04-05-2024 | Knee-TEP | Knee Arthroplasty

The peak force to push a trial tibial insert into position cannot be used to select the correct thickness in total knee arthroplasty

Authors: Gabriel Santana, S. M. Howell, M. L. Hull

Published in: Archives of Orthopaedic and Trauma Surgery

Login to get access

Abstract

An objective of a total knee arthroplasty (TKA) is to restore native (i.e. healthy) function, and a crucial step is determining the correct insert thickness for each patient. If the insert is too thick, then stiffness results, and if too thin, then instability results. Two methods to determine the insert thickness are by manually assessing the joint laxity and by using a trial insert with goniometric markings that measures the internal-external rotation of the trial with respect to a mark on the femoral component. The former is qualitative and depends on the surgeon’s experience and ‘feel’ and while the latter is quantitative, it can be used only with an insert with medial ball-in-socket conformity. An unexplored method is to measure the force required to push a trial insert into position. To determine whether this method has merit, the push force was measured in 30 patients undergoing unrestricted kinematically aligned TKA using an insert with ball-in-socket medial conformity, a flat lateral surface, and retention of the posterior cruciate ligament. During surgery, the surgeon determined three appropriate thicknesses to test from a selection ranging from 10 mm to 14 mm in 1 mm increments. The peak push forces going from an insert 1 mm thinner than the correct thickness as determined by an insert goniometer and from the correct thickness to 1 mm thicker were measured. Mean peak forces for the different insert thicknesses were 127 ± 104 N, 127 ± 95 N, and 144 ± 96 N for 1 mm thinner, correct, and 1 mm thicker, respectively, and did not differ (p = 0.3210). As a result, measurement of peak force during trial positioning of a tibial insert cannot be used to identify the correct thickness for all insert designs.
Appendix
Available only for authorised users
Literature
1.
go back to reference Aratake M, Mitsugi N, Taki N et al (2016) Impact of the polyethylene thickness on range of motion in total knee arthroplasty. Orthop Proc, 98-B(SUPP-1):22 Aratake M, Mitsugi N, Taki N et al (2016) Impact of the polyethylene thickness on range of motion in total knee arthroplasty. Orthop Proc, 98-B(SUPP-1):22
2.
go back to reference Buckwalter JA, Woo SLY (1996) Age-related changes in ligaments and joint capsules: implications for participation in sports. Sports Med Arthrosc Rev 4(3):250–262CrossRef Buckwalter JA, Woo SLY (1996) Age-related changes in ligaments and joint capsules: implications for participation in sports. Sports Med Arthrosc Rev 4(3):250–262CrossRef
3.
go back to reference Elmasry SS, Sculco PK, Kahlenberg CA et al (2022) Arthroplasty surgeons differ in their intraoperative soft tissue assessments: a study in human cadavers to quantify surgical decision-making in TKA. Clin Orthop Relat Res 480(8):1604–1615CrossRefPubMedPubMedCentral Elmasry SS, Sculco PK, Kahlenberg CA et al (2022) Arthroplasty surgeons differ in their intraoperative soft tissue assessments: a study in human cadavers to quantify surgical decision-making in TKA. Clin Orthop Relat Res 480(8):1604–1615CrossRefPubMedPubMedCentral
4.
go back to reference Elorza SP, O’Donnell E, Delman C, Howell SM, Hull ML (2023) Posterior cruciate ligament retention with medial ball-in-socket conformity promotes internal tibial rotation and knee flexion while providing high clinical outcome scores. Knee 43:153–162CrossRef Elorza SP, O’Donnell E, Delman C, Howell SM, Hull ML (2023) Posterior cruciate ligament retention with medial ball-in-socket conformity promotes internal tibial rotation and knee flexion while providing high clinical outcome scores. Knee 43:153–162CrossRef
5.
go back to reference Manning WA, Blain A, Longstaff L, Deehan DJ (2019) A load-measuring device can achieve fine-tuning of mediolateral load at knee arthroplasty but may lead to a more lax knee state. Knee Surg Sports Traumatol Arthrosc 27(7):2238–2250CrossRefPubMed Manning WA, Blain A, Longstaff L, Deehan DJ (2019) A load-measuring device can achieve fine-tuning of mediolateral load at knee arthroplasty but may lead to a more lax knee state. Knee Surg Sports Traumatol Arthrosc 27(7):2238–2250CrossRefPubMed
6.
go back to reference Meneghini RM, Ziemba-Davis MM, Lovro LR, Ireland PH, Damer BM (2016) Can intraoperative sensors determine the target ligament balance? Early outcomes in total knee arthroplasty. J Arthroplasty 31(10):2181–2187CrossRefPubMed Meneghini RM, Ziemba-Davis MM, Lovro LR, Ireland PH, Damer BM (2016) Can intraoperative sensors determine the target ligament balance? Early outcomes in total knee arthroplasty. J Arthroplasty 31(10):2181–2187CrossRefPubMed
7.
go back to reference Nedopil AJ, Howell SM, Hull ML (2022) Measurement of tibial axial orientation helps select the optimal insert thickness to personalize ligament and PCL tension in a medial ball-in-socket TKA. J Pers Med 12(9):1427CrossRefPubMedPubMedCentral Nedopil AJ, Howell SM, Hull ML (2022) Measurement of tibial axial orientation helps select the optimal insert thickness to personalize ligament and PCL tension in a medial ball-in-socket TKA. J Pers Med 12(9):1427CrossRefPubMedPubMedCentral
8.
go back to reference Nedopil AJ, Singh AK, Howell SM, Hull ML (2018) Does calipered kinematically aligned TKA restore native left to right symmetry of the lower limb and improve function? J Arthroplasty 33(2):398–406CrossRefPubMed Nedopil AJ, Singh AK, Howell SM, Hull ML (2018) Does calipered kinematically aligned TKA restore native left to right symmetry of the lower limb and improve function? J Arthroplasty 33(2):398–406CrossRefPubMed
9.
go back to reference Roth JD, Howell SM, Hull ML (2021) Tibial forces are more useful than varus-valgus laxities for identifying and correcting overstuffing in kinematically aligned total knee arthroplasty. J Orthop Res 39(6):1271–1280CrossRefPubMed Roth JD, Howell SM, Hull ML (2021) Tibial forces are more useful than varus-valgus laxities for identifying and correcting overstuffing in kinematically aligned total knee arthroplasty. J Orthop Res 39(6):1271–1280CrossRefPubMed
10.
go back to reference Shelton TJ, Howell SM, Hull ML (2019) A total knee arthroplasty is stiffer when the intraoperative tibial force is greater than the native knee. J Knee Surg 32(10):1008–1014CrossRefPubMed Shelton TJ, Howell SM, Hull ML (2019) A total knee arthroplasty is stiffer when the intraoperative tibial force is greater than the native knee. J Knee Surg 32(10):1008–1014CrossRefPubMed
11.
go back to reference Shelton TJ, Nedopil AJ, Howell SM, Hull ML (2017) Do varus or valgus outliers have higher forces in the medial or lateral compartments than those which are in-range after a kinematically aligned total knee arthroplasty? Bone Joint J 99(10):1319–1328CrossRefPubMed Shelton TJ, Nedopil AJ, Howell SM, Hull ML (2017) Do varus or valgus outliers have higher forces in the medial or lateral compartments than those which are in-range after a kinematically aligned total knee arthroplasty? Bone Joint J 99(10):1319–1328CrossRefPubMed
12.
go back to reference Wadugodapitiya S, Sakamoto M, Tanaka M, Sakagami Y, Morise Y, Kobayashi K (2022) Assessment of knee collateral ligament stiffness by strain ultrasound elastography. Biomed Mater Eng 33(5):337–349PubMedPubMedCentral Wadugodapitiya S, Sakamoto M, Tanaka M, Sakagami Y, Morise Y, Kobayashi K (2022) Assessment of knee collateral ligament stiffness by strain ultrasound elastography. Biomed Mater Eng 33(5):337–349PubMedPubMedCentral
13.
go back to reference Wilson WT, Deakin AH, Payne AP, Picard F, Wearing SC (2012) Comparative analysis of the structural properties of the collateral ligaments of the human knee. J Orthop Sports Phys Ther 42(4):345–351CrossRefPubMed Wilson WT, Deakin AH, Payne AP, Picard F, Wearing SC (2012) Comparative analysis of the structural properties of the collateral ligaments of the human knee. J Orthop Sports Phys Ther 42(4):345–351CrossRefPubMed
14.
go back to reference Santana G, Howell SM, Hull ML (2024) A new handheld tool to select the correct tibial insert thickness by measuring the force required to push a trial insert into position during total knee arthroplasty. J Med Dev 18(3):011002-1–011002-5 Santana G, Howell SM, Hull ML (2024) A new handheld tool to select the correct tibial insert thickness by measuring the force required to push a trial insert into position during total knee arthroplasty. J Med Dev 18(3):011002-1–011002-5
Metadata
Title
The peak force to push a trial tibial insert into position cannot be used to select the correct thickness in total knee arthroplasty
Authors
Gabriel Santana
S. M. Howell
M. L. Hull
Publication date
04-05-2024
Publisher
Springer Berlin Heidelberg
Keywords
Knee-TEP
Knee-TEP
Published in
Archives of Orthopaedic and Trauma Surgery
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-024-05313-6