Published in: Surgical Endoscopy 6/2005

01-06-2005 | Original article

Evaluation of adhesion formation, mesh fixation strength, and hydroxyproline content after intraabdominal placement of polytetrafluoroethylene mesh secured using titanium spiral tacks, nitinol anchors, and polypropylene suture or polyglactin 910 suture

Authors: C. S. Joels, B. D. Matthews, K. W. Kercher, C. Austin, H. J. Norton, T. C. Williams, B. T. Heniford

Published in: Surgical Endoscopy | Issue 6/2005

Login to get access

Abstract

Background

The purpose of this study is to evaluate fixation methods for polytetrafluoroethylene (ePTFE) mesh with an in vivo model of laparoscopic ventral hernia repair.

Methods

In 40 New Zealand white rabbits, a 4 × 4-cm ePTFE mesh (n = 80, two per animal) was attached to an intact peritoneum with polyglactin 910 (PG 910) (n = 20) or polypropylene (PP) (n = 20) suture, titanium spiral tacks (TS) (n = 20), or nitinol anchors (NA) (n = 20). Mesh was harvested at 8 and 16 weeks for fixation strength testing, adhesion assessment, and collagen (hydroxyproline) content. Fixation strength on day 0 was determined with mesh attached to harvested abdominal wall. Statistical significance was determined as p < 0.05.

Results

There was no difference in fixation strength between PP (39.1 N) and PG 910 (40.0 N) sutures at time zero. At week 8, PP (25.7 N) was significantly stronger (p < 0.05) than PG 910 (11.4 N) suture, but not at week 16. The fixation strength of TS and NA (day 0, 15.4 vs 7.4 N; week 8, 17.5 vs 15.3 N; week 16, 19.1 vs 13.8 N) was not significantly different. Fixation with PP suture was significantly (p < 0.05) stronger than that with TS and NA at day 0 (39.1, 15.4, and 7.4 N, respectively) but not at weeks 8 or 16. The fixation strength of suture decreased significantly (p < 0.05) from day 0 to week 16 (PP: day 0 = 39.1 N, week 8 = 25.7 N, week 16 = 21.4 N; PG 910: day 0 = 40.0 N, week 8 = 11.4 N, week 16 = 12.8 N). The fixation strength of NA and TS did not change significantly (NA: day 0 = 7.4 N, week 8 = 15.3 N, week 16 = 13.8 N; TS: week 0 = 15.4 N, week 8 = 17.5 N, week 16 = 19.1 N). There were no differences in adhesion area based on fixation device used; however, there were more (p < 0.05) mesh samples using NA with adhesions compared to TS and adhesion tenacity was greater (p < 0.05) compared to that of TS, PP, and PG. Hydroxyproline content at weeks 8 and 16 was similar for all fixation devices.

Conclusions

The initial fixation strength for nonabsorbable suture is significantly greater than that of the metallic fixation devices, but after 8 weeks there is no difference. Laparoscopic ventral hernia repair without transabdominal suture fixation may be predisposed to acute failure. The metallic devices have similar fixation strength, although the incidence of adhesions and tenacity of adhesions appear to be greater with the nitinol anchors. Since these devices have similar fixation strengths and most likely provide adequate supplementation to transabdominal sutures for mesh fixation after laparoscopic ventral hernia repair, their use should be based on other factors, such as their propensity for adhesions, ease of application, and cost.
Literature
1.
go back to reference Carbajo, MA, Martin Olmo, JC, Blanco, JI 1999Laparoscopic treatment vs open surgery in the solution of major incisional and abdominal wall hernias with meshSurg Endosc13250252CrossRefPubMed Carbajo, MA, Martin Olmo, JC, Blanco, JI 1999Laparoscopic treatment vs open surgery in the solution of major incisional and abdominal wall hernias with meshSurg Endosc13250252CrossRefPubMed
2.
go back to reference Carbonell, AM, Harold, KL, Mahmutovic, AJ, Hassan, R, Matthews, BD, Kercher, KW, Sing, RF, Heniford, BT 2003Local injection for the treatment of suture site pain after laparoscopic ventral hernia repairAm Surg69688691PubMed Carbonell, AM, Harold, KL, Mahmutovic, AJ, Hassan, R, Matthews, BD, Kercher, KW, Sing, RF, Heniford, BT 2003Local injection for the treatment of suture site pain after laparoscopic ventral hernia repairAm Surg69688691PubMed
3.
go back to reference Committee on Care and Use of Laboratory Animals1996Guide for the care and use of laboratory animalsU.S. Department of Health and Human Services, Public Health Services, National Institute of HealthBethesda, MD Committee on Care and Use of Laboratory Animals1996Guide for the care and use of laboratory animalsU.S. Department of Health and Human Services, Public Health Services, National Institute of HealthBethesda, MD
4.
go back to reference Garrard, CL, Clements, RH, Nanney, L, Davidson, JM, Richards, WO 1999Adhesion formation is reduced after laparoscopic surgerySurg Endosc131013CrossRefPubMed Garrard, CL, Clements, RH, Nanney, L, Davidson, JM, Richards, WO 1999Adhesion formation is reduced after laparoscopic surgerySurg Endosc131013CrossRefPubMed
5.
go back to reference Heniford, BT, Park, A, Ramshaw, BJ, Voeller, G 2003Laparoscopic repair of ventral hernias: nine year’s experience with 850 consecutive herniasAnn Surg238391399PubMed Heniford, BT, Park, A, Ramshaw, BJ, Voeller, G 2003Laparoscopic repair of ventral hernias: nine year’s experience with 850 consecutive herniasAnn Surg238391399PubMed
6.
go back to reference Karokousis, C, Volpe, C, Tanski, J 1995Use of a mesh for musculoaponeurotic effects of the abdominal wall in cancer surgery and the risk of bowel fistulasJ Am Coll Surg1811116PubMed Karokousis, C, Volpe, C, Tanski, J 1995Use of a mesh for musculoaponeurotic effects of the abdominal wall in cancer surgery and the risk of bowel fistulasJ Am Coll Surg1811116PubMed
7.
go back to reference Katkhouda, N, Mavor, E 2001Use of fibrin sealant for mesh fixation in laparoscopic extraperitoneal inguinal hernia repairAnn Surg233111PubMed Katkhouda, N, Mavor, E 2001Use of fibrin sealant for mesh fixation in laparoscopic extraperitoneal inguinal hernia repairAnn Surg233111PubMed
8.
go back to reference Kaufman, Z, Engelberg, M, Zager, M 1981Fecal fistula: a late complication of Marlex mesh repairDis Colon Rectum24543544PubMed Kaufman, Z, Engelberg, M, Zager, M 1981Fecal fistula: a late complication of Marlex mesh repairDis Colon Rectum24543544PubMed
9.
go back to reference Van’t Riet, M, Vos Steenwijk, PJ, Kleinrensink, GJ, Steyerberg, EW, Bonjer, HJ 2002Tensile strength of mesh fixation methods in laparoscopic incisional hernia repairSurg Endosc1617131716CrossRefPubMed Van’t Riet, M, Vos Steenwijk, PJ, Kleinrensink, GJ, Steyerberg, EW, Bonjer, HJ 2002Tensile strength of mesh fixation methods in laparoscopic incisional hernia repairSurg Endosc1617131716CrossRefPubMed
10.
go back to reference Winslow, ER, Diaz, S, Desai, K, Meininger, T, Soper, NJ, Klingensmith, ME 2004Laparoscopic incisional hernia repair in a porcine model: what do transfixation sutures add?Surg Endosc17.online Winslow, ER, Diaz, S, Desai, K, Meininger, T, Soper, NJ, Klingensmith, ME 2004Laparoscopic incisional hernia repair in a porcine model: what do transfixation sutures add?Surg Endosc17.online
Metadata
Title
Evaluation of adhesion formation, mesh fixation strength, and hydroxyproline content after intraabdominal placement of polytetrafluoroethylene mesh secured using titanium spiral tacks, nitinol anchors, and polypropylene suture or polyglactin 910 suture
Authors
C. S. Joels
B. D. Matthews
K. W. Kercher
C. Austin
H. J. Norton
T. C. Williams
B. T. Heniford
Publication date
01-06-2005
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 6/2005
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-004-8927-5

Other articles of this Issue 6/2005

Surgical Endoscopy 6/2005 Go to the issue