01-02-2015 | Original Article

Acetabular Version Measurement in Total Hip Arthroplasty: the Impact of Inclination and the Value of Multi-Planar CT Reformation

Authors: Michael Loftus, MD, MBA, Yan Ma, PhD, Bernard Ghelman, MD

Published in: HSS Journal ® | Issue 1/2015

Login to get access

Abstract

Background

The orientation of the acetabular cup component of a total hip arthroplasty can be evaluated in a number of ways, utilizing a myriad of imaging techniques and measurement parameters, including intraoperative surgical estimates, postoperative radiographs, and cross-sectional imaging such as computed tomography (CT) and magnetic resonance imaging (MRI).

Questions/Purposes

How do traditional versus corrected measurements of acetabular version vary from one another based on the inclination of the cup? What is the reliability of the corrected acetabular version measurements based on interobserver and intraobserver consistency?

Patients and Methods

Two fellowship-trained musculoskeletal radiologists reviewed CT scans on 60 total hip arthroplasties. Acetabular inclination, traditional CT acetabular version, and CT acetabular version corrected for inclination (by utilizing multi-planar reformations to measure in the plane of the cup face) were each measured. The difference was then calculated between the “traditional” axial CT and “corrected” acetabular version measurements, and the association between this difference and the acetabular inclination was assessed.

Results

The “traditional” axial CT and “corrected” acetabular version measurements differed from one another in every case, with the traditional method yielding a version measurement that was on average 9.5° higher than the corrected technique. However, as the acetabular cup inclination angle decreased, the “traditional” measurement became more variable and increasingly discordant with the “corrected” version measurement.

Conclusions

There is inherent variability between the many methods utilized for defining and measuring acetabular version, with axial CT measurements often used as an accepted proxy for true cup anteversion. However, the variability between different measurement techniques is correlated with acetabular inclination, and this variability is most pronounced when acetabular inclination is low, ultimately leading to potential confusion in measurement terminology. The increasingly widespread availability of multi-planar CT reformations provides an opportunity to standardize methodology, eliminate the impact of inclination on acetabular version measurements, and potentially provide a more reliable comparison of the impact of cup orientation on surgical outcomes.
Appendix
Available only for authorised users
Literature
1.
go back to reference Ackland MK, Bourne WB, Uhthoff HK. Anteversion of the acetabular cup. Measurement of angle after total hip replacement. J Bone Jt Surger. 1986; 68(3): 409-413. Ackland MK, Bourne WB, Uhthoff HK. Anteversion of the acetabular cup. Measurement of angle after total hip replacement. J Bone Jt Surger. 1986; 68(3): 409-413.
2.
go back to reference Barmeir E, Dubowitz B, Roffman M. Computed tomography in the assessment and planning of complicated total hip replacement. Acta Orthop Scand. 1982; 53: 597-604.CrossRefPubMed Barmeir E, Dubowitz B, Roffman M. Computed tomography in the assessment and planning of complicated total hip replacement. Acta Orthop Scand. 1982; 53: 597-604.CrossRefPubMed
3.
go back to reference Blendea S, Eckman K, Jaramaz B, Levison TJ, Digioia AM. Measurements of acetabular cup position and pelvic spatial orientation after total hip arthroplasty using computed tomography/radiography matching. Comput aided Surg Off J Int Soc Comput Aided Surg. 2005; 10(1): 37-43. Available at: http://www.ncbi.nlm.nih.gov/pubmed/16199380.CrossRef Blendea S, Eckman K, Jaramaz B, Levison TJ, Digioia AM. Measurements of acetabular cup position and pelvic spatial orientation after total hip arthroplasty using computed tomography/radiography matching. Comput aided Surg Off J Int Soc Comput Aided Surg. 2005; 10(1): 37-43. Available at: http://​www.​ncbi.​nlm.​nih.​gov/​pubmed/​16199380.CrossRef
4.
go back to reference Chandler DR, Tarr RR, Gruen TA, Sarmiento A. Radiographic assessment of acetabular cup orientation. A new design concept. Clin Orthop Relat Res. 1984; 186: 60-64.PubMed Chandler DR, Tarr RR, Gruen TA, Sarmiento A. Radiographic assessment of acetabular cup orientation. A new design concept. Clin Orthop Relat Res. 1984; 186: 60-64.PubMed
5.
11.
go back to reference Mian SW, Truchly G, Pflum FA. Computed tomography measurement of acetabular cup anteversion and retroversion in total hip arthroplasty. Clin Orthop Relat Res. 1992; 276: 206-209.PubMed Mian SW, Truchly G, Pflum FA. Computed tomography measurement of acetabular cup anteversion and retroversion in total hip arthroplasty. Clin Orthop Relat Res. 1992; 276: 206-209.PubMed
16.
go back to reference Olivecrona H, Weidenhielm L, Olivecrona L, et al. Spatial component position in total hip arthroplasty. Accuracy and repeatability with a new CT method. Acta radiol. 2003; 44(1): 84-91.PubMed Olivecrona H, Weidenhielm L, Olivecrona L, et al. Spatial component position in total hip arthroplasty. Accuracy and repeatability with a new CT method. Acta radiol. 2003; 44(1): 84-91.PubMed
17.
go back to reference Van Bosse HJP, Lee D, Henderson ER, Sala DA, Feldman DS. Pelvic positioning creates error in CT acetabular measurements. Clinical Orthopaedics and Related Research. 2011; 469(6): 1683-91.CrossRefPubMedCentralPubMed Van Bosse HJP, Lee D, Henderson ER, Sala DA, Feldman DS. Pelvic positioning creates error in CT acetabular measurements. Clinical Orthopaedics and Related Research. 2011; 469(6): 1683-91.CrossRefPubMedCentralPubMed
Metadata
Title
Acetabular Version Measurement in Total Hip Arthroplasty: the Impact of Inclination and the Value of Multi-Planar CT Reformation
Authors
Michael Loftus, MD, MBA
Yan Ma, PhD
Bernard Ghelman, MD
Publication date
01-02-2015
Publisher
Springer US
Published in
HSS Journal ® / Issue 1/2015
Print ISSN: 1556-3316
Electronic ISSN: 1556-3324
DOI
https://doi.org/10.1007/s11420-014-9416-6

Other articles of this Issue 1/2015

HSS Journal ® 1/2015 Go to the issue