Published in: Journal of Clinical Monitoring and Computing 6/2017

01-12-2017 | Original Research

Continuous blood pressure monitoring via non-invasive radial artery applanation tonometry and invasive arterial catheter demonstrates good agreement in patients undergoing colon carcinoma surgery

Authors: Jing Sun, Hanjian Chen, Jun Zheng, Bin Mao, Shengmei Zhu, Jingyi Feng

Published in: Journal of Clinical Monitoring and Computing | Issue 6/2017

Login to get access

Abstract

Purpose

Radial artery applanation tonometry (RAAT) has been developed and utilized for continuous arterial pressure monitoring. However, evidence is lacking to clinically verify the RAAT technology and identify appropriate patient groups before routine clinical use. This study aims to evaluate the RAAT technology by comparing systolic blood pressure (SBP), mean blood pressure (MBP) and diastolic blood pressure (DBP) values in patients undergoing colon carcinoma surgery.

Methods

Blood Pressure (BP) values obtained via RAAT (TL-300, Tensys Medical Inc., San Diego, CA, USA) and conventional arterial catheterization from 30 colon carcinoma surgical patients were collected and compared via Bland-Atman method, linear regression and 4-quadrant plot concordance analysis.

Results

For SBPs, MBPs and DBPs, means of the differences (±standard deviation; 95% limits of agreement) were −0.9 (±7.6; −15.7 to 13.9) mmHg, 3.1 (±6.5; −9.6 to 15.8) mmHg and 4.3 (±7.4; −10.3 to 18.8) mmHg, respectively. Linear regression coefficients of determination were 0.8706 for SBPs, 0.8353 for MBPs and 0.6858 for DBPs. Four-quadrant concordance correlation coefficients were 0.8740, 0.8522 and 0.7108 for SBPs, MBPs and DBPs, respectively.

Conclusions

A highly selected patient collective undergoing colon carcinoma surgery was studied. BP measurements obtained via the TL-300 had clinically acceptable agreement with that acquired invasively using an arterial catheter. For use in clinical routine, it is necessary to take measures for improvement regarding movement artifacts and dilution of noise. A large sample size of patients under various conditions is also needed to further evaluate the RAAT technology before clinically routine use.
Literature
2.
go back to reference Scheer B, Perel A, Pfeiffer UJ. Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Crit Care. 2002;6:198–204.CrossRef Scheer B, Perel A, Pfeiffer UJ. Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Crit Care. 2002;6:198–204.CrossRef
7.
go back to reference Meidert AS, Huber W, Hapfelmeier A, et al. Evaluation of the radial artery applanation tonometry technology for continuous noninvasive blood pressure monitoring compared with central aortic blood pressure measurements in patients with multiple organ dysfunction syndrome. J Crit Care. 2013;28(6):908–12. doi:10.1016/j.jcrc.2013.06.012.CrossRefPubMed Meidert AS, Huber W, Hapfelmeier A, et al. Evaluation of the radial artery applanation tonometry technology for continuous noninvasive blood pressure monitoring compared with central aortic blood pressure measurements in patients with multiple organ dysfunction syndrome. J Crit Care. 2013;28(6):908–12. doi:10.​1016/​j.​jcrc.​2013.​06.​012.CrossRefPubMed
8.
go back to reference Meidert AS, Huber W, Müller JN, et al. Radial artery applanation tonometry for continuous non-invasive arterial pressure monitoring in intensive care unit patients: comparison with invasively assessed radial arterial pressure. Br J Anaesth. 2014;112(3):521–8. doi:10.1093/bja/aet400.CrossRefPubMed Meidert AS, Huber W, Müller JN, et al. Radial artery applanation tonometry for continuous non-invasive arterial pressure monitoring in intensive care unit patients: comparison with invasively assessed radial arterial pressure. Br J Anaesth. 2014;112(3):521–8. doi:10.​1093/​bja/​aet400.CrossRefPubMed
10.
go back to reference Lin WQ, Wu HH, Su CS, et al. Comparison of continuous noninvasive blood pressure monitoring by TL-300 with standard invasive blood pressure measurement in patients undergoing elective neurosurgery. J Neurosurg Anesthesiol. 2015. doi:10.1097/ANA.0000000000000245. Lin WQ, Wu HH, Su CS, et al. Comparison of continuous noninvasive blood pressure monitoring by TL-300 with standard invasive blood pressure measurement in patients undergoing elective neurosurgery. J Neurosurg Anesthesiol. 2015. doi:10.​1097/​ANA.​0000000000000245​.
11.
go back to reference Saugel B, Meidert AS, Hapfelmeier A, Eyer F, Schmid RM, Huber W. Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device. Br J Anaesth. 2013;111(2):185–90. doi:10.1093/bja/aet025.CrossRefPubMed Saugel B, Meidert AS, Hapfelmeier A, Eyer F, Schmid RM, Huber W. Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device. Br J Anaesth. 2013;111(2):185–90. doi:10.​1093/​bja/​aet025.CrossRefPubMed
12.
go back to reference Saugel B, Fassio F, Hapfelmeier A, Meidert AS, Schmid RM, Huber W. The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients. Intensive Care Med. 2012;38(9):1471–7. doi:10.1007/s00134-012-2617-x.CrossRefPubMed Saugel B, Fassio F, Hapfelmeier A, Meidert AS, Schmid RM, Huber W. The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients. Intensive Care Med. 2012;38(9):1471–7. doi:10.​1007/​s00134-012-2617-x.CrossRefPubMed
14.
go back to reference Ilies C, BauerM, Berg P et al (2012) Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. Br J Anaesth 108:202–210. doi:10.1093/bja/aer394.CrossRefPubMed Ilies C, BauerM, Berg P et al (2012) Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. Br J Anaesth 108:202–210. doi:10.​1093/​bja/​aer394.CrossRefPubMed
15.
16.
go back to reference ANSI/AAMI SP10:2002 American national standard for manual, electronic, or automated sphygmomanometers. Association for the Advancement of Medical Instrumentation 2002. ANSI/AAMI SP10:2002 American national standard for manual, electronic, or automated sphygmomanometers. Association for the Advancement of Medical Instrumentation 2002.
Metadata
Title
Continuous blood pressure monitoring via non-invasive radial artery applanation tonometry and invasive arterial catheter demonstrates good agreement in patients undergoing colon carcinoma surgery
Authors
Jing Sun
Hanjian Chen
Jun Zheng
Bin Mao
Shengmei Zhu
Jingyi Feng
Publication date
01-12-2017
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 6/2017
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-016-9967-9

Other articles of this Issue 6/2017

Journal of Clinical Monitoring and Computing 6/2017 Go to the issue