Comparing the reliability of a new fluorescence camera with conventional laser fluorescence devices in detecting caries lesions in occlusal and smooth surfaces of primary teeth

Authors: Monique Saveriano De Benedetto, Caroline Carvalho Morais, Tatiane Fernandes Novaes, Jonas de Almeida Rodrigues, Mariana Minatel Braga, Fausto Medeiros Mendes

Published in: Lasers in Medical Science | Issue 2/2011

Login to get access

Abstract

The aims of this in vitro study were (1) to compare the reliability of a new fluorescence camera (FC; VistaProof) with that of two other laser fluorescence devices (DIAGNOdent and DIAGNOdent pen) in assessing smooth and occlusal caries lesions in primary teeth and (2) to determine the influence of the spacer, the protective cover and air drying on the FC readings. We selected 129 primary molars which were divided into two groups: smooth surfaces (n=68) and occlusal surfaces (n=61). Two trained examiners performed the evaluations using the FC and the other laser fluorescence devices separately. One examiner repeated all assessments after 1 week. Inter- and intraexaminer reproducibility was calculated from the absolute values derived from the devices using intraclass correlation coefficient (ICC) and Bland-Altman analysis. Pearson correlation was performed among the values obtained with the different devices. For aim 2, only teeth from the occlusal surface group were used. Another examiner performed measurements with the FC, varying the size of the spacer, the use of the plastic protective cover and the drying time (0, 3 and 15 s). Wilcoxon’s test was used to compare these measurements. High values of interexaminer ICC were observed with the FC in both experiments. However, the highest intraexaminer ICC for smooth surfaces was found for the DIAGNOdent (ICC 0.918). A positive correlation between all methods was observed in both studies. Only drying for 15 s significantly increased the FC readings. In conclusion, the FC showed high reliability for occlusal and smooth surfaces, similar to the other fluorescence methods tested. Additionally, prolonged air drying increased the FC measurements.
Literature
1.
go back to reference Bader JD, Shugars DA (2004) A systematic review of the performance of a laser fluorescence device for detecting caries. J Am Dent Assoc 135:1413–1426PubMed Bader JD, Shugars DA (2004) A systematic review of the performance of a laser fluorescence device for detecting caries. J Am Dent Assoc 135:1413–1426PubMed
2.
go back to reference Hibst R, Paulus R, Lussi A (2001) Detection of occlusal caries by laser fluorescence: basic and clinical investigations. Med Laser Appl 16:205–213CrossRef Hibst R, Paulus R, Lussi A (2001) Detection of occlusal caries by laser fluorescence: basic and clinical investigations. Med Laser Appl 16:205–213CrossRef
3.
go back to reference Thoms M (2006) Detection of intraoral lesions using a fluorescence camera. In: Rechmann P, Fried D (eds) Proceedings of the SPIE, vol. 6137 – Lasers in Dentistry XII, 613705, pp 37–43 Thoms M (2006) Detection of intraoral lesions using a fluorescence camera. In: Rechmann P, Fried D (eds) Proceedings of the SPIE, vol. 6137 – Lasers in Dentistry XII, 613705, pp 37–43
4.
go back to reference Lussi A, Hellwig E (2006) Performance of a new laser fluorescence device for the detection of occlusal caries in vitro. J Dent 34:467–471CrossRefPubMed Lussi A, Hellwig E (2006) Performance of a new laser fluorescence device for the detection of occlusal caries in vitro. J Dent 34:467–471CrossRefPubMed
5.
go back to reference Novaes TF, Matos R, Braga MM, Imparato JC, Raggio DP, Mendes FM (2009) Performance of a pen-type laser fluorescence device and conventional methods in detecting approximal caries lesions in primary teeth – in vivo study. Caries Res 43:36–42CrossRefPubMed Novaes TF, Matos R, Braga MM, Imparato JC, Raggio DP, Mendes FM (2009) Performance of a pen-type laser fluorescence device and conventional methods in detecting approximal caries lesions in primary teeth – in vivo study. Caries Res 43:36–42CrossRefPubMed
6.
go back to reference Rodrigues JA, Hug I, Diniz MB, Lussi A (2008) Performance of fluorescence methods, radiographic examination and ICDAS II on occlusal surfaces in vitro. Caries Res 42:297–304CrossRefPubMed Rodrigues JA, Hug I, Diniz MB, Lussi A (2008) Performance of fluorescence methods, radiographic examination and ICDAS II on occlusal surfaces in vitro. Caries Res 42:297–304CrossRefPubMed
7.
go back to reference Diniz MB, Rodrigues JA, de Paula AB, Cordeiro Rde C (2009) In vivo evaluation of laser fluorescence performance using different cut-off limits for occlusal caries detection. Lasers Med Sci 24:295–300CrossRefPubMed Diniz MB, Rodrigues JA, de Paula AB, Cordeiro Rde C (2009) In vivo evaluation of laser fluorescence performance using different cut-off limits for occlusal caries detection. Lasers Med Sci 24:295–300CrossRefPubMed
8.
go back to reference Rodrigues JA, Diniz MB, Josgrilberg EB, Cordeiro RC (2009) In vitro comparison of laser fluorescence performance with visual examination for detection of occlusal caries in permanent and primary molars. Lasers Med Sci 24:501–506CrossRefPubMed Rodrigues JA, Diniz MB, Josgrilberg EB, Cordeiro RC (2009) In vitro comparison of laser fluorescence performance with visual examination for detection of occlusal caries in permanent and primary molars. Lasers Med Sci 24:501–506CrossRefPubMed
9.
go back to reference Francescut P, Zimmerli B, Lussi A (2006) Influence of different storage methods on laser fluorescence values: a two-year study. Caries Res 40:181–185CrossRefPubMed Francescut P, Zimmerli B, Lussi A (2006) Influence of different storage methods on laser fluorescence values: a two-year study. Caries Res 40:181–185CrossRefPubMed
10.
go back to reference Mendes FM, Hissadomi M, Imparato JCP (2004) Effects of drying time and the presence of plaque on the in vitro performance of laser fluorescence in occlusal caries of primary teeth. Caries Res 38:104–108CrossRefPubMed Mendes FM, Hissadomi M, Imparato JCP (2004) Effects of drying time and the presence of plaque on the in vitro performance of laser fluorescence in occlusal caries of primary teeth. Caries Res 38:104–108CrossRefPubMed
11.
go back to reference Baelum V, Machiulskiene V, Nyvad B, Richards A, Vaeth M (2003) Application of survival analysis to carious lesion transitions in intervention trials. Community Dent Oral Epidemiol 31:252–260CrossRefPubMed Baelum V, Machiulskiene V, Nyvad B, Richards A, Vaeth M (2003) Application of survival analysis to carious lesion transitions in intervention trials. Community Dent Oral Epidemiol 31:252–260CrossRefPubMed
12.
go back to reference Aljehani A, Yang L, Shi XQ (2007) In vitro quantification of smooth surface caries with DIAGNOdent and the DIAGNOdent pen. Acta Odontol Scand 65:60–63CrossRefPubMed Aljehani A, Yang L, Shi XQ (2007) In vitro quantification of smooth surface caries with DIAGNOdent and the DIAGNOdent pen. Acta Odontol Scand 65:60–63CrossRefPubMed
13.
go back to reference Diniz MB, Leme AF, Cardoso KD, Rodrigues JD, Cordeiro RD (2009) The efficacy of laser fluorescence to detect in vitro demineralization and remineralization of smooth enamel surfaces. Photomed Laser Surg 27:57–61CrossRefPubMed Diniz MB, Leme AF, Cardoso KD, Rodrigues JD, Cordeiro RD (2009) The efficacy of laser fluorescence to detect in vitro demineralization and remineralization of smooth enamel surfaces. Photomed Laser Surg 27:57–61CrossRefPubMed
14.
go back to reference Kuhnisch J, Bucher K, Henschel V, Hickel R (2007) Reproducibility of DIAGNOdent 2095 and DIAGNOdent Pen measurements: results from an in vitro study on occlusal sites. Eur J Oral Sci 115:206–211CrossRefPubMed Kuhnisch J, Bucher K, Henschel V, Hickel R (2007) Reproducibility of DIAGNOdent 2095 and DIAGNOdent Pen measurements: results from an in vitro study on occlusal sites. Eur J Oral Sci 115:206–211CrossRefPubMed
15.
go back to reference Lin LI (1989) A concordance correlation coefficient to evaluate reproducibility. Biometrics 45:255–268CrossRefPubMed Lin LI (1989) A concordance correlation coefficient to evaluate reproducibility. Biometrics 45:255–268CrossRefPubMed
16.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
17.
go back to reference Kuhnisch J, Ziehe A, Brandstadt A, Heinrich-Weltzien R (2004) An in vitro study of the reliability of DIAGNOdent measurements. J Oral Rehabil 31:895–899CrossRefPubMed Kuhnisch J, Ziehe A, Brandstadt A, Heinrich-Weltzien R (2004) An in vitro study of the reliability of DIAGNOdent measurements. J Oral Rehabil 31:895–899CrossRefPubMed
18.
go back to reference Kuhnisch J, Bucher K, Hickel R (2007) The intra/inter-examiner reproducibility of the new DIAGNOdent Pen on occlusal sites. J Dent 35:509–512CrossRefPubMed Kuhnisch J, Bucher K, Hickel R (2007) The intra/inter-examiner reproducibility of the new DIAGNOdent Pen on occlusal sites. J Dent 35:509–512CrossRefPubMed
19.
go back to reference Buchalla W (2005) Comparative fluorescence spectroscopy shows differences in noncavitated enamel lesions. Caries Res 39:150–156CrossRefPubMed Buchalla W (2005) Comparative fluorescence spectroscopy shows differences in noncavitated enamel lesions. Caries Res 39:150–156CrossRefPubMed
20.
go back to reference Mendes FM, Pinheiro SL, Bengtson AL (2004) Effect of alteration in organic material of the occlusal caries on DIAGNOdent readings. Braz Oral Res 18:141–144CrossRefPubMed Mendes FM, Pinheiro SL, Bengtson AL (2004) Effect of alteration in organic material of the occlusal caries on DIAGNOdent readings. Braz Oral Res 18:141–144CrossRefPubMed
21.
go back to reference Alencar CJ, Braga MM, de Oliveira E, Nicolau J, Mendes FM (2009) Dye-enhanced laser fluorescence detection of caries lesions around brackets. Lasers Med Sci 24:865–870CrossRefPubMed Alencar CJ, Braga MM, de Oliveira E, Nicolau J, Mendes FM (2009) Dye-enhanced laser fluorescence detection of caries lesions around brackets. Lasers Med Sci 24:865–870CrossRefPubMed
22.
go back to reference Wenzel A, Frandsen E, Hintze H (1999) Patient discomfort and cross-infection control in bitewing examination with a storage phosphor plate and a CCD-based sensor. J Dent 27:243–246CrossRefPubMed Wenzel A, Frandsen E, Hintze H (1999) Patient discomfort and cross-infection control in bitewing examination with a storage phosphor plate and a CCD-based sensor. J Dent 27:243–246CrossRefPubMed
23.
go back to reference Cabral RM, Mendes FM, Nicolau J, Zezell DM (2006) The influence of PVC seal wrap and probe tips autoclaving on the in vitro performance of laser fluorescence device in occlusal caries in primary teeth. J Clin Pediatr Dent 30:306–309PubMed Cabral RM, Mendes FM, Nicolau J, Zezell DM (2006) The influence of PVC seal wrap and probe tips autoclaving on the in vitro performance of laser fluorescence device in occlusal caries in primary teeth. J Clin Pediatr Dent 30:306–309PubMed
24.
go back to reference Lussi A, Longbottom C, Gygax M, Braig F (2005) Influence of professional cleaning and drying of occlusal surfaces on laser fluorescence in vivo. Caries Res 39:284–286CrossRefPubMed Lussi A, Longbottom C, Gygax M, Braig F (2005) Influence of professional cleaning and drying of occlusal surfaces on laser fluorescence in vivo. Caries Res 39:284–286CrossRefPubMed
Metadata
Title
Comparing the reliability of a new fluorescence camera with conventional laser fluorescence devices in detecting caries lesions in occlusal and smooth surfaces of primary teeth
Authors
Monique Saveriano De Benedetto
Caroline Carvalho Morais
Tatiane Fernandes Novaes
Jonas de Almeida Rodrigues
Mariana Minatel Braga
Fausto Medeiros Mendes
Publication date
01-03-2011
Publisher
Springer-Verlag
Published in
Lasers in Medical Science / Issue 2/2011
Print ISSN: 0268-8921
Electronic ISSN: 1435-604X
DOI
https://doi.org/10.1007/s10103-010-0757-1

Other articles of this Issue 2/2011

Lasers in Medical Science 2/2011 Go to the issue