Published in: Lasers in Medical Science 2/2011

01-03-2011 | Original Article

Effect of low fluency dentin conditioning on tensile bond strength of composite bonded to Er:YAG laser-prepared dentin: a preliminary study

Authors: B. Bahrami, N. Askari, M. Tielemans, D. Heysselaer, L. Lamard, A. Peremans, C. Nyssen-Behets, S. Nammour

Published in: Lasers in Medical Science | Issue 2/2011

Login to get access

Abstract

Several studies in the literature have previously shown that the bond strength of a composite bonded to dentin is almost equivalent as when dentin is prepared by either bur or Er:YAG laser. The aim of this preliminary study is to assess the hypothesis that dentin conditioning at low fluency by means of Er:YAG laser can improve the value of adhesion of composites resin to dentin. Sixty surfaces of caries-free human third molars extracted for orthodontic purposes were randomly divided into five groups of 12 teeth. The bur group was the control, prepared using bur, group L was prepared using Er:YAG 200 mJ, SSP (50 µs), 20 Hz, 15 seconds of sweeping, for groups L80, L100, L120, they were prepared first, with the same parameters of the group L 200, and then they received a conditioning, which is, respectively, 15 s of irradiations at: 80 mJ (SSP, 10 Hz), 100 mJ (SSP, 10 Hz), and 120 mJ (SSP, 10 Hz). All samples were restored in a single-component adhesive system: Xenon (DENTSPLY), and ceramX (DENTSPLY) as the resin composite. The specimens were submitted to tensile bond strength test using a universal testing machine. Data were submitted to statistical analysis using ANOVA coupled to a Tukey-Kramer test at the 95% level. The mean values in MPa were 33.3 for group B, 36.73 for group L 200, 41.7 for group L80, 37.9 for group L100, and 39.1 for group L120. Our results showed that dentin conditioning at a low fluency of 12.58 J/cm2 per pulse, with 80 mJ output energy and 50-µs pulse duration can significantly improve tensile bond strength of a composite bonded to Er:YAG laser-prepared dentine.
Literature
1.
go back to reference Hibst R, Keller U (1989) Experimental studies of the application of the Er:YAG laser on dental hard substances. I. Measurement of the ablation rate. Laser Surg Med 9(4):338–344CrossRef Hibst R, Keller U (1989) Experimental studies of the application of the Er:YAG laser on dental hard substances. I. Measurement of the ablation rate. Laser Surg Med 9(4):338–344CrossRef
2.
go back to reference Ceballos L, Toledano M (2002) Bonding to Er:YAG-laser-treated dentin. J Dent Res 81:119–122CrossRef Ceballos L, Toledano M (2002) Bonding to Er:YAG-laser-treated dentin. J Dent Res 81:119–122CrossRef
3.
go back to reference Visuri SR, Wright DD, Wigdor HA, Walsh TJ (1996) Shear strength of composite bonded to Er:YAG laser prepared dentine. J Dent Res 75:599–655CrossRefPubMed Visuri SR, Wright DD, Wigdor HA, Walsh TJ (1996) Shear strength of composite bonded to Er:YAG laser prepared dentine. J Dent Res 75:599–655CrossRefPubMed
4.
go back to reference Corona SA, De Souza AE (2007) Effect of energy and pulse repetition rate of Er:YAG laser on dentin ablation ability and morphological analysis of the laser-irradiated substrate. Photomed Laser Surg 25(1):26–33CrossRefPubMed Corona SA, De Souza AE (2007) Effect of energy and pulse repetition rate of Er:YAG laser on dentin ablation ability and morphological analysis of the laser-irradiated substrate. Photomed Laser Surg 25(1):26–33CrossRefPubMed
5.
go back to reference Corona SA, Atoui JA (2005) Composite resin’s adhesive resistance to dentin: influence of Er:YAG laser focal distance variation. Photomed Laser Surg 23(2):229–232CrossRefPubMed Corona SA, Atoui JA (2005) Composite resin’s adhesive resistance to dentin: influence of Er:YAG laser focal distance variation. Photomed Laser Surg 23(2):229–232CrossRefPubMed
6.
go back to reference Delme KI, Deman PJ (2006) Influence of different Er:YAG laser energies and frequencies on the surface morphology of dentin and enamel. J Oral Laser Appl 6:43–52 Delme KI, Deman PJ (2006) Influence of different Er:YAG laser energies and frequencies on the surface morphology of dentin and enamel. J Oral Laser Appl 6:43–52
7.
go back to reference Goncalves M, Corona SA (2008) Influence of pulse repetition rate of Er:YAG laser and dentin depth on tensile bond strength of dentin-resin interface. J Biomed Mater Res A 86(2):477–482PubMed Goncalves M, Corona SA (2008) Influence of pulse repetition rate of Er:YAG laser and dentin depth on tensile bond strength of dentin-resin interface. J Biomed Mater Res A 86(2):477–482PubMed
8.
go back to reference Aizawa K, Kameyama A (2006) Resin bonding to dentin irradiated by high repetition rate Er:YAG laser. Photomed Laser Surg 24(3):397–401CrossRefPubMed Aizawa K, Kameyama A (2006) Resin bonding to dentin irradiated by high repetition rate Er:YAG laser. Photomed Laser Surg 24(3):397–401CrossRefPubMed
9.
go back to reference Dunn WJ, Davis JT (2005) Shear bond strength and SEM evaluation of composite bonded to Er:YAG laser-prepared dentin and enamel. Dent Mater 21(7):616–624CrossRefPubMed Dunn WJ, Davis JT (2005) Shear bond strength and SEM evaluation of composite bonded to Er:YAG laser-prepared dentin and enamel. Dent Mater 21(7):616–624CrossRefPubMed
10.
go back to reference Goncalves M, Corona SA (2002) Tensile bond strength of dentin-resinous system interfaces conditioned with Er:YAG laser irradiation. J Clin Laser Med Surg 20(2):89–93CrossRefPubMed Goncalves M, Corona SA (2002) Tensile bond strength of dentin-resinous system interfaces conditioned with Er:YAG laser irradiation. J Clin Laser Med Surg 20(2):89–93CrossRefPubMed
11.
go back to reference Hossain M, Kinoshita JI (2006) Compositional and structural changes of human dentin following Er, Cr:YSGG laser irradiation in vitro. J Oral Laser Appl 1:23–28 Hossain M, Kinoshita JI (2006) Compositional and structural changes of human dentin following Er, Cr:YSGG laser irradiation in vitro. J Oral Laser Appl 1:23–28
12.
go back to reference Aoki A, Ishikawa I, Yamada T, Otsuki M, Watanabe H, Tagami J (1998) Comparison between Er:YAG laser and conventional technique for root caries treatment in vitro. J Dent Res 77(10):1766CrossRef Aoki A, Ishikawa I, Yamada T, Otsuki M, Watanabe H, Tagami J (1998) Comparison between Er:YAG laser and conventional technique for root caries treatment in vitro. J Dent Res 77(10):1766CrossRef
13.
go back to reference Dostalova T, Jelinkova H, Kucerova H (1998) Non-contact Er:YAG laser ablation: clinical evaluation. J Clin Laser Med Surg 16(5):273–282PubMed Dostalova T, Jelinkova H, Kucerova H (1998) Non-contact Er:YAG laser ablation: clinical evaluation. J Clin Laser Med Surg 16(5):273–282PubMed
14.
go back to reference Tokonabe H, Kouji R, Watanabe H (1999) Morphological changes of human teeth with Er:YAG laser irradiation. J Clin Laser Med Surg 17(1):7–12PubMed Tokonabe H, Kouji R, Watanabe H (1999) Morphological changes of human teeth with Er:YAG laser irradiation. J Clin Laser Med Surg 17(1):7–12PubMed
15.
go back to reference Bertrand MF, Hessleyer D (2004) Scanning electron microscopic evaluation of resin-dentin interface after Er:YAG laser preparation. Lasers Surg Med 35(1):51–57CrossRefPubMed Bertrand MF, Hessleyer D (2004) Scanning electron microscopic evaluation of resin-dentin interface after Er:YAG laser preparation. Lasers Surg Med 35(1):51–57CrossRefPubMed
16.
go back to reference Ramos RP, Chimello DT (2002) Effect of Er:YAG laser on bond strength to dentin of a self-etching primer and two single-bottle adhesive systems. Lasers Surg Med 31(3):164–170CrossRefPubMed Ramos RP, Chimello DT (2002) Effect of Er:YAG laser on bond strength to dentin of a self-etching primer and two single-bottle adhesive systems. Lasers Surg Med 31(3):164–170CrossRefPubMed
17.
go back to reference Sassi JF, Chimello DT (2004) Comparative study of the dentin/adhesive systems interface after treatment with Er:YAG laser and acid etching using scanning electron microscope. Lasers Surg Med 34(5):385–390CrossRefPubMed Sassi JF, Chimello DT (2004) Comparative study of the dentin/adhesive systems interface after treatment with Er:YAG laser and acid etching using scanning electron microscope. Lasers Surg Med 34(5):385–390CrossRefPubMed
18.
go back to reference Delme KI, De Moor RJ (2007) Scanning electron microscopic evaluation of enamel and dentin surfaces after Er:YAG Laser. Photomed Laser Surg 25(5):393–401CrossRefPubMed Delme KI, De Moor RJ (2007) Scanning electron microscopic evaluation of enamel and dentin surfaces after Er:YAG Laser. Photomed Laser Surg 25(5):393–401CrossRefPubMed
Metadata
Title
Effect of low fluency dentin conditioning on tensile bond strength of composite bonded to Er:YAG laser-prepared dentin: a preliminary study
Authors
B. Bahrami
N. Askari
M. Tielemans
D. Heysselaer
L. Lamard
A. Peremans
C. Nyssen-Behets
S. Nammour
Publication date
01-03-2011
Publisher
Springer-Verlag
Published in
Lasers in Medical Science / Issue 2/2011
Print ISSN: 0268-8921
Electronic ISSN: 1435-604X
DOI
https://doi.org/10.1007/s10103-010-0767-z

Other articles of this Issue 2/2011

Lasers in Medical Science 2/2011 Go to the issue