Published in: Molecular Imaging and Biology 3/2009

01-05-2009 | Research Article

Toward Quantitative Small Animal Pinhole SPECT: Assessment of Quantitation Accuracy Prior to Image Compensations

Authors: Chia-Lin Chen, Yuchuan Wang, Jason J. S. Lee, Benjamin M. W. Tsui

Published in: Molecular Imaging and Biology | Issue 3/2009

Login to get access

Abstract

Purpose

We assessed the quantitation accuracy of small animal pinhole single photon emission computed tomography (SPECT) under the current preclinical settings, where image compensations are not routinely applied.

Procedures

The effects of several common image-degrading factors and imaging parameters on quantitation accuracy were evaluated using Monte-Carlo simulation methods. Typical preclinical imaging configurations were modeled, and quantitative analyses were performed based on image reconstructions without compensating for attenuation, scatter, and limited system resolution.

Results

Using mouse-sized phantom studies as examples, attenuation effects alone degraded quantitation accuracy by up to −18% (Tc-99m or In-111) or −41% (I-125). The inclusion of scatter effects changed the above numbers to −12% (Tc-99m or In-111) and −21% (I-125), respectively, indicating the significance of scatter in quantitative I-125 imaging. Region-of-interest (ROI) definitions have greater impacts on regional quantitation accuracy for small sphere sources as compared to attenuation and scatter effects. For the same ROI, SPECT acquisitions using pinhole apertures of different sizes could significantly affect the outcome, whereas the use of different radii-of-rotation yielded negligible differences in quantitation accuracy for the imaging configurations simulated.

Conclusions

We have systematically quantified the influence of several factors affecting the quantitation accuracy of small animal pinhole SPECT. In order to consistently achieve accurate quantitation within 5% of the truth, comprehensive image compensation methods are needed.
Literature
1.
go back to reference Weissleder R, Mahmood U (2001) Molecular imaging. Radiology 219:316–33. MayPubMed Weissleder R, Mahmood U (2001) Molecular imaging. Radiology 219:316–33. MayPubMed
3.
go back to reference Pomper MG (2005) Translational molecular imaging for cancer. Canc Imag 5:S16–26. Spec No ACrossRef Pomper MG (2005) Translational molecular imaging for cancer. Canc Imag 5:S16–26. Spec No ACrossRef
4.
go back to reference Beekman F, van der Have F (2007) The pinhole: gateway to ultra-high-resolution three-dimensional radionuclide imaging. Eur J Nucl Med Mol Imaging 34:151–161. FebPubMedCrossRef Beekman F, van der Have F (2007) The pinhole: gateway to ultra-high-resolution three-dimensional radionuclide imaging. Eur J Nucl Med Mol Imaging 34:151–161. FebPubMedCrossRef
5.
go back to reference Meikle SR, Kench P, Kassiou M, Banati RB (2005) Small animal SPECT and its place in the matrix of molecular imaging technologies. Phys Med Biol 50:R45–R61. Nov 21PubMedCrossRef Meikle SR, Kench P, Kassiou M, Banati RB (2005) Small animal SPECT and its place in the matrix of molecular imaging technologies. Phys Med Biol 50:R45–R61. Nov 21PubMedCrossRef
6.
go back to reference Acton PD, Choi SR, Plossl K, Kung HF (2002) Quantification of dopamine transporters in the mouse brain using ultra-high resolution single-photon emission tomography. Eur J Nucl Med Mol Imaging 29:691–698. MayPubMedCrossRef Acton PD, Choi SR, Plossl K, Kung HF (2002) Quantification of dopamine transporters in the mouse brain using ultra-high resolution single-photon emission tomography. Eur J Nucl Med Mol Imaging 29:691–698. MayPubMedCrossRef
7.
go back to reference Acton PD, Thomas D, Zhou R (2006) Quantitative imaging of myocardial infarct in rats with high resolution pinhole SPECT. Int J Cardiovasc Imaging 22:429–434. JunPubMedCrossRef Acton PD, Thomas D, Zhou R (2006) Quantitative imaging of myocardial infarct in rats with high resolution pinhole SPECT. Int J Cardiovasc Imaging 22:429–434. JunPubMedCrossRef
8.
go back to reference Alvarez-Fischer D, Blessmann G, Trosowski C et al (2007) Quantitative [I-123]FP-CIT pinhole SPECT imaging predicts striatal dopamine levels, but not number of nigral neurons in different mouse models of Parkinson’s disease. Neuroimage 38:5–12. Oct 15PubMedCrossRef Alvarez-Fischer D, Blessmann G, Trosowski C et al (2007) Quantitative [I-123]FP-CIT pinhole SPECT imaging predicts striatal dopamine levels, but not number of nigral neurons in different mouse models of Parkinson’s disease. Neuroimage 38:5–12. Oct 15PubMedCrossRef
9.
go back to reference Liu ZL, Kastis GA, Stevenson GD et al (2002) Quantitative analysis of acute myocardial infarct in rat hearts with ischemia–reperfusion using a high-resolution stationary SPECT system. J Nucl Med 43:933–939. JulPubMed Liu ZL, Kastis GA, Stevenson GD et al (2002) Quantitative analysis of acute myocardial infarct in rat hearts with ischemia–reperfusion using a high-resolution stationary SPECT system. J Nucl Med 43:933–939. JulPubMed
10.
go back to reference Tsui BMW, Frey EC, Zhao X, Lalush DS, Johnston RE, Mccartney WH (1994) The importance and implementation of accurate 3d compensation methods for quantitative SPECT. Phys Med Biol 39:509–530. MarPubMedCrossRef Tsui BMW, Frey EC, Zhao X, Lalush DS, Johnston RE, Mccartney WH (1994) The importance and implementation of accurate 3d compensation methods for quantitative SPECT. Phys Med Biol 39:509–530. MarPubMedCrossRef
11.
go back to reference Tsui BMW, Zhao XD, Frey EC, Mccartney WH (1994) Quantitative single-photon emission computed-tomography—basics and clinical considerations. Semin Nucl Med 24:38–65. JanPubMedCrossRef Tsui BMW, Zhao XD, Frey EC, Mccartney WH (1994) Quantitative single-photon emission computed-tomography—basics and clinical considerations. Semin Nucl Med 24:38–65. JanPubMedCrossRef
12.
go back to reference Deloar HM, Watabe H, Aoi T, Iida H (2003) Evaluation of penetration and scattering components in conventional pinhole SPECT: phantom studies using Monte Carlo simulation. Phys Med Biol 48:995–1008. Apr 21PubMedCrossRef Deloar HM, Watabe H, Aoi T, Iida H (2003) Evaluation of penetration and scattering components in conventional pinhole SPECT: phantom studies using Monte Carlo simulation. Phys Med Biol 48:995–1008. Apr 21PubMedCrossRef
13.
go back to reference Li J, Jaszczak RJ, Coleman RE (1995) Quantitative small field-of-view pinhole spect imaging—initial evaluation. IEEE Trans Nucl Sci 42:1109–1113. AugCrossRef Li J, Jaszczak RJ, Coleman RE (1995) Quantitative small field-of-view pinhole spect imaging—initial evaluation. IEEE Trans Nucl Sci 42:1109–1113. AugCrossRef
14.
go back to reference Hwang AB, Taylor CC, VanBrocklin HF, Dae MW, Hasegawa BH (2006) Attenuation correction of small animal SPECT images acquired with I-125-iodorotenone. IEEE Trans Nucl Sci 53:1213–1220. JunCrossRef Hwang AB, Taylor CC, VanBrocklin HF, Dae MW, Hasegawa BH (2006) Attenuation correction of small animal SPECT images acquired with I-125-iodorotenone. IEEE Trans Nucl Sci 53:1213–1220. JunCrossRef
15.
go back to reference Chen C-L, Wang Y, Lee JJS, Tsui BMW (2008) Integration of SimSET photon history generator in GATE for efficient Monte Carlo simulations of Pinhole SPECT. Med Phys 35:3278–3284. JulyPubMedCrossRef Chen C-L, Wang Y, Lee JJS, Tsui BMW (2008) Integration of SimSET photon history generator in GATE for efficient Monte Carlo simulations of Pinhole SPECT. Med Phys 35:3278–3284. JulyPubMedCrossRef
16.
go back to reference Harrison RL, Haynor DR, Gillispie SB, Vannoy SD, Kaplan MS, Lewellen TK (1993) A public-domain simulation system for emission tomography-photon tracking through heterogeneous attenuation using importance sampling. J Nucl Med 34:P60–P60. May Harrison RL, Haynor DR, Gillispie SB, Vannoy SD, Kaplan MS, Lewellen TK (1993) A public-domain simulation system for emission tomography-photon tracking through heterogeneous attenuation using importance sampling. J Nucl Med 34:P60–P60. May
17.
go back to reference Jan S, Santin G, Strul D et al (2004) GATE: a simulation toolkit for PET and SPECT. Phys Med Biol 49:4543–4561. Oct 7PubMedCrossRef Jan S, Santin G, Strul D et al (2004) GATE: a simulation toolkit for PET and SPECT. Phys Med Biol 49:4543–4561. Oct 7PubMedCrossRef
18.
go back to reference Wang YC, Tsui BMW (2007) Pinhole SPECT with different data acquisition geometries: Usefulness of unified projection operators in homogeneous coordinates. IEEE Trans Med Imag IEEE 26:298–308. MarCrossRef Wang YC, Tsui BMW (2007) Pinhole SPECT with different data acquisition geometries: Usefulness of unified projection operators in homogeneous coordinates. IEEE Trans Med Imag IEEE 26:298–308. MarCrossRef
19.
go back to reference Segars WP, Tsui BMW, Frey EC, Johnson GA, Berr SS (2004) Development of a 4-D digital mouse phantom for molecular imaging research. Mol Imag Biol 6:149–159. May–JunCrossRef Segars WP, Tsui BMW, Frey EC, Johnson GA, Berr SS (2004) Development of a 4-D digital mouse phantom for molecular imaging research. Mol Imag Biol 6:149–159. May–JunCrossRef
Metadata
Title
Toward Quantitative Small Animal Pinhole SPECT: Assessment of Quantitation Accuracy Prior to Image Compensations
Authors
Chia-Lin Chen
Yuchuan Wang
Jason J. S. Lee
Benjamin M. W. Tsui
Publication date
01-05-2009
Publisher
Springer-Verlag
Published in
Molecular Imaging and Biology / Issue 3/2009
Print ISSN: 1536-1632
Electronic ISSN: 1860-2002
DOI
https://doi.org/10.1007/s11307-008-0181-0

Other articles of this Issue 3/2009

Molecular Imaging and Biology 3/2009 Go to the issue