Published in: Journal of Medical Ultrasonics 3/2016

01-07-2016 | Review Article

Twinkling artifact on color Doppler ultrasound: an advantage or a pitfall?

Authors: Ebru Ozan, Gokce Kaan Atac, Sadi Gundogdu

Published in: Journal of Medical Ultrasonics | Issue 3/2016

Login to get access

Abstract

The twinkling artifact (TA) or color comet-tail artifact is characterized by a rapidly changing mixture of red and blue color Doppler signals. Even though many diseases and clinical conditions have been shown to produce this artifact, its source is not clearly understood yet. The TA may provide additional information to gray-scale ultrasound findings in several clinical situations. However, there may be pitfalls to keep in mind. We must first be aware of the TA to benefit from the advantages and avoid the pitfalls. In this review, we aim to give practicing radiologists an overview of the mechanisms and clinical applications of the TA by illustrating sample cases we have encountered.
Literature
1.
go back to reference Rahmouni A, Bargoin R, Herment A, et al. Color Doppler twinkling artifact in hyperechoic regions. Radiology. 1996;199:269–71.CrossRefPubMed Rahmouni A, Bargoin R, Herment A, et al. Color Doppler twinkling artifact in hyperechoic regions. Radiology. 1996;199:269–71.CrossRefPubMed
2.
go back to reference Chelfouh N, Grenier N, Higueret D, et al. Characterization of urinary calculi: in vitro study of “twinkling artifact” revealed by color-flow sonography. AJR Am J Roentgenol. 1998;171:1055–60.CrossRefPubMed Chelfouh N, Grenier N, Higueret D, et al. Characterization of urinary calculi: in vitro study of “twinkling artifact” revealed by color-flow sonography. AJR Am J Roentgenol. 1998;171:1055–60.CrossRefPubMed
3.
go back to reference Aytac SK, Ozcan H. Effect of color Doppler system on the twinkling sign associated with urinary tract calculi. J Clin Ultrasound. 1999;27:433–9.CrossRefPubMed Aytac SK, Ozcan H. Effect of color Doppler system on the twinkling sign associated with urinary tract calculi. J Clin Ultrasound. 1999;27:433–9.CrossRefPubMed
4.
go back to reference Rubaltelli L, Khadivi Y, Stramare R, et al. Power Doppler signals produced by static structures: a frequent cause of interpretation errors in the study of slow flows. Radiol Med. 2000;99:161–4.PubMed Rubaltelli L, Khadivi Y, Stramare R, et al. Power Doppler signals produced by static structures: a frequent cause of interpretation errors in the study of slow flows. Radiol Med. 2000;99:161–4.PubMed
5.
go back to reference Gao J, Hentel K, Rubin JM. Correlation between twinkling artifact and color Doppler carrier frequency: preliminary observations in renal calculi. Ultrasound Med Biol. 2012;38:1534–9.CrossRefPubMed Gao J, Hentel K, Rubin JM. Correlation between twinkling artifact and color Doppler carrier frequency: preliminary observations in renal calculi. Ultrasound Med Biol. 2012;38:1534–9.CrossRefPubMed
6.
go back to reference Kamaya A, Tuthill T, Rubin JM. Twinkling artifact on color Doppler sonography: dependence on machine parameters and underlying cause. AJR Am J Roentgenol. 2003;180:215–22.CrossRefPubMed Kamaya A, Tuthill T, Rubin JM. Twinkling artifact on color Doppler sonography: dependence on machine parameters and underlying cause. AJR Am J Roentgenol. 2003;180:215–22.CrossRefPubMed
7.
go back to reference Lu W, Sapozhnikov OA, Bailey MR, et al. Evidence for trapped surface bubbles as the cause for the twinkling artifact in ultrasound imaging. Ultrasound Med Biol. 2013;39:1026–38.CrossRefPubMedPubMedCentral Lu W, Sapozhnikov OA, Bailey MR, et al. Evidence for trapped surface bubbles as the cause for the twinkling artifact in ultrasound imaging. Ultrasound Med Biol. 2013;39:1026–38.CrossRefPubMedPubMedCentral
8.
go back to reference Liu L, Funamoto K, Ozawa K, et al. In vitro study of ultrasound radiation force-driven twinkling sign using PVA-H gel and glass beads tissue-mimicking phantom. J Med Ultrasonics. 2013;40:197–203.CrossRef Liu L, Funamoto K, Ozawa K, et al. In vitro study of ultrasound radiation force-driven twinkling sign using PVA-H gel and glass beads tissue-mimicking phantom. J Med Ultrasonics. 2013;40:197–203.CrossRef
9.
go back to reference Campbell SC, Cullinan JA, Rubens DJ. Slow flow or no flow? Color and power Doppler US pitfalls in the abdomen and pelvis. Radiographics. 2004;24:497–506.CrossRefPubMed Campbell SC, Cullinan JA, Rubens DJ. Slow flow or no flow? Color and power Doppler US pitfalls in the abdomen and pelvis. Radiographics. 2004;24:497–506.CrossRefPubMed
10.
go back to reference Kim HJ, Lee JY, Jang JY, et al. Color Doppler twinkling artifacts from gallstones: in vitro analysis regarding their compositions and architectures. Ultrasound Med Biol. 2010;36:2117–22.CrossRefPubMed Kim HJ, Lee JY, Jang JY, et al. Color Doppler twinkling artifacts from gallstones: in vitro analysis regarding their compositions and architectures. Ultrasound Med Biol. 2010;36:2117–22.CrossRefPubMed
11.
go back to reference Kim HC, Yang DM, Jin W, et al. Color Doppler twinkling artifacts in various conditions during abdominal and pelvic sonography. J Ultrasound Med. 2010;29:621–32.PubMed Kim HC, Yang DM, Jin W, et al. Color Doppler twinkling artifacts in various conditions during abdominal and pelvic sonography. J Ultrasound Med. 2010;29:621–32.PubMed
12.
go back to reference Yoon JH, Cha SS, Han SS, et al. Gallbladder adenomyomatosis: imaging findings. Abdom Imaging. 2006;31:555–63.CrossRefPubMed Yoon JH, Cha SS, Han SS, et al. Gallbladder adenomyomatosis: imaging findings. Abdom Imaging. 2006;31:555–63.CrossRefPubMed
13.
go back to reference Ghersin E, Soudack M, Gaitini D. Twinkling artifact in gallbladder adenomyomatosis. J Ultrasound Med. 2003;22:229–31.PubMed Ghersin E, Soudack M, Gaitini D. Twinkling artifact in gallbladder adenomyomatosis. J Ultrasound Med. 2003;22:229–31.PubMed
14.
go back to reference Tchelepi H, Ralls PW. Color comet-tail artifact: clinical applications. AJR Am J Roentgenol. 2009;192:11–8.CrossRefPubMed Tchelepi H, Ralls PW. Color comet-tail artifact: clinical applications. AJR Am J Roentgenol. 2009;192:11–8.CrossRefPubMed
15.
go back to reference Jeon SJ, Yoon SE. Color Doppler twinkling artifact in hepatic bile duct hamartomas (von Meyenburg complexes). J Ultrasound Med. 2006;25:399–402.PubMed Jeon SJ, Yoon SE. Color Doppler twinkling artifact in hepatic bile duct hamartomas (von Meyenburg complexes). J Ultrasound Med. 2006;25:399–402.PubMed
16.
go back to reference Dillman JR, Kappil M, Weadock WJ, et al. Sonographic twinkling artifact for renal calculus detection: correlation with CT. Radiology. 2011;259:911–6.CrossRefPubMed Dillman JR, Kappil M, Weadock WJ, et al. Sonographic twinkling artifact for renal calculus detection: correlation with CT. Radiology. 2011;259:911–6.CrossRefPubMed
17.
go back to reference Korkmaz M, Aras B, Sanal B, et al. Investigating the clinical significance of twinkling artifacts in patients with urolithiasis smaller than 5 mm. Jpn J Radiol. 2014;32:482–6.CrossRefPubMed Korkmaz M, Aras B, Sanal B, et al. Investigating the clinical significance of twinkling artifacts in patients with urolithiasis smaller than 5 mm. Jpn J Radiol. 2014;32:482–6.CrossRefPubMed
18.
go back to reference Yavuz A, Ceken K, Alimoglu E, Kabaalioglu A. The reliability of color doppler “twinkling” artifact for diagnosing millimetrical nephrolithiasis: comparison with B-Mode US and CT scanning results. J Med Ultrason. 2015;42:215–22.CrossRef Yavuz A, Ceken K, Alimoglu E, Kabaalioglu A. The reliability of color doppler “twinkling” artifact for diagnosing millimetrical nephrolithiasis: comparison with B-Mode US and CT scanning results. J Med Ultrason. 2015;42:215–22.CrossRef
19.
go back to reference Ripollés T, Martínez-Pérez MJ, Vizuete J, et al. Sonographic diagnosis of symptomatic ureteral calculi: usefulness of the twinkling artifact. Abdom Imaging. 2013;38:863–9.CrossRefPubMed Ripollés T, Martínez-Pérez MJ, Vizuete J, et al. Sonographic diagnosis of symptomatic ureteral calculi: usefulness of the twinkling artifact. Abdom Imaging. 2013;38:863–9.CrossRefPubMed
20.
go back to reference Sharma G, Sharma A. Clinical implications and applications of the twinkling sign in ureteral calculus: a preliminary study. J Urol. 2013;189:2132–5.CrossRefPubMed Sharma G, Sharma A. Clinical implications and applications of the twinkling sign in ureteral calculus: a preliminary study. J Urol. 2013;189:2132–5.CrossRefPubMed
21.
go back to reference Wang LC, Sullivan M, Du H, et al. US appearance of ductal carcinoma in situ. Radiographics. 2013;33:213–28.CrossRefPubMed Wang LC, Sullivan M, Du H, et al. US appearance of ductal carcinoma in situ. Radiographics. 2013;33:213–28.CrossRefPubMed
22.
go back to reference Hashimoto Y, Murata A, Miyamoto N, et al. Clinical significance of microcalcifications detection in invasive breast carcinoma. Yonago Acta Med. 2015;58:89–93.PubMedPubMedCentral Hashimoto Y, Murata A, Miyamoto N, et al. Clinical significance of microcalcifications detection in invasive breast carcinoma. Yonago Acta Med. 2015;58:89–93.PubMedPubMedCentral
23.
go back to reference Tsujimoto F. Microcalcifications in the breast detected by a color Doppler method using twinkling artifacts: some important discussions based on clinical cases and experiments with a new ultrasound modality called multidetector-ultrasonography (MD-US). J Med Ultrasonics. 2014;41:99–108.CrossRef Tsujimoto F. Microcalcifications in the breast detected by a color Doppler method using twinkling artifacts: some important discussions based on clinical cases and experiments with a new ultrasound modality called multidetector-ultrasonography (MD-US). J Med Ultrasonics. 2014;41:99–108.CrossRef
24.
go back to reference Frates MC, Benson CB, Charboneau JW, et al. Management of thyroid nodules detected at US: Society of Radiologists in Ultrasound consensus conference statement. Ultrasound Q. 2006;22:231–8 (discussion 239–40). Frates MC, Benson CB, Charboneau JW, et al. Management of thyroid nodules detected at US: Society of Radiologists in Ultrasound consensus conference statement. Ultrasound Q. 2006;22:231–8 (discussion 239–40).
25.
go back to reference Shin LK, Olcott EW, Jeffrey RB, et al. Sonographic evaluation of cervical lymph nodes in papillary thyroid cancer. Ultrasound Q. 2013;29:25–32.CrossRefPubMed Shin LK, Olcott EW, Jeffrey RB, et al. Sonographic evaluation of cervical lymph nodes in papillary thyroid cancer. Ultrasound Q. 2013;29:25–32.CrossRefPubMed
26.
go back to reference Chiou HJ, Chou YH, Chiou SY, et al. High-resolution ultrasonography in superficial soft tissue tumors. J Med Ultrasound. 2007;15:152–74.CrossRef Chiou HJ, Chou YH, Chiou SY, et al. High-resolution ultrasonography in superficial soft tissue tumors. J Med Ultrasound. 2007;15:152–74.CrossRef
28.
29.
go back to reference Oktar SO, Yücel C, Erbaş G, et al. Use of twinkling artifact in sonographic detection of intestinal pneumatosis. Abdom Imaging. 2006;31:293–6.CrossRefPubMed Oktar SO, Yücel C, Erbaş G, et al. Use of twinkling artifact in sonographic detection of intestinal pneumatosis. Abdom Imaging. 2006;31:293–6.CrossRefPubMed
Metadata
Title
Twinkling artifact on color Doppler ultrasound: an advantage or a pitfall?
Authors
Ebru Ozan
Gokce Kaan Atac
Sadi Gundogdu
Publication date
01-07-2016
Publisher
Springer Japan
Published in
Journal of Medical Ultrasonics / Issue 3/2016
Print ISSN: 1346-4523
Electronic ISSN: 1613-2254
DOI
https://doi.org/10.1007/s10396-016-0715-z

Other articles of this Issue 3/2016

Journal of Medical Ultrasonics 3/2016 Go to the issue