Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 11/2010

01-11-2010 | Miscellaneous

Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography

Authors: Jehn-Yu Huang, Melike Pekmezci, Stephanie Yaplee, Shan Lin

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 11/2010

Login to get access

Abstract

Background

To assess the intra-examiner repeatability and agreement of corneal pachymetry maps measured by time-domain (Visante OCT) and Fourier-domain optical coherence tomography (RTVue OCT).

Methods

This observational cross-sectional study enrolled 72 eyes from 72 volunteers. Only one eye of each patient was chosen randomly to receive repeated scanning with both devices by the same examiner, in order to test the intra-observer repeatability. The first scan by each device from all enrolled eyes was used to analyze the difference and agreement between the two devices. The agreement between the two devices was analyzed by the Bland–Altman method. Intra-observer repeatability of each OCT device was analyzed by intra-class correlation (ICC).

Results

The mean corneal thickness of the central 2 mm zone was 524.3 ± 35.7 µm and 525.4 ± 35.3 µm by the Visante OCT and the RTVue OCT respectively (p = 0.089). The intra-observer repeatability of the RTVue OCT (ICC = 0.994) was superior to that of the Visante OCT (ICC = 0.989) in the central 2 mm zone. However, the intra-observer repeatability of the RTVue OCT in the pericentral 2 to 5 mm zone was not superior to that of the Visante OCT. Both OCT devices had similar repeatability in the pericentral 2 to 5 mm zone [ICC of the Visante OCT = 0.991; 95% confidence interval (CI) 0.986–0.995), ICC of the RTVue OCT = 0.991; 95% CI 0.985–0.994)).

Conclusions

The difference in CCT measurement by the Visante OCT and the RTVue OCT is probably too small to influence clinical decision making for refractive surgery and glaucoma management. The RTVue OCT demonstrated better intra-observer repeatability in the central 2 mm zone, which probably was related to its rapid image acquisition capability.
Literature
2.
go back to reference Johnson M, Kass MA, Moses RA, Grodzki WJ (1978) Increased corneal thickness simulating elevated intraocular pressure. Arch Ophthalmol 96:664–665PubMed Johnson M, Kass MA, Moses RA, Grodzki WJ (1978) Increased corneal thickness simulating elevated intraocular pressure. Arch Ophthalmol 96:664–665PubMed
3.
go back to reference Whitacre MM, Stein RA, Hassanein K (1993) The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 115:592–596PubMed Whitacre MM, Stein RA, Hassanein K (1993) The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 115:592–596PubMed
4.
go back to reference Stodtmeister R (1998) Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand 76:319–324CrossRefPubMed Stodtmeister R (1998) Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand 76:319–324CrossRefPubMed
5.
go back to reference Feltgen N, Leifert D, Funk J (2001) Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol 85:85–87CrossRefPubMed Feltgen N, Leifert D, Funk J (2001) Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol 85:85–87CrossRefPubMed
6.
go back to reference Gordon MO, Beiser JA, Brandt JD, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK 2nd, Wilson MR, Kass MA (2002) The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol 120:714–720; discussion 829–730, doi:ecs20044 [pii]PubMed Gordon MO, Beiser JA, Brandt JD, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK 2nd, Wilson MR, Kass MA (2002) The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol 120:714–720; discussion 829–730, doi:ecs20044 [pii]PubMed
10.
go back to reference Drexler W, Baumgartner A, Findl O, Hitzenberger CK, Sattmann H, Fercher AF (1997) Submicrometer precision biometry of the anterior segment of the human eye. Invest Ophthalmol Vis Sci 38:1304–1313PubMed Drexler W, Baumgartner A, Findl O, Hitzenberger CK, Sattmann H, Fercher AF (1997) Submicrometer precision biometry of the anterior segment of the human eye. Invest Ophthalmol Vis Sci 38:1304–1313PubMed
11.
go back to reference Bechmann M, Thiel MJ, Neubauer AS, Ullrich S, Ludwig K, Kenyon KR, Ulbig MW (2001) Central corneal thickness measurement with a retinal optical coherence tomography device versus standard ultrasonic pachymetry. Cornea 20:50–54CrossRefPubMed Bechmann M, Thiel MJ, Neubauer AS, Ullrich S, Ludwig K, Kenyon KR, Ulbig MW (2001) Central corneal thickness measurement with a retinal optical coherence tomography device versus standard ultrasonic pachymetry. Cornea 20:50–54CrossRefPubMed
15.
go back to reference Wong AC, Wong CC, Yuen NS, Hui SP (2002) Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetry. Eye 16:715–721. doi:10.1038/sj.eye.6700211 CrossRefPubMed Wong AC, Wong CC, Yuen NS, Hui SP (2002) Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetry. Eye 16:715–721. doi:10.​1038/​sj.​eye.​6700211 CrossRefPubMed
16.
17.
go back to reference Viestenz A, Vogt S, Langenbucher A, Walter S, Behrens-Baumann W (2009) Biometry of the anterior segment using optical coherence tomography: Evaluation of different devices and analysis programs. Ophthalmologe 106:723–728, doi:10.1007/s00347-008-1836-6 CrossRefPubMed Viestenz A, Vogt S, Langenbucher A, Walter S, Behrens-Baumann W (2009) Biometry of the anterior segment using optical coherence tomography: Evaluation of different devices and analysis programs. Ophthalmologe 106:723–728, doi:10.​1007/​s00347-008-1836-6 CrossRefPubMed
18.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
19.
go back to reference Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMed Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMed
20.
24.
go back to reference de Boer JF, Cense B, Park BH, Pierce MC, Tearney GJ, Bouma BE (2003) Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography. Opt Lett 28:2067–2069CrossRefPubMed de Boer JF, Cense B, Park BH, Pierce MC, Tearney GJ, Bouma BE (2003) Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography. Opt Lett 28:2067–2069CrossRefPubMed
29.
go back to reference Brandt JD, Beiser JA, Kass MA, Gordon MO (2001) Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS). Ophthalmology 108:1779–1788CrossRefPubMed Brandt JD, Beiser JA, Kass MA, Gordon MO (2001) Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS). Ophthalmology 108:1779–1788CrossRefPubMed
30.
go back to reference Shimmyo M, Ross AJ, Moy A, Mostafavi R (2003) Intraocular pressure, Goldmann applanation tension, corneal thickness, and corneal curvature in Caucasians, Asians, Hispanics, and African Americans. Am J Ophthalmol 136:603–613, doi:S0002939403004240 [pii]CrossRefPubMed Shimmyo M, Ross AJ, Moy A, Mostafavi R (2003) Intraocular pressure, Goldmann applanation tension, corneal thickness, and corneal curvature in Caucasians, Asians, Hispanics, and African Americans. Am J Ophthalmol 136:603–613, doi:S000293940300424​0 [pii]CrossRefPubMed
31.
go back to reference La Rosa FA, Gross RL, Orengo-Nania S (2001) Central corneal thickness of Caucasians and African Americans in glaucomatous and nonglaucomatous populations. Arch Ophthalmol 119:23–27,. doi:ecs90288 [pii]PubMed La Rosa FA, Gross RL, Orengo-Nania S (2001) Central corneal thickness of Caucasians and African Americans in glaucomatous and nonglaucomatous populations. Arch Ophthalmol 119:23–27,. doi:ecs90288 [pii]PubMed
Metadata
Title
Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography
Authors
Jehn-Yu Huang
Melike Pekmezci
Stephanie Yaplee
Shan Lin
Publication date
01-11-2010
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 11/2010
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-010-1360-7

Other articles of this Issue 11/2010

Graefe's Archive for Clinical and Experimental Ophthalmology 11/2010 Go to the issue