In vivo comparison of the coagulation capability of SonoSurg and Harmonic Ace on 4 mm and 5 mm arteries

Authors: Ronald H. Clements, Rajendra Palepu

Published in: Surgical Endoscopy | Issue 12/2007

Login to get access

Abstract

Introduction

Ultrasonic dissectors are useful to ligate arteries, but their effectiveness on larger arteries has not been thoroughly documented in vivo. The purpose of this study is to determine the bursting pressure of sealed arteries and the incidence of failure to adequately seal 4 mm and 5 mm arteries in a porcine model using two commercially available ultrasonic coagulators.

Methods

The splenic, renal, superior mesenteric and carotid arteries of 60 kg swine were dissected and sealed alternating between SonoSurg (SS), 70% output (Olympus Surgical America) and Harmonic Ace (HA), level 3, (Ethicon Endosurgery).The burst pressures of each sealed segment was measured by subjecting them to infusion of water at 50 ml/hr until the seal burst. The mean burst pressure, incidence of burst pressure less than 360mm Hg and incidence of not sealing were calculated and compared with the t and chi-square tests. Significance was assumed at p = 0.05.

Results

Each instrument was used to divide 44 arteries measuring between 4 and 5 mm (4.7 ± 0.48 mm vs. 4.7 ± 0.44 mm, p = NS, SS vs. HA, respectively) in diameter. The burst pressure (900.2 ± 574.9 mmHg vs. 896.6 ± 481.0 mmHg, p = NS) was not different comparing SS with. HA. Four vessels (9.1%) failed to seal with each instrument and resulted in immediate hemorrhage. Including the vessel that failed to seal initially, five vessels (11.4%) ligated with SS and six (13.6%) ligated with HA had burst pressure less than 360 mm Hg. This incidence of inadequate sealing was not significantly different between the two instruments when compared with chi-square analysis.

Conclusion

SS and HA seal 4–5 mm arteries with similar burst pressures. There is no difference in the incidences of failure to initially seal the arteries or in bursting pressure <360 mmHg in 4–5 mm arteries in the porcine model.
Literature
1.
go back to reference Amaral JF (1994) The experimental development of an ultrasonically activated scalpel for laparoscopic use. Surg Laparosc Endoscopy 4: 92–9 Amaral JF (1994) The experimental development of an ultrasonically activated scalpel for laparoscopic use. Surg Laparosc Endoscopy 4: 92–9
2.
go back to reference Gossot D, Buess G, Cuschieri A, Leporte E, Lirici M, Marvik R, Meijer D, Melzer A, Schurr MO, The EAES Technology Group (1999) Ultrasonic dissection for endoscopic surgery. Surg Endosc 13: 412–417CrossRefPubMed Gossot D, Buess G, Cuschieri A, Leporte E, Lirici M, Marvik R, Meijer D, Melzer A, Schurr MO, The EAES Technology Group (1999) Ultrasonic dissection for endoscopic surgery. Surg Endosc 13: 412–417CrossRefPubMed
3.
go back to reference Abe K, Terashima M, Fujiwara H, Takagane A, Uesugi N, Saito K (2005) Experimental Evaluation of Bursting Pressure in Lymphatic Vessels with Ultrasonically Activated Shears. World J Surg 29: 106–9CrossRefPubMed Abe K, Terashima M, Fujiwara H, Takagane A, Uesugi N, Saito K (2005) Experimental Evaluation of Bursting Pressure in Lymphatic Vessels with Ultrasonically Activated Shears. World J Surg 29: 106–9CrossRefPubMed
4.
go back to reference Nezhat F, Yadav J, Rahaman J, Gretz H 3rd, Gardner GJ, Cohen CJ (2005) Laparoscopic lymphadenectomy for gynecologic malignancies using ultrasonically activated shears: Analysis of first 100 cases. Gynecol Oncol 97: 813–9CrossRefPubMed Nezhat F, Yadav J, Rahaman J, Gretz H 3rd, Gardner GJ, Cohen CJ (2005) Laparoscopic lymphadenectomy for gynecologic malignancies using ultrasonically activated shears: Analysis of first 100 cases. Gynecol Oncol 97: 813–9CrossRefPubMed
5.
go back to reference Spivak H, Richardson WS, Hunter JG (1998) The use of bipolar cautery, laparosonic coagulating shears, and vascular clips for hemostasis of small and medium-sized vessels. Surg Endosc 12: 183–5CrossRefPubMed Spivak H, Richardson WS, Hunter JG (1998) The use of bipolar cautery, laparosonic coagulating shears, and vascular clips for hemostasis of small and medium-sized vessels. Surg Endosc 12: 183–5CrossRefPubMed
6.
go back to reference Harold KL, Pollinger H, Matthews BD, Kercher KW, Sing RF, Heniford BT (2003) Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries. Surg Endosc 17: 1228–30; Epub 2003 Jun 13CrossRefPubMed Harold KL, Pollinger H, Matthews BD, Kercher KW, Sing RF, Heniford BT (2003) Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries. Surg Endosc 17: 1228–30; Epub 2003 Jun 13CrossRefPubMed
Metadata
Title
In vivo comparison of the coagulation capability of SonoSurg and Harmonic Ace on 4 mm and 5 mm arteries
Authors
Ronald H. Clements
Rajendra Palepu
Publication date
01-12-2007
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 12/2007
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-007-9345-2

Other articles of this Issue 12/2007

Surgical Endoscopy 12/2007 Go to the issue