Published in: Magnetic Resonance Materials in Physics, Biology and Medicine 3/2015

01-06-2015 | Research Article

In vivo proton magnetic resonance spectroscopic imaging of the healthy human brain at 9.4 T: initial experience

Authors: Grzegorz L. Chadzynski, Rolf Pohmann, Gunamony Shajan, Rupert Kolb, Sotirios Bisdas, Uwe Klose, Klaus Scheffler

Published in: Magnetic Resonance Materials in Physics, Biology and Medicine | Issue 3/2015

Login to get access

Abstract

Object

In this study, the feasibility of in vivo proton magnetic resonance spectroscopic imaging (1H MRSI) of the healthy human brain at a field strength of 9.4 T, using conventional acquisition techniques, is examined and the initial experience is summarized.

Materials and methods

MRSI measurements were performed on a 9.4 T MR scanner (Siemens, Erlangen, Germany) equipped with head-only gradient insert (AC84, Siemens) and custom-developed, 8-channel transmit/24-channel receive, and 16-channel transmit/31-channel receive coils. Spectra were acquired from the superior part of the human brain with a modified STEAM sequence. Spectral quantification was done with LCModel software.

Results

Reasonable quality and signal-to-noise ratio of the acquired spectra allowed reliable quantification of 12 metabolites (Cramer-Rao lower bounds < 20 %), some of which may be difficult to quantify at field strengths below 7 T due to overlapping resonances or low concentrations.

Conclusion

While further developments are necessary to minimize chemical shift displacement and homogeneity of the transmit field, it is demonstrated that in vivo 1H MRSI at a field strength of 9.4 T is possible. However, further studies applying up-to-date techniques to overcome high-field specific problems are needed in order to assess the potential gain in sensitivity that may be offered by MRSI at 9.4 T.
Literature
1.
go back to reference Pohmann R, Shajan G, Balla DZ (2011) Contrast at ultra-high field: relaxation times, magnetization transfer and phase in the rat brain at 16.4 T. Magn Reson Med 66:1572–1581CrossRefPubMed Pohmann R, Shajan G, Balla DZ (2011) Contrast at ultra-high field: relaxation times, magnetization transfer and phase in the rat brain at 16.4 T. Magn Reson Med 66:1572–1581CrossRefPubMed
3.
go back to reference Marjanska M, Auerbach EJ, Valabregue R, de Moortele Van, Adriany G, Garwood M (2012) Localized 1H NMR spectroscopy in different regions of human brain in vivo at 7 T: T2 relaxation times and concentrations of cerebral metabolites. NMR Biomed 25:332–339CrossRefPubMedCentralPubMed Marjanska M, Auerbach EJ, Valabregue R, de Moortele Van, Adriany G, Garwood M (2012) Localized 1H NMR spectroscopy in different regions of human brain in vivo at 7 T: T2 relaxation times and concentrations of cerebral metabolites. NMR Biomed 25:332–339CrossRefPubMedCentralPubMed
4.
go back to reference Henning A, Fuchs A, Murdoch JB, Boesinger P (2009) Slice-selective FID acquisition, localized by outer volume suppression (FIDLOVS) for 1H-MRSI of the human brain at 7 T with minimal signal loss. NMR Biomed 22:683–696CrossRefPubMed Henning A, Fuchs A, Murdoch JB, Boesinger P (2009) Slice-selective FID acquisition, localized by outer volume suppression (FIDLOVS) for 1H-MRSI of the human brain at 7 T with minimal signal loss. NMR Biomed 22:683–696CrossRefPubMed
5.
go back to reference Tkac I, Oez G, Adriany G, Ugurbil K, Gruetter R (2009) In-vivo 1H NMR spectroscopy of the human brain at high magnetic fields: metabolite quantification at 4T vs. 7T. Magn Reson Med 62:868–879CrossRefPubMedCentralPubMed Tkac I, Oez G, Adriany G, Ugurbil K, Gruetter R (2009) In-vivo 1H NMR spectroscopy of the human brain at high magnetic fields: metabolite quantification at 4T vs. 7T. Magn Reson Med 62:868–879CrossRefPubMedCentralPubMed
6.
go back to reference Mekle R, Mlynarik V, Gambarota G, Hergt M, Krueger G, Gruetter R (2009) MR spectroscopy of the human brain with enhanced signal intensity using ultrashort echo times on a clinical platform at 3T and 7T. Magn Reson Med 61:1279–1285CrossRefPubMed Mekle R, Mlynarik V, Gambarota G, Hergt M, Krueger G, Gruetter R (2009) MR spectroscopy of the human brain with enhanced signal intensity using ultrashort echo times on a clinical platform at 3T and 7T. Magn Reson Med 61:1279–1285CrossRefPubMed
7.
go back to reference Yang S, Hu J, Kou Z, Yang Y (2008) Spectral simplification for resolved glutamate and glutamine measurement using a standard STEAM sequence with optimized timing parameters at 3, 4, 4.7, 7 and 9.4 T. Magn Reson Med 59:236–244CrossRefPubMed Yang S, Hu J, Kou Z, Yang Y (2008) Spectral simplification for resolved glutamate and glutamine measurement using a standard STEAM sequence with optimized timing parameters at 3, 4, 4.7, 7 and 9.4 T. Magn Reson Med 59:236–244CrossRefPubMed
8.
go back to reference Choi C, Dimitrov I, Douglas D, Zhao C, Hawesa H, Ghose S, Tamminga CA (2009) In vivo detection of serine in the human brain by proton magnetic resonance spectroscopy (1H MRS) at 7 T. Magn Reson Med 62:1042–1046CrossRefPubMedCentralPubMed Choi C, Dimitrov I, Douglas D, Zhao C, Hawesa H, Ghose S, Tamminga CA (2009) In vivo detection of serine in the human brain by proton magnetic resonance spectroscopy (1H MRS) at 7 T. Magn Reson Med 62:1042–1046CrossRefPubMedCentralPubMed
9.
go back to reference Banerjee A, Ganji S, Hulsey K, Dimitrov I, Maher E, Ghose S, Tamminga C, Choi C (2012) Measurement of glycine in gray and white matter in the human brain in vivo by 1H MRS at 7 T. Magn Reson Med 68:325–331CrossRefPubMedCentralPubMed Banerjee A, Ganji S, Hulsey K, Dimitrov I, Maher E, Ghose S, Tamminga C, Choi C (2012) Measurement of glycine in gray and white matter in the human brain in vivo by 1H MRS at 7 T. Magn Reson Med 68:325–331CrossRefPubMedCentralPubMed
10.
11.
go back to reference Deelchand DK, Van de Moortele PF, Adriany G, Iltis I, Andersen P, Strupp JP, Vaughan T, Ugurbil K, Henry PG (2010) In-vivo 1H NMR spectroscopy of the human brain at 9.4 T: initial results. J Magn Reson 206:74–80CrossRefPubMedCentralPubMed Deelchand DK, Van de Moortele PF, Adriany G, Iltis I, Andersen P, Strupp JP, Vaughan T, Ugurbil K, Henry PG (2010) In-vivo 1H NMR spectroscopy of the human brain at 9.4 T: initial results. J Magn Reson 206:74–80CrossRefPubMedCentralPubMed
12.
go back to reference Tkac I, Andersen P, Adriany G, Merkle H, Ugurbil K, Gruetter R (2001) In vivo 1H NMR spectroscopy of the human brain at 7T. Magn Reson Med 46:451–456CrossRefPubMed Tkac I, Andersen P, Adriany G, Merkle H, Ugurbil K, Gruetter R (2001) In vivo 1H NMR spectroscopy of the human brain at 7T. Magn Reson Med 46:451–456CrossRefPubMed
13.
go back to reference Boer VO, Siero JCW, Hoogduin H, van Gorp JS, Luijten PR, Klomp DWJ (2011) High-field MRS of the human brain at short TE and TR. NMR Biomed 24:1081–1088CrossRefPubMed Boer VO, Siero JCW, Hoogduin H, van Gorp JS, Luijten PR, Klomp DWJ (2011) High-field MRS of the human brain at short TE and TR. NMR Biomed 24:1081–1088CrossRefPubMed
14.
go back to reference Bogner W, Gruber S, Trattnig S, Chmielnik M (2012) High-resolution mapping of human brain metabolites by free induction decay 1H MRSI at 7T. NMR Biomed 25:873–882CrossRefPubMed Bogner W, Gruber S, Trattnig S, Chmielnik M (2012) High-resolution mapping of human brain metabolites by free induction decay 1H MRSI at 7T. NMR Biomed 25:873–882CrossRefPubMed
15.
go back to reference Shajan G, Kozlov M, Hoffman J, Turner R, Scheffler K, Pohmann R (2013) A 16-channel dual-row transmit array in combination with a 31-element receive array for human brain imaging at 9.4 T. Magn Reson Med 71:870–879CrossRef Shajan G, Kozlov M, Hoffman J, Turner R, Scheffler K, Pohmann R (2013) A 16-channel dual-row transmit array in combination with a 31-element receive array for human brain imaging at 9.4 T. Magn Reson Med 71:870–879CrossRef
16.
go back to reference Ogg RJ, Kingsley RB, Taylor JS (1994) WET, a T1 and B1 insensitive water suppression method for in vivo localized 1H NMR spectroscopy. J Magn Reson B 104(1):1–10CrossRefPubMed Ogg RJ, Kingsley RB, Taylor JS (1994) WET, a T1 and B1 insensitive water suppression method for in vivo localized 1H NMR spectroscopy. J Magn Reson B 104(1):1–10CrossRefPubMed
17.
go back to reference Yarnykh VL (2007) Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med 57:192–200CrossRefPubMed Yarnykh VL (2007) Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med 57:192–200CrossRefPubMed
18.
go back to reference Pohmann R, Scheffler K (2013) A theoretical and experimental comparison of different techniques for B1 mapping at very high fields. NMR Biomed 26:265–275CrossRefPubMed Pohmann R, Scheffler K (2013) A theoretical and experimental comparison of different techniques for B1 mapping at very high fields. NMR Biomed 26:265–275CrossRefPubMed
19.
go back to reference De Graaf RA (2007) Radiofrequency Pulses. In: De Graaf RA (ed) In vivo NMR spectroscopy, 2nd edn. Wiley, Chichester, pp 233–293CrossRef De Graaf RA (2007) Radiofrequency Pulses. In: De Graaf RA (ed) In vivo NMR spectroscopy, 2nd edn. Wiley, Chichester, pp 233–293CrossRef
20.
go back to reference Provencher SW (1993) Estimation of metabolite concentrations from localized in vivo proton NMR spectra. Magn Reson Med 30:672–679CrossRefPubMed Provencher SW (1993) Estimation of metabolite concentrations from localized in vivo proton NMR spectra. Magn Reson Med 30:672–679CrossRefPubMed
21.
go back to reference Govindaraju V, Young K, Maudsley AA (2000) Proton NMR chemical shifts and coupling constants for brain metabolites. NMR Biomed 13:129–153CrossRefPubMed Govindaraju V, Young K, Maudsley AA (2000) Proton NMR chemical shifts and coupling constants for brain metabolites. NMR Biomed 13:129–153CrossRefPubMed
22.
go back to reference Tkac I (2008) Refinement of simulated basis set for LCModel analysis. In: Proceedings of the 16th scientific meeting, International Society for Magnetic Resonance in Medicine, Toronto, p 1624 Tkac I (2008) Refinement of simulated basis set for LCModel analysis. In: Proceedings of the 16th scientific meeting, International Society for Magnetic Resonance in Medicine, Toronto, p 1624
23.
go back to reference Ernst T, Kreis R, Ross BD (1993) Absolute quantitation of water and metabolites in the human brain. I. Compartments and water. J Magn Reson Ser B 102:1–8CrossRef Ernst T, Kreis R, Ross BD (1993) Absolute quantitation of water and metabolites in the human brain. I. Compartments and water. J Magn Reson Ser B 102:1–8CrossRef
24.
go back to reference Hong ST, Balla DZ, Shajan G, Choi C, Ugurbil K, Pohmann R (2011) Enhanced neurochemical profile of the rat brain using in vivo 1H NMR spectroscopy at 16.4 T. Magn Reson Med 65:28–34CrossRefPubMed Hong ST, Balla DZ, Shajan G, Choi C, Ugurbil K, Pohmann R (2011) Enhanced neurochemical profile of the rat brain using in vivo 1H NMR spectroscopy at 16.4 T. Magn Reson Med 65:28–34CrossRefPubMed
25.
go back to reference Terpstra M, Ugurbil K, Gruetter R (2002) Direct in vivo measurement of human cerebral GABA concentration using MEGA-editing at 7T. Magn Reson Med 47:1009–1012CrossRefPubMed Terpstra M, Ugurbil K, Gruetter R (2002) Direct in vivo measurement of human cerebral GABA concentration using MEGA-editing at 7T. Magn Reson Med 47:1009–1012CrossRefPubMed
26.
go back to reference Andreychenko A, Boer VO, de Castro CSA, Luijten PR, Klomp DWJ (2012) Efficient spectral editing at 7T: GABA detection with MEGA-sLASER. Magn Reson Med 68:1018–1025CrossRefPubMed Andreychenko A, Boer VO, de Castro CSA, Luijten PR, Klomp DWJ (2012) Efficient spectral editing at 7T: GABA detection with MEGA-sLASER. Magn Reson Med 68:1018–1025CrossRefPubMed
27.
go back to reference Helms G (2000) A precise and user-independent quantification technique for regional comparison of single volume proton MR spectroscopy of the human brain. NMR Biomed 13:398–406CrossRefPubMed Helms G (2000) A precise and user-independent quantification technique for regional comparison of single volume proton MR spectroscopy of the human brain. NMR Biomed 13:398–406CrossRefPubMed
28.
go back to reference Wang Y, Li SJ (1998) Differentiation of metabolic concentrations between gray matter and white matter of human brain by in vivo 1H magnetic resonance spectroscopy. Magn Reson Med 39:28–33CrossRefPubMed Wang Y, Li SJ (1998) Differentiation of metabolic concentrations between gray matter and white matter of human brain by in vivo 1H magnetic resonance spectroscopy. Magn Reson Med 39:28–33CrossRefPubMed
29.
go back to reference Pan JW, Lo KM, Hetherington HP (2012) Role of high order and degree B0 shimming for spectroscopic imaging of the human brain at 7 Tesla. Magn Reson Med 68:1007–1017CrossRefPubMedCentralPubMed Pan JW, Lo KM, Hetherington HP (2012) Role of high order and degree B0 shimming for spectroscopic imaging of the human brain at 7 Tesla. Magn Reson Med 68:1007–1017CrossRefPubMedCentralPubMed
30.
go back to reference Hetherington HP, Chu WJ, Gonen O, Pan JW (2006) Robust fully automated shimming of the human brain for high-field 1H spectroscopic imaging. Magn Reson Med 56:26–33CrossRefPubMed Hetherington HP, Chu WJ, Gonen O, Pan JW (2006) Robust fully automated shimming of the human brain for high-field 1H spectroscopic imaging. Magn Reson Med 56:26–33CrossRefPubMed
31.
go back to reference Shah S, Kellerman P, Greiser A, Weale PJ, Zuehlsdorff S, Jerecic R (2009) Rapid fieldmap estimation for cardiac shimming. In: Proceedings of the 17th scientific meeting, International Society for Magnetic Resonance in Medicine, Honolulu, Hawaii, USA, p 566 Shah S, Kellerman P, Greiser A, Weale PJ, Zuehlsdorff S, Jerecic R (2009) Rapid fieldmap estimation for cardiac shimming. In: Proceedings of the 17th scientific meeting, International Society for Magnetic Resonance in Medicine, Honolulu, Hawaii, USA, p 566
32.
go back to reference Gruetter R (1993) Automatic, localized in vivo adjustment of all first- and second-order shim coils. Magn Reson Med 29:804–811CrossRefPubMed Gruetter R (1993) Automatic, localized in vivo adjustment of all first- and second-order shim coils. Magn Reson Med 29:804–811CrossRefPubMed
33.
go back to reference Gruetter R, Tkac I (2000) Field mapping without reference scan using asymmetric echo-planar techniques. Magn Reson Med 43:319–323CrossRefPubMed Gruetter R, Tkac I (2000) Field mapping without reference scan using asymmetric echo-planar techniques. Magn Reson Med 43:319–323CrossRefPubMed
34.
go back to reference Zhong X, Lyubich YM, De Vito T, Shah S, Knight-Scott J (2012) Improving in vivo 1H-MRS with robust automated shim techniques: a comparison study of FASTESTMAP and GRESHIM. In: Proceedings of the 20th scientific meeting, International Society for Magnetic Resonance in Medicine, Melbourne, Victoria, Australia, p 4397 Zhong X, Lyubich YM, De Vito T, Shah S, Knight-Scott J (2012) Improving in vivo 1H-MRS with robust automated shim techniques: a comparison study of FASTESTMAP and GRESHIM. In: Proceedings of the 20th scientific meeting, International Society for Magnetic Resonance in Medicine, Melbourne, Victoria, Australia, p 4397
35.
go back to reference Zhong X, Lyubich YM, De Vito T, Shah S, Knight-Scott J (2013) Quantitative comparison of shim algorithms for in vivo 1H-MRS. In: Proceedings of the 21th scientific meeting, International Society for Magnetic Resonance in Medicine, Salt Lake City, Utah, USA, p 3988 Zhong X, Lyubich YM, De Vito T, Shah S, Knight-Scott J (2013) Quantitative comparison of shim algorithms for in vivo 1H-MRS. In: Proceedings of the 21th scientific meeting, International Society for Magnetic Resonance in Medicine, Salt Lake City, Utah, USA, p 3988
36.
go back to reference Avdievich NI, Pan JW, Beahring JM, Spencer DD, Hetherington HP (2009) Short echo spectroscopic imaging of the human brain at 7 T using transceiver arrays. Magn Reson Med 62:17–25CrossRefPubMedCentralPubMed Avdievich NI, Pan JW, Beahring JM, Spencer DD, Hetherington HP (2009) Short echo spectroscopic imaging of the human brain at 7 T using transceiver arrays. Magn Reson Med 62:17–25CrossRefPubMedCentralPubMed
37.
go back to reference Andronesi OC, Gagoski BA, Sorensen AG (2012) Neurologic 3D MR spectroscopic imaging with low-power adiabatic pulses and fast spiral acquisition. Radiology 262(2):647–661CrossRefPubMedCentralPubMed Andronesi OC, Gagoski BA, Sorensen AG (2012) Neurologic 3D MR spectroscopic imaging with low-power adiabatic pulses and fast spiral acquisition. Radiology 262(2):647–661CrossRefPubMedCentralPubMed
38.
go back to reference Hetherington HP, Avdievich NI, Kuznetsov AM, Pan JW (2010) RF shimming for spectroscopic localization in the human brain at 7 T. Magn Reson Med 63:9–19PubMedCentralPubMed Hetherington HP, Avdievich NI, Kuznetsov AM, Pan JW (2010) RF shimming for spectroscopic localization in the human brain at 7 T. Magn Reson Med 63:9–19PubMedCentralPubMed
39.
go back to reference Boer VO, van Lier ALHMW, Hoogduin JM, Wijnen JP, Luijten PR, Klomp DWJ (2011) 7-T 1H MRS with adiabatic refocusing at short TE using radiofrequency focusing with a dual-channel volume transmit coil. NMR Biomed 24:1038–1046CrossRefPubMed Boer VO, van Lier ALHMW, Hoogduin JM, Wijnen JP, Luijten PR, Klomp DWJ (2011) 7-T 1H MRS with adiabatic refocusing at short TE using radiofrequency focusing with a dual-channel volume transmit coil. NMR Biomed 24:1038–1046CrossRefPubMed
40.
go back to reference Emir UE, Auerbach EJ, Van de Moortele PF, Marjanska M, Ugurbil K, Terpstra M, Tkac I, Oez G (2012) Regional neurochemical profiles in the human brain measured by 1H MRS at 7 T using local B1 shimming. NMR Biomed 25:152–160CrossRefPubMedCentralPubMed Emir UE, Auerbach EJ, Van de Moortele PF, Marjanska M, Ugurbil K, Terpstra M, Tkac I, Oez G (2012) Regional neurochemical profiles in the human brain measured by 1H MRS at 7 T using local B1 shimming. NMR Biomed 25:152–160CrossRefPubMedCentralPubMed
41.
go back to reference Boer VO, Klomp DWJ, Juchem C, Luijten PR, de Graaf RA (2012) Multislice 1H MRSI of the human brain at 7 T using dynamic B0 and B1 shimming. Magn Reson Med 68:662–670CrossRefPubMedCentralPubMed Boer VO, Klomp DWJ, Juchem C, Luijten PR, de Graaf RA (2012) Multislice 1H MRSI of the human brain at 7 T using dynamic B0 and B1 shimming. Magn Reson Med 68:662–670CrossRefPubMedCentralPubMed
42.
go back to reference Zhu H, Soher B, Ouwerkerk R, Schaer M, Barker PB (2013) Spin-echo magnetic resonance spectroscopic imaging at 7T with frequency-modulated refocusing pulses. Magn Reson Med 69:1217–1225CrossRefPubMedCentralPubMed Zhu H, Soher B, Ouwerkerk R, Schaer M, Barker PB (2013) Spin-echo magnetic resonance spectroscopic imaging at 7T with frequency-modulated refocusing pulses. Magn Reson Med 69:1217–1225CrossRefPubMedCentralPubMed
43.
go back to reference Tkac I, Starcuk Z, Choi IY, Gruetter R (1999) In vivo 1H NMR spectroscopy of rat brain at 1 ms echo time. Magn Reson Med 41:649–656CrossRefPubMed Tkac I, Starcuk Z, Choi IY, Gruetter R (1999) In vivo 1H NMR spectroscopy of rat brain at 1 ms echo time. Magn Reson Med 41:649–656CrossRefPubMed
Metadata
Title
In vivo proton magnetic resonance spectroscopic imaging of the healthy human brain at 9.4 T: initial experience
Authors
Grzegorz L. Chadzynski
Rolf Pohmann
Gunamony Shajan
Rupert Kolb
Sotirios Bisdas
Uwe Klose
Klaus Scheffler
Publication date
01-06-2015
Publisher
Springer Berlin Heidelberg
Published in
Magnetic Resonance Materials in Physics, Biology and Medicine / Issue 3/2015
Print ISSN: 0968-5243
Electronic ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-014-0460-5

Other articles of this Issue 3/2015

Magnetic Resonance Materials in Physics, Biology and Medicine 3/2015 Go to the issue