Published in: European Spine Journal 6/2011

01-06-2011 | Original Article

Optimal entry points and trajectories for cervical pedicle screw placement into subaxial cervical vertebrae

Authors: Dong-Ho Lee, Sung-Woo Lee, Suk Jung Kang, Chang Ju Hwang, Nam Heun Kim, Ju-Yul Bae, Yung-Tae Kim, Choon Sung Lee, K. Daniel Riew

Published in: European Spine Journal | Issue 6/2011

Login to get access

Abstract

The present study was performed to determine the optimal entry points and trajectories for cervical pedicle screw insertion into C3–7. The study involved 40 patients (M:F = 20:20) with various cervical diseases. A surgical simulation program was used to construct three-dimensional spine models from cervical spine axial CT images. Axial, sagittal, and coronal plane data were simultaneously processed to determine the ideal pedicle trajectory (a line passing through the center of the pedicle on coronal, sagittal, and transverse CT images). The optimal entry points on the lateral masses were then identified. Horizontal offsets and vertical offsets of the optimal entry points were measured from three different anatomical landmarks: the lateral notch, the center of the superior edge and the center of lateral mass. The transverse angle and sagittal angles of the ideal pedicle trajectory were measured. Using those entry points and trajectory results, virtual screws were placed into the pedicles using the simulation program, and the outcomes were evaluated. We found that at C3–6, the optimal entry point was located 2.0–2.4 mm medial and 0–0.8 mm inferior to the lateral notch. Since the difference of 1 mm is difficult to discern intra-operatively, for ease of remembrance, we recommend rounding off our findings to arrive at a starting point for the C3–6 pedicle screws to be 2 mm directly medial to the lateral notch. At C7, by contrast, the optimal entry point was 1.6 mm lateral and 2.5 mm superior to the center of lateral mass. Again, for ease of remembrance, we recommend rounding off these numbers to use a starting point for the C7 pedicle screws to be 2 mm lateral and 2 mm superior to the center of lateral mass. The average transverse angles were 45° at C3–5, 38° at C6, and 28° at C7. The entry points for each vertebra should be adjusted according to the transverse angles of pedicles. The mean sagittal angles were 7° upward at C3, and parallel to the upper end plate at C4–7. The simulation study showed that the entry point and ideal pedicle trajectory led to screw placements that were safer than those used in other studies.
Literature
1.
go back to reference Reinhold M, Bach C, Audige L, Bale R, Attal R, Blauth M, Magerl F (2008) Comparison of two novel fluoroscopy-based stereotactic methods for cervical pedicle screw placement and review of the literature. Eur Spine J 17:564–575. doi:10.1007/s00586-008-0584-2 Reinhold M, Bach C, Audige L, Bale R, Attal R, Blauth M, Magerl F (2008) Comparison of two novel fluoroscopy-based stereotactic methods for cervical pedicle screw placement and review of the literature. Eur Spine J 17:564–575. doi:10.​1007/​s00586-008-0584-2
2.
go back to reference Neo M, Sakamoto T, Fujibayashi S, Nakamura T (2005) The clinical risk of vertebral artery injury from cervical pedicle screws inserted in degenerative vertebrae. Spine (Phila Pa 1976) 30:2800–2805CrossRef Neo M, Sakamoto T, Fujibayashi S, Nakamura T (2005) The clinical risk of vertebral artery injury from cervical pedicle screws inserted in degenerative vertebrae. Spine (Phila Pa 1976) 30:2800–2805CrossRef
3.
go back to reference Abumi K, Shono Y, Ito M, Taneichi H, Kotani Y, Kaneda K (2000) Complications of pedicle screw fixation in reconstructive surgery of the cervical spine. Spine (Phila Pa 1976) 25:962–969CrossRef Abumi K, Shono Y, Ito M, Taneichi H, Kotani Y, Kaneda K (2000) Complications of pedicle screw fixation in reconstructive surgery of the cervical spine. Spine (Phila Pa 1976) 25:962–969CrossRef
4.
go back to reference Ugur HC, Attar A, Uz A, Tekdemir I, Egemen N, Caglar S, Genc Y (2000) Surgical anatomic evaluation of the cervical pedicle and adjacent neural structures. Neurosurgery 47:1162–1168 (discussion 1168–1169) Ugur HC, Attar A, Uz A, Tekdemir I, Egemen N, Caglar S, Genc Y (2000) Surgical anatomic evaluation of the cervical pedicle and adjacent neural structures. Neurosurgery 47:1162–1168 (discussion 1168–1169)
5.
go back to reference Abumi K, Itoh H, Taneichi H, Kaneda K (1994) Transpedicular screw fixation for traumatic lesions of the middle and lower cervical spine: description of the techniques and preliminary report. J Spinal Disord 7:19–28PubMedCrossRef Abumi K, Itoh H, Taneichi H, Kaneda K (1994) Transpedicular screw fixation for traumatic lesions of the middle and lower cervical spine: description of the techniques and preliminary report. J Spinal Disord 7:19–28PubMedCrossRef
6.
go back to reference Abumi K, Kaneda K (1997) Pedicle screw fixation for nontraumatic lesions of the cervical spine. Spine (Phila Pa 1976) 22:1853–1863CrossRef Abumi K, Kaneda K (1997) Pedicle screw fixation for nontraumatic lesions of the cervical spine. Spine (Phila Pa 1976) 22:1853–1863CrossRef
7.
go back to reference Karaikovic EE, Kunakornsawat S, Daubs MD, Madsen TW, Gaines RW (2000) Surgical anatomy of the cervical pedicles: landmarks for posterior cervical pedicle entrance localization. J Spinal Disord 13:63–72PubMedCrossRef Karaikovic EE, Kunakornsawat S, Daubs MD, Madsen TW, Gaines RW (2000) Surgical anatomy of the cervical pedicles: landmarks for posterior cervical pedicle entrance localization. J Spinal Disord 13:63–72PubMedCrossRef
8.
go back to reference Ludwig SC, Kramer DL, Balderston RA, Vaccaro AR, Foley KF, Albert TJ (2000) Placement of pedicle screws in the human cadaveric cervical spine: comparative accuracy of three techniques. Spine (Phila Pa 1976) 25:1655–1667CrossRef Ludwig SC, Kramer DL, Balderston RA, Vaccaro AR, Foley KF, Albert TJ (2000) Placement of pedicle screws in the human cadaveric cervical spine: comparative accuracy of three techniques. Spine (Phila Pa 1976) 25:1655–1667CrossRef
9.
go back to reference Ludwig SC, Kramer DL, Vaccaro AR, Albert TJ (1999) Transpedicle screw fixation of the cervical spine. Clin Orthop Relat Res 359:77–88PubMedCrossRef Ludwig SC, Kramer DL, Vaccaro AR, Albert TJ (1999) Transpedicle screw fixation of the cervical spine. Clin Orthop Relat Res 359:77–88PubMedCrossRef
10.
go back to reference Zheng X, Chaudhari R, Wu C, Mehbod AA, Transfeldt EE (2010) Subaxial cervical pedicle screw insertion with newly defined entry point and trajectory: accuracy evaluation in cadavers. Eur Spine J 19:105–112. doi:10.1007/s00586-009-1213-4 Zheng X, Chaudhari R, Wu C, Mehbod AA, Transfeldt EE (2010) Subaxial cervical pedicle screw insertion with newly defined entry point and trajectory: accuracy evaluation in cadavers. Eur Spine J 19:105–112. doi:10.​1007/​s00586-009-1213-4
11.
go back to reference Johnston TL, Karaikovic EE, Lautenschlager EP, Marcu D (2006) Cervical pedicle screws vs. lateral mass screws: uniplanar fatigue analysis and residual pullout strengths. Spine J 6:667–672PubMedCrossRef Johnston TL, Karaikovic EE, Lautenschlager EP, Marcu D (2006) Cervical pedicle screws vs. lateral mass screws: uniplanar fatigue analysis and residual pullout strengths. Spine J 6:667–672PubMedCrossRef
12.
go back to reference Jones EL, Heller JG, Silcox DH, Hutton WC (1997) Cervical pedicle screws versus lateral mass screws. Anatomic feasibility and biomechanical comparison. Spine (Phila Pa 1976) 22:977–982CrossRef Jones EL, Heller JG, Silcox DH, Hutton WC (1997) Cervical pedicle screws versus lateral mass screws. Anatomic feasibility and biomechanical comparison. Spine (Phila Pa 1976) 22:977–982CrossRef
13.
go back to reference Rhee JM, Kraiwattanapong C, Hutton WC (2005) A comparison of pedicle and lateral mass screw construct stiffnesses at the cervicothoracic junction: a biomechanical study. Spine (Phila Pa 1976) 30:E636–E640CrossRef Rhee JM, Kraiwattanapong C, Hutton WC (2005) A comparison of pedicle and lateral mass screw construct stiffnesses at the cervicothoracic junction: a biomechanical study. Spine (Phila Pa 1976) 30:E636–E640CrossRef
14.
go back to reference Dunlap BJ, Karaikovic EE, Park HS, Sokolowski MJ, Zhang LQ (2010) Load sharing properties of cervical pedicle screw-rod constructs versus lateral mass screw-rod constructs. Eur Spine J. doi:10.1007/s00586-010-1278-0 Dunlap BJ, Karaikovic EE, Park HS, Sokolowski MJ, Zhang LQ (2010) Load sharing properties of cervical pedicle screw-rod constructs versus lateral mass screw-rod constructs. Eur Spine J. doi:10.​1007/​s00586-010-1278-0
15.
go back to reference Abumi K, Ito M, Kaneda K (2000) Surgical treatment of cervical destructive spondyloarthropathy (DSA). Spine (Phila Pa 1976) 25:2899–2905CrossRef Abumi K, Ito M, Kaneda K (2000) Surgical treatment of cervical destructive spondyloarthropathy (DSA). Spine (Phila Pa 1976) 25:2899–2905CrossRef
16.
go back to reference Abumi K, Kaneda K, Shono Y, Fujiya M (1999) One-stage posterior decompression and reconstruction of the cervical spine by using pedicle screw fixation systems. J Neurosurg 90:19–26PubMed Abumi K, Kaneda K, Shono Y, Fujiya M (1999) One-stage posterior decompression and reconstruction of the cervical spine by using pedicle screw fixation systems. J Neurosurg 90:19–26PubMed
17.
go back to reference Abumi K, Shono Y, Taneichi H, Ito M, Kaneda K (1999) Correction of cervical kyphosis using pedicle screw fixation systems. Spine (Phila Pa 1976) 24:2389–2396CrossRef Abumi K, Shono Y, Taneichi H, Ito M, Kaneda K (1999) Correction of cervical kyphosis using pedicle screw fixation systems. Spine (Phila Pa 1976) 24:2389–2396CrossRef
18.
go back to reference Oda I, Abumi K, Ito M, Kotani Y, Oya T, Hasegawa K, Minami A (2006) Palliative spinal reconstruction using cervical pedicle screws for metastatic lesions of the spine: a retrospective analysis of 32 cases. Spine (Phila Pa 1976) 31:1439–1444. doi:10.1097/01.brs.0000219952.40906.1f Oda I, Abumi K, Ito M, Kotani Y, Oya T, Hasegawa K, Minami A (2006) Palliative spinal reconstruction using cervical pedicle screws for metastatic lesions of the spine: a retrospective analysis of 32 cases. Spine (Phila Pa 1976) 31:1439–1444. doi:10.​1097/​01.​brs.​0000219952.​40906.​1f
19.
go back to reference Yukawa Y, Kato F, Ito K, Horie Y, Hida T, Nakashima H, Machino M (2009) Placement and complications of cervical pedicle screws in 144 cervical trauma patients using pedicle axis view techniques by fluoroscope. Eur Spine J 18:1293–1299. doi:10.1007/s00586-009-1032-7 Yukawa Y, Kato F, Ito K, Horie Y, Hida T, Nakashima H, Machino M (2009) Placement and complications of cervical pedicle screws in 144 cervical trauma patients using pedicle axis view techniques by fluoroscope. Eur Spine J 18:1293–1299. doi:10.​1007/​s00586-009-1032-7
20.
go back to reference Karaikovic EE, Daubs MD, Madsen RW, Gaines RW (1997) Morphologic characteristics of human cervical pedicles. Spine (Phila Pa 1976) 22:493–500CrossRef Karaikovic EE, Daubs MD, Madsen RW, Gaines RW (1997) Morphologic characteristics of human cervical pedicles. Spine (Phila Pa 1976) 22:493–500CrossRef
21.
go back to reference Panjabi MM, Duranceau J, Goel V, Oxland T, Takata K (1991) Cervical human vertebrae. Quantitative three-dimensional anatomy of the middle and lower regions. Spine (Phila Pa 1976) 16:861–869CrossRef Panjabi MM, Duranceau J, Goel V, Oxland T, Takata K (1991) Cervical human vertebrae. Quantitative three-dimensional anatomy of the middle and lower regions. Spine (Phila Pa 1976) 16:861–869CrossRef
22.
go back to reference Shin EK, Panjabi MM, Chen NC, Wang JL (2000) The anatomic variability of human cervical pedicles: considerations for transpedicular screw fixation in the middle and lower cervical spine. Eur Spine J 9:61–66PubMedCrossRef Shin EK, Panjabi MM, Chen NC, Wang JL (2000) The anatomic variability of human cervical pedicles: considerations for transpedicular screw fixation in the middle and lower cervical spine. Eur Spine J 9:61–66PubMedCrossRef
23.
go back to reference Onibokun A, Khoo LT, Bistazzoni S, Chen NF, Sassi M (2009) Anatomical considerations for cervical pedicle screw insertion: the use of multiplanar computerized tomography measurements in 122 consecutive clinical cases. Spine J 9:729–734PubMedCrossRef Onibokun A, Khoo LT, Bistazzoni S, Chen NF, Sassi M (2009) Anatomical considerations for cervical pedicle screw insertion: the use of multiplanar computerized tomography measurements in 122 consecutive clinical cases. Spine J 9:729–734PubMedCrossRef
24.
go back to reference Ruofu Z, Huilin Y, Xiaoyun H, Xishun H, Tiansi T, Liang C, Xigong L (2008) CT evaluation of cervical pedicle in a Chinese population for surgical application of transpedicular screw placement. Surg Radiol Anat 30:389–396. doi:10.1007/s00276-008-0339-0 Ruofu Z, Huilin Y, Xiaoyun H, Xishun H, Tiansi T, Liang C, Xigong L (2008) CT evaluation of cervical pedicle in a Chinese population for surgical application of transpedicular screw placement. Surg Radiol Anat 30:389–396. doi:10.​1007/​s00276-008-0339-0
25.
go back to reference Sakamoto T, Neo M, Nakamura T (2004) Transpedicular screw placement evaluated by axial computed tomography of the cervical pedicle. Spine (Phila Pa 1976) 29:2510–2514 (discussion 2515) Sakamoto T, Neo M, Nakamura T (2004) Transpedicular screw placement evaluated by axial computed tomography of the cervical pedicle. Spine (Phila Pa 1976) 29:2510–2514 (discussion 2515)
26.
go back to reference Karaikovic EE, Yingsakmongkol W, Gaines RW (2001) Accuracy of cervical pedicle screw placement using the funnel technique. Spine (Phila Pa 1976) 26:2456–2462CrossRef Karaikovic EE, Yingsakmongkol W, Gaines RW (2001) Accuracy of cervical pedicle screw placement using the funnel technique. Spine (Phila Pa 1976) 26:2456–2462CrossRef
27.
go back to reference Kotani Y, Abumi K, Ito M, Minami A (2003) Improved accuracy of computer-assisted cervical pedicle screw insertion. J Neurosurg 99:257–263PubMed Kotani Y, Abumi K, Ito M, Minami A (2003) Improved accuracy of computer-assisted cervical pedicle screw insertion. J Neurosurg 99:257–263PubMed
28.
go back to reference Lu S, Xu YQ, Lu WW, Ni GX, Li YB, Shi JH, Li DP, Chen GP, Chen YB, Zhang YZ (2009) A novel patient-specific navigational template for cervical pedicle screw placement. Spine 34:E959–E966. doi:10.1097/BRS.0b013e3181c09985 Lu S, Xu YQ, Lu WW, Ni GX, Li YB, Shi JH, Li DP, Chen GP, Chen YB, Zhang YZ (2009) A novel patient-specific navigational template for cervical pedicle screw placement. Spine 34:E959–E966. doi:10.​1097/​BRS.​0b013e3181c09985​
29.
go back to reference Chazono M, Soshi S, Inoue T, Kida Y, Ushiku C (2006) Anatomical considerations for cervical pedicle screw insertion: the use of multiplanar computerized tomography reconstruction measurements. J Neurosurg Spine 4:472–477PubMedCrossRef Chazono M, Soshi S, Inoue T, Kida Y, Ushiku C (2006) Anatomical considerations for cervical pedicle screw insertion: the use of multiplanar computerized tomography reconstruction measurements. J Neurosurg Spine 4:472–477PubMedCrossRef
30.
go back to reference Yukawa Y, Kato F, Yoshihara H, Yanase M, Ito K (2006) Cervical pedicle screw fixation in 100 cases of unstable cervical injuries: pedicle axis views obtained using fluoroscopy. J Neurosurg Spine 5:488–493PubMedCrossRef Yukawa Y, Kato F, Yoshihara H, Yanase M, Ito K (2006) Cervical pedicle screw fixation in 100 cases of unstable cervical injuries: pedicle axis views obtained using fluoroscopy. J Neurosurg Spine 5:488–493PubMedCrossRef
Metadata
Title
Optimal entry points and trajectories for cervical pedicle screw placement into subaxial cervical vertebrae
Authors
Dong-Ho Lee
Sung-Woo Lee
Suk Jung Kang
Chang Ju Hwang
Nam Heun Kim
Ju-Yul Bae
Yung-Tae Kim
Choon Sung Lee
K. Daniel Riew
Publication date
01-06-2011
Publisher
Springer-Verlag
Published in
European Spine Journal / Issue 6/2011
Print ISSN: 0940-6719
Electronic ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-010-1655-8

Other articles of this Issue 6/2011

European Spine Journal 6/2011 Go to the issue