Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 4/2011

01-04-2011 | Miscellaneous

Effects of sweep VEP parameters on visual acuity and contrast thresholds in children and adults

Authors: Fahad M. Almoqbel, Naveen K. Yadav, Susan J. Leat, Liseann M. Head, Elizabeth L. Irving

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 4/2011

Login to get access

Abstract

Background

There are many parameters that may impact the thresholds obtained with sweep visually evoked potentials (sVEP), yet a number of these parameters have not been systematically studied, and there is no recognised standard for sVEP recording. In this study, the effects of electrode placement, temporal frequency, sweep direction, presence of a fixation target, stimulus area, and sweep duration on visual acuity (VA) and contrast thresholds of the sVEP were investigated. Additionally, the effect of these parameters on the number of viable threshold readings obtained from five active electrodes was investigated.

Methods

Participants were six children (aged 6-8 years) and six adults (aged 17-30 years) with normal vision. Binocular sVEP VA and contrast thresholds were measured for two electrode placements (ISCEV and PowerDiva) of five active electrodes, three temporal frequencies (6, 7.5, and 10 Hz), two sweep directions (low to high and high to low), presence or absence of a fixation target, three stimulus areas, and three sweep durations.

Results

There were differences between adults and children with respect to visual acuity, the adults having better VA than the children (p = 0.033 in experiment 2). Overall, there were more viable readings at 7.5 Hz than at either 10 or 6 Hz (p = 0.0014 for VA and 0.001 for contrast thresholds). The adults performed better (in terms of viable readings) with the fixation target than without it (p = 0.04). The smallest stimulus size used gave rise to fewer viable readings in both adults and children (p = 0.022 for VA and 0.0001 for contrast thresholds). The other parameters (electrode placement, sweep direction and sweep duration) did not give rise to significant differences.

Conclusions

A temporal frequency of 7.5 Hz, a stimulus area of 4° or larger for VA and 10° or larger for contrast thresholds, and the use of a fixation target gave more viable readings, and may be indicated for future application. Consideration of the number of viable readings showed more differences between parameters than the actual thresholds, and it is suggested that more readings presumably would yield more reliable threshold measurements.
Literature
1.
go back to reference Almoqbel F, Leat SJ, Irving E (2008) The technique, validity, and clinical use of the sweep VEP. Ophthalmic Physiol Opt 28:393–403PubMedCrossRef Almoqbel F, Leat SJ, Irving E (2008) The technique, validity, and clinical use of the sweep VEP. Ophthalmic Physiol Opt 28:393–403PubMedCrossRef
2.
go back to reference Odom JV, Bach M, Brigell M, Holder GE, McCulloch DL, Tormene AP, Vaegan (2010) ISCEV standard for clinical visual evoked potentials (2009 update). Doc Ophthalmol 120:111–119PubMedCrossRef Odom JV, Bach M, Brigell M, Holder GE, McCulloch DL, Tormene AP, Vaegan (2010) ISCEV standard for clinical visual evoked potentials (2009 update). Doc Ophthalmol 120:111–119PubMedCrossRef
3.
go back to reference Good WV, Hou C (2006) Sweep visual evoked potential grating acuity thresholds paradoxically improve in low-luminance conditions in children with cortical visual impairment. Investig Ophthalmol Vis Sci 47:3220–3224CrossRef Good WV, Hou C (2006) Sweep visual evoked potential grating acuity thresholds paradoxically improve in low-luminance conditions in children with cortical visual impairment. Investig Ophthalmol Vis Sci 47:3220–3224CrossRef
4.
go back to reference Arai M, Katsumi O, Paranhos FRL, Lopes de Faria JM, Hirose T (1997) Comparison of Snellen acuity and objective assessment using the spatial frequency sweep PVER. Graefes Arch Clin Exp Ophthalmol 235:442–447PubMedCrossRef Arai M, Katsumi O, Paranhos FRL, Lopes de Faria JM, Hirose T (1997) Comparison of Snellen acuity and objective assessment using the spatial frequency sweep PVER. Graefes Arch Clin Exp Ophthalmol 235:442–447PubMedCrossRef
5.
go back to reference John FM, Bromham NR, Woodhouse JM, Candy TR (2004) Spatial vision deficits in infants and children with Down syndrome. Investig Ophthalmol Vis Sci 45:1566–1572CrossRef John FM, Bromham NR, Woodhouse JM, Candy TR (2004) Spatial vision deficits in infants and children with Down syndrome. Investig Ophthalmol Vis Sci 45:1566–1572CrossRef
6.
go back to reference Peterzell DH, Kelly JP (1997) Development of spatial frequency tuned "covariance" channels: individual differences in the electrophysiological (VEP) contrast sensitivity function. Optom Vis Sci 74:800–807PubMedCrossRef Peterzell DH, Kelly JP (1997) Development of spatial frequency tuned "covariance" channels: individual differences in the electrophysiological (VEP) contrast sensitivity function. Optom Vis Sci 74:800–807PubMedCrossRef
7.
go back to reference Prager TC, Zou YL, Jensen CL, Fraley JK, Anderson RE, Heird WC (1999) Evaluation of methods for assessing visual function of infants. J AAPOS 3:275–282PubMedCrossRef Prager TC, Zou YL, Jensen CL, Fraley JK, Anderson RE, Heird WC (1999) Evaluation of methods for assessing visual function of infants. J AAPOS 3:275–282PubMedCrossRef
8.
go back to reference Riddell PM, Ladenheim B, Mast J, Catalano T, Nobile R, Hainline L (1997) Comparison of measures of visual acuity in infants: teller acuity cards and sweep visual evoked potentials. Optom Vis Sci 74:702–707PubMedCrossRef Riddell PM, Ladenheim B, Mast J, Catalano T, Nobile R, Hainline L (1997) Comparison of measures of visual acuity in infants: teller acuity cards and sweep visual evoked potentials. Optom Vis Sci 74:702–707PubMedCrossRef
9.
go back to reference Norcia AM, Tyler CW (1985) Spatial frequency sweep VEP: visual acuity during the first year of life. Vis Res 25:1399–1408PubMedCrossRef Norcia AM, Tyler CW (1985) Spatial frequency sweep VEP: visual acuity during the first year of life. Vis Res 25:1399–1408PubMedCrossRef
10.
go back to reference Norcia AM, Tyler CW (1985) Infant VEP acuity measurements: analysis of individual differences and measurement error. Electroencephalogr Clin Neurophysiol 61:359–369PubMedCrossRef Norcia AM, Tyler CW (1985) Infant VEP acuity measurements: analysis of individual differences and measurement error. Electroencephalogr Clin Neurophysiol 61:359–369PubMedCrossRef
11.
go back to reference Norcia AM, Tyler CW, Hamer RD (1988) High visual contrast sensitivity in the young human infant. Investig Ophthalmol Vis Sci 29:44–49 Norcia AM, Tyler CW, Hamer RD (1988) High visual contrast sensitivity in the young human infant. Investig Ophthalmol Vis Sci 29:44–49
12.
go back to reference Norcia AM, Tyler CW, Hamer RD (1989) Measurement of spatial contrast sensitivity with the swept contrast VEP. Vis Res 29:627–637PubMedCrossRef Norcia AM, Tyler CW, Hamer RD (1989) Measurement of spatial contrast sensitivity with the swept contrast VEP. Vis Res 29:627–637PubMedCrossRef
13.
go back to reference Norcia AM, Tyler CW, Hamer RD (1990) Development of contrast sensitivity in the human infant. Vis Res 30:1475–1486PubMedCrossRef Norcia AM, Tyler CW, Hamer RD (1990) Development of contrast sensitivity in the human infant. Vis Res 30:1475–1486PubMedCrossRef
14.
go back to reference Hamer RD, Norcia AM, Tyler CW, Hsu-Winges C (1989) The development of monocular and binocular VEP acuity. Vis Res 29:397–408PubMedCrossRef Hamer RD, Norcia AM, Tyler CW, Hsu-Winges C (1989) The development of monocular and binocular VEP acuity. Vis Res 29:397–408PubMedCrossRef
15.
go back to reference Bach M, Maurer JP, Wolf ME (2008) Visual evoked potential-based acuity assessment in normal vision, artificially degraded vision, and in patients. Br J Ophthalmol 9:396–403CrossRef Bach M, Maurer JP, Wolf ME (2008) Visual evoked potential-based acuity assessment in normal vision, artificially degraded vision, and in patients. Br J Ophthalmol 9:396–403CrossRef
16.
go back to reference da Costa MF, Salomão SR, Berezovsky A, de Haro FM, Ventura DF (2004) Relationship between vision and motor impairment in children with spastic cerebral palsy: new evidence from electrophysiology. Behav Brain Res 149:145–150PubMedCrossRef da Costa MF, Salomão SR, Berezovsky A, de Haro FM, Ventura DF (2004) Relationship between vision and motor impairment in children with spastic cerebral palsy: new evidence from electrophysiology. Behav Brain Res 149:145–150PubMedCrossRef
17.
go back to reference Oliveira AGF, Costa MF, de Souza JM, Ventura DF (2004) Contrast sensitivity threshold measured by sweep-visual evoked potential in term and preterm infants at 3 and 10 months of age. Braz J Med Biol Res 37:1389–1396PubMed Oliveira AGF, Costa MF, de Souza JM, Ventura DF (2004) Contrast sensitivity threshold measured by sweep-visual evoked potential in term and preterm infants at 3 and 10 months of age. Braz J Med Biol Res 37:1389–1396PubMed
18.
go back to reference Crow RW, Levin LB, LaBree L, Rubin R, Feldon SE (2003) Sweep visual evoked potential evaluation of contrast sensitivity in Alzheimer’s dementia. Investig Ophthalmol Vis Sci 44:875–878CrossRef Crow RW, Levin LB, LaBree L, Rubin R, Feldon SE (2003) Sweep visual evoked potential evaluation of contrast sensitivity in Alzheimer’s dementia. Investig Ophthalmol Vis Sci 44:875–878CrossRef
19.
go back to reference Gottlob I, Fendick MG, Guo S, Zubcov AA, Odom JV, Reinecke RD (1990) Visual acuity measurement by swept spatial frequency visual-evoked-cortical potentials (VECPS): clinical application in children with various visual disorders. J Pediatr Ophthalmol Strabismus 27:40–47PubMed Gottlob I, Fendick MG, Guo S, Zubcov AA, Odom JV, Reinecke RD (1990) Visual acuity measurement by swept spatial frequency visual-evoked-cortical potentials (VECPS): clinical application in children with various visual disorders. J Pediatr Ophthalmol Strabismus 27:40–47PubMed
20.
go back to reference Gottlob I, Wizov SS, Odom JV, Reinecke RD (1993) Predicting optotype visual acuity by swept spatial visual-evoked potential. Clin Vis Sci 8:417–423 Gottlob I, Wizov SS, Odom JV, Reinecke RD (1993) Predicting optotype visual acuity by swept spatial visual-evoked potential. Clin Vis Sci 8:417–423
21.
go back to reference Lauritzen L, Jørgensen MH, Michaelsen KF (2004) Test-retest reliability of swept visual evoked potential measurements of infant visual acuity and contrast sensitivity. Pediatr Res 55:701–708PubMedCrossRef Lauritzen L, Jørgensen MH, Michaelsen KF (2004) Test-retest reliability of swept visual evoked potential measurements of infant visual acuity and contrast sensitivity. Pediatr Res 55:701–708PubMedCrossRef
22.
go back to reference Ridder WH III, Rouse MW (2007) Predicting potential acuities in amblyopes. Doc Ophthalmol 114:135–145PubMedCrossRef Ridder WH III, Rouse MW (2007) Predicting potential acuities in amblyopes. Doc Ophthalmol 114:135–145PubMedCrossRef
23.
go back to reference Seiple WH, Holopigian K (1989) An examination of VEP response phase. Electroencephalogr Clin Neurophysiol 73:520–531PubMedCrossRef Seiple WH, Holopigian K (1989) An examination of VEP response phase. Electroencephalogr Clin Neurophysiol 73:520–531PubMedCrossRef
24.
go back to reference Tyler CW, Apkarian P, Levi DM, Nakayama K (1979) Rapid assessment of visual function: an electronic sweep technique for the pattern visual evoked potential. Investig Ophthalmol Vis Sci 18:703–713 Tyler CW, Apkarian P, Levi DM, Nakayama K (1979) Rapid assessment of visual function: an electronic sweep technique for the pattern visual evoked potential. Investig Ophthalmol Vis Sci 18:703–713
25.
go back to reference Zemon V, Hartmann EE, Gordon J, Prünte-Glowazki A (1997) An electrophysiological technique for assessment of the development of spatial vision. Optom Vis Sci 74:708–716PubMedCrossRef Zemon V, Hartmann EE, Gordon J, Prünte-Glowazki A (1997) An electrophysiological technique for assessment of the development of spatial vision. Optom Vis Sci 74:708–716PubMedCrossRef
26.
go back to reference Zhou P, Zhao MW, Li XX, Hu XF, Wu X, Niu LJ, Yu WZ, Xu XL (2008) A new method for extrapolating the sweep pattern visual evoked potential acuity. Doc Ophthalmol 117:85–91PubMedCrossRef Zhou P, Zhao MW, Li XX, Hu XF, Wu X, Niu LJ, Yu WZ, Xu XL (2008) A new method for extrapolating the sweep pattern visual evoked potential acuity. Doc Ophthalmol 117:85–91PubMedCrossRef
27.
go back to reference Allen D, Norcia AM, Tyler CW (1986) Comparative study of electrophysiological and psychophysical measurement of the contrast sensitivity function in humans. Am J Optom Physiol Opt 63:442–449PubMed Allen D, Norcia AM, Tyler CW (1986) Comparative study of electrophysiological and psychophysical measurement of the contrast sensitivity function in humans. Am J Optom Physiol Opt 63:442–449PubMed
28.
go back to reference Sokol S, Moskowitz A, McCormack G (1992) Infant VEP and preferential looking acuity measured with phase alternating gratings. Investig Ophthalmol Vis Sci 33:3156–3161 Sokol S, Moskowitz A, McCormack G (1992) Infant VEP and preferential looking acuity measured with phase alternating gratings. Investig Ophthalmol Vis Sci 33:3156–3161
29.
go back to reference Allen D, Tyler CW, Norcia AM (1996) Development of grating acuity and contrast sensitivity in the central and peripheral visual field of the human infant. Vis Res 36:1945–1953PubMedCrossRef Allen D, Tyler CW, Norcia AM (1996) Development of grating acuity and contrast sensitivity in the central and peripheral visual field of the human infant. Vis Res 36:1945–1953PubMedCrossRef
30.
go back to reference Nelson JI, Seiple WH, Kupersmith MJ, Carr RE (1984) A rapid evoked potential index of cortical adaptation. Electroencephalogr Clin Neurophysiol 59:454–464PubMedCrossRef Nelson JI, Seiple WH, Kupersmith MJ, Carr RE (1984) A rapid evoked potential index of cortical adaptation. Electroencephalogr Clin Neurophysiol 59:454–464PubMedCrossRef
31.
go back to reference Brigell MG, Peachey NS, Seiple WH (1987) Pattern electroretinogram threshold does not show contrast adaptation. Investig Ophthalmol Vis Sci 28:1614–1616 Brigell MG, Peachey NS, Seiple WH (1987) Pattern electroretinogram threshold does not show contrast adaptation. Investig Ophthalmol Vis Sci 28:1614–1616
32.
go back to reference Seiple WH, Kupersmith MJ, Nelson JI, Carr RE (1988) Evoked potential assessment of cortical adaptation. Appl Opt 27:1089–1093PubMedCrossRef Seiple WH, Kupersmith MJ, Nelson JI, Carr RE (1988) Evoked potential assessment of cortical adaptation. Appl Opt 27:1089–1093PubMedCrossRef
33.
go back to reference Spinelli D, Pirchio M, Sandini G (1983) Visual acuity in the young infant is highest in a small retinal area. Vis Res 23:1133–1136PubMedCrossRef Spinelli D, Pirchio M, Sandini G (1983) Visual acuity in the young infant is highest in a small retinal area. Vis Res 23:1133–1136PubMedCrossRef
34.
go back to reference Katsumi O, Hirose T, Tsukada T (1988) Effect of number of elements and size of stimulus field on recordability of pattern reversal visual evoked response. Investig Ophthalmol Vis Sci 29:922–927 Katsumi O, Hirose T, Tsukada T (1988) Effect of number of elements and size of stimulus field on recordability of pattern reversal visual evoked response. Investig Ophthalmol Vis Sci 29:922–927
35.
go back to reference Sakaue H, Katsumi O, Mehta M, Hirose T (1990) Simultaneous pattern reversal ERG and VER recordings. Investig Ophthalmol Vis Sci 31:506–511 Sakaue H, Katsumi O, Mehta M, Hirose T (1990) Simultaneous pattern reversal ERG and VER recordings. Investig Ophthalmol Vis Sci 31:506–511
36.
go back to reference Norcia AM, Tyler CW, Allen D (1986) Electrophysiological assessment of contrast sensitivity in human infants. Am J Optom Physiol Opt 61:12–15 Norcia AM, Tyler CW, Allen D (1986) Electrophysiological assessment of contrast sensitivity in human infants. Am J Optom Physiol Opt 61:12–15
37.
go back to reference Sokol S, Moskowitz A, McCormack G, Augliere R (1988) Infant grating acuity is temporally tuned. Vis Res 28:1357–1366PubMedCrossRef Sokol S, Moskowitz A, McCormack G, Augliere R (1988) Infant grating acuity is temporally tuned. Vis Res 28:1357–1366PubMedCrossRef
38.
go back to reference Katsumi O, Denno S, Arai M, de Faria JML, Hirose T (1997) Comparison of preferential looking acuity and pattern reversal visual evoked response acuity in pediatric patients. Graefes Arch Clin Exp Ophthalmol 235:684–690PubMedCrossRef Katsumi O, Denno S, Arai M, de Faria JML, Hirose T (1997) Comparison of preferential looking acuity and pattern reversal visual evoked response acuity in pediatric patients. Graefes Arch Clin Exp Ophthalmol 235:684–690PubMedCrossRef
39.
go back to reference Ridder WH III, McCulloch D, Herbert AM (1998) Stimulus duration, neural adaptation, and sweep visual evoked potential acuity estimates. Investig Ophthalmol Vis Sci 39:2759–2768 Ridder WH III, McCulloch D, Herbert AM (1998) Stimulus duration, neural adaptation, and sweep visual evoked potential acuity estimates. Investig Ophthalmol Vis Sci 39:2759–2768
40.
go back to reference Yadav NK, Almoqbel F, Head L, Irving EL, Leat SJ (2009) Threshold determination in sweep VEP and the effects of criterion. Doc Ophthalmol 119:109–121PubMedCrossRef Yadav NK, Almoqbel F, Head L, Irving EL, Leat SJ (2009) Threshold determination in sweep VEP and the effects of criterion. Doc Ophthalmol 119:109–121PubMedCrossRef
41.
go back to reference Tang Y, Norcia AM (1995) An adaptive filter for steady-state evoked responses. Electroencephalogr Clin Neurophysiol 96:268–277PubMedCrossRef Tang Y, Norcia AM (1995) An adaptive filter for steady-state evoked responses. Electroencephalogr Clin Neurophysiol 96:268–277PubMedCrossRef
42.
go back to reference Winer B (1971) Multifactor experiments having repeated measures on the same elements. Statistical principles in experimental design. McGraw-Hill, New York, pp 514–603 Winer B (1971) Multifactor experiments having repeated measures on the same elements. Statistical principles in experimental design. McGraw-Hill, New York, pp 514–603
43.
go back to reference Winer B (1971) Single-factor experiments having repeated measures on the same elements. Statistical principles in experimental design. McGraw-Hill, New York, pp 261–305 Winer B (1971) Single-factor experiments having repeated measures on the same elements. Statistical principles in experimental design. McGraw-Hill, New York, pp 261–305
44.
go back to reference Nelson JI, Kupersmith MJ, Seiple WH, Weiss PA, Carr RE (1984) Spatio-temporal conditions which elicit or abolish the oblique effect in man: direct measurement with swept evoked potential. Vis Res 24:579–586PubMedCrossRef Nelson JI, Kupersmith MJ, Seiple WH, Weiss PA, Carr RE (1984) Spatio-temporal conditions which elicit or abolish the oblique effect in man: direct measurement with swept evoked potential. Vis Res 24:579–586PubMedCrossRef
45.
go back to reference Seiple WH, Kupersmith MJ, Nelson JI, Carr RE (1984) The assessment of evoked potential contrast thresholds using real-time retrieval. Investig Ophthalmol Vis Sci 25:627–631 Seiple WH, Kupersmith MJ, Nelson JI, Carr RE (1984) The assessment of evoked potential contrast thresholds using real-time retrieval. Investig Ophthalmol Vis Sci 25:627–631
Metadata
Title
Effects of sweep VEP parameters on visual acuity and contrast thresholds in children and adults
Authors
Fahad M. Almoqbel
Naveen K. Yadav
Susan J. Leat
Liseann M. Head
Elizabeth L. Irving
Publication date
01-04-2011
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 4/2011
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-010-1469-8

Other articles of this Issue 4/2011

Graefe's Archive for Clinical and Experimental Ophthalmology 4/2011 Go to the issue