Guided ultrasound calibration: where, how, and how many calibration fiducials

Authors: Elvis C. S. Chen, Terry M. Peters, Burton Ma

Published in: International Journal of Computer Assisted Radiology and Surgery | Issue 6/2016

Login to get access

Abstract

Purpose

Many image-guided interventions rely on tracked ultrasound where the transducer is augmented with a tracking device. The relationship between the ultrasound image coordinate system and the tracking sensor must be determined accurately via probe calibration. We introduce a novel calibration framework guided by the prediction of target registration error (TRE): Between successive measurements of the calibration phantom, our framework guides the user in choosing the pose of the calibration phantom by optimizing TRE.

Methods

We introduced an oriented line calibration phantom and modeled the ultrasound calibration process as a point-to-line registration problem. We then derived a spatial stiffness model of point-to-line registration for estimating TRE magnitude at any target. Assuming isotropic, identical localization error, we used the model to estimate TRE for each pixel using the current calibration estimate. We then searched through the calibration tool space to find the pose for the next fiducial which maximally minimized TRE.

Results

Both simulation and experimental results suggested that TRE decreases monotonically, reaching an asymptote when a sufficient number of measurements (typically around 12) are made. Independent point reconstruction accuracy assessment showed sub-millimeter accuracy of the calibration framework.

Conclusion

We have introduced the first TRE-guided ultrasound calibration framework. Using a hollow straw as an oriented line phantom, we virtually constructed a rigid lines phantom and modeled the calibration process as a point-to-line registration. Highly accurate calibration was achieved with minimal measurements by using a spatial stiffness model of TRE to strategically choose the pose of the calibration phantom between successive measurements.
Literature
1.
go back to reference Ackerman M, Cheng A, Boctor E, Chirikjian G (2014) Online ultrasound sensor calibration using gradient descent on the Euclidean group. In: Robotics and automation (ICRA), 2014 IEEE international conference on, pp 4900–4905 Ackerman M, Cheng A, Boctor E, Chirikjian G (2014) Online ultrasound sensor calibration using gradient descent on the Euclidean group. In: Robotics and automation (ICRA), 2014 IEEE international conference on, pp 4900–4905
2.
go back to reference Ameri G, McLeod AJ, Baxter JSH, Chen ECS, Peters TM (2015) Line fiducial material and thickness considerations for ultrasound calibration. In: Proceedings of the SPIE, vol 9415 Ameri G, McLeod AJ, Baxter JSH, Chen ECS, Peters TM (2015) Line fiducial material and thickness considerations for ultrasound calibration. In: Proceedings of the SPIE, vol 9415
3.
go back to reference Besl P, McKay ND (1992) A method for registration of 3-D shapes. Pattern analysis and machine intelligence. IEEE Trans 14(2):239–256 Besl P, McKay ND (1992) A method for registration of 3-D shapes. Pattern analysis and machine intelligence. IEEE Trans 14(2):239–256
4.
go back to reference Chen EC, McLeod AJ, Baxter JS, Peters TM (2015) An iterative closest point framework for ultrasound calibration. In: Augmented environments for computer-assisted interventions, vol 9365, Springer International Publishing, pp 69–79 Chen EC, McLeod AJ, Baxter JS, Peters TM (2015) An iterative closest point framework for ultrasound calibration. In: Augmented environments for computer-assisted interventions, vol 9365, Springer International Publishing, pp 69–79
5.
go back to reference Chen EC, McLeod AJ, Baxter JS, Peters TM (2015) Registration of 3D shapes under anisotropic scaling. Int J Comput Assist Radiol Surg 10(6):867–878CrossRefPubMed Chen EC, McLeod AJ, Baxter JS, Peters TM (2015) Registration of 3D shapes under anisotropic scaling. Int J Comput Assist Radiol Surg 10(6):867–878CrossRefPubMed
6.
go back to reference Chen TK, Thurston AD, Ellis RE, Abolmaesumi P (2009) A real-time freehand ultrasound calibration system with automatic accuracy feedback and control. Ultrasound Med Biol 35(1):79–93CrossRefPubMed Chen TK, Thurston AD, Ellis RE, Abolmaesumi P (2009) A real-time freehand ultrasound calibration system with automatic accuracy feedback and control. Ultrasound Med Biol 35(1):79–93CrossRefPubMed
7.
go back to reference Comeau RM, Fenster A, Peters TM (1998) Integrated MR and ultrasound imaging for improved image guidance in neurosurgery. In: Proceedings of the SPIE, vol 3338, pp 747–754 Comeau RM, Fenster A, Peters TM (1998) Integrated MR and ultrasound imaging for improved image guidance in neurosurgery. In: Proceedings of the SPIE, vol 3338, pp 747–754
8.
go back to reference Detmer PR, Bashein G, Hodges T, Beach KW, Filer EP, Burns DH, Strandness DE Jr (1994) 3D ultrasonic image feature localization based on magnetic scanhead tracking: in vitro calibration and validation. Ultrasound Med Biol 20:923–936CrossRefPubMed Detmer PR, Bashein G, Hodges T, Beach KW, Filer EP, Burns DH, Strandness DE Jr (1994) 3D ultrasonic image feature localization based on magnetic scanhead tracking: in vitro calibration and validation. Ultrasound Med Biol 20:923–936CrossRefPubMed
9.
go back to reference Fitzpatrick J, West J, Jr Maurer CR (1998) Predicting error in rigid-body point-based registration. Med Imaging IEEE Trans 17(5):694–702CrossRef Fitzpatrick J, West J, Jr Maurer CR (1998) Predicting error in rigid-body point-based registration. Med Imaging IEEE Trans 17(5):694–702CrossRef
10.
go back to reference Gower JC, Dijksterhuis GB (2004) Procrustes problems. Oxford University Press, OxfordCrossRef Gower JC, Dijksterhuis GB (2004) Procrustes problems. Oxford University Press, OxfordCrossRef
11.
go back to reference Guo X, Cheng A, Zhang H, Kang HJ, Etienne-Cummings R, Boctor E (2014) Active echo: A new paradigm for ultrasound calibration. In: Medical image computing and computer-assisted intervention, Springer, pp 397–404 Guo X, Cheng A, Zhang H, Kang HJ, Etienne-Cummings R, Boctor E (2014) Active echo: A new paradigm for ultrasound calibration. In: Medical image computing and computer-assisted intervention, Springer, pp 397–404
12.
go back to reference Horn BK (1987) Closed-form solution of absolute orientation using unit quaternions. J Opt Soc Am 4:629–642CrossRef Horn BK (1987) Closed-form solution of absolute orientation using unit quaternions. J Opt Soc Am 4:629–642CrossRef
13.
go back to reference Hsu PW, Prager RW, Gee AH, Treece GM (2009) Freehand 3D ultrasound calibration: a review. In: Advanced imaging in biology and medicine, Springer, Berlin, pp 47–84 Hsu PW, Prager RW, Gee AH, Treece GM (2009) Freehand 3D ultrasound calibration: a review. In: Advanced imaging in biology and medicine, Springer, Berlin, pp 47–84
14.
go back to reference Hsu PW, Treece GM, Prager RW, Houghton NE, Gee AH (2008) Comparison of freehand 3-D ultrasound calibration techniques using a stylus. Ultrasound Med Biol 34(10):1610–1621CrossRefPubMed Hsu PW, Treece GM, Prager RW, Houghton NE, Gee AH (2008) Comparison of freehand 3-D ultrasound calibration techniques using a stylus. Ultrasound Med Biol 34(10):1610–1621CrossRefPubMed
15.
go back to reference Khamene A, Sauer F (2005) A novel phantom-less spatial and temporal ultrasound calibration method. In: Medical image computing and computer-assisted intervention, Springer, pp 65–72 Khamene A, Sauer F (2005) A novel phantom-less spatial and temporal ultrasound calibration method. In: Medical image computing and computer-assisted intervention, Springer, pp 65–72
16.
go back to reference Ma B, Ellis RE (2003) Robust registration for computer-integrated orthopedic surgery: laboratory validation and clinical experience. Med Image Anal 7(3):237–250CrossRefPubMed Ma B, Ellis RE (2003) Robust registration for computer-integrated orthopedic surgery: laboratory validation and clinical experience. Med Image Anal 7(3):237–250CrossRefPubMed
17.
go back to reference Ma B, Ellis RE (2004) Spatial-stiffness analysis of surface-based registration. In: Medical image computing and computer-assisted intervention, Springer, pp 623–630 Ma B, Ellis RE (2004) Spatial-stiffness analysis of surface-based registration. In: Medical image computing and computer-assisted intervention, Springer, pp 623–630
18.
go back to reference Ma B, Moghari M, Ellis R, Abolmaesumi P (2010) Estimation of optimal fiducial target registration error in the presence of heteroscedastic noise. Med Imaging IEEE Trans 29(3):708–723CrossRef Ma B, Moghari M, Ellis R, Abolmaesumi P (2010) Estimation of optimal fiducial target registration error in the presence of heteroscedastic noise. Med Imaging IEEE Trans 29(3):708–723CrossRef
19.
go back to reference Ma B, Peters TM, Chen ECS (2016) Estimation of line-based target registration error. Proc SPIE 9786:978626-1–978626-9 Ma B, Peters TM, Chen ECS (2016) Estimation of line-based target registration error. Proc SPIE 9786:978626-1–978626-9
20.
go back to reference Mercier L, Langø T, Lindseth F, Collins LD (2005) A review of calibration techniques for freehand 3-D ultrasound systems. Ultrasound Med Biol 31(2):143–165CrossRefPubMed Mercier L, Langø T, Lindseth F, Collins LD (2005) A review of calibration techniques for freehand 3-D ultrasound systems. Ultrasound Med Biol 31(2):143–165CrossRefPubMed
21.
go back to reference Muratore DM, Galloway RL Jr (2001) Beam calibration without a phantom for creating a 3-D freehand ultrasound system. Ultrasound Med Biol 27(11):1557–1566CrossRefPubMed Muratore DM, Galloway RL Jr (2001) Beam calibration without a phantom for creating a 3-D freehand ultrasound system. Ultrasound Med Biol 27(11):1557–1566CrossRefPubMed
22.
go back to reference Prager RW, Rohling RN, Gee AH, Berman L (1998) Rapid calibration for 3-D freehand ultrasound. Ultrasound Med Biol 24(6):855–869CrossRefPubMed Prager RW, Rohling RN, Gee AH, Berman L (1998) Rapid calibration for 3-D freehand ultrasound. Ultrasound Med Biol 24(6):855–869CrossRefPubMed
23.
go back to reference Schönemann PH (1966) A generalized solution of the orthogonal procrustes problem. Psychometrika 31(1):1–10CrossRef Schönemann PH (1966) A generalized solution of the orthogonal procrustes problem. Psychometrika 31(1):1–10CrossRef
24.
go back to reference Welch J, Bax M, Mori K, Krummel T, Shahidi R, Maurer C (2002) A fast and accurate method of ultrasound probe calibration for image-guided surgery. In: Proceedings of computer assisted radiology and surgery, p 1078 Welch J, Bax M, Mori K, Krummel T, Shahidi R, Maurer C (2002) A fast and accurate method of ultrasound probe calibration for image-guided surgery. In: Proceedings of computer assisted radiology and surgery, p 1078
25.
go back to reference Yaniv Z (2015) Which pivot calibration? In: Proceedings of the SPIE, vol 9415 Yaniv Z (2015) Which pivot calibration? In: Proceedings of the SPIE, vol 9415
Metadata
Title
Guided ultrasound calibration: where, how, and how many calibration fiducials
Authors
Elvis C. S. Chen
Terry M. Peters
Burton Ma
Publication date
01-06-2016
Publisher
Springer Berlin Heidelberg
Published in
International Journal of Computer Assisted Radiology and Surgery / Issue 6/2016
Print ISSN: 1861-6410
Electronic ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-016-1390-7

Other articles of this Issue 6/2016

International Journal of Computer Assisted Radiology and Surgery 6/2016 Go to the issue