Skip to main content
Top
Published in: Journal of Artificial Organs 3/2014

01-09-2014 | Original Article

Mid-term results of 17-mm St. Jude Medical Regent prosthetic valves in elder patients with small aortic annuli: comparison with 19-mm bioprosthetic valves

Authors: Hideki Teshima, Masahiko Ikebuchi, Toshikazu Sano, Ryuta Tai, Naohiro Horio, Hiroyuki Irie

Published in: Journal of Artificial Organs | Issue 3/2014

Login to get access

Abstract

This study was designed to compare the mid-term outcomes after aortic valve replacement (AVR) between 17-mm mechanical heart valves (MV) and 19-mm bioprosthetic valves (BV) in elderly patients with small aortic annuli. Between 2000 and 2011, 127 consecutive patients (mean age 79 years; 87 % female) underwent AVR for aortic valve stenosis with a small aortic annulus. 19-mm BV (n = 67) was implanted. When the 19-mm BV did not fit the annulus, 17-mm St. Jude Medical Regent prosthetic mechanical valve (n = 60) was used instead of an aortic root-enlargement procedure. The follow-up rate was 94.0 % in the BV group, and 98.5 % in the MV group. No significant differences in survival rate and valve-related complications were found between the 2 groups. In-hospital mortality rates were 1.5 % (n = 1) in the BV group and 5.0 % (n = 3) in the MV group. Late mortality rates were 3.9 % per patient-years (p-y; n = 8) in the BV group, and 6.0 % per p-y (n = 10) in the MV group. Five-year Kaplan–Meier survival rates were 62 % in the BV group, and 72 % in the MV group (log-rank P = 0.280). Freedom from major adverse valve-related stroke and cerebral bleeding events was 92.5 and 98.5 % in the BV group, and 94.7 and 100 % in the MV group. AVR using 17-mm MV in elder patients with small aortic annuli provided equivalent mid-term clinical results to that with 19-mm BV.
Literature
1.
go back to reference Bonow RO, Carabello BA, Chatterjee K, De Leon AC Jr, Faxon DP, Freed MD, Gaasch WH, Lytle BW, Nishimura RA, O’Gara PT, O’Rourke RA, Otto CM, Shah PM, Shanewise JS, Smith SC Jr, Jacobs AK, Adams CD, Anderson JL, Antman EM, Fuster V, Halperin JL, Hiratzka LF, Hunt SA, Lytle BW, Nishimura R, Page RL, Riegel B. ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2006;48:e1–148.PubMedCrossRef Bonow RO, Carabello BA, Chatterjee K, De Leon AC Jr, Faxon DP, Freed MD, Gaasch WH, Lytle BW, Nishimura RA, O’Gara PT, O’Rourke RA, Otto CM, Shah PM, Shanewise JS, Smith SC Jr, Jacobs AK, Adams CD, Anderson JL, Antman EM, Fuster V, Halperin JL, Hiratzka LF, Hunt SA, Lytle BW, Nishimura R, Page RL, Riegel B. ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2006;48:e1–148.PubMedCrossRef
2.
go back to reference van Geldorp MW, Eric Jamieson WR, Kappetein AP, Ye J, Fradet GJ, Eijkemans MJ, Grunkemeier GL, Bogers AJ, Takkenberg JJ. Patient outcome after aortic valve replacement with a mechanical or biological prosthesis: weighing lifetime anticoagulant-related event risk against reoperation risk. J Thorac Cardiovasc Surg. 2009;137:881–6.PubMedCrossRef van Geldorp MW, Eric Jamieson WR, Kappetein AP, Ye J, Fradet GJ, Eijkemans MJ, Grunkemeier GL, Bogers AJ, Takkenberg JJ. Patient outcome after aortic valve replacement with a mechanical or biological prosthesis: weighing lifetime anticoagulant-related event risk against reoperation risk. J Thorac Cardiovasc Surg. 2009;137:881–6.PubMedCrossRef
3.
go back to reference Brown ML, Schaff HV, Lahr BD, Mullany CJ, Sundt TM, Dearani JA, McGregor CG, Orszulak TA. Aortic valve replacement in patients aged 50 to 70 years: improved outcome with mechanical versus biologic prostheses. J Thorac Cardiovasc Surg. 2008;135:878–84.PubMedCrossRef Brown ML, Schaff HV, Lahr BD, Mullany CJ, Sundt TM, Dearani JA, McGregor CG, Orszulak TA. Aortic valve replacement in patients aged 50 to 70 years: improved outcome with mechanical versus biologic prostheses. J Thorac Cardiovasc Surg. 2008;135:878–84.PubMedCrossRef
4.
go back to reference Yoshikawa K, Fukunaga S, Arinaga K, Hori H, Nakamura E, Ueda T, Tayama E, Aoyagi S. Long-term results of aortic valve replacement with a small St. Jude Medical valve in Japanese patients. Ann Thorac Surg. 2008;85:1303–9.PubMedCrossRef Yoshikawa K, Fukunaga S, Arinaga K, Hori H, Nakamura E, Ueda T, Tayama E, Aoyagi S. Long-term results of aortic valve replacement with a small St. Jude Medical valve in Japanese patients. Ann Thorac Surg. 2008;85:1303–9.PubMedCrossRef
5.
go back to reference Okamura H, Yamaguchi A, Nagano H, Itoh S, Morita H, Naito K, Yuri K, Adachi H. Mid-term outcomes after aortic valve replacement with the 17-mm St. Jude Medical Regent valve. Circ J. 2012;76:365–71.PubMedCrossRef Okamura H, Yamaguchi A, Nagano H, Itoh S, Morita H, Naito K, Yuri K, Adachi H. Mid-term outcomes after aortic valve replacement with the 17-mm St. Jude Medical Regent valve. Circ J. 2012;76:365–71.PubMedCrossRef
6.
go back to reference Kobayashi Y, Fukushima Y, Hayase T, Kojima K, Endo G. Clinical outcome of aortic valve replacement with 16-mm ATS-Advanced Performance valve for small aortic annulus. Ann Thorac Surg. 2010;89:1195–9.PubMedCrossRef Kobayashi Y, Fukushima Y, Hayase T, Kojima K, Endo G. Clinical outcome of aortic valve replacement with 16-mm ATS-Advanced Performance valve for small aortic annulus. Ann Thorac Surg. 2010;89:1195–9.PubMedCrossRef
7.
go back to reference Sawaki S, Usui A, Abe T, Yoshikawa M, Akita T, Ueda Y. Late mortality and morbidity in elderly patients with mechanical heart valves. Asian Cardiovasc Thorac Ann. 2006;14:189–94.PubMedCrossRef Sawaki S, Usui A, Abe T, Yoshikawa M, Akita T, Ueda Y. Late mortality and morbidity in elderly patients with mechanical heart valves. Asian Cardiovasc Thorac Ann. 2006;14:189–94.PubMedCrossRef
8.
go back to reference Hammermeister K, Sethi GK, Henderson WG, Grover FL, Oprian C, Rahimtoola SH. Outcomes 15 years after valve replacement with a mechanical versus a bioprosthetic valve: final report of the Veterans Affairs randomized trial. J Am Coll Cardiol. 2000;36:1152–8.PubMedCrossRef Hammermeister K, Sethi GK, Henderson WG, Grover FL, Oprian C, Rahimtoola SH. Outcomes 15 years after valve replacement with a mechanical versus a bioprosthetic valve: final report of the Veterans Affairs randomized trial. J Am Coll Cardiol. 2000;36:1152–8.PubMedCrossRef
10.
go back to reference Kahn SS, Trento A, DeRobertis M, Kass RM, Sandhu M, Czer LS, Blanche C, Raissi S, Fontana GP, Cheng W, Chaux A, Matloff JM. Twenty-year comparison of tissue and mechanical valve replacement. J Thorac Cardiovasc Surg. 2001;122:257–69.CrossRef Kahn SS, Trento A, DeRobertis M, Kass RM, Sandhu M, Czer LS, Blanche C, Raissi S, Fontana GP, Cheng W, Chaux A, Matloff JM. Twenty-year comparison of tissue and mechanical valve replacement. J Thorac Cardiovasc Surg. 2001;122:257–69.CrossRef
11.
go back to reference Chan V, Jamieson WR, Germann E, Chang F, Miyagishima RT, Burr LH, Janusz MT, Ling H, Fradet GJ. Performance of bioprostheses and mechanical prostheses assessed by composites of valve-related complications to 15 years after aortic valve replacement. J Thorac Cardiovasc Surg. 2006;131:1267–73.PubMedCrossRef Chan V, Jamieson WR, Germann E, Chang F, Miyagishima RT, Burr LH, Janusz MT, Ling H, Fradet GJ. Performance of bioprostheses and mechanical prostheses assessed by composites of valve-related complications to 15 years after aortic valve replacement. J Thorac Cardiovasc Surg. 2006;131:1267–73.PubMedCrossRef
12.
go back to reference Kulik A, Bedard P, Lam BK, Rubens FD, Hendry PJ, Masters RG, Mesana TG, Ruel M. Mechanical versus bioprosthetic valve replacements in middle-aged patients. Eur J Cardiothorac Surg. 2006;30:485–91.PubMedCrossRef Kulik A, Bedard P, Lam BK, Rubens FD, Hendry PJ, Masters RG, Mesana TG, Ruel M. Mechanical versus bioprosthetic valve replacements in middle-aged patients. Eur J Cardiothorac Surg. 2006;30:485–91.PubMedCrossRef
13.
go back to reference Pibarot P, Dumesnil J. Patient–prosthesis mismatch and the predictive use of indexed effective orifice area: is it relevant? Cardiac Surg Today. 2003;1:143–51. Pibarot P, Dumesnil J. Patient–prosthesis mismatch and the predictive use of indexed effective orifice area: is it relevant? Cardiac Surg Today. 2003;1:143–51.
14.
go back to reference Rao V, Jamieson WR, Ivanov J, Armstrong S, David TE. Prosthesis–patient mismatch affects survival after aortic valve replacement. Circulation. 2000;102:115–9. Rao V, Jamieson WR, Ivanov J, Armstrong S, David TE. Prosthesis–patient mismatch affects survival after aortic valve replacement. Circulation. 2000;102:115–9.
15.
go back to reference Hanayama N, Christakis GT, Mallidi HR, Joyner CD, Fremes SE, Morgan CD, Mitoff PRR, Goldman BS. Patient prosthesis mismatch is rare after aortic valve replacement: valve size may be irrelevant. Ann Thoracic Surg. 2002;73:1822–9.CrossRef Hanayama N, Christakis GT, Mallidi HR, Joyner CD, Fremes SE, Morgan CD, Mitoff PRR, Goldman BS. Patient prosthesis mismatch is rare after aortic valve replacement: valve size may be irrelevant. Ann Thoracic Surg. 2002;73:1822–9.CrossRef
16.
go back to reference Amarelli C, Della Corte A, Romano G, Iasevoli G, Dialetto G, De Santo LS, De Feo M, Torella M, Scardone M, Cotrufo M. Left ventricular mass regression after aortic valve replacement with 17-mm St Jude Medical mechanical prostheses in isolated aortic stenosis. J Thorac Cardiovasc Surg. 2005;129:512–7.PubMedCrossRef Amarelli C, Della Corte A, Romano G, Iasevoli G, Dialetto G, De Santo LS, De Feo M, Torella M, Scardone M, Cotrufo M. Left ventricular mass regression after aortic valve replacement with 17-mm St Jude Medical mechanical prostheses in isolated aortic stenosis. J Thorac Cardiovasc Surg. 2005;129:512–7.PubMedCrossRef
17.
go back to reference Minardi G, Manzara C, Creazzo V, Maselli D, Casali G, Pulignano G, Musumeci F. Evaluation of 17-mm St. Jude Medical Regent prosthetic aortic heart valves by rest and dobutamine stress echocardiography. J Cardiothorac Surg. 2006;1:27.PubMedCentralPubMedCrossRef Minardi G, Manzara C, Creazzo V, Maselli D, Casali G, Pulignano G, Musumeci F. Evaluation of 17-mm St. Jude Medical Regent prosthetic aortic heart valves by rest and dobutamine stress echocardiography. J Cardiothorac Surg. 2006;1:27.PubMedCentralPubMedCrossRef
18.
go back to reference Garatti A, Mori F, Innocente F, Canziani A, Gagliardotto P, Mossuto E, Santoro T, Montericcio V, Frigiola A, Menicanti L. Aortic valve replacement with 17-mm mechanical prostheses: is patient–prosthesis mismatch a relevant phenomenon? Ann Thorac Surg. 2011;91:71–7.PubMedCrossRef Garatti A, Mori F, Innocente F, Canziani A, Gagliardotto P, Mossuto E, Santoro T, Montericcio V, Frigiola A, Menicanti L. Aortic valve replacement with 17-mm mechanical prostheses: is patient–prosthesis mismatch a relevant phenomenon? Ann Thorac Surg. 2011;91:71–7.PubMedCrossRef
19.
go back to reference Akins CW, Miller DC, Turina MI, Kouchoukos NT, Blackstone EH, Grunkemeier GL, Takkenberg JJ, David TE, Butchart EG, Adams DH, Shahian DM, Hagl S, Mayer JE, Lytle BW, Councils of the American Association for Thoracic Surgery, Society of Thoracic Surgeons, European Association for Cardio-Thoracic Surgery, Ad Hoc Liaison Committee for Standardizing Definitions of Prosthetic Heart Valve Morbidity. Guidelines for reporting mortality and morbidity after cardiac valve interventions. J Thorac Cardiovasc Surg. 2008;135:732–8.PubMedCrossRef Akins CW, Miller DC, Turina MI, Kouchoukos NT, Blackstone EH, Grunkemeier GL, Takkenberg JJ, David TE, Butchart EG, Adams DH, Shahian DM, Hagl S, Mayer JE, Lytle BW, Councils of the American Association for Thoracic Surgery, Society of Thoracic Surgeons, European Association for Cardio-Thoracic Surgery, Ad Hoc Liaison Committee for Standardizing Definitions of Prosthetic Heart Valve Morbidity. Guidelines for reporting mortality and morbidity after cardiac valve interventions. J Thorac Cardiovasc Surg. 2008;135:732–8.PubMedCrossRef
20.
go back to reference Kvidal P, Bergstrom R, Malm T, Stahle E. Long-term follow-up of morbidity and mortality after aortic valve replacement with a mechanical valve prosthesis. Eur Heart J. 2000;21:1099–111.PubMedCrossRef Kvidal P, Bergstrom R, Malm T, Stahle E. Long-term follow-up of morbidity and mortality after aortic valve replacement with a mechanical valve prosthesis. Eur Heart J. 2000;21:1099–111.PubMedCrossRef
21.
go back to reference Edmunds LH Jr, Clark RE, Cohn LH, Grunkemeier GL, Miller DC, Weisel RD. Guidelines for reporting morbidity and mortality after cardiac valvular operations. Ann Thorac Surg. 1996;62:932–5.PubMedCrossRef Edmunds LH Jr, Clark RE, Cohn LH, Grunkemeier GL, Miller DC, Weisel RD. Guidelines for reporting morbidity and mortality after cardiac valvular operations. Ann Thorac Surg. 1996;62:932–5.PubMedCrossRef
22.
go back to reference Jamieson WR, Janusz MT, Burr LH, Ling H, Miyagishima RT, Germann E. Carpentier-Edwards supraannular porcine bioprosthesis: second-generation prosthesis in aortic valve replacement. Ann Thorac Surg. 2001;71:S224–7.PubMedCrossRef Jamieson WR, Janusz MT, Burr LH, Ling H, Miyagishima RT, Germann E. Carpentier-Edwards supraannular porcine bioprosthesis: second-generation prosthesis in aortic valve replacement. Ann Thorac Surg. 2001;71:S224–7.PubMedCrossRef
24.
go back to reference Matsuyama K, Matsumoto M, Sugita T, Nishizawa J, Yoshida K, Tokuda Y, Matsuo T. Anticoagulant therapy in Japanese patients with mechanical mitral valves. Circ J. 2002;66:668–70.PubMedCrossRef Matsuyama K, Matsumoto M, Sugita T, Nishizawa J, Yoshida K, Tokuda Y, Matsuo T. Anticoagulant therapy in Japanese patients with mechanical mitral valves. Circ J. 2002;66:668–70.PubMedCrossRef
25.
go back to reference Vicchio M, Della Corte A, De Santo LS, De Feo M, Caianiello G, Scardone M, Cotrufo M. Tissue versus mechanical prostheses: quality of life in octogenerians. Ann Thorac Surg. 2008;85:1290–5.PubMedCrossRef Vicchio M, Della Corte A, De Santo LS, De Feo M, Caianiello G, Scardone M, Cotrufo M. Tissue versus mechanical prostheses: quality of life in octogenerians. Ann Thorac Surg. 2008;85:1290–5.PubMedCrossRef
26.
go back to reference Mori T, Asano M, Ohtake H, Bitoh A, Sekiguchi S, Matsuo Y, Aiba M, Yamada M, Kawada T, Takaba T. Anticoagulant therapy after prosthetic valve replacement -optimal PT-INR in Japanese patients-. Ann Thorac Cardiovasc Surg. 2002;8:83–7.PubMed Mori T, Asano M, Ohtake H, Bitoh A, Sekiguchi S, Matsuo Y, Aiba M, Yamada M, Kawada T, Takaba T. Anticoagulant therapy after prosthetic valve replacement -optimal PT-INR in Japanese patients-. Ann Thorac Cardiovasc Surg. 2002;8:83–7.PubMed
27.
go back to reference Butchart EG, Payne N, Li HH, Buchan K, Mandana K, Grunkemeier GL. Better anticoagulation control improves survival after valve replacement. J Thorac Cardiovasc Surg. 2002;123:715–23.PubMedCrossRef Butchart EG, Payne N, Li HH, Buchan K, Mandana K, Grunkemeier GL. Better anticoagulation control improves survival after valve replacement. J Thorac Cardiovasc Surg. 2002;123:715–23.PubMedCrossRef
Metadata
Title
Mid-term results of 17-mm St. Jude Medical Regent prosthetic valves in elder patients with small aortic annuli: comparison with 19-mm bioprosthetic valves
Authors
Hideki Teshima
Masahiko Ikebuchi
Toshikazu Sano
Ryuta Tai
Naohiro Horio
Hiroyuki Irie
Publication date
01-09-2014
Publisher
Springer Japan
Published in
Journal of Artificial Organs / Issue 3/2014
Print ISSN: 1434-7229
Electronic ISSN: 1619-0904
DOI
https://doi.org/10.1007/s10047-014-0770-4

Other articles of this Issue 3/2014

Journal of Artificial Organs 3/2014 Go to the issue