Published in: Surgical Endoscopy 6/2011

01-06-2011 | New Technology

A novel ultralow-illumination endoscope system

Authors: Keri Kim, Misao Kubota, Yuji Ohkawa, Takashi Shiraishi, Teruo Kawai, Akira Kobayashi, Hiromasa Yamashita, Toshio Chiba

Published in: Surgical Endoscopy | Issue 6/2011

Login to get access

Abstract

Background

Endoscopic surgery has become an accepted major type of minimally invasive surgery. However, complications arising from heat generated by sources of endoscopic illumination can include surgical fire or burns, and intense illumination during ob-gyn/fetoscopic surgery might damage fetal ocular development. Fiber-optic bundles for illumination within the endoscope essentially double the outer diameter of the endoscope, which is a major obstacle to miniaturization and decreasing costs. Light cables also decrease the maneuverability of the endoscope

Methods

We developed a novel endoscope with ultralow illumination to visualize dark body cavities and investigated its feasibility in vivo. An adaptor was created to connect a conventional endoscope to an ultrahigh-sensitivity camera developed by the Japan Broadcasting Corporation (NHK) for broadcasting. The ability to visualize rabbit visceral blood vessels in vivo by the new prototype and by a current endoscope under ultralow illumination provided by a standard light source was compared. In addition, the performance of the two endoscopes was compared using only an extracorporeal flashlight without any specific light source placed within body cavities.

Results

The new endoscope could visualize the target under ultralow illumination of approximately 100 lx. Very little could be visualized using the current endoscope, whereas the prototype generated clear images of the rabbit blood vessels under both ultralow illumination and extracorporeal illumination provided by a flashlight.

Conclusions

The potential for damage caused by a light source can be minimized using our new endoscope, which results in safer and less invasive procedures. Further studies are under way to develop a nonilluminated endoscope without a light cable or source and to miniaturize the camera to decrease costs and improve the maneuverability of the entire endoscope system.
Literature
2.
go back to reference Williams DM, Littwin S, Patterson AJ, Brock-Utne JG (2006) Fiberoptic light source-induced surgical fires—the contribution of forced-air warming blankets. Acta Anaesthesiol Scand 50(4):505–508PubMedCrossRef Williams DM, Littwin S, Patterson AJ, Brock-Utne JG (2006) Fiberoptic light source-induced surgical fires—the contribution of forced-air warming blankets. Acta Anaesthesiol Scand 50(4):505–508PubMedCrossRef
4.
go back to reference Ito M, Harada T, Yamauchi N, Tsudo T, Mizuta M, Terakawa N (2006) Small bowel perforation from a thermal burn caused by contact with the end of a laparoscope during ovarian cystectomy. J Obstet Gynaecol Res 32(4):434–436PubMedCrossRef Ito M, Harada T, Yamauchi N, Tsudo T, Mizuta M, Terakawa N (2006) Small bowel perforation from a thermal burn caused by contact with the end of a laparoscope during ovarian cystectomy. J Obstet Gynaecol Res 32(4):434–436PubMedCrossRef
5.
go back to reference Eggen MA, Brock-Utne JG (1994) Fiberoptic illumination systems can serve as a source of smoldering fires. J Clin Monit 10(4):244–246PubMedCrossRef Eggen MA, Brock-Utne JG (1994) Fiberoptic illumination systems can serve as a source of smoldering fires. J Clin Monit 10(4):244–246PubMedCrossRef
6.
go back to reference Smith LP, Roy S (2008) Fire/burn risk with electrosurgical devices and endoscopy fiberoptic cables. Am J Otolaryngol 29(3):171–176PubMedCrossRef Smith LP, Roy S (2008) Fire/burn risk with electrosurgical devices and endoscopy fiberoptic cables. Am J Otolaryngol 29(3):171–176PubMedCrossRef
7.
go back to reference Aige-Gil V, Murillo-Ferrol N (1992) Effects of white light on the pineal gland of the chick embryo. Histol Histopathol 7(1):1–6PubMed Aige-Gil V, Murillo-Ferrol N (1992) Effects of white light on the pineal gland of the chick embryo. Histol Histopathol 7(1):1–6PubMed
8.
go back to reference Sánchez del Campo F, Puchades Orts A, Panchón Ruíz A, Herrera Lara M (1989) Action of laser light on the ocular development of chick embryos. Anat Anz 169(4):253–259PubMed Sánchez del Campo F, Puchades Orts A, Panchón Ruíz A, Herrera Lara M (1989) Action of laser light on the ocular development of chick embryos. Anat Anz 169(4):253–259PubMed
9.
go back to reference Tanioka K, Matsubara T, Ohkawa Y, Miyakawa K, Suzuki S, Takahata T, Egami N, Ogusu K, Kobayashi A, Hirai T, Kawai T, Hombo M, Yoshida T (2003) Ultra-high-sensitivity new super-HARP pickup tube and its camera. IEICE Trans Electron E86-C(9):1790–1795 Tanioka K, Matsubara T, Ohkawa Y, Miyakawa K, Suzuki S, Takahata T, Egami N, Ogusu K, Kobayashi A, Hirai T, Kawai T, Hombo M, Yoshida T (2003) Ultra-high-sensitivity new super-HARP pickup tube and its camera. IEICE Trans Electron E86-C(9):1790–1795
10.
go back to reference Tanioka K, Yamazaki J, Shidara K, Taketoshi K, Kawamura T, Ishioka S, Takasaki Y (1987) An avalanche-mode amorphous selenium photoconductive layer for use as a camera tube target. IEEE Electron Device Lett 8(9):392–394CrossRef Tanioka K, Yamazaki J, Shidara K, Taketoshi K, Kawamura T, Ishioka S, Takasaki Y (1987) An avalanche-mode amorphous selenium photoconductive layer for use as a camera tube target. IEEE Electron Device Lett 8(9):392–394CrossRef
Metadata
Title
A novel ultralow-illumination endoscope system
Authors
Keri Kim
Misao Kubota
Yuji Ohkawa
Takashi Shiraishi
Teruo Kawai
Akira Kobayashi
Hiromasa Yamashita
Toshio Chiba
Publication date
01-06-2011
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 6/2011
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-010-1498-8