Published in: International Ophthalmology 2/2016

01-04-2016 | Original Paper

The effect of pupil dilation on AL-Scan biometric parameters

Authors: Ertuğrul Can, Mustafa Duran, Tuğba Çetinkaya, Nurşen Arıtürk

Published in: International Ophthalmology | Issue 2/2016

Login to get access

Abstract

The purpose of this study was to investigate the effects of pupil dilation on the parameters of the AL-Scan (Nidek Co., Ltd, Gamagori, Japan). We compared the measurements of axial length (AL), anterior chamber depth (ACD), central corneal keratometry reading, pupil diameter, and intraocular lens (IOL) power of 72 eyes of 72 healthy volunteers and patients scheduled for cataract surgery before and 45 min after instillation of cyclopentolate hydrochloride 1 % using the AL-Scan. Intraobserver repeatability was assessed by taking three consecutive recordings of ACD and AL. Only ACD readings were significantly different between predilation and postdilation (P < 0.001). The difference of the other measurements between two sessions was not statistically significant (P > 0.001). Only two cases in the study demonstrated changes in IOL power higher than 0.5 D. The intraobserver repeatability of both devices was good (CV values for ACD and AL were 0.16 and 0.20 %, respectively). Dilated pupil size did not affect the measurement of IOL power using the A-Scan optical biometer, but increase in ACD after dilation should be taken into account when performing refractive surgeries in which ACD is very important such as phakic anterior chamber IOL implantation.
Literature
1.
go back to reference Haigis W, Lege B, Miller N, Schneider B (2000) Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol 238:765–773CrossRefPubMed Haigis W, Lege B, Miller N, Schneider B (2000) Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol 238:765–773CrossRefPubMed
2.
go back to reference Gallinaro C, Toulemont PJ, Cochener B (1996) Excimer laser photorefractive keratectomy to correct astigmatism. J Cataract Refract Surg 22:557–563CrossRefPubMed Gallinaro C, Toulemont PJ, Cochener B (1996) Excimer laser photorefractive keratectomy to correct astigmatism. J Cataract Refract Surg 22:557–563CrossRefPubMed
3.
go back to reference Buckhurst PJ, Wolffsohn JS, Shah S (2009) A new optical low coherence reflectometry device for ocular biometry in cataract patients. Br J Ophthalmol 93:949–953CrossRefPubMed Buckhurst PJ, Wolffsohn JS, Shah S (2009) A new optical low coherence reflectometry device for ocular biometry in cataract patients. Br J Ophthalmol 93:949–953CrossRefPubMed
4.
go back to reference Holzer MP, Mamusa M, Auffarth GU (2009) Accuracy of a new partial coherence interferometry analyser for biometric measurements. Br J Ophthalmol 93:807–810CrossRefPubMed Holzer MP, Mamusa M, Auffarth GU (2009) Accuracy of a new partial coherence interferometry analyser for biometric measurements. Br J Ophthalmol 93:807–810CrossRefPubMed
5.
go back to reference Tehrani M, Krummenauer F, Blom E (2003) Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes. J Cataract Refract Surg 29:741–746CrossRefPubMed Tehrani M, Krummenauer F, Blom E (2003) Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes. J Cataract Refract Surg 29:741–746CrossRefPubMed
6.
go back to reference Drexler W, Findl O, Menapace R (1998) Partial coherence interferometry: a novel approach to biometry in cataract surgery. Am J Ophthalmol 126:524–534CrossRefPubMed Drexler W, Findl O, Menapace R (1998) Partial coherence interferometry: a novel approach to biometry in cataract surgery. Am J Ophthalmol 126:524–534CrossRefPubMed
7.
go back to reference Sadiq SA, McElvanney AM (1996) Pupillary dilation and axial length measurement for preoperative assessment of intraocular lens power. Eur J Ophthalmol 6:147–149PubMed Sadiq SA, McElvanney AM (1996) Pupillary dilation and axial length measurement for preoperative assessment of intraocular lens power. Eur J Ophthalmol 6:147–149PubMed
8.
go back to reference Bakbak B, Koktekir BE, Gedik S (2013) The effect of pupil dilation on biometric parameters of the Lenstar 900. Cornea 32(4):e21–e24CrossRefPubMed Bakbak B, Koktekir BE, Gedik S (2013) The effect of pupil dilation on biometric parameters of the Lenstar 900. Cornea 32(4):e21–e24CrossRefPubMed
9.
go back to reference Wendt M, Croft MA, McDonald J (2008) Lens diameter and thickness as a function of age and pharmacologically stimulated accommodation in rhesus monkeys. Exp Eye Res 86:746–752CrossRefPubMedPubMedCentral Wendt M, Croft MA, McDonald J (2008) Lens diameter and thickness as a function of age and pharmacologically stimulated accommodation in rhesus monkeys. Exp Eye Res 86:746–752CrossRefPubMedPubMedCentral
11.
go back to reference Palamar M, Egrilmez S, Uretmen O (2011) Influences of cyclopentolate hydrochloride on anterior segment parameters with Pentacam in children. Acta Ophthalmol 89:e461–e465CrossRefPubMed Palamar M, Egrilmez S, Uretmen O (2011) Influences of cyclopentolate hydrochloride on anterior segment parameters with Pentacam in children. Acta Ophthalmol 89:e461–e465CrossRefPubMed
12.
go back to reference Arriola-Villalobos P, Diaz-Valle D (2014) Effect of pharmacologic pupil dilation on OLCR optical biometry measurements for IOL predictions. Eur J Ophthalmol 24:53–57CrossRefPubMed Arriola-Villalobos P, Diaz-Valle D (2014) Effect of pharmacologic pupil dilation on OLCR optical biometry measurements for IOL predictions. Eur J Ophthalmol 24:53–57CrossRefPubMed
13.
go back to reference Holladay JT (1997) Standardizing constants for ultrasonic biometry, keratometry, and intraocular lens power calculations. J Cataract Refract Surg 23:1356–1370CrossRefPubMed Holladay JT (1997) Standardizing constants for ultrasonic biometry, keratometry, and intraocular lens power calculations. J Cataract Refract Surg 23:1356–1370CrossRefPubMed
14.
go back to reference Olsen T, Corydon L, Gimbel H (1995) Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm. J Cataract Refract Surg 21:313–319CrossRefPubMed Olsen T, Corydon L, Gimbel H (1995) Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm. J Cataract Refract Surg 21:313–319CrossRefPubMed
15.
go back to reference Norn M (1985) The effects of drugs on tear flow. Trans Ophthalmol Soc UK 104(Pt 4):410–414PubMed Norn M (1985) The effects of drugs on tear flow. Trans Ophthalmol Soc UK 104(Pt 4):410–414PubMed
Metadata
Title
The effect of pupil dilation on AL-Scan biometric parameters
Authors
Ertuğrul Can
Mustafa Duran
Tuğba Çetinkaya
Nurşen Arıtürk
Publication date
01-04-2016
Publisher
Springer Netherlands
Published in
International Ophthalmology / Issue 2/2016
Print ISSN: 0165-5701
Electronic ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-015-0097-z

Other articles of this Issue 2/2016

International Ophthalmology 2/2016 Go to the issue