Published in: Japanese Journal of Radiology 5/2013

01-05-2013 | Original Article

Measurement consistency of MR elastography at 3.0 T: comparison among three different region-of-interest placement methods

Authors: Toshimichi Mitsufuji, Yoshinobu Shinagawa, Ritsuko Fujimitsu, Hiroshi Urakawa, Kimiyo Inoue, Koichi Takano, Kengo Yoshimitsu

Published in: Japanese Journal of Radiology | Issue 5/2013

Login to get access

Abstract

Purpose

To test inter- and intraobserver consistency of liver stiffness measurement on MR elastography (MRE) at 3.0 T.

Materials and methods

Two abdominal radiologists independently measured stiffness of the liver on MRE in three volunteers and seven patients with chronic liver diseases using three different region-of-interest (ROI) placement methods. Methods 1 and 2 involved placing circular and free-hand-drawn ROIs, respectively, visually referring to anatomical (three-dimensional T1-weighted) and wave images. Method 3 involved placing ROIs on the fused images of MRE and anatomical images developed on a work station, visually referring to wave images. The inter- and intraobserver consistency was assessed with regression and Bland–Altman analysis.

Results

Thirty-eight images were available for measurement in total. As for interobserver consistency, method 3 showed the best regression coefficient, correlation coefficient, and y intercept. The absolute values of the interobserver differences for method 3 were significantly smaller than those of method 1 or method 2 (p < 0.05, each). Intraobserver consistency of method 3 was excellent for both observers.

Conclusion

Stiffness measurement of the liver on MRE performed with the fusion method at 3.0 T provides the highest inter- and intraobserver consistency.
Literature
1.
go back to reference Rouviere O, Yin M, Dresner MA, et al. MR elastography of the liver: preliminary results. Radiology. 2006;240:440–8.PubMedCrossRef Rouviere O, Yin M, Dresner MA, et al. MR elastography of the liver: preliminary results. Radiology. 2006;240:440–8.PubMedCrossRef
2.
go back to reference Huwart L, Sempoux C, Salameh N, et al. Liver fibrosis: noninvasive assessment with MR elastography versus aspartate aminotransferase-to-platelet ratio index. Radiology. 2007;245:458–66.PubMedCrossRef Huwart L, Sempoux C, Salameh N, et al. Liver fibrosis: noninvasive assessment with MR elastography versus aspartate aminotransferase-to-platelet ratio index. Radiology. 2007;245:458–66.PubMedCrossRef
3.
go back to reference Klatt D, Asbach P, Rump J, et al. In vivo determination of hepatic stiffness using steady-state free precession magnetic resonance elastography. Invest Radiol. 2006;41:841–8.PubMedCrossRef Klatt D, Asbach P, Rump J, et al. In vivo determination of hepatic stiffness using steady-state free precession magnetic resonance elastography. Invest Radiol. 2006;41:841–8.PubMedCrossRef
4.
go back to reference Huwart L, Sempoux C, Vicaut E, et al. Magnetic resonance elastography for the noninvasive staging of liver fibrosis. Gastroenterology. 2008;135:32–40.PubMedCrossRef Huwart L, Sempoux C, Vicaut E, et al. Magnetic resonance elastography for the noninvasive staging of liver fibrosis. Gastroenterology. 2008;135:32–40.PubMedCrossRef
5.
go back to reference Motosugi U, Ichikawa T, Sou H, et al. Magnetic resonance elastography of the liver: preliminary results and estimation of interrater reliability. Jpn J Radiol. 2010;28:623–7.PubMedCrossRef Motosugi U, Ichikawa T, Sou H, et al. Magnetic resonance elastography of the liver: preliminary results and estimation of interrater reliability. Jpn J Radiol. 2010;28:623–7.PubMedCrossRef
6.
go back to reference Asbach P, Klatt D, Schlosser B, et al. Viscoelasticity-based staging of hepatic fibrosis with multifrequency MR elastography. Radiology. 2010;257:80–6.PubMedCrossRef Asbach P, Klatt D, Schlosser B, et al. Viscoelasticity-based staging of hepatic fibrosis with multifrequency MR elastography. Radiology. 2010;257:80–6.PubMedCrossRef
7.
go back to reference Shire NJ, Yi M, Chen J, et al. Test–retest repeatability of MR elastography for noninvasive liver fibrosis assessment in hepatitis C. J Magn Reson Imaging. 2011;34:947–55.PubMedCrossRef Shire NJ, Yi M, Chen J, et al. Test–retest repeatability of MR elastography for noninvasive liver fibrosis assessment in hepatitis C. J Magn Reson Imaging. 2011;34:947–55.PubMedCrossRef
8.
go back to reference Hines CDG, Bley TA, Lindstrom MJ, Reeder SB. Repeatability of magnetic resonance elastography for quantification of hepatic stiffness. J Magn Reson Imaging. 2010;31:725–31.PubMedCrossRef Hines CDG, Bley TA, Lindstrom MJ, Reeder SB. Repeatability of magnetic resonance elastography for quantification of hepatic stiffness. J Magn Reson Imaging. 2010;31:725–31.PubMedCrossRef
9.
go back to reference Kim BH, Lee JM, Lee YJ, et al. MR elastography for noninvasive assessment of hepatic fibrosis: experience from a tertiary center in Asia. J Magn Reson Imaging. 2011;34:1110–6.PubMedCrossRef Kim BH, Lee JM, Lee YJ, et al. MR elastography for noninvasive assessment of hepatic fibrosis: experience from a tertiary center in Asia. J Magn Reson Imaging. 2011;34:1110–6.PubMedCrossRef
10.
go back to reference Lee DH, Lee JM, Han JK, Choi BI. MR elastography of healthy liver parenchyma: normal value and reliability of the liver stiffness value measurement. JMRI, 2013 [pEpub ahead of print]. Lee DH, Lee JM, Han JK, Choi BI. MR elastography of healthy liver parenchyma: normal value and reliability of the liver stiffness value measurement. JMRI, 2013 [pEpub ahead of print].
11.
go back to reference Sonn GA, Natarajan S, Margolis DJ, et al. Targeted biopsy in the detection of prostate cancer using an office based magnetic resonance ultrasound fusion device. J Urol. 2012;189:86–9.PubMedCrossRef Sonn GA, Natarajan S, Margolis DJ, et al. Targeted biopsy in the detection of prostate cancer using an office based magnetic resonance ultrasound fusion device. J Urol. 2012;189:86–9.PubMedCrossRef
12.
go back to reference Nishie A, Stolpen AH, Obuchi M, Kuehn DM, Dagit A, Andresen K. Evaluation of locally recurrent pelvic malignancy: performance of T2- and diffusion-weighted MRI with image fusion. J Magn Reson Imaging. 2008;28(3):705–13.PubMedCrossRef Nishie A, Stolpen AH, Obuchi M, Kuehn DM, Dagit A, Andresen K. Evaluation of locally recurrent pelvic malignancy: performance of T2- and diffusion-weighted MRI with image fusion. J Magn Reson Imaging. 2008;28(3):705–13.PubMedCrossRef
Metadata
Title
Measurement consistency of MR elastography at 3.0 T: comparison among three different region-of-interest placement methods
Authors
Toshimichi Mitsufuji
Yoshinobu Shinagawa
Ritsuko Fujimitsu
Hiroshi Urakawa
Kimiyo Inoue
Koichi Takano
Kengo Yoshimitsu
Publication date
01-05-2013
Publisher
Springer Japan
Published in
Japanese Journal of Radiology / Issue 5/2013
Print ISSN: 1867-1071
Electronic ISSN: 1867-108X
DOI
https://doi.org/10.1007/s11604-013-0195-7

Other articles of this Issue 5/2013

Japanese Journal of Radiology 5/2013 Go to the issue