Jr, Kregor PJ, Simonian PT, Mayo KA. Early results of percutaneous iliosacral screws placed with the patient in the supine position. J Orthop Trauma. 1995;9:207–214.PubMedCrossRef
20.
go back to reference Routt ML Jr, Simonian PT. Closed reduction and percutaneous skeletal fixation of sacral fractures. Clin Orthop Relat Res. 1996;329:121–128.PubMedCrossRef Routt ML Jr, Simonian PT. Closed reduction and percutaneous skeletal fixation of sacral fractures. Clin Orthop Relat Res. 1996;329:121–128.PubMedCrossRef
21.
go back to reference Schep NW, Haverlag R, van Vugt AB. Computer-assisted versus conventional surgery for insertion of 96 cannulated iliosacral screws in patients with postpartum pelvic pain. J Trauma. 2004;57:1299–1302.PubMedCrossRef Schep NW, Haverlag R, van Vugt AB. Computer-assisted versus conventional surgery for insertion of 96 cannulated iliosacral screws in patients with postpartum pelvic pain. J Trauma. 2004;57:1299–1302.PubMedCrossRef
22.
go back to reference Shuler TE, Boone DC, Gruen GS, Peitzman AB. Percutaneous iliosacral screw fixation: early treatment for unstable posterior pelvic ring disruptions. J Trauma. 1995;38:453–458.PubMedCrossRef Shuler TE, Boone DC, Gruen GS, Peitzman AB. Percutaneous iliosacral screw fixation: early treatment for unstable posterior pelvic ring disruptions. J Trauma. 1995;38:453–458.PubMedCrossRef
23.
go back to reference Smith HE, Yuan PS, Sasso R, Papadopolous S, Vaccaro AR. An evaluation of image-guided technologies in the placement of percutaneous iliosacral screws. Spine. 2006;31:234–238.PubMedCrossRef Smith HE, Yuan PS, Sasso R, Papadopolous S, Vaccaro AR. An evaluation of image-guided technologies in the placement of percutaneous iliosacral screws. Spine. 2006;31:234–238.PubMedCrossRef
24.
go back to reference Stephen DJ. Pseudoaneurysm of the superior gluteal arterial system: an unusual cause of pain after a pelvic fracture. J Trauma. 1997;43:146–149.PubMedCrossRef Stephen DJ. Pseudoaneurysm of the superior gluteal arterial system: an unusual cause of pain after a pelvic fracture. J Trauma. 1997;43:146–149.PubMedCrossRef
25.
go back to reference Stöckle U, König B, Hofstetter R, Nolte LP, Haas NP. [Navigation assisted by image conversion: an experimental study on pelvic screw fixation] [in German]. Unfallchirurg. 2001;104:215–220.PubMedCrossRef Stöckle U, König B, Hofstetter R, Nolte LP, Haas NP. [Navigation assisted by image conversion: an experimental study on pelvic screw fixation] [in German]. Unfallchirurg. 2001;104:215–220.PubMedCrossRef
26.
go back to reference Stöckle U, König B, Schäffler A, Zschernack T, Haas NP. [Clinical experience with the Siremobil Iso-C(3D) imaging system in pelvic surgery] [in German]. Unfallchirurg. 2006;109:30–40.PubMedCrossRef Stöckle U, König B, Schäffler A, Zschernack T, Haas NP. [Clinical experience with the Siremobil Iso-C(3D) imaging system in pelvic surgery] [in German]. Unfallchirurg. 2006;109:30–40.PubMedCrossRef
27.
go back to reference Stöckle U, Krettek C, Pohlemann T, Messmer P. Clinical applications: pelvis. Injury. 2004;35(suppl 1):46–56.CrossRef Stöckle U, Krettek C, Pohlemann T, Messmer P. Clinical applications: pelvis. Injury. 2004;35(suppl 1):46–56.CrossRef
28.
go back to reference Templeman D, Schmidt A, Freese J, Weisman I. Proximity of iliosacral screws to neurovascular structures after internal fixation. Clin Orthop Relat Res. 1996;329:194–198.PubMedCrossRef Templeman D, Schmidt A, Freese J, Weisman I. Proximity of iliosacral screws to neurovascular structures after internal fixation. Clin Orthop Relat Res. 1996;329:194–198.PubMedCrossRef
29.
go back to reference Tile M, Pennal GF. Pelvic disruption: principles of management. Clin Orthop Relat Res. 1980;151:56–64.PubMed Tile M, Pennal GF. Pelvic disruption: principles of management. Clin Orthop Relat Res. 1980;151:56–64.PubMed
30.
go back to reference Tonetti J, Carrat L, Blendea S, Merloz P, Troccaz J, Lavallee S, Chirossel JP. Clinical results of percutaneous pelvic surgery: computer assisted surgery using ultrasound compared to standard fluoroscopy. Comput Aided Surg. 2001;6:204–211.PubMedCrossRef Tonetti J, Carrat L, Blendea S, Merloz P, Troccaz J, Lavallee S, Chirossel JP. Clinical results of percutaneous pelvic surgery: computer assisted surgery using ultrasound compared to standard fluoroscopy. Comput Aided Surg. 2001;6:204–211.PubMedCrossRef
31.
go back to reference Tonetti J, Carrat L, Lavallee S, Pittet L, Merloz P, Chirossel JP. Percutaneous iliosacral screw placement using image guided techniques. Clin Orthop Relat Res. 1998;354:103–110.PubMedCrossRef Tonetti J, Carrat L, Lavallee S, Pittet L, Merloz P, Chirossel JP. Percutaneous iliosacral screw placement using image guided techniques. Clin Orthop Relat Res. 1998;354:103–110.PubMedCrossRef
32.
go back to reference van den Bosch EW, van Zwienen CM, van Vugt AB. Fluoroscopic positioning of sacroiliac screws in 88 patients. J Trauma. 2002;53:44–48.PubMedCrossRef van den Bosch EW, van Zwienen CM, van Vugt AB. Fluoroscopic positioning of sacroiliac screws in 88 patients. J Trauma. 2002;53:44–48.PubMedCrossRef
Metadata
Title
Computer-navigated Iliosacral Screw Insertion Reduces Malposition Rate and Radiation Exposure
Authors
Jörn Zwingmann, MD
Gerhard Konrad, MD
Elmar Kotter, MD
Norbert P. Südkamp, Prof
Michael Oberst, MD
Publication date
01-07-2009
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 7/2009
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-008-0632-6

Other articles of this Issue 7/2009

Clinical Orthopaedics and Related Research® 7/2009 Go to the issue

Symposium: Papers Presented at the 2008 Meeting of the Musculoskeletal Infection Society

Periprosthetic Infection Due to Resistant Staphylococci: Serious Problems on the Horizon

Symposium: Papers Presented at the 2008 Meeting of the Musculoskeletal Infection Society

Staged Revision for Knee Arthroplasty Infection: What Is the Role of Serologic Tests Before Reimplantation?