01-12-2009 | Technical Report

The effect of voxel size on image reconstruction in cone-beam computed tomography

Authors: Hideyuki Tanimoto, Yoshinori Arai

Published in: Oral Radiology | Issue 2/2009

Login to get access

Abstract

Objectives

This study evaluated the effects of changing the voxel size on the resolution and noise of cone-beam computed tomography (CBCT) reconstruction images.

Methods

The voxel sizes used for reconstruction were 160, 80, and 40 μm using prototype software for the Accuitomo F8 (J. Morita, Kyoto, Japan). The resolution was measured using a modulation transfer function (MTF), and CBCT images of a 1-mm-thick, 10-mm-diameter aluminum pipe set slightly inclined from the vertical were taken with a field of view of 8 cm. To measure the noise, a tomographic image of an 8-cm-diameter water phantom was taken and reconstructed at the three voxel sizes. The standard deviation (SD) of the noise was then determined.

Results

The MTF at 2 lp/mm was 0.05, 0.12, and 0.12 for voxel sizes of 160, 80, and 40 μm, respectively, and the SD of the noise was 10.0, 13.8, and 17.1% for the same respective voxel sizes.

Conclusions

The limit of resolution was determined to be the 80-μm voxel size. When the voxels were smaller, the noise increased.
Literature
1.
go back to reference Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shinoda K. Development of a compact computed tomographic apparatus for dental use. Dentomaxillofac Radiol. 1999;28:245–8.CrossRefPubMed Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shinoda K. Development of a compact computed tomographic apparatus for dental use. Dentomaxillofac Radiol. 1999;28:245–8.CrossRefPubMed
2.
go back to reference Terakado M, Hashimoto K, Arai Y, Honda M, Sekiwa T, Sato H. Diagnostic imaging with newly developed ortho cubic super-high resolution computed tomography (Ortho-CT). Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2000;89:509–18.CrossRefPubMed Terakado M, Hashimoto K, Arai Y, Honda M, Sekiwa T, Sato H. Diagnostic imaging with newly developed ortho cubic super-high resolution computed tomography (Ortho-CT). Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2000;89:509–18.CrossRefPubMed
3.
go back to reference Hashimoto K, Arai Y, Iwai K, Kawashima S, Terakado M. A comparison of a new limited cone beam computed tomography machine for dental use with a multidetector row helical CT machine. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;95:371–7.CrossRefPubMed Hashimoto K, Arai Y, Iwai K, Kawashima S, Terakado M. A comparison of a new limited cone beam computed tomography machine for dental use with a multidetector row helical CT machine. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;95:371–7.CrossRefPubMed
4.
go back to reference Honda K, Arai Y, Kashima M, Takano Y, Sawada K, Ejima K, et al. Evaluation of the usefulness of the limited cone-beam CT (3DX) in the assessment of the thickness of the roof of the glenoid fossa of the temporomandibular joint. Dentomaxillofac Radiol. 2004;33:391–5.CrossRefPubMed Honda K, Arai Y, Kashima M, Takano Y, Sawada K, Ejima K, et al. Evaluation of the usefulness of the limited cone-beam CT (3DX) in the assessment of the thickness of the roof of the glenoid fossa of the temporomandibular joint. Dentomaxillofac Radiol. 2004;33:391–5.CrossRefPubMed
5.
go back to reference Araki M, Hashimoto K, Kawashima S, Matsumoto K, Akiyama Y. Radiographic features of enostosis determined with limited cone-beam computer tomography in comparison with rotational panoramic radiography. Oral Radiol. 2006;22:27–33.CrossRef Araki M, Hashimoto K, Kawashima S, Matsumoto K, Akiyama Y. Radiographic features of enostosis determined with limited cone-beam computer tomography in comparison with rotational panoramic radiography. Oral Radiol. 2006;22:27–33.CrossRef
6.
go back to reference Kagawa T, Fukunari F, Shiraishi T, Yamasaki M, Ichihara T, Kihara Y, et al. Development of a simple image designed for small x-ray field CT equipment 3DX. Oral Radiol. 2006;22:47–51.CrossRef Kagawa T, Fukunari F, Shiraishi T, Yamasaki M, Ichihara T, Kihara Y, et al. Development of a simple image designed for small x-ray field CT equipment 3DX. Oral Radiol. 2006;22:47–51.CrossRef
7.
go back to reference Sakabe J, Kuroki Y, Fujimaki S, Nakajima I, Honda K. Reproducibility and accuracy of measuring unerupted teeth using limited cone beam x-ray CT. Dentomaxillofac Radiol. 2007;36:2–6.CrossRefPubMed Sakabe J, Kuroki Y, Fujimaki S, Nakajima I, Honda K. Reproducibility and accuracy of measuring unerupted teeth using limited cone beam x-ray CT. Dentomaxillofac Radiol. 2007;36:2–6.CrossRefPubMed
8.
go back to reference Suomalainen AK, Salo A, Robinson S, Peltola JS. The 3DX multi-image micro-CT device in clinical dental practice. Dentomaxillofac Radiol. 2007;36:80–5.CrossRefPubMed Suomalainen AK, Salo A, Robinson S, Peltola JS. The 3DX multi-image micro-CT device in clinical dental practice. Dentomaxillofac Radiol. 2007;36:80–5.CrossRefPubMed
9.
go back to reference Lofthag-Hansen S, Thilander-Klang A, Ekestubbe A, Helmrot E, Gröndahl K. Calculating effective dose on a cone beam computed tomography device: 3D Accuitomo and 3D Accuitomo FPD. Dentomaxillofac Radiol. 2008;37:72–9.CrossRefPubMed Lofthag-Hansen S, Thilander-Klang A, Ekestubbe A, Helmrot E, Gröndahl K. Calculating effective dose on a cone beam computed tomography device: 3D Accuitomo and 3D Accuitomo FPD. Dentomaxillofac Radiol. 2008;37:72–9.CrossRefPubMed
10.
go back to reference Hirsch E, Wolf U, Heinicke F, Silva MAG. Dosimetry of the cone beam computed tomography Veraviewepocs 3D compared with the 3D Accuitomo in different fields of view. Dentomaxillofac Radiol. 2008;37:268–73.CrossRefPubMed Hirsch E, Wolf U, Heinicke F, Silva MAG. Dosimetry of the cone beam computed tomography Veraviewepocs 3D compared with the 3D Accuitomo in different fields of view. Dentomaxillofac Radiol. 2008;37:268–73.CrossRefPubMed
11.
go back to reference Araki K, Maki K, Seki K, Sakamaki K, Harata Y, Sakaino R, et al. Characteristics of a newly developed dentomaxillofacial x-ray cone beam CT scanner (CB MercuRay™): system configuration and physical properties. Dentomaxillofac Radiol. 2004;33:51–9.CrossRefPubMed Araki K, Maki K, Seki K, Sakamaki K, Harata Y, Sakaino R, et al. Characteristics of a newly developed dentomaxillofacial x-ray cone beam CT scanner (CB MercuRay™): system configuration and physical properties. Dentomaxillofac Radiol. 2004;33:51–9.CrossRefPubMed
12.
go back to reference Marmulla R, Wörtche R, Mühling J, Hassfeld S. Geometric accuracy of the NewTom 9000 cone beam CT. Dentomaxillofac Radiol. 2005;34:28–31.CrossRefPubMed Marmulla R, Wörtche R, Mühling J, Hassfeld S. Geometric accuracy of the NewTom 9000 cone beam CT. Dentomaxillofac Radiol. 2005;34:28–31.CrossRefPubMed
13.
go back to reference Arnheiter C, Scarfe WC, Farman AG. Trends in maxillofacial cone-beam computed tomography usage. Oral Radiol. 2006;22:80–5.CrossRef Arnheiter C, Scarfe WC, Farman AG. Trends in maxillofacial cone-beam computed tomography usage. Oral Radiol. 2006;22:80–5.CrossRef
14.
go back to reference Bartling SH, Majdani O, Gupta R, Rodt T, Dullin C, Fitzgerald PF, et al. Large scan field, high spatial resolution flat-panel detector based volumetric CT of the whole human skull base and for maxillofacial imaging. Dentomaxillofac Radiol. 2007;36:317–27.CrossRefPubMed Bartling SH, Majdani O, Gupta R, Rodt T, Dullin C, Fitzgerald PF, et al. Large scan field, high spatial resolution flat-panel detector based volumetric CT of the whole human skull base and for maxillofacial imaging. Dentomaxillofac Radiol. 2007;36:317–27.CrossRefPubMed
15.
go back to reference Stratemann SA, Huang JC, Maki K, Miller AJ, Hatcher DC. Comparison of cone beam computed tomography imaging with physical measures. Dentomaxillofac Radiol. 2008;37:80–93.CrossRefPubMed Stratemann SA, Huang JC, Maki K, Miller AJ, Hatcher DC. Comparison of cone beam computed tomography imaging with physical measures. Dentomaxillofac Radiol. 2008;37:80–93.CrossRefPubMed
16.
go back to reference Stewart CB. Radiologic science for technologists: physics, biology, and protection. St. Louis: C.V. Mosby; 1988. Stewart CB. Radiologic science for technologists: physics, biology, and protection. St. Louis: C.V. Mosby; 1988.
17.
go back to reference Samei E, Ranger NT, Dobbins JT III, et al. Intercomparison of methods for image quality characterization. I. Modulation transfer function. Med Phys. 2006;33(5):1454–65.CrossRefPubMed Samei E, Ranger NT, Dobbins JT III, et al. Intercomparison of methods for image quality characterization. I. Modulation transfer function. Med Phys. 2006;33(5):1454–65.CrossRefPubMed
18.
go back to reference Higashide R, Ichikawa K, Kunimoto H, Sagawa M. Proposal and verification of presampled MTF measurement by the simple analysis method using the edge method (in Japanese). Dent Radiol. 2008;64:417–25. Higashide R, Ichikawa K, Kunimoto H, Sagawa M. Proposal and verification of presampled MTF measurement by the simple analysis method using the edge method (in Japanese). Dent Radiol. 2008;64:417–25.
19.
go back to reference Berland L. Practical CT technology and techniques. New York: Raven Press; 1987. Berland L. Practical CT technology and techniques. New York: Raven Press; 1987.
21.
go back to reference ICRP. ICRP publication 60: 1990 recommendations of the International Commission on Radiological Protection. Annals of the ICRP, vol 21. Oxford: Pergamon Press; 1991. ICRP. ICRP publication 60: 1990 recommendations of the International Commission on Radiological Protection. Annals of the ICRP, vol 21. Oxford: Pergamon Press; 1991.
23.
go back to reference Hayakawa Y, Farman AG, Kelly MS, Kuroyanagi K. Signal-to noise ratio: computed dental radiography versus Sens-A-Ray. Oral Radiol. 1995;11(2):109–13.CrossRef Hayakawa Y, Farman AG, Kelly MS, Kuroyanagi K. Signal-to noise ratio: computed dental radiography versus Sens-A-Ray. Oral Radiol. 1995;11(2):109–13.CrossRef
Metadata
Title
The effect of voxel size on image reconstruction in cone-beam computed tomography
Authors
Hideyuki Tanimoto
Yoshinori Arai
Publication date
01-12-2009
Publisher
Springer Japan
Published in
Oral Radiology / Issue 2/2009
Print ISSN: 0911-6028
Electronic ISSN: 1613-9674
DOI
https://doi.org/10.1007/s11282-009-0019-8

Other articles of this Issue 2/2009

Oral Radiology 2/2009 Go to the issue