Published in: Sleep and Breathing 2/2016

01-05-2016 | Sleep Breathing Physiology and Disorders • Original Article

Diagnostic validity of the use of a portable single-channel electromyography device for sleep bruxism

Authors: Juliana Stuginski-Barbosa, André Luís Porporatti, Yuri Martins Costa, Peter Svensson, Paulo César Rodrigues Conti

Published in: Sleep and Breathing | Issue 2/2016

Login to get access

Abstract

Purpose

The aim of this study was to determine an appropriate cutoff value and the number of nights of sleep with the portable single-channel EMG device (GrindCare) necessary for a valid sleep bruxism (SB) diagnosis.

Methods

Twenty consecutive post-graduate students and staff at Bauru School of Dentistry composed the sample. Each participant underwent the GrindCare for five consecutive nights and the polysomnography (PSG). The discrimination between bruxers and non-bruxers was based only on the PSG analysis. Data about electromyography per hour with GrindCare (EMG/h) and PSG (bursts/h) were scored.

Results

There were positive correlations between the two devices for EMG/h and bursts/h in three and five consecutive nights. Bland-Altman analysis of the EMG bursts/h showed positive agreement between the methods. The receiver operating characteristic (ROC) analyses also showed that using a minimum of 18 EMG/h for three nights and 19 EMG/h for five nights in GrindCare as cutoffs resulted in a 90 % specificity and positive likelihood ratio equal to 5.

Conclusions

GrindCare is able to discriminate SB diagnosed by PSG and gold standard criteria, when used for three or five consecutive nights, and it may be a valid choice in clinical practice for SB assessment.
Literature
1.
go back to reference American Academy of Sleep Medicine (2014) International classification of sleep disorders, 3rd edn. American Academy of Sleep Disorders, Darien American Academy of Sleep Medicine (2014) International classification of sleep disorders, 3rd edn. American Academy of Sleep Disorders, Darien
4.
go back to reference Rompré PH, Daigle-Landry D, Guitard F et al (2007) Identification of a sleep bruxism subgroup with a higher risk of pain. J Dent Res 86:837–842CrossRefPubMed Rompré PH, Daigle-Landry D, Guitard F et al (2007) Identification of a sleep bruxism subgroup with a higher risk of pain. J Dent Res 86:837–842CrossRefPubMed
5.
go back to reference Lavigne GJ, Rompré PH, Montplaisir JY (1996) Sleep bruxism: validity of clinical research diagnostic criteria in a controlled polysomnographic study. J Dent Res 75:546–552CrossRefPubMed Lavigne GJ, Rompré PH, Montplaisir JY (1996) Sleep bruxism: validity of clinical research diagnostic criteria in a controlled polysomnographic study. J Dent Res 75:546–552CrossRefPubMed
6.
go back to reference Kato T, Dal-Fabbro C, Lavigne GJ (2003) Current knowledge on awake and sleep bruxism: overview. Alpha Omegan 96:24–32PubMed Kato T, Dal-Fabbro C, Lavigne GJ (2003) Current knowledge on awake and sleep bruxism: overview. Alpha Omegan 96:24–32PubMed
10.
go back to reference Conti PCR, Stuginski-Barbosa J, Bonjardim LR et al (2014) Contingent electrical stimulation inhibits jaw muscle activity during sleep but not pain intensity or masticatory muscle pressure pain threshold in self-reported bruxers: a pilot study. Oral Surg Oral Med Oral Pathol Oral Radiol 117:45–52. doi:10.1016/j.oooo.2013.08.015 CrossRefPubMed Conti PCR, Stuginski-Barbosa J, Bonjardim LR et al (2014) Contingent electrical stimulation inhibits jaw muscle activity during sleep but not pain intensity or masticatory muscle pressure pain threshold in self-reported bruxers: a pilot study. Oral Surg Oral Med Oral Pathol Oral Radiol 117:45–52. doi:10.​1016/​j.​oooo.​2013.​08.​015 CrossRefPubMed
11.
go back to reference Haugland MS, Huynh N, Lavigne G (2011) Single-channel EMG versus PSG scoring of RMMA in sleep bruxers Haugland MS, Huynh N, Lavigne G (2011) Single-channel EMG versus PSG scoring of RMMA in sleep bruxers
12.
go back to reference Dworkin SF, LeResche L (1992) Research diagnostic criteria for temporomandibular disorders: review, criteria, examinations and specifications, critique. J Craniomandib Disord 6:301–355PubMed Dworkin SF, LeResche L (1992) Research diagnostic criteria for temporomandibular disorders: review, criteria, examinations and specifications, critique. J Craniomandib Disord 6:301–355PubMed
13.
go back to reference Raphael KG, Janal MN, Sirois DA, Svensson P (2013) Effect of contingent electrical stimulation on masticatory muscle activity and pain in patients with a myofascial temporomandibular disorder and sleep bruxism. J Orofac Pain 27:21–31CrossRefPubMed Raphael KG, Janal MN, Sirois DA, Svensson P (2013) Effect of contingent electrical stimulation on masticatory muscle activity and pain in patients with a myofascial temporomandibular disorder and sleep bruxism. J Orofac Pain 27:21–31CrossRefPubMed
14.
go back to reference Iber C, Anacoli-Israel S, Chesson A, Quen S (2007) The AASM manual for the scoring of sleep and associated events: rules, terminology and technical specifications, 1st edn. American Academy of Sleep Medicine, Westchester Iber C, Anacoli-Israel S, Chesson A, Quen S (2007) The AASM manual for the scoring of sleep and associated events: rules, terminology and technical specifications, 1st edn. American Academy of Sleep Medicine, Westchester
17.
22.
go back to reference Widmer CG, Lund JP, Feine JS (1990) Evaluation of diagnostic tests for TMD. J Calif Dent Assoc 18:53–60PubMed Widmer CG, Lund JP, Feine JS (1990) Evaluation of diagnostic tests for TMD. J Calif Dent Assoc 18:53–60PubMed
23.
24.
go back to reference Lavigne GJ, Guitard F, Rompré PH, Montplaisir JY (2001) Variability in sleep bruxism activity over time. J Sleep Res 10:237–244CrossRefPubMed Lavigne GJ, Guitard F, Rompré PH, Montplaisir JY (2001) Variability in sleep bruxism activity over time. J Sleep Res 10:237–244CrossRefPubMed
25.
go back to reference Bernhardt O, Hawali S, Sümnig W, Meyer G (2012) Electrical stimulation of the temporalis muscle during sleep of myofascial pain—a pilot study. J Craniomandib Funct 4:197–210 Bernhardt O, Hawali S, Sümnig W, Meyer G (2012) Electrical stimulation of the temporalis muscle during sleep of myofascial pain—a pilot study. J Craniomandib Funct 4:197–210
Metadata
Title
Diagnostic validity of the use of a portable single-channel electromyography device for sleep bruxism
Authors
Juliana Stuginski-Barbosa
André Luís Porporatti
Yuri Martins Costa
Peter Svensson
Paulo César Rodrigues Conti
Publication date
01-05-2016
Publisher
Springer Berlin Heidelberg
Published in
Sleep and Breathing / Issue 2/2016
Print ISSN: 1520-9512
Electronic ISSN: 1522-1709
DOI
https://doi.org/10.1007/s11325-015-1283-y

Other articles of this Issue 2/2016

Sleep and Breathing 2/2016 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine