01-02-2008 | International Hip Society Symposium

Dual Mobility Cemented Cups Have Low Dislocation Rates in THA Revisions

Authors: Frantz L. Langlais, MD, Mickaël Ropars, MD, François Gaucher, MD, Thierry Musset, MD, Olivier Chaix, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 2/2008

Login to get access

Abstract

THA revisions using standard cups are at risk of dislocation (5.1% to 14.4% incidence), especially in patients over 70 years of age. Constrained tripolar cups have reduced this risk (6% incidence) but are associated with substantial loosening rates (9%). The nonconstrained dual mobility cup was designed to improve prosthetic stability (polyethylene head ≥ 40 mm diameter) without increasing loosening rates by reducing wear and limiting impingement (rotation range of 108°). We implanted 88 cemented dual mobility cups for THA revisions in 82 patients at high risk of dislocation. Average patient age was 72 years (range, 65–86 years). Eighty-five of the 88 hips were reviewed at 2 to 5 years followup. One patient (1.1%) had a traumatic dislocation at 2 years postoperatively. Two patients (2.3%) had asymptomatic early loosening and three patients (3.5%) had localized radiographic lucencies. These results confirm those with press-fit dual mobility cups suggesting a low dislocation rate at 5 years and a cup survival of 94.6%. At middle term followup, cemented dual mobility cup achieved better results than constrained cups in cases at risk of dislocation and recurrent loosening.
Level of Evidence: Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.
Literature
1.
go back to reference Adam P, Farizon F, Fessy MH. Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2005;91:627–636.PubMed Adam P, Farizon F, Fessy MH. Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2005;91:627–636.PubMed
2.
go back to reference Alberton GM, High WA, Morrey BF. Dislocation after revision total hip arthroplasty: an analysis of risk factors and treatment options. J Bone Joint Surg Am. 2002;84:1788–1792.PubMed Alberton GM, High WA, Morrey BF. Dislocation after revision total hip arthroplasty: an analysis of risk factors and treatment options. J Bone Joint Surg Am. 2002;84:1788–1792.PubMed
3.
go back to reference Beaulé PE, Roussignol X, Schmalzried TP, Udomkiat P, Amstutz HC, Dujardin FH. Tripolar arthroplasty for recurrent total hip prosthesis dislocation [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2003;89:242–249.PubMed Beaulé PE, Roussignol X, Schmalzried TP, Udomkiat P, Amstutz HC, Dujardin FH. Tripolar arthroplasty for recurrent total hip prosthesis dislocation [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2003;89:242–249.PubMed
4.
go back to reference Berend KR, Lombardi AV Jr, Mallory TH, Adams JB, Russell JH, Groseth KL. Long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty. J Arthroplasty. 2005;20(7 Suppl 3):93–102.PubMedCrossRef Berend KR, Lombardi AV Jr, Mallory TH, Adams JB, Russell JH, Groseth KL. Long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty. J Arthroplasty. 2005;20(7 Suppl 3):93–102.PubMedCrossRef
5.
go back to reference Berend KR, Lombardi AV Jr, Welch M, Adams JB. A constrained device with increased range of motion prevents early dislocation. Clin Orthop Relat Res. 2006;447:70–75.PubMedCrossRef Berend KR, Lombardi AV Jr, Welch M, Adams JB. A constrained device with increased range of motion prevents early dislocation. Clin Orthop Relat Res. 2006;447:70–75.PubMedCrossRef
6.
go back to reference Bremner RB, Goetz DO, Callaghan JJ, Capello WN, Johnston RC. Use of constrained acetabular components for hip instability: an average 10-year follow-up study. J Arthroplasty. 2003;18(7 Suppl 1):131–137.PubMedCrossRef Bremner RB, Goetz DO, Callaghan JJ, Capello WN, Johnston RC. Use of constrained acetabular components for hip instability: an average 10-year follow-up study. J Arthroplasty. 2003;18(7 Suppl 1):131–137.PubMedCrossRef
7.
go back to reference Brytröm S, Espehaug B, Furnes O, Havelin LI. Femoral size is a risk factor for total hip dislocation (a study of 42987 primary THR from the Norwegian Arthroplasty Register). Acta Orthop Scand. 2003;74:514–524.CrossRef Brytröm S, Espehaug B, Furnes O, Havelin LI. Femoral size is a risk factor for total hip dislocation (a study of 42987 primary THR from the Norwegian Arthroplasty Register). Acta Orthop Scand. 2003;74:514–524.CrossRef
8.
go back to reference Callaghan JJ, O’Rourke MR, Goetz DD, Lewallen DG, Johnston RC, Capello WN. Use of a constrained tripolar acetabular liner to treat intraoperative instability and postoperative dislocation after total hip arthroplasty. Clin Orthop Relat Res. 2004;429:117–123.PubMedCrossRef Callaghan JJ, O’Rourke MR, Goetz DD, Lewallen DG, Johnston RC, Capello WN. Use of a constrained tripolar acetabular liner to treat intraoperative instability and postoperative dislocation after total hip arthroplasty. Clin Orthop Relat Res. 2004;429:117–123.PubMedCrossRef
9.
go back to reference Cooke CC, Hozack W, Lavernia C, Sharkey P, Shastri S, Rothman RH. Early failure mechanisms of constrained tripolar acetabular sockets used in revision total hip arthroplasty. J Arthroplasty. 2003;18:827–833.PubMedCrossRef Cooke CC, Hozack W, Lavernia C, Sharkey P, Shastri S, Rothman RH. Early failure mechanisms of constrained tripolar acetabular sockets used in revision total hip arthroplasty. J Arthroplasty. 2003;18:827–833.PubMedCrossRef
10.
go back to reference Crowninshield RD, Maloney WJ, Wentz DH, Humphrey SM, Blanchard CR. Biomechanics of large femoral heads: what they do and don’t do. Clin Orthop Relat Res. 2004;429:102–107.PubMedCrossRef Crowninshield RD, Maloney WJ, Wentz DH, Humphrey SM, Blanchard CR. Biomechanics of large femoral heads: what they do and don’t do. Clin Orthop Relat Res. 2004;429:102–107.PubMedCrossRef
11.
go back to reference D’Antonio JA, Capello WN, Borden LS, Bargar WL, Bierbaum BF, Boettcher WG, Steinberg ME, Stulberg SD, Wedge JH. Classification and management of acetabular abnormalities in total hip arthroplasty. Clin Orthop Relat Res. 1989;243:126–137.PubMed D’Antonio JA, Capello WN, Borden LS, Bargar WL, Bierbaum BF, Boettcher WG, Steinberg ME, Stulberg SD, Wedge JH. Classification and management of acetabular abnormalities in total hip arthroplasty. Clin Orthop Relat Res. 1989;243:126–137.PubMed
12.
go back to reference DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed
13.
go back to reference Farizon F, de Lavison R, Azoulai JJ, Bousquet G. Results with a cementless alumina-coated cup wear mobility. Int Orthop. 1998;22:219–224.PubMedCrossRef Farizon F, de Lavison R, Azoulai JJ, Bousquet G. Results with a cementless alumina-coated cup wear mobility. Int Orthop. 1998;22:219–224.PubMedCrossRef
14.
go back to reference Geller JA, Malchau H, Bragdon C, Greene M, Harris WH, Freiberg AA. Large diameter femoral heads on highly cross-linked polyethylene: minimum 3-year results. Clin Orthop Relat Res. 2006;447:53–59.PubMedCrossRef Geller JA, Malchau H, Bragdon C, Greene M, Harris WH, Freiberg AA. Large diameter femoral heads on highly cross-linked polyethylene: minimum 3-year results. Clin Orthop Relat Res. 2006;447:53–59.PubMedCrossRef
15.
go back to reference Goetz DD, Bremner BR, Callaghan JJ, Capello WN, Johnston RC. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. J Bone Joint Surg Am. 2004;86:2419–2423.PubMed Goetz DD, Bremner BR, Callaghan JJ, Capello WN, Johnston RC. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. J Bone Joint Surg Am. 2004;86:2419–2423.PubMed
16.
go back to reference Harris WH, McCarthy JC Jr, O’Neill DA. Femoral component loosening using contemporary techniques of femoral cement fixation. J Bone Joint Surg Am. 1982;64:1063–1067.PubMed Harris WH, McCarthy JC Jr, O’Neill DA. Femoral component loosening using contemporary techniques of femoral cement fixation. J Bone Joint Surg Am. 1982;64:1063–1067.PubMed
17.
go back to reference Jolles BM, Zangger P, Leyvraz PF. Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty. 2002;17:282–288.PubMedCrossRef Jolles BM, Zangger P, Leyvraz PF. Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty. 2002;17:282–288.PubMedCrossRef
18.
go back to reference Kerboull M, Hamadouche M, Kerboull L. The Kerboull acetabular reinforcement device in major acetabular reconstructions. Clin Orthop Relat Res. 2000;378:155–168.PubMedCrossRef Kerboull M, Hamadouche M, Kerboull L. The Kerboull acetabular reinforcement device in major acetabular reconstructions. Clin Orthop Relat Res. 2000;378:155–168.PubMedCrossRef
19.
go back to reference Khatod M, Barber T, Paxton E, Namba R, Fithian D. An analysis of the risk of hip dislocation with a contemporary total joint registry. Clin Orthop Relat Res. 2006;447:19–23.PubMedCrossRef Khatod M, Barber T, Paxton E, Namba R, Fithian D. An analysis of the risk of hip dislocation with a contemporary total joint registry. Clin Orthop Relat Res. 2006;447:19–23.PubMedCrossRef
20.
go back to reference Lagrange J, Letournel E, Pupin P. Review of patients with total hip replacement after a period of more than ten years (author’s transl) [in French]. Chirurgie. 1979;105:733–737.PubMed Lagrange J, Letournel E, Pupin P. Review of patients with total hip replacement after a period of more than ten years (author’s transl) [in French]. Chirurgie. 1979;105:733–737.PubMed
21.
go back to reference Langlais F, Belot N, Ropars M, Thomazeau H, Lambotte JC, Cathelineau G. Antibiotic cements in articular prostheses: current orthopaedic concepts. Int J Antimicrob Agents. 2006;28:84–89.PubMedCrossRef Langlais F, Belot N, Ropars M, Thomazeau H, Lambotte JC, Cathelineau G. Antibiotic cements in articular prostheses: current orthopaedic concepts. Int J Antimicrob Agents. 2006;28:84–89.PubMedCrossRef
22.
go back to reference Langlais F, Kerboull M, Sedel L, Ling RS. The “French paradox.”. J Bone Joint Surg Br. 2003;85:17–20.PubMedCrossRef Langlais F, Kerboull M, Sedel L, Ling RS. The “French paradox.”. J Bone Joint Surg Br. 2003;85:17–20.PubMedCrossRef
23.
go back to reference Langlais F, Lambotte JC, Collin PH, Langlois F, Fontaine JW, Thomazeau H. Trochanteric slide osteotomy in revision total hip arthroplasty for loosenings. J Bone Joint Surg Br. 2003,85:510–516.PubMedCrossRef Langlais F, Lambotte JC, Collin PH, Langlois F, Fontaine JW, Thomazeau H. Trochanteric slide osteotomy in revision total hip arthroplasty for loosenings. J Bone Joint Surg Br. 2003,85:510–516.PubMedCrossRef
24.
go back to reference Leclercq S, el Blidi S, Aubriot JH. Bousquet’s device in the treatment of recurrent dislocation of a total hip prosthesis: apropos of 13 cases [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1995;81:389–394.PubMed Leclercq S, el Blidi S, Aubriot JH. Bousquet’s device in the treatment of recurrent dislocation of a total hip prosthesis: apropos of 13 cases [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1995;81:389–394.PubMed
25.
go back to reference Lecuire F, Benareau I, Rubini J, Basso M. Intra-prosthetic dislocation of the Bousquet dual mobility socket [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2004;90:249–255.PubMed Lecuire F, Benareau I, Rubini J, Basso M. Intra-prosthetic dislocation of the Bousquet dual mobility socket [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2004;90:249–255.PubMed
26.
go back to reference Lunn JV, Kearns SS, Quinlan W, Murray P, Byrne JO. Impaction allografting and the Kerboull acetabular reinforcement device: 35 hips followed for 3–7 years. Acta Orthop. 2005;76:296–302.PubMed Lunn JV, Kearns SS, Quinlan W, Murray P, Byrne JO. Impaction allografting and the Kerboull acetabular reinforcement device: 35 hips followed for 3–7 years. Acta Orthop. 2005;76:296–302.PubMed
27.
go back to reference Merle d’Aubigné R, Postel M. Functional results of hip arthroplasty with acrylic prosthesis. J Bone Joint Surg Am. 1954;36:451–475. Merle d’Aubigné R, Postel M. Functional results of hip arthroplasty with acrylic prosthesis. J Bone Joint Surg Am. 1954;36:451–475.
28.
go back to reference Muratoglu OK, Bragdon CR, O’Connor D, Perinchief RS, Estok DM 2nd, Jasty M, Harris WH. Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. J Arthroplasty. 2001;16(8 Suppl 1):24–30.PubMedCrossRef Muratoglu OK, Bragdon CR, O’Connor D, Perinchief RS, Estok DM 2nd, Jasty M, Harris WH. Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. J Arthroplasty. 2001;16(8 Suppl 1):24–30.PubMedCrossRef
29.
go back to reference Peraldi P, Vandenbussche E, Augereau B. Bad clinical results of cemented caps with metal-backed acetabular components: 124 cases with 21 months follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1997;83:561–568.PubMed Peraldi P, Vandenbussche E, Augereau B. Bad clinical results of cemented caps with metal-backed acetabular components: 124 cases with 21 months follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1997;83:561–568.PubMed
30.
go back to reference Philippot R, Adam P, Farizon F, Fessy MH, Bousquet G. Survival of cementless dual mobility sockets: ten-year follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2006;92:326–331.PubMed Philippot R, Adam P, Farizon F, Fessy MH, Bousquet G. Survival of cementless dual mobility sockets: ten-year follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2006;92:326–331.PubMed
31.
go back to reference Phillips CB, Barrett JA, Losina E, Mahomed NN, Lingard EA, Guadagnoli E, Baron JA, Harris WH, Poss R, Katz JN. Incidence rates of dislocation, pulmonary embolism, and deep infection during the fist six months after elective total hip replacement. J Bone Joint Surg Am. 2003;85:20–26.PubMedCrossRef Phillips CB, Barrett JA, Losina E, Mahomed NN, Lingard EA, Guadagnoli E, Baron JA, Harris WH, Poss R, Katz JN. Incidence rates of dislocation, pulmonary embolism, and deep infection during the fist six months after elective total hip replacement. J Bone Joint Surg Am. 2003;85:20–26.PubMedCrossRef
32.
go back to reference Ritter MA. The cemented acetabular component of a total hip replacement: all polyethylene versus metal backing. Clin Orthop Relat Res. 1995;311:69–75.PubMed Ritter MA. The cemented acetabular component of a total hip replacement: all polyethylene versus metal backing. Clin Orthop Relat Res. 1995;311:69–75.PubMed
33.
go back to reference Shapiro GS, Weiland DE, Markel DC, Padgett DE, Sculco TP, Pellicci PM. The use of constrained acetabular component for recurrent dislocation. J Arthroplasty. 2003;18:250–258.PubMedCrossRef Shapiro GS, Weiland DE, Markel DC, Padgett DE, Sculco TP, Pellicci PM. The use of constrained acetabular component for recurrent dislocation. J Arthroplasty. 2003;18:250–258.PubMedCrossRef
34.
go back to reference Shrader MW, Parvizi J, Lewallen DG. The use of a constrained acetabular component to treat instability after THR. J Bone Joint Surg Am. 2003;85:2179–2183.PubMed Shrader MW, Parvizi J, Lewallen DG. The use of a constrained acetabular component to treat instability after THR. J Bone Joint Surg Am. 2003;85:2179–2183.PubMed
35.
go back to reference Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64:1295–1306.PubMed Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64:1295–1306.PubMed
36.
go back to reference Zahiri CA, Schmalzried TP, Ebramzadeh E, Szuszczewicz ES, Salib D, Kim C, Amstutz HC. Lessons learned from loosening of the McKee-Farrar metal-on-metal total hip replacement. J Arthroplasty. 1999;14:326–332.PubMedCrossRef Zahiri CA, Schmalzried TP, Ebramzadeh E, Szuszczewicz ES, Salib D, Kim C, Amstutz HC. Lessons learned from loosening of the McKee-Farrar metal-on-metal total hip replacement. J Arthroplasty. 1999;14:326–332.PubMedCrossRef
Metadata
Title
Dual Mobility Cemented Cups Have Low Dislocation Rates in THA Revisions
Authors
Frantz L. Langlais, MD
Mickaël Ropars, MD
François Gaucher, MD
Thierry Musset, MD
Olivier Chaix, MD
Publication date
01-02-2008
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 2/2008
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-007-0047-9

Other articles of this Issue 2/2008

Clinical Orthopaedics and Related Research® 2/2008 Go to the issue

Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society

Minimum 10-year Survival of Kerboull Cemented Stems According to Surface Finish

Symposium: Papers Presented at the Closed Meeting of the International Hip Society

Long-term Radiographic Assessment of Cemented Polyethylene Acetabular Cups

Symposium: Papers Presented at the Closed Meeting of the International Hip Society

Papers Presented at the Closed Meeting of the International Hip Society