Comparison of four energy-based vascular sealing and cutting instruments: A porcine model
Authors:
Benjamin Person, David A. Vivas, Dan Ruiz, Michael Talcott, James E. Coad, Steven D. Wexner
Published in:
Surgical Endoscopy
|
Issue 2/2008
Login to get access
Abstract
Aim
To compare the safety and efficacy of four energy-based vascular sealing and cutting instruments.
Methods
Blood vessels of various types and diameters were harvested from four pigs using four instruments: Harmonic ACE™ (Ethicon Endo-Surgery, Cincinnati, OH), LigaSure™ V and LigaSure Atlas™ (Valleylab, Inc., Boulder, CO; a division of Tyco Healthcare), and EnSeal™ vessel fusion system (SurgRx, Inc. Redwood City, CA). The diameters of the vessels, speed and adequacy of the cutting and sealing process, and bursting pressures were compared. An additional set of specimens was sealed and left in situ for up to 4 h after which the vessels were harvested and histopathologically analyzed for the degree of thermal injury.
Results
The bursting pressures were significantly higher with EnSeal™ compared to all other instruments (p < 0.0001). The sealing process was significantly shorter with Harmonic ACE™ and significantly longer with LigaSure Atlas™ (p <0.0001). The mean seal width was larger with the LigaSure Atlas™ compared to the other instruments, and it was smaller with EnSeal™ and Harmonic ACE™. Less radial adventitial collagen denaturation was present with EnSeal™ and LigaSure™ V than with the other two instruments; there were no significant differences in collagen denaturation although proximal thermal injury to the smooth muscle in the media of the vessel wall was less common with LigaSure Atlas™ than with the other instruments; however, the numbers were too small for statistical analysis.
Conclusions
The bursting pressures with EnSeal™ were significantly higher than with all the other instruments. Harmonic ACE™ was the fastest sealing instrument and LigaSure Atlas™ was slowest. EnSeal™ created less radial thermal damage to the adventitial collagen of the vessels and LigaSure Atlas™ created less thermal damage to the media of the vessels. The clinical significance of these findings is unknown.