Published in: Journal of Interventional Cardiac Electrophysiology 1/2008

01-10-2008

Advances in remote monitoring of implantable pacemakers, cardioverter defibrillators and cardiac resynchronization therapy systems

Authors: Werner Jung, Andreas Rillig, Ralf Birkemeyer, Tomislav Miljak, Udo Meyerfeldt

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 1/2008

Login to get access

Abstract

Unlike the standard quarterly or semi-annual direct ambulatory device interrogations procedures, state-of-the-art implantable electronic cardiovascular devices (IECD) enable their wireless remote interrogation and monitoring, and automatically send reports and special alerts on a daily basis. This allows physicians to respond more proactively to changes in patient or device status, more appropriately triage patient care, and more efficiently perform the post-implant ambulatory follow-ups. This review presents the-state-of the-art technology of remote IECD monitoring and summarizes the main clinical observations published through June 2008. Cardiovascular remote monitoring systems made by several manufacturers are currently in various phases of development, clinical investigation, and medical applications. Data collected in several completed and ongoing studies strongly suggest that this new technology will make important contributions, particularly with respect to the facilitation of IECD follow-ups, enhancement of patient safety and quality of life, and lowering of medical costs. Further technological advances and a more clear understanding and appreciation of the clinical and economic benefits of telecardiology, will likely increase sharply the use of remote IECD monitoring in upcoming years.
Literature
1.
go back to reference Wilkoff, B. L., Auricchio, A., Brugada, J., Cowie, M., Ellenbogen, K. A., Gillis, A. M., et al. (2008). HRS/EHRA Expert Consensus on the Monitoring of Cardiovascular Implantable Electronic Devices (IECDs): description of techniques, indications, personnel, frequency and ethical considerations: developed in partnership with the Heart Rhythm Society (HRS) and the European Heart Rhythm Association (EHRA); and in collaboration with the American College of Cardiology (ACC), the American Heart Association (AHA), the European Society of Cardiology (ESC), the Heart Failure Association of ESC (HFA), and the Heart Failure Society of America (HFSA). Endorsed by the Heart Rhythm Society, the European Heart Rhythm Association (a registered branch of the ESC), the American College of Cardiology, the American Heart Association. Europace, 10(6), 707–725.PubMedCrossRef Wilkoff, B. L., Auricchio, A., Brugada, J., Cowie, M., Ellenbogen, K. A., Gillis, A. M., et al. (2008). HRS/EHRA Expert Consensus on the Monitoring of Cardiovascular Implantable Electronic Devices (IECDs): description of techniques, indications, personnel, frequency and ethical considerations: developed in partnership with the Heart Rhythm Society (HRS) and the European Heart Rhythm Association (EHRA); and in collaboration with the American College of Cardiology (ACC), the American Heart Association (AHA), the European Society of Cardiology (ESC), the Heart Failure Association of ESC (HFA), and the Heart Failure Society of America (HFSA). Endorsed by the Heart Rhythm Society, the European Heart Rhythm Association (a registered branch of the ESC), the American College of Cardiology, the American Heart Association. Europace, 10(6), 707–725.PubMedCrossRef
2.
go back to reference Epstein, A. E., Dimarco, J. P., Ellenbogen, K. A., Estes, N. A., Freedman, R. A., Gettes, L. S., et al. (2008). ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation, 117(21), e350–e408.PubMedCrossRef Epstein, A. E., Dimarco, J. P., Ellenbogen, K. A., Estes, N. A., Freedman, R. A., Gettes, L. S., et al. (2008). ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation, 117(21), e350–e408.PubMedCrossRef
3.
go back to reference McAlister, F.A., Ezekowitz, J., Hooton, N., Vandermeer, B., Spooner, C., Dryden, D. M., et al. (2007). Cardiac resynchronization therapy for patients with left ventricular systolic dysfunction: A systematic review. Journal of the American Medical Association, 297(22), 2502–2514.PubMedCrossRef McAlister, F.A., Ezekowitz, J., Hooton, N., Vandermeer, B., Spooner, C., Dryden, D. M., et al. (2007). Cardiac resynchronization therapy for patients with left ventricular systolic dysfunction: A systematic review. Journal of the American Medical Association, 297(22), 2502–2514.PubMedCrossRef
4.
go back to reference Krahn, A. D., Klein, G. J., Skanes, A. C., & Yee, R. (2004). Insertable loop recorder use for detection of intermittent arrhythmias. Pacing and Clinical Electrophysiology, 27(5), 657–664.PubMedCrossRef Krahn, A. D., Klein, G. J., Skanes, A. C., & Yee, R. (2004). Insertable loop recorder use for detection of intermittent arrhythmias. Pacing and Clinical Electrophysiology, 27(5), 657–664.PubMedCrossRef
5.
go back to reference Abraham, W. T. (2006). New approaches to monitoring heart failure before symptoms appear. Reviews in Cardiovascular Medicine, 7(Suppl 1), S33–S41.PubMed Abraham, W. T. (2006). New approaches to monitoring heart failure before symptoms appear. Reviews in Cardiovascular Medicine, 7(Suppl 1), S33–S41.PubMed
6.
go back to reference Lazarus, A. (2007). Remote, wireless, ambulatory monitoring of implantable pacemakers, cardioverter defibrillators, and cardiac resynchronization therapy systems: analysis of a worldwide database. Pacing and Clinical Electrophysiology, 30(Suppl 1), S2–S12.PubMed Lazarus, A. (2007). Remote, wireless, ambulatory monitoring of implantable pacemakers, cardioverter defibrillators, and cardiac resynchronization therapy systems: analysis of a worldwide database. Pacing and Clinical Electrophysiology, 30(Suppl 1), S2–S12.PubMed
7.
go back to reference Nielsen, J. C., Kottkamp, H., Zabel, M., Aliot, E., Kreutzer, U., Bauer, A., et al. (2008). Automatic home monitoring of implantable cardioverter defibrillators. Europace, 10(6), 729–735.PubMedCrossRef Nielsen, J. C., Kottkamp, H., Zabel, M., Aliot, E., Kreutzer, U., Bauer, A., et al. (2008). Automatic home monitoring of implantable cardioverter defibrillators. Europace, 10(6), 729–735.PubMedCrossRef
8.
go back to reference Ellery, S., Pakrashi, T., Paul, V., & Sack, S. (2006). Predicting mortality and rehospitalization in heart failure patients with Home Monitoring: the Home CARE pilot study. Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(Suppl 3), iii29–iii35.PubMed Ellery, S., Pakrashi, T., Paul, V., & Sack, S. (2006). Predicting mortality and rehospitalization in heart failure patients with Home Monitoring: the Home CARE pilot study. Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(Suppl 3), iii29–iii35.PubMed
9.
go back to reference Perings, C., Klein, G., Toft, E., Moro, C., Klug, D., Bocker, D., et al. (2006). The RIONI study rationale and design: Validation of the first stored electrograms transmitted via Home Monitoring in patients with implantable defibrillators. Europace, 8(4), 288–292.PubMedCrossRef Perings, C., Klein, G., Toft, E., Moro, C., Klug, D., Bocker, D., et al. (2006). The RIONI study rationale and design: Validation of the first stored electrograms transmitted via Home Monitoring in patients with implantable defibrillators. Europace, 8(4), 288–292.PubMedCrossRef
10.
go back to reference Ritter, O., & Bauer, W. R. (2006). Use of “IEGM online” in ICD patients—early detection of inappropriate classified ventricular tachycardia via Home Monitoring. Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(7), 368–372.PubMed Ritter, O., & Bauer, W. R. (2006). Use of “IEGM online” in ICD patients—early detection of inappropriate classified ventricular tachycardia via Home Monitoring. Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(7), 368–372.PubMed
11.
go back to reference Schoenfeld, M. H., Compton, S. J., Mead, R. H., Weiss, D. N., Sherfesee, L., Englund, J., et al. (2004). Remote monitoring of implantable cardioverter defibrillators: A prospective analysis. Pacing and Clinical Electrophysiology, 27(6 Pt 1), 757–763.PubMedCrossRef Schoenfeld, M. H., Compton, S. J., Mead, R. H., Weiss, D. N., Sherfesee, L., Englund, J., et al. (2004). Remote monitoring of implantable cardioverter defibrillators: A prospective analysis. Pacing and Clinical Electrophysiology, 27(6 Pt 1), 757–763.PubMedCrossRef
12.
go back to reference Ypenburg, C., Bax, J. J., van der Wall, E. E., Schalij, M. J., & van Ereven, L. (2007). Intrathoracic impedance monitoring to predict decompensated heart failure. The American Journal of Cardiology, 99(4), 554–557.PubMedCrossRef Ypenburg, C., Bax, J. J., van der Wall, E. E., Schalij, M. J., & van Ereven, L. (2007). Intrathoracic impedance monitoring to predict decompensated heart failure. The American Journal of Cardiology, 99(4), 554–557.PubMedCrossRef
13.
go back to reference Small, R. (2006). Integrating monitoring into the infrastructure and workflow of routine practice: OptiVol. Reviews in Cardiovascular Medicine, 7(Suppl 1), S47–S55.PubMed Small, R. (2006). Integrating monitoring into the infrastructure and workflow of routine practice: OptiVol. Reviews in Cardiovascular Medicine, 7(Suppl 1), S47–S55.PubMed
14.
go back to reference Wang, L. (2007). Fundamentals of intrathoracic impedance monitoring in heart failure. The American Journal of Cardiology, 99(10A), 3G–10G.PubMedCrossRef Wang, L. (2007). Fundamentals of intrathoracic impedance monitoring in heart failure. The American Journal of Cardiology, 99(10A), 3G–10G.PubMedCrossRef
15.
go back to reference Joseph, G. K., Wilkoff, B. L., Dresing, T., Burkhardt, J., & Khaykin, Y. (2004). Remote interrogation and monitoring of implantable cardioverter defibrillators. Journal of Interventional Cardiac Electrophysiology, 11(2), 161–166.PubMedCrossRef Joseph, G. K., Wilkoff, B. L., Dresing, T., Burkhardt, J., & Khaykin, Y. (2004). Remote interrogation and monitoring of implantable cardioverter defibrillators. Journal of Interventional Cardiac Electrophysiology, 11(2), 161–166.PubMedCrossRef
16.
go back to reference Clark, R. A., Inglis, S. C., McAlister, F. A., Cleland, J. G., & Stewart, S. (2007). Telemonitoring or structured telephone support programmes for patients with chronic heart failure: Systematic review and meta-analysis. BMJ (Clinical Research Ed), 334(7600), 942.CrossRef Clark, R. A., Inglis, S. C., McAlister, F. A., Cleland, J. G., & Stewart, S. (2007). Telemonitoring or structured telephone support programmes for patients with chronic heart failure: Systematic review and meta-analysis. BMJ (Clinical Research Ed), 334(7600), 942.CrossRef
17.
go back to reference Brugada, P. (2006). What evidence do we have to replace in-hospital implantable cardioverter defibrillator follow-up? Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(Suppl 3), iii3–iii9.PubMed Brugada, P. (2006). What evidence do we have to replace in-hospital implantable cardioverter defibrillator follow-up? Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(Suppl 3), iii3–iii9.PubMed
18.
go back to reference Ricci, R. P., Russo, M., & Santini, M. (2006). Management of atrial fibrillation: What are the possibilities of early detection with Home Monitoring? Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(Suppl 3), iii10–iii16.PubMed Ricci, R. P., Russo, M., & Santini, M. (2006). Management of atrial fibrillation: What are the possibilities of early detection with Home Monitoring? Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(Suppl 3), iii10–iii16.PubMed
19.
go back to reference Varma, N., Stambler, B., & Chun, S. (2005). Detection of atrial fibrillation by implanted devices with wireless data transmission capability. Pacing and Clinical Electrophysiology, 28(Suppl. 1), S133–S136.PubMedCrossRef Varma, N., Stambler, B., & Chun, S. (2005). Detection of atrial fibrillation by implanted devices with wireless data transmission capability. Pacing and Clinical Electrophysiology, 28(Suppl. 1), S133–S136.PubMedCrossRef
20.
go back to reference Ricci, R. P., Morichelli, L., & Santini, M. (2008). Home monitoring remote control of pacemaker and implantable cardioverter defibrillator patients in clinical practice: Impact on medical management and health-care resource utilization. Europace, 10(2), 164–170.PubMedCrossRef Ricci, R. P., Morichelli, L., & Santini, M. (2008). Home monitoring remote control of pacemaker and implantable cardioverter defibrillator patients in clinical practice: Impact on medical management and health-care resource utilization. Europace, 10(2), 164–170.PubMedCrossRef
21.
go back to reference Calcagnini, G., Censi, F., Floris, M., Pignalberi, C., Ricci, R., Biancalana, G., et al. (2006). Evaluation of electromagnetic interference of GSM mobile phones with pacemakers featuring remote monitoring functions. Pacing and Clinical Electrophysiology, 29(4), 380–385.PubMedCrossRef Calcagnini, G., Censi, F., Floris, M., Pignalberi, C., Ricci, R., Biancalana, G., et al. (2006). Evaluation of electromagnetic interference of GSM mobile phones with pacemakers featuring remote monitoring functions. Pacing and Clinical Electrophysiology, 29(4), 380–385.PubMedCrossRef
22.
go back to reference Elsner, C., Sommer, P., Piorkowski, C., Taborsky, M., Neuser, H., Bytesnik, J., et al. (2006). A prospective multicenter comparison trial of Home Monitoring against regular follow-up in MADIT II patients: additional visits and cost impact. Computers in Cardiology, 33, 241–244. Elsner, C., Sommer, P., Piorkowski, C., Taborsky, M., Neuser, H., Bytesnik, J., et al. (2006). A prospective multicenter comparison trial of Home Monitoring against regular follow-up in MADIT II patients: additional visits and cost impact. Computers in Cardiology, 33, 241–244.
23.
go back to reference Varma, N. (2007). Rationale and design of a prospective study of the efficacy of a remote monitoring system used in implantable cardioverter defibrillator follow-up: The Lumos-T Reduces Routine Office Device Follow-Up Study (TRUST) study. American Heart Journal, 154(6), 1029–1034.PubMedCrossRef Varma, N. (2007). Rationale and design of a prospective study of the efficacy of a remote monitoring system used in implantable cardioverter defibrillator follow-up: The Lumos-T Reduces Routine Office Device Follow-Up Study (TRUST) study. American Heart Journal, 154(6), 1029–1034.PubMedCrossRef
24.
go back to reference Fauchier, L., Sadoul, N., Kouakam, C., Briand, F., Chauvin, M., Babuty, D., et al. (2005). Potential cost savings by telemedicine-assisted long-term care of implantable cardioverter defibrillator recipients. Pacing and Clinical Electrophysiology, 28(Suppl. 1), S255–S259.PubMedCrossRef Fauchier, L., Sadoul, N., Kouakam, C., Briand, F., Chauvin, M., Babuty, D., et al. (2005). Potential cost savings by telemedicine-assisted long-term care of implantable cardioverter defibrillator recipients. Pacing and Clinical Electrophysiology, 28(Suppl. 1), S255–S259.PubMedCrossRef
26.
go back to reference Neuzil, P., Taborsky, M., Holy, F., & Wallbrueck, K. (2008). Early automatic remote detection of combined lead insulation defect and ICD damage. Europace, 10(5), 556–557.PubMedCrossRef Neuzil, P., Taborsky, M., Holy, F., & Wallbrueck, K. (2008). Early automatic remote detection of combined lead insulation defect and ICD damage. Europace, 10(5), 556–557.PubMedCrossRef
27.
go back to reference Spencker, S., Mueller, D., Marek, A., & Zabel, M. (2007). Severe pacemaker lead perforation detected by an automatic home-monitoring system. European Heart Journal, 28(12), 1432.PubMedCrossRef Spencker, S., Mueller, D., Marek, A., & Zabel, M. (2007). Severe pacemaker lead perforation detected by an automatic home-monitoring system. European Heart Journal, 28(12), 1432.PubMedCrossRef
28.
go back to reference Zartner, P. A., Handke, R. P., Brecher, A. M., & Schneider, M. B. (2007). Integrated Home Monitoring predicts lead failure in a pacemaker dependent 4-year-old girl. Europace, 9(3), 192–193.PubMedCrossRef Zartner, P. A., Handke, R. P., Brecher, A. M., & Schneider, M. B. (2007). Integrated Home Monitoring predicts lead failure in a pacemaker dependent 4-year-old girl. Europace, 9(3), 192–193.PubMedCrossRef
29.
go back to reference Scholten, M. F., Thornton, A. S., Theuns, D. A., Res, J., & Jordaens, L. J. (2004). Twiddler’s syndrome detected by Home Monitoring device. Pacing and Clinical Electrophysiology, 27(8), 1151–1152.PubMedCrossRef Scholten, M. F., Thornton, A. S., Theuns, D. A., Res, J., & Jordaens, L. J. (2004). Twiddler’s syndrome detected by Home Monitoring device. Pacing and Clinical Electrophysiology, 27(8), 1151–1152.PubMedCrossRef
30.
go back to reference Siaplaouras, S., Buob, A., Neuberger, H. R., & Mewis, C. (2006). Remote detection of incessant slow VT with an ICD capable of Home Monitoring. Europace, 8(7), 512–514.PubMedCrossRef Siaplaouras, S., Buob, A., Neuberger, H. R., & Mewis, C. (2006). Remote detection of incessant slow VT with an ICD capable of Home Monitoring. Europace, 8(7), 512–514.PubMedCrossRef
31.
go back to reference Koos, R., Sinha, A. M., & Stellbrink, C. (2005). Home Monitoring in an ICD patient with incessant ventricular tachycardia. Zeitschrift fur Kardiologie, 94(7), 461–464.PubMedCrossRef Koos, R., Sinha, A. M., & Stellbrink, C. (2005). Home Monitoring in an ICD patient with incessant ventricular tachycardia. Zeitschrift fur Kardiologie, 94(7), 461–464.PubMedCrossRef
32.
go back to reference Chen, J., Wilkoff, B. L., Choucair, W., Cohen, T. J., Crossley, G. H., Johnson, W. B., et al. (2008). Design of the Pacemaker REmote Follow-up Evaluation and Review (PREFER) trial to assess the clinical value of the remote pacemaker interrogation in the management of pacemaker patients. Trials, 9, 18.PubMedCrossRef Chen, J., Wilkoff, B. L., Choucair, W., Cohen, T. J., Crossley, G. H., Johnson, W. B., et al. (2008). Design of the Pacemaker REmote Follow-up Evaluation and Review (PREFER) trial to assess the clinical value of the remote pacemaker interrogation in the management of pacemaker patients. Trials, 9, 18.PubMedCrossRef
34.
go back to reference Yu, C. M., Wang, L., Chau, E., Chan, R. H., Kong, S. L., Tang, M. O., et al. (2005). Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation, 112(6), 841–848.PubMedCrossRef Yu, C. M., Wang, L., Chau, E., Chan, R. H., Kong, S. L., Tang, M. O., et al. (2005). Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation, 112(6), 841–848.PubMedCrossRef
37.
go back to reference Saxon, L. A., Boehmer, J. P., Neuman, S., & Mullin, C. M. (2007). Remote Active Monitoring in Patients with Heart Failure (RAPID-RF): Design and rationale. Journal of Cardiac Failure, 13(4), 241–246.PubMedCrossRef Saxon, L. A., Boehmer, J. P., Neuman, S., & Mullin, C. M. (2007). Remote Active Monitoring in Patients with Heart Failure (RAPID-RF): Design and rationale. Journal of Cardiac Failure, 13(4), 241–246.PubMedCrossRef
38.
go back to reference Carlson, M. D., Wilkoff, B. L., Maisel, W. H., Carlson, M. D., Ellenbogen, K. A., Saxon, L. A., et al. (2006). Recommendations from the Heart Rhythm Society Task Force on Device Performance Policies and Guidelines Endorsed by the American College of Cardiology Foundation (ACCF) and the American Heart Association (AHA) and the International Coalition of Pacing and Electrophysiology Organizations (COPE). Heart Rhythm, 3(10), 1250–1273.PubMedCrossRef Carlson, M. D., Wilkoff, B. L., Maisel, W. H., Carlson, M. D., Ellenbogen, K. A., Saxon, L. A., et al. (2006). Recommendations from the Heart Rhythm Society Task Force on Device Performance Policies and Guidelines Endorsed by the American College of Cardiology Foundation (ACCF) and the American Heart Association (AHA) and the International Coalition of Pacing and Electrophysiology Organizations (COPE). Heart Rhythm, 3(10), 1250–1273.PubMedCrossRef
39.
go back to reference Ruffy, R. (2007). The device recalls nightmare: Help might be on the way. Pacing and Clinical Electrophysiology, 30(Suppl 1), S1.PubMed Ruffy, R. (2007). The device recalls nightmare: Help might be on the way. Pacing and Clinical Electrophysiology, 30(Suppl 1), S1.PubMed
40.
go back to reference Boriani, G., Diemberger, I., Martignani, C., Biffi, M., Valzania, C., Bertini, M., et al. (2008). Telecardiology and remote monitoring of implanted electrical devices: The potential for fresh clinical care perspectives. Journal of General Internal Medicine, 23(Suppl 1), 73–77.PubMedCrossRef Boriani, G., Diemberger, I., Martignani, C., Biffi, M., Valzania, C., Bertini, M., et al. (2008). Telecardiology and remote monitoring of implanted electrical devices: The potential for fresh clinical care perspectives. Journal of General Internal Medicine, 23(Suppl 1), 73–77.PubMedCrossRef
41.
go back to reference Heidbuchel, H., Lioen, P., Foulon, S., Huybrechts, W., Ector, J., Willems, R., et al. (2008). Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator. Europace, 10(3), 351–357.PubMedCrossRef Heidbuchel, H., Lioen, P., Foulon, S., Huybrechts, W., Ector, J., Willems, R., et al. (2008). Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator. Europace, 10(3), 351–357.PubMedCrossRef
42.
go back to reference Res, J. C., Theuns, D. A., & Jordaens, L. (2006). The role of remote monitoring in the reduction of inappropriate implantable cardioverter defibrillator therapies. Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(Suppl 3), iii17–iii21.PubMed Res, J. C., Theuns, D. A., & Jordaens, L. (2006). The role of remote monitoring in the reduction of inappropriate implantable cardioverter defibrillator therapies. Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(Suppl 3), iii17–iii21.PubMed
43.
go back to reference Page, E., Cazeau, S., Ritter, P., Galley, D., & Casset, C. (2007). Physiological approach to monitor patients in congestive heart failure: Application of a new implantable device-based system to monitor daily life activity and ventilation. Europace, 9(8), 687–693.PubMedCrossRef Page, E., Cazeau, S., Ritter, P., Galley, D., & Casset, C. (2007). Physiological approach to monitor patients in congestive heart failure: Application of a new implantable device-based system to monitor daily life activity and ventilation. Europace, 9(8), 687–693.PubMedCrossRef
44.
go back to reference Halimi, F., Clementy, J., Attuel, P., Dessenne, X., & Amara, W. (2007). Post-implant fast hospital discharge: The Oedipe trial. Giornale Italiano di Aritmologia e Cardiostimolazione, 10(3, Suppl 2), 20, Abstract. Halimi, F., Clementy, J., Attuel, P., Dessenne, X., & Amara, W. (2007). Post-implant fast hospital discharge: The Oedipe trial. Giornale Italiano di Aritmologia e Cardiostimolazione, 10(3, Suppl 2), 20, Abstract.
45.
go back to reference Deharo, J. C., & Djiane, P. (2006). Home Monitoring: what can we expect in the future? Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(Suppl 3), iii36–iii39.PubMed Deharo, J. C., & Djiane, P. (2006). Home Monitoring: what can we expect in the future? Clinical Research in Cardiology; Official Journal of the German Cardiac Society, 95(Suppl 3), iii36–iii39.PubMed
46.
go back to reference Steinhaus, D., Reynolds, D. W., Gadler, F., Kay, G. N., Hess, M. F., & Bennett, T. (2005). Implant experience with an implantable hemodynamic monitor for the management of symptomatic heart failure. Pacing and Clinical Electrophysiology, 28(8), 747–753.PubMedCrossRef Steinhaus, D., Reynolds, D. W., Gadler, F., Kay, G. N., Hess, M. F., & Bennett, T. (2005). Implant experience with an implantable hemodynamic monitor for the management of symptomatic heart failure. Pacing and Clinical Electrophysiology, 28(8), 747–753.PubMedCrossRef
47.
go back to reference Bourge, R. C., Abraham, W. T., Adamson, P. B., Aaron, M. F., Aranda, J. M., Magalski, A., et al. (2008). Randomized controlled trial of an implantable continuous hemodynamic monitor in patients with advanced heart failure: The COMPASS-HF study. Journal of the American College of Cardiology, 51(11), 1073–1079.PubMedCrossRef Bourge, R. C., Abraham, W. T., Adamson, P. B., Aaron, M. F., Aranda, J. M., Magalski, A., et al. (2008). Randomized controlled trial of an implantable continuous hemodynamic monitor in patients with advanced heart failure: The COMPASS-HF study. Journal of the American College of Cardiology, 51(11), 1073–1079.PubMedCrossRef
48.
go back to reference Zima, E., Lippert, M., Czygan, G., & Merkely, B. (2006). Determination of left ventricular volume changes by intracardiac conductance using a biventricular electrode configuration. Europace, 8(7), 537–544.PubMedCrossRef Zima, E., Lippert, M., Czygan, G., & Merkely, B. (2006). Determination of left ventricular volume changes by intracardiac conductance using a biventricular electrode configuration. Europace, 8(7), 537–544.PubMedCrossRef
49.
go back to reference Stahl, C., Beierlein, W., Walker, T., Straub, A., Nagy, Z., Knubben, K., et al. (2007). Intracardiac impedance monitors hemodynamic deterioration in a chronic heart failure pig model. Journal of Cardiovascular Electrophysiology, 18(9), 985–990.PubMedCrossRef Stahl, C., Beierlein, W., Walker, T., Straub, A., Nagy, Z., Knubben, K., et al. (2007). Intracardiac impedance monitors hemodynamic deterioration in a chronic heart failure pig model. Journal of Cardiovascular Electrophysiology, 18(9), 985–990.PubMedCrossRef
50.
go back to reference Kjellstrom, B., Linde, C., Bennett, T., Ohlsson, A., & Ryden, L. (2004). Six years follow-up of an implanted SvO(2) sensor in the right ventricle. European Journal of Heart Failure, 6(5), 627–634.PubMed Kjellstrom, B., Linde, C., Bennett, T., Ohlsson, A., & Ryden, L. (2004). Six years follow-up of an implanted SvO(2) sensor in the right ventricle. European Journal of Heart Failure, 6(5), 627–634.PubMed
51.
go back to reference Delnoy, P. P., Marcelli, E., Oudeluttikhuis, H., Nicastia, D., Renesto, F., Cercenelli, L., et al. (2008). Validation of a peak endocardial acceleration-based algorithm to optimize cardiac resynchronization: Early clinical results. Europace, 10(7), 801–808.PubMedCrossRef Delnoy, P. P., Marcelli, E., Oudeluttikhuis, H., Nicastia, D., Renesto, F., Cercenelli, L., et al. (2008). Validation of a peak endocardial acceleration-based algorithm to optimize cardiac resynchronization: Early clinical results. Europace, 10(7), 801–808.PubMedCrossRef
52.
go back to reference Asbach, S., Weiss, I., Wenzel, B., Bode, C., & Zehender, M. (2006). Intrathoracic far-field electrocardiogram allows continuous monitoring of ischemia after total coronary occlusion. Pacing and Clinical Electrophysiology, 29(12), 1334–1340.PubMedCrossRef Asbach, S., Weiss, I., Wenzel, B., Bode, C., & Zehender, M. (2006). Intrathoracic far-field electrocardiogram allows continuous monitoring of ischemia after total coronary occlusion. Pacing and Clinical Electrophysiology, 29(12), 1334–1340.PubMedCrossRef
Metadata
Title
Advances in remote monitoring of implantable pacemakers, cardioverter defibrillators and cardiac resynchronization therapy systems
Authors
Werner Jung
Andreas Rillig
Ralf Birkemeyer
Tomislav Miljak
Udo Meyerfeldt
Publication date
01-10-2008
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 1/2008
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-008-9311-5

Other articles of this Issue 1/2008

Journal of Interventional Cardiac Electrophysiology 1/2008 Go to the issue