Molecular leakage at implant-abutment connection—in vitro investigation of tightness of internal conical implant-abutment connections against endotoxin penetration

Authors: Sönke Harder, Birka Dimaczek, Yaha Açil, Hendrik Terheyden, Sandra Freitag-Wolf, Matthias Kern

Published in: Clinical Oral Investigations | Issue 4/2010

Login to get access

Abstract

Microleakage has been discussed as a major contributing factor for inflammatory reactions at the implant-abutment connection. In previous studies, the tightness against corpuscular bodies (viable bacteria) has been successfully investigated under static and dynamic conditions. The aim of this study was to investigate the tightness against endotoxins of two implant systems (AstraTech and Ankylos) with conical internal connections under static conditions. The inner parts of eight implants of each system were inoculated with endotoxin. Implants were screwed together with the respective abutments and stored under isostatic conditions in a supernatant of pyrogen-free water for 168 h. Supernatant samples were taken after 5 min, 24 h, 72 h, and 168 h, and endotoxin contamination was determined by the amebocyte-lysate test. Only one implant in the AstraTech group showed no sign of endotoxin contamination after 168 h, while the other implants showed contamination after varying storage times, respectively. The implants in the Ankylos group showed endotoxin contamination after only 5 min of storage in the supernatant solution. The tested internal conical implant-abutment connections appear to be unable to prevent endotoxin leakage. In average, Astra implants showed a higher tightness than Ankylos implants.
Literature
1.
go back to reference Zipprich H, Weigl P, Lange B, Lauer HC (2007) Erfassung, Ursachen und Folgen von Mikrobewegungen am Implantat-Abutment-Interface. Implantologie 15:31–46 Zipprich H, Weigl P, Lange B, Lauer HC (2007) Erfassung, Ursachen und Folgen von Mikrobewegungen am Implantat-Abutment-Interface. Implantologie 15:31–46
2.
go back to reference Broggini N, McManus LM, Hermann JS, Medina RU, Oates TW, Schenk RK, Buser D, Mellonig JT, Cochran DL (2003) Persistent acute inflammation at the implant-abutment interface. J Dent Res 82:232–237CrossRefPubMed Broggini N, McManus LM, Hermann JS, Medina RU, Oates TW, Schenk RK, Buser D, Mellonig JT, Cochran DL (2003) Persistent acute inflammation at the implant-abutment interface. J Dent Res 82:232–237CrossRefPubMed
3.
go back to reference Hermann JS, Schoolfield JD, Schenk RK, Buser D, Cochran DL (2001) Influence of the size of the microgap on crestal bone changes around titanium implants. A histometric evaluation of unloaded non-submerged implants in the canine mandible. J Periodontol 72:1372–1383CrossRefPubMed Hermann JS, Schoolfield JD, Schenk RK, Buser D, Cochran DL (2001) Influence of the size of the microgap on crestal bone changes around titanium implants. A histometric evaluation of unloaded non-submerged implants in the canine mandible. J Periodontol 72:1372–1383CrossRefPubMed
4.
go back to reference Besimo CE, Guindy JS, Lewetag D, Meyer J (1999) Prevention of bacterial leakage into and from prefabricated screw-retained crowns on implants in vitro. Int J Oral Maxillofac Implants 14:654–660PubMed Besimo CE, Guindy JS, Lewetag D, Meyer J (1999) Prevention of bacterial leakage into and from prefabricated screw-retained crowns on implants in vitro. Int J Oral Maxillofac Implants 14:654–660PubMed
5.
go back to reference Quirynen M, Bollen CM, Eyssen H, van Steenberghe D (1994) Microbial penetration along the implant components of the Branemark system. An in vitro study. Clin Oral Implants Res 5:239–244CrossRefPubMed Quirynen M, Bollen CM, Eyssen H, van Steenberghe D (1994) Microbial penetration along the implant components of the Branemark system. An in vitro study. Clin Oral Implants Res 5:239–244CrossRefPubMed
6.
go back to reference Quirynen M, van Steenberghe D (1993) Bacterial colonization of the internal part of two-stage implants. An in vivo study. Clin Oral Implants Res 4:158–161CrossRefPubMed Quirynen M, van Steenberghe D (1993) Bacterial colonization of the internal part of two-stage implants. An in vivo study. Clin Oral Implants Res 4:158–161CrossRefPubMed
7.
go back to reference Steinebrunner L, Wolfart S, Bossmann K, Kern M (2005) In vitro evaluation of bacterial leakage along the implant-abutment interface of different implant systems. Int J Oral Maxillofac Implants 20:875–881PubMed Steinebrunner L, Wolfart S, Bossmann K, Kern M (2005) In vitro evaluation of bacterial leakage along the implant-abutment interface of different implant systems. Int J Oral Maxillofac Implants 20:875–881PubMed
8.
go back to reference Wahl G, Muller F, Schaal KP (1992) The microbial colonization of implant elements made of plastics and titanium. Schweiz Monatsschr Zahnmed 102:1321–1326PubMed Wahl G, Muller F, Schaal KP (1992) The microbial colonization of implant elements made of plastics and titanium. Schweiz Monatsschr Zahnmed 102:1321–1326PubMed
9.
go back to reference Nair SP, Meghji S, Wilson M, Reddi K, White P, Henderson B (1996) Bacterially induced bone destruction: mechanisms and misconceptions. Infect Immun 64:2371–2380PubMed Nair SP, Meghji S, Wilson M, Reddi K, White P, Henderson B (1996) Bacterially induced bone destruction: mechanisms and misconceptions. Infect Immun 64:2371–2380PubMed
11.
go back to reference Nakata T (1994) Destruction of challenged endotoxin in a dry heat oven. J Pharm Sci Technol 48:59–63PubMed Nakata T (1994) Destruction of challenged endotoxin in a dry heat oven. J Pharm Sci Technol 48:59–63PubMed
12.
go back to reference Bang FB (1956) A bacterial disease of Limulus polyphemus. Bull Johns Hopkins Hosp 98:325–351PubMed Bang FB (1956) A bacterial disease of Limulus polyphemus. Bull Johns Hopkins Hosp 98:325–351PubMed
13.
go back to reference Levin J, Bang FB (1968) Clottable protein in Limulus: its localization and kinetics of its coagulation by endotoxin. Thromb Diath Haemorrh 19:186–197PubMed Levin J, Bang FB (1968) Clottable protein in Limulus: its localization and kinetics of its coagulation by endotoxin. Thromb Diath Haemorrh 19:186–197PubMed
14.
go back to reference Young NS, Levin J, Prendergast RA (1972) An invertebrate coagulation system activated by endotoxin: evidence for enzymatic mediation. J Clin Invest 51:1790–1797CrossRefPubMed Young NS, Levin J, Prendergast RA (1972) An invertebrate coagulation system activated by endotoxin: evidence for enzymatic mediation. J Clin Invest 51:1790–1797CrossRefPubMed
15.
go back to reference Scully MF, Newman YM, Clark SE, Kakkar VV (1980) Evaluation of a chromogenic method for endotoxin measurement. Thromb Res 20:263–270CrossRefPubMed Scully MF, Newman YM, Clark SE, Kakkar VV (1980) Evaluation of a chromogenic method for endotoxin measurement. Thromb Res 20:263–270CrossRefPubMed
16.
go back to reference Hurley JC (1995) Endotoxemia: methods of detection and clinical correlates. Clin Microbiol Rev 8:268–292PubMed Hurley JC (1995) Endotoxemia: methods of detection and clinical correlates. Clin Microbiol Rev 8:268–292PubMed
17.
go back to reference Jansen VK, Conrads G, Richter EJ (1997) Microbial leakage and marginal fit of the implant-abutment interface. Int J Oral Maxillofac Implants 12:527–540PubMed Jansen VK, Conrads G, Richter EJ (1997) Microbial leakage and marginal fit of the implant-abutment interface. Int J Oral Maxillofac Implants 12:527–540PubMed
18.
go back to reference Persson LG, Lekholm U, Leonhardt A, Dahlen G, Lindhe J (1996) Bacterial colonization on internal surfaces of Branemark system implant components. Clin Oral Implants Res 7:90–95CrossRefPubMed Persson LG, Lekholm U, Leonhardt A, Dahlen G, Lindhe J (1996) Bacterial colonization on internal surfaces of Branemark system implant components. Clin Oral Implants Res 7:90–95CrossRefPubMed
19.
go back to reference O’Mahony A, MacNeill SR, Cobb CM (2000) Design features that may influence bacterial plaque retention: a retrospective analysis of failed implants. Quintessence Int 31:249–256PubMed O’Mahony A, MacNeill SR, Cobb CM (2000) Design features that may influence bacterial plaque retention: a retrospective analysis of failed implants. Quintessence Int 31:249–256PubMed
20.
go back to reference Gross M, Abramovich I, Weiss EI (1999) Microleakage at the abutment-implant interface of osseointegrated implants: a comparative study. Int J Oral Maxillofac Implants 14:94–100PubMed Gross M, Abramovich I, Weiss EI (1999) Microleakage at the abutment-implant interface of osseointegrated implants: a comparative study. Int J Oral Maxillofac Implants 14:94–100PubMed
21.
go back to reference Jann B, Reske K, Jann K (1975) Heterogeneity of lipopolysaccharides. Analysis of polysaccharide chain lengths by sodium dodecylsulfate-polyacrylamide gel electrophoresis. Eur J Biochem 60:239–246CrossRefPubMed Jann B, Reske K, Jann K (1975) Heterogeneity of lipopolysaccharides. Analysis of polysaccharide chain lengths by sodium dodecylsulfate-polyacrylamide gel electrophoresis. Eur J Biochem 60:239–246CrossRefPubMed
Metadata
Title
Molecular leakage at implant-abutment connection—in vitro investigation of tightness of internal conical implant-abutment connections against endotoxin penetration
Authors
Sönke Harder
Birka Dimaczek
Yaha Açil
Hendrik Terheyden
Sandra Freitag-Wolf
Matthias Kern
Publication date
01-08-2010
Publisher
Springer-Verlag
Published in
Clinical Oral Investigations / Issue 4/2010
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-009-0317-x

Other articles of this Issue 4/2010

Clinical Oral Investigations 4/2010 Go to the issue