01-02-2014 | Original Paper

Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers

Authors: Paola De la Parra-Colín, Manuel Garza-León, Tonatiuh Barrientos-Gutierrez

Published in: International Ophthalmology | Issue 1/2014

Login to get access

Abstract

To assess the repeatability and comparability of six anterior segment biometry parameters obtained with a novel Scheimpflug camera with a Placido disc topographer (Sirius) and slit-scanning tomography with a Scheimpflug camera (Pentacam), in a sample of 16 unoperated eyes of healthy subjects. The anterior segment was analyzed by a single examiner using the Sirius and the Pentacam analyzers. Mean simulated keratometry (Sim K), flat and steep axis keratometry (K f and K s ), central and thinnest corneal thicknesses (CCT and TCT), and anterior chamber depth (ACD) measurements were evaluated. Repeatability of three sets of measurements from each device was assessed using the coefficient of variation (CV), within-subject standard deviation, and intraclass correlation coefficient (ICC). Bonferroni-adjusted t-tests, and Bland and Altman plots were used to establish agreement between devices. For both devices the CV of repeated measurements was <0.79 %. The ICC was >0.95 in all measurements except for Sirius K s (ICC = 0.869). For all parameters evaluated, the Pentacam systematically yielded higher values, although differences were statistically significant in only three parameters—0.31 diopters for K s , 10.1 μm for CCT and 12.4 μm for TCT. In the assessment of normal corneas both devices showed overall high repeatability. Although good agreement was found in three parameters (Sim K, K f and ACD) these devices do not seem to be interchangeable for pachymetric determination.
Literature
5.
go back to reference Kawamorita T, Uozato H, Kamiya K, Bax L, Tsutsui K, Aizawa D, Shimizu K (2009) Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement. J Cataract Refract Surg 35:127–133. doi:10.1016/j.jcrs.2008.10.019 PubMedCrossRef Kawamorita T, Uozato H, Kamiya K, Bax L, Tsutsui K, Aizawa D, Shimizu K (2009) Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement. J Cataract Refract Surg 35:127–133. doi:10.​1016/​j.​jcrs.​2008.​10.​019 PubMedCrossRef
6.
go back to reference De Sanctis U, Missolungi A, Mutani B, Richiardi L, Grignolo FM (2007) Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol 144:712–718. doi:10.1016/j.ajo.2007.07.021 PubMedCrossRef De Sanctis U, Missolungi A, Mutani B, Richiardi L, Grignolo FM (2007) Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol 144:712–718. doi:10.​1016/​j.​ajo.​2007.​07.​021 PubMedCrossRef
11.
go back to reference Al-Mezaine HS, Al-Amro SA, Kangave D, Sadaawy A, Wehaib TA, Al-Obeidan S (2008) Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry. Int Ophthalmol 28:333–338. doi:10.1007/s10792-007-9143-9 PubMedCrossRef Al-Mezaine HS, Al-Amro SA, Kangave D, Sadaawy A, Wehaib TA, Al-Obeidan S (2008) Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry. Int Ophthalmol 28:333–338. doi:10.​1007/​s10792-007-9143-9 PubMedCrossRef
17.
go back to reference Menassa N, Kaufmann C, Goggin M, Job OM, Bachmann LM, Thiel MA (2008) Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems. J Cataract Refract Surg 34:1742–1747. doi:10.1016/j.jcrs.2008.06.024 PubMedCrossRef Menassa N, Kaufmann C, Goggin M, Job OM, Bachmann LM, Thiel MA (2008) Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems. J Cataract Refract Surg 34:1742–1747. doi:10.​1016/​j.​jcrs.​2008.​06.​024 PubMedCrossRef
18.
Metadata
Title
Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers
Authors
Paola De la Parra-Colín
Manuel Garza-León
Tonatiuh Barrientos-Gutierrez
Publication date
01-02-2014
Publisher
Springer Netherlands
Published in
International Ophthalmology / Issue 1/2014
Print ISSN: 0165-5701
Electronic ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-013-9780-0

Other articles of this Issue 1/2014

International Ophthalmology 1/2014 Go to the issue