A strategy to optimize radiation exposure for non-contrast head CT: comparison with the Japanese diagnostic reference levels

Authors: Kanako K. Kumamaru, Yosuke Kogure, Michimasa Suzuki, Masaaki Hori, Atsushi Nakanishi, Koji Kamagata, Akifumi Hagiwara, Christina Andica, Keiken Ri, Naoyoshi Houshido, Shigeki Aoki

Published in: Japanese Journal of Radiology | Issue 6/2016

Login to get access

Abstract

Purpose

To describe how we performed a protocol review, analyzed data, identified opportunities to reduce radiation exposure, and then implemented a new imaging protocol for non-contrast adult head CT at our institution with reduced radiation exposure, using the Japanese diagnostic reference levels (DRLs) as the reference.

Methods

After analyzing the CT dose index (CTDIvol) and dose length product (DLP) in all non-contrast adult head CT examinations performed during a 3-month period (September to November 2015) in order to identify a specific protocol that contributed to the above-DRL-level radiation exposure observed for non-contrast adult head CT at our institution, phantom studies with objective and subjective image quality analyses were performed to develop a new imaging protocol. After implementing the new protocol, its feasibility was evaluated in terms of radiation exposure, prevalence of significant disease detection, and subjective image quality.

Results

The review of 2040 examinations revealed that a helical protocol (CTDIvol = 93.4 mGy) with one of four CT scanners mainly contributed to the above-DRL-level radiation exposure (mean DLP at this scanner = 1401.2 mGy cm) in non-contrast adult head CT at our institution. To replace this protocol, the phantom study identified a wide-volume scan using 120 kVp, 350 mAs, a 4-cm detector, a slice thickness of 5 mm, and a CTDIvol of 69.8 mGy as a new protocol that yielded comparable image quality to the existing protocol. After the implementation of the new protocol, the overall mean DLP reduced to 1365 mGy cm without any apparent degradation of image quality. No significant decrease in the prevalence of significant findings after protocol revision was noted.

Conclusion

We report the successful implementation of a new protocol with reduced radiation exposure for non-contrast head CT examinations.
Literature
1.
go back to reference International Commission on Radiological Protection. Radiological protection and safety in medicine. A report of the International Commission on Radiological Protection. Ann ICRP. 1996;26:1–47. International Commission on Radiological Protection. Radiological protection and safety in medicine. A report of the International Commission on Radiological Protection. Ann ICRP. 1996;26:1–47.
3.
go back to reference Nishizawa K, Matsumoto M, Iwai K, Maruyama T. Survey of CT practice in Japan and collective effective dose estimation. Nihon Igaku Hoshasen Gakkai Zasshi Nippon Acta Radiologica. 2004;64:151–8. Nishizawa K, Matsumoto M, Iwai K, Maruyama T. Survey of CT practice in Japan and collective effective dose estimation. Nihon Igaku Hoshasen Gakkai Zasshi Nippon Acta Radiologica. 2004;64:151–8.
6.
go back to reference McCollough C, Branham T, Herlihy V, Bhargavan M, Robbins L, Bush K, McNitt-Gray M, Payne JT, Ruckdeschel T, Pfeiffer D, Cody D, Zeman R. Diagnostic reference levels from the ACR CT Accreditation Program. J Am Coll Radiol. 2011;8(11):795–803.CrossRefPubMed McCollough C, Branham T, Herlihy V, Bhargavan M, Robbins L, Bush K, McNitt-Gray M, Payne JT, Ruckdeschel T, Pfeiffer D, Cody D, Zeman R. Diagnostic reference levels from the ACR CT Accreditation Program. J Am Coll Radiol. 2011;8(11):795–803.CrossRefPubMed
7.
go back to reference Cohnen M, Fischer H, Hamacher J, Lins E, Kotter R, Modder U. CT of the head by use of reduced current and kilovoltage: relationship between image quality and dose reduction. AJNR Am J Neuroradiol. 2000;21:1654–60.PubMed Cohnen M, Fischer H, Hamacher J, Lins E, Kotter R, Modder U. CT of the head by use of reduced current and kilovoltage: relationship between image quality and dose reduction. AJNR Am J Neuroradiol. 2000;21:1654–60.PubMed
Metadata
Title
A strategy to optimize radiation exposure for non-contrast head CT: comparison with the Japanese diagnostic reference levels
Authors
Kanako K. Kumamaru
Yosuke Kogure
Michimasa Suzuki
Masaaki Hori
Atsushi Nakanishi
Koji Kamagata
Akifumi Hagiwara
Christina Andica
Keiken Ri
Naoyoshi Houshido
Shigeki Aoki
Publication date
01-06-2016
Publisher
Springer Japan
Published in
Japanese Journal of Radiology / Issue 6/2016
Print ISSN: 1867-1071
Electronic ISSN: 1867-108X
DOI
https://doi.org/10.1007/s11604-016-0545-3

Other articles of this Issue 6/2016

Japanese Journal of Radiology 6/2016 Go to the issue