01-03-2010 | Neuro-ophthalmology

Optimizing the calibration and interpretation of dynamic ocular force measurements

Authors: William H. Morgan, Stephen J. Cringle, Min H. Kang, Surinder Pandav, Chandrakumar Balaratnasingam, Don Ezekial, Dao-Yi Yu

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 3/2010

Login to get access

Abstract

Background

Ophthalmodynamometric studies can provide useful clinical information regarding glaucoma, cerebrospinal fluid pressure, and vascular disease but are affected by variable reproducibility and unknown calibration of force in terms of intraocular pressure (IOP). The aim of this study was to calculate calibration factors and identify key design principles using three different types of ophthalmodynamometers.

Methods

We constructed a modified ophthalmodynamometer named OcuDyn using a large contact lens inside a ring force transducer with continuous signal acquisition and averaging. OcuDyn and Sisler ophthalmodynamometers were applied to isolated pig eyes which were cannulated and connected to a fluid reservoir and a pressure transducer to measure induced IOP at increasing force application. Using these two devices and a Meditron ophthalmodynamometer, we measured the minimum ophthalmodynamometric force (ODF) required to induce pulsation in the hemi-veins and diastolic retinal artery of each eye in glaucoma patient volunteers and family members.

Results

Blood pressure was measured with the sphygmomanometer cuff held at eye level. In pigs, the relationship between induced IOP and ODF was strong (minimum r > 0.98, p < 0.001). In humans, the pressure increment (blood pressure – baseline IOP) correlated highly with human arterial ODF (r = 0.83, n = 75, p < 0.001, mean slope 0.32). Mean coefficients of variation were 12.4% (n = 66) for veins and 5.6% (n = 39) for arteries in diastole; these results did not differ significantly with those from the Meditron but were significantly lower than results with the Sisler (20%, p = 0.0014).

Conclusions

Linearity between induced IOP and ODF is strong, suggesting that calibration of 0.32 mmHg/g for OcuDyn or using two point blood pressure measures allows for valid interpolation. The better optics of the Meditron and OcuDyn result in more repeatable end-point determination and outweigh benefits conferred by signal averaging or slit lamp mounting. We calculate calibration factors of 0.89 mmHg/arbitrary unit for Meditron and 0.82 mmHg/g for Sisler. These factors may be useful in the estimation of ocular vascular, orbital, and cerebrospinal fluid pressures.
Literature
1.
go back to reference Firsching R, Schutze M, Motschmann M, Behrends-Baumann W (2000) Venous ophthalmodynamometry: a noninvasive method for assessment of intracranial pressure. J Neurosurg 93:33–36CrossRefPubMed Firsching R, Schutze M, Motschmann M, Behrends-Baumann W (2000) Venous ophthalmodynamometry: a noninvasive method for assessment of intracranial pressure. J Neurosurg 93:33–36CrossRefPubMed
2.
go back to reference Firsching R, Schutze M, Motschmann M, Behrens-Baumann W, Meyer-Schwickerath R (1998) Non-invasive measurement of intracranial pressure. Lancet 351:523–524CrossRefPubMed Firsching R, Schutze M, Motschmann M, Behrens-Baumann W, Meyer-Schwickerath R (1998) Non-invasive measurement of intracranial pressure. Lancet 351:523–524CrossRefPubMed
3.
go back to reference Querfurth HW, Arms SW, Lichy CM, Irwin WT, Steiner T (2004) Prediction of intracranial pressure from noninvasive transocular venous and arterial hemodynamic measurements: a pilot study. Neurocrit Care 1:183–194CrossRefPubMed Querfurth HW, Arms SW, Lichy CM, Irwin WT, Steiner T (2004) Prediction of intracranial pressure from noninvasive transocular venous and arterial hemodynamic measurements: a pilot study. Neurocrit Care 1:183–194CrossRefPubMed
4.
go back to reference Jonas JB, Pfeil K, Chatzikonstantinou A, Rensch F (2008) Ophthalmodynamometric measurement of central retinal vein pressure as surrogate of intracranial pressure in idiopathic intracranial hypertension. Graefes Arch Clin Exp Ophthalmol 246:1059–1060CrossRefPubMed Jonas JB, Pfeil K, Chatzikonstantinou A, Rensch F (2008) Ophthalmodynamometric measurement of central retinal vein pressure as surrogate of intracranial pressure in idiopathic intracranial hypertension. Graefes Arch Clin Exp Ophthalmol 246:1059–1060CrossRefPubMed
5.
go back to reference Beaumont PE, Kang HK (1994) Ophthalmodynamometry and corticosteroids in central retinal vein occlusion. Aust NZ J Ophthalmol 22:271–274CrossRef Beaumont PE, Kang HK (1994) Ophthalmodynamometry and corticosteroids in central retinal vein occlusion. Aust NZ J Ophthalmol 22:271–274CrossRef
6.
go back to reference Jonas JB (2004) Ophthalmodynamometric measurement of orbital tissue pressure in thyroid-associated orbitopathy. Acta Ophthalmol Scand 82:239CrossRefPubMed Jonas JB (2004) Ophthalmodynamometric measurement of orbital tissue pressure in thyroid-associated orbitopathy. Acta Ophthalmol Scand 82:239CrossRefPubMed
7.
go back to reference Jonas JB, Kaltakji F (2008) Ophthalmodynamometry and ischemic ophthalmopathy. Graefes Arch Clin Exp Ophthalmol 246:783–784CrossRefPubMed Jonas JB, Kaltakji F (2008) Ophthalmodynamometry and ischemic ophthalmopathy. Graefes Arch Clin Exp Ophthalmol 246:783–784CrossRefPubMed
8.
go back to reference Jonas JB, Harder B (2008) Central retinal artery and vein collapse pressure in giant cell arteritis versus nonarteritic anterior ischaemic optic neuropathy. Eye 22:556–558CrossRefPubMed Jonas JB, Harder B (2008) Central retinal artery and vein collapse pressure in giant cell arteritis versus nonarteritic anterior ischaemic optic neuropathy. Eye 22:556–558CrossRefPubMed
9.
go back to reference Morgan WH, Hazelton ML, Azar SL, House PH, Yu DY, Cringle SJ, Balaratnasingam C (2004) Retinal venous pulsation in glaucoma and glaucoma suspects. Opthalmology 111:1489–1494CrossRef Morgan WH, Hazelton ML, Azar SL, House PH, Yu DY, Cringle SJ, Balaratnasingam C (2004) Retinal venous pulsation in glaucoma and glaucoma suspects. Opthalmology 111:1489–1494CrossRef
10.
go back to reference Jonas JB (2003) Central retinal artery and vein collapse pressure in eyes with chronic open-angle glaucoma. Br J Ophthalmol 87:949–951CrossRefPubMed Jonas JB (2003) Central retinal artery and vein collapse pressure in eyes with chronic open-angle glaucoma. Br J Ophthalmol 87:949–951CrossRefPubMed
11.
go back to reference Jonas JB (2003) Reproducibility of ophthalmodynamometric measurements of central retinal artery and vein collapse pressure. Br J Ophthalmol 87:577–579CrossRefPubMed Jonas JB (2003) Reproducibility of ophthalmodynamometric measurements of central retinal artery and vein collapse pressure. Br J Ophthalmol 87:577–579CrossRefPubMed
12.
go back to reference Sisler HA (1972) Optical-corneal pressure ophthalmodynamometer. Am J Ophthalmol 74:987–988PubMed Sisler HA (1972) Optical-corneal pressure ophthalmodynamometer. Am J Ophthalmol 74:987–988PubMed
13.
go back to reference Morgan WH, Balaratnasingam C, Hazelton ML, House PH, Cringle SJ, Yu DY (2005) The force required to induce hemivein pulsation is associated with the site of maximal field loss in glaucoma. Invest Ophthalmol Vis Sci 46:1307–1312CrossRefPubMed Morgan WH, Balaratnasingam C, Hazelton ML, House PH, Cringle SJ, Yu DY (2005) The force required to induce hemivein pulsation is associated with the site of maximal field loss in glaucoma. Invest Ophthalmol Vis Sci 46:1307–1312CrossRefPubMed
14.
go back to reference Jonas JB, Harder B (2007) Ophthalmodynamometric differences between ischemic vs nonischemic retinal vein occlusion. Am J Ophthalmol 143:112–116CrossRefPubMed Jonas JB, Harder B (2007) Ophthalmodynamometric differences between ischemic vs nonischemic retinal vein occlusion. Am J Ophthalmol 143:112–116CrossRefPubMed
15.
go back to reference Jonas JB (2004) Ophthalmodynamometric determination of the central retinal vessel collapse pressure correlated with systemic blood pressure. Br J Ophthalmol 88:501–504CrossRefPubMed Jonas JB (2004) Ophthalmodynamometric determination of the central retinal vessel collapse pressure correlated with systemic blood pressure. Br J Ophthalmol 88:501–504CrossRefPubMed
16.
go back to reference Meyer-Schwickerath R, Stodtmeister R, Hartmann K (2004) Non-invasive determination of intracranial pressure. Physiological basis and practical procedure. Klin Monatsbl Augenheilkd 221:1007–1011CrossRefPubMed Meyer-Schwickerath R, Stodtmeister R, Hartmann K (2004) Non-invasive determination of intracranial pressure. Physiological basis and practical procedure. Klin Monatsbl Augenheilkd 221:1007–1011CrossRefPubMed
17.
go back to reference Lowe RD (1962) Calibration of Bailliart's ophthalmodynamometer. Arch Ophthalmol 67:424–427PubMed Lowe RD (1962) Calibration of Bailliart's ophthalmodynamometer. Arch Ophthalmol 67:424–427PubMed
18.
go back to reference Pierscionek BK, Ashkenazi I, Schachter J (2007) The effect of changing intraocular pressure on the corneal and scleral curvatures in the fresh porcine eye. Br J Ophthalmol 91:801–803CrossRefPubMed Pierscionek BK, Ashkenazi I, Schachter J (2007) The effect of changing intraocular pressure on the corneal and scleral curvatures in the fresh porcine eye. Br J Ophthalmol 91:801–803CrossRefPubMed
19.
go back to reference Bartholomew LR, Pang DX, Sam DA, Cavender JC (1997) Ultrasound biomicroscopy of globes from young adult pigs. Am J Vet Res 58:942–948PubMed Bartholomew LR, Pang DX, Sam DA, Cavender JC (1997) Ultrasound biomicroscopy of globes from young adult pigs. Am J Vet Res 58:942–948PubMed
20.
go back to reference Morgan WH, Yu DY, Alder VA, Cringle SJ, Constable IJ (1998) Relation between pressure determined by ophthalmodynamometry and aortic pressure in the dog. Br J Ophthalmol 82:821–825CrossRefPubMed Morgan WH, Yu DY, Alder VA, Cringle SJ, Constable IJ (1998) Relation between pressure determined by ophthalmodynamometry and aortic pressure in the dog. Br J Ophthalmol 82:821–825CrossRefPubMed
21.
go back to reference Fung YC (1984) Biodynamics: circulation. Springer, Berlin Heidelberg New York, pp 132–134 Fung YC (1984) Biodynamics: circulation. Springer, Berlin Heidelberg New York, pp 132–134
22.
go back to reference Feigl EO (1974) The arterial system. In: Ruch TD, Patton HD (eds) Physiology and biophysics; circulation, respiration and fluid balance. W B Saunders, Philadelphia, pp 117–125 Feigl EO (1974) The arterial system. In: Ruch TD, Patton HD (eds) Physiology and biophysics; circulation, respiration and fluid balance. W B Saunders, Philadelphia, pp 117–125
23.
go back to reference Lovasik JV, Kothe AC, Kergoat H (1993) Comparison of noninvasive methods to derive the mean central retinal artery pressure in man. Optom Vis Sci 70:1005–1011CrossRefPubMed Lovasik JV, Kothe AC, Kergoat H (1993) Comparison of noninvasive methods to derive the mean central retinal artery pressure in man. Optom Vis Sci 70:1005–1011CrossRefPubMed
24.
25.
go back to reference Jonas JB (2004) Retinal arterial collapse pressure in eyes with retinal arterial occlusive diseases. Br J Ophthalmol 88:589CrossRefPubMed Jonas JB (2004) Retinal arterial collapse pressure in eyes with retinal arterial occlusive diseases. Br J Ophthalmol 88:589CrossRefPubMed
26.
go back to reference Morgan WH, Chauhan BC, Yu DY, Cringle SJ, Alder VA, House PH (2002) Optic disc movement with variations in intraocular and cerebrospinal fluid pressure. Inv Ophthalmol Vis Sci 43:3236–3242 Morgan WH, Chauhan BC, Yu DY, Cringle SJ, Alder VA, House PH (2002) Optic disc movement with variations in intraocular and cerebrospinal fluid pressure. Inv Ophthalmol Vis Sci 43:3236–3242
27.
go back to reference Inoue R, Hangai M, Kotera Y, Nakanishi H, Mori S, Morishita S, Yoshimura N (2009) Three-dimensional high-speed optical coherence tomography imaging of lamina cribrosa in glaucoma. Ophthalmology 116:214–222CrossRefPubMed Inoue R, Hangai M, Kotera Y, Nakanishi H, Mori S, Morishita S, Yoshimura N (2009) Three-dimensional high-speed optical coherence tomography imaging of lamina cribrosa in glaucoma. Ophthalmology 116:214–222CrossRefPubMed
28.
go back to reference Morgan WH, Hazelton ML, Balaratnasingam C, Chan H, House PH, Barry CJ, Cringle SJ, Yu DY (2009) The association between retinal vein ophthalmodynamometric force change and optic disc excavation. Br J Ophthalmol 93:594–596CrossRefPubMed Morgan WH, Hazelton ML, Balaratnasingam C, Chan H, House PH, Barry CJ, Cringle SJ, Yu DY (2009) The association between retinal vein ophthalmodynamometric force change and optic disc excavation. Br J Ophthalmol 93:594–596CrossRefPubMed
Metadata
Title
Optimizing the calibration and interpretation of dynamic ocular force measurements
Authors
William H. Morgan
Stephen J. Cringle
Min H. Kang
Surinder Pandav
Chandrakumar Balaratnasingam
Don Ezekial
Dao-Yi Yu
Publication date
01-03-2010
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 3/2010
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-009-1288-y

Other articles of this Issue 3/2010

Graefe's Archive for Clinical and Experimental Ophthalmology 3/2010 Go to the issue