Skip to main content
Top
Published in: Archives of Orthopaedic and Trauma Surgery 5/2007

01-07-2007 | Orthopaedic Surgery

Preoperative templating of femoral components on plain X-rays

Rotational evaluation with synthetic X-rays on ORTHODOC

Authors: Takehito Hananouchi, Nobuhiko Sugano, Nobuo Nakamura, Takashi Nishii, Hidenobu Miki, Mitsuyoshi Yamamura, Hideki Yoshikawa

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 5/2007

Login to get access

Abstract

Introduction

Preoperative femoral templating for total hip arthroplasty (THA) has been performed by antero-posterior radiographs of both hips which should be taken with the femur rotated internally to reduce the effect of femoral anteversion. However, there is no criterion to evaluate the optimal rotated radiograph. Here, we quantitatively investigated 50 femora whether the thickness of the lesser trochanter (TLT) was useful when evaluating the femoral rotation, and assessed the effect of prosthetic templating on the inappropriate rotated radiographs.

Materials and methods

We assessed 50 femora of 46 consecutive osteoarthritic patients using “synthetic X-ray” based on computer tomography (CT) images, which can be displayed as virtual plain radiographs with any magnification and any projected direction such as plain radiographs. We made four femoral radiograph groups of different rotation prepared (neutral, 15°, 30°, and 45° external rotation), and measured the TLT. We also templated femoral stem on the different rotated synthetic X-ray of each patient, and investigated the effect of inappropriate rotated radiograph by the measurement of the positon and the size of the stem.

Results

Seventy-four percent of the neutral rotational group had less than 5 mm of the TLT. While 30° and 45° external rotation group had few cases with less than 5 mm of the TLT. Compared to the neutral group, smaller stem size was selected in more than 80% cases of 30° and 45° external rotational group, and the stem position in these two groups was more than 5 mm proximally.

Conclusion

We suggest that when templating femoral stem, this criterion “the TLT is less than 5 mm” reduces some risks by inappropriate rotated radiograph.
Literature
1.
go back to reference Bell AL, Brand RA (1989) Roentgenographic changes in proximal femoral dimensions due to hip rotation. Clin Orthop Relat Res (240): 194–199 Bell AL, Brand RA (1989) Roentgenographic changes in proximal femoral dimensions due to hip rotation. Clin Orthop Relat Res (240): 194–199
2.
go back to reference Carter LW, Stovall DO, Young TR (1995) Determination of accuracy of preoperative templating of noncemented femoral prostheses. J Arthroplasty 10(4): 507–513PubMed Carter LW, Stovall DO, Young TR (1995) Determination of accuracy of preoperative templating of noncemented femoral prostheses. J Arthroplasty 10(4): 507–513PubMed
3.
go back to reference Conn KS, Clarke MT, Hallett JP (2002) A simple guide to determine the magnification of radiographs and to improve the accuracy of preoperative templating. J Bone Joint Surg Br 84(2): 269–272PubMedCrossRef Conn KS, Clarke MT, Hallett JP (2002) A simple guide to determine the magnification of radiographs and to improve the accuracy of preoperative templating. J Bone Joint Surg Br 84(2): 269–272PubMedCrossRef
4.
go back to reference D’Antonio JA (1994) Preoperative templating and choosing the implant for primary tha in the young patient. Instr Course Lect 43:339–346PubMed D’Antonio JA (1994) Preoperative templating and choosing the implant for primary tha in the young patient. Instr Course Lect 43:339–346PubMed
5.
go back to reference Dore DD, Rubash HE (1994) Primary total hip arthroplasty in the older patient: optimizing the results. Instr Course Lect 43:347–357PubMed Dore DD, Rubash HE (1994) Primary total hip arthroplasty in the older patient: optimizing the results. Instr Course Lect 43:347–357PubMed
6.
go back to reference Eckrich SG, Noble PC, Tullos HS (1994) Effect of rotation on the radiographic appearance of the femoral canal. J Arthroplasty 9(4): 419–426PubMedCrossRef Eckrich SG, Noble PC, Tullos HS (1994) Effect of rotation on the radiographic appearance of the femoral canal. J Arthroplasty 9(4): 419–426PubMedCrossRef
7.
go back to reference Eggli S, Pisan M, Muller ME (1998) The value of preoperative planning for total hip arthroplasty. J Bone Joint Surg Br 80(3): 382–390PubMedCrossRef Eggli S, Pisan M, Muller ME (1998) The value of preoperative planning for total hip arthroplasty. J Bone Joint Surg Br 80(3): 382–390PubMedCrossRef
8.
go back to reference Engh CA (1991) Recent advances in cementless total hip arthroplasty using the aml prosthesis. Tech Orthop 6(3): 59–72CrossRef Engh CA (1991) Recent advances in cementless total hip arthroplasty using the aml prosthesis. Tech Orthop 6(3): 59–72CrossRef
9.
go back to reference Gonzalez Della Valle A, Slullitel G, Piccaluga F, Salvati EA (2005) The precision and usefulness of preoperative planning for cemented and hybrid primary total hip arthroplasty. J Arthroplasty 20(1): 51–58PubMedCrossRef Gonzalez Della Valle A, Slullitel G, Piccaluga F, Salvati EA (2005) The precision and usefulness of preoperative planning for cemented and hybrid primary total hip arthroplasty. J Arthroplasty 20(1): 51–58PubMedCrossRef
10.
go back to reference Goodman SB, Huene DS, Imrie S (1992) Preoperative templating for the equalization of leg lengths in total hip arthroplasty. Contemp Orthop 24(6): 703–710PubMed Goodman SB, Huene DS, Imrie S (1992) Preoperative templating for the equalization of leg lengths in total hip arthroplasty. Contemp Orthop 24(6): 703–710PubMed
11.
go back to reference Jolles BM, Leyvraz PF (2005) Computer-assisted two-dimensional preoperative planning for total hip arthroplasty: clinical evaluation. Orthopedics 28(11): 1356–1359PubMed Jolles BM, Leyvraz PF (2005) Computer-assisted two-dimensional preoperative planning for total hip arthroplasty: clinical evaluation. Orthopedics 28(11): 1356–1359PubMed
12.
go back to reference Knight JL, Atwater RD (1992) Preoperative planning for total hip arthroplasty. Quant Utility Precision J Arthroplasty 7 suppl: 403–409 Knight JL, Atwater RD (1992) Preoperative planning for total hip arthroplasty. Quant Utility Precision J Arthroplasty 7 suppl: 403–409
13.
go back to reference Laine HJ, Lehto MU, Moilanen T (2000) Diversity of proximal femoral medullary canal. J Arthroplasty 15(1): 86–92PubMedCrossRef Laine HJ, Lehto MU, Moilanen T (2000) Diversity of proximal femoral medullary canal. J Arthroplasty 15(1): 86–92PubMedCrossRef
14.
go back to reference Linclau L, Dokter G, Peene P (1993) Radiological aspects in preoperative planning and postoperative assessment of cementless total hip arthroplasty. Acta Orthop Belg 59(2): 163–167PubMed Linclau L, Dokter G, Peene P (1993) Radiological aspects in preoperative planning and postoperative assessment of cementless total hip arthroplasty. Acta Orthop Belg 59(2): 163–167PubMed
15.
go back to reference McAlindon RJ, Kruse RW (1997) Measurement of rib vertebral angle difference. Intraobs Error Interobs Variation Spine 22(2): 198–199 McAlindon RJ, Kruse RW (1997) Measurement of rib vertebral angle difference. Intraobs Error Interobs Variation Spine 22(2): 198–199
16.
go back to reference Nishihara S, Sugano N, Nishii T, Tanaka H, Nakamura N, Yoshikawa H, Ochi T (2004) Clinical accuracy evaluation of femoral canal preparation using the robodoc system. J Orthop Sci 9(5): 452–461PubMedCrossRef Nishihara S, Sugano N, Nishii T, Tanaka H, Nakamura N, Yoshikawa H, Ochi T (2004) Clinical accuracy evaluation of femoral canal preparation using the robodoc system. J Orthop Sci 9(5): 452–461PubMedCrossRef
17.
go back to reference Noble PC, Kamaric E, Sugano N, Matsubara M, Harada Y, Ohzono K, Paravic V (2003) Three-dimensional shape of the dysplastic femur: implications for THR. Clin Orthop Relat Res (417): 27–40 Noble PC, Kamaric E, Sugano N, Matsubara M, Harada Y, Ohzono K, Paravic V (2003) Three-dimensional shape of the dysplastic femur: implications for THR. Clin Orthop Relat Res (417): 27–40
18.
go back to reference Seel EH, Verrill CL, Mehta RL, Davies EM (2005) Measurement of fracture kyphosis with the oxford cobbometer: intra-and interobserver reliabilities and comparison with other techniques. Spine 30(8): 964–968PubMedCrossRef Seel EH, Verrill CL, Mehta RL, Davies EM (2005) Measurement of fracture kyphosis with the oxford cobbometer: intra-and interobserver reliabilities and comparison with other techniques. Spine 30(8): 964–968PubMedCrossRef
19.
go back to reference Sugano N, Noble PC, Kamaric E, Salama JK, Ochi T, Tullos HS (1998) The morphology of the femur in developmental dysplasia of the hip. J Bone Joint Surg Br 80(4): 711–719PubMedCrossRef Sugano N, Noble PC, Kamaric E, Salama JK, Ochi T, Tullos HS (1998) The morphology of the femur in developmental dysplasia of the hip. J Bone Joint Surg Br 80(4): 711–719PubMedCrossRef
20.
go back to reference Sugano N, Ohzono K, Nishii T, Haraguchi K, Sakai T, Ochi T (1998) Computed-tomography-based computer preoperative planning for total hip arthroplasty. Comput Aided Surg 3(6): 320–324PubMedCrossRef Sugano N, Ohzono K, Nishii T, Haraguchi K, Sakai T, Ochi T (1998) Computed-tomography-based computer preoperative planning for total hip arthroplasty. Comput Aided Surg 3(6): 320–324PubMedCrossRef
21.
go back to reference Suh KT, Cheon SJ, Kim DW (2004) Comparison of preoperative templating with postoperative assessment in cementless total hip arthroplasty. Acta Orthop Scand 75(1): 40–44PubMedCrossRef Suh KT, Cheon SJ, Kim DW (2004) Comparison of preoperative templating with postoperative assessment in cementless total hip arthroplasty. Acta Orthop Scand 75(1): 40–44PubMedCrossRef
22.
go back to reference Viceconti M, Lattanzi R, Antonietti B, Paderni S, Olmi R, Sudanese A, Toni A (2003) Ct-based surgical planning software improves the accuracy of total hip replacement preoperative planning. Med Eng Phys 25(5): 371–377PubMedCrossRef Viceconti M, Lattanzi R, Antonietti B, Paderni S, Olmi R, Sudanese A, Toni A (2003) Ct-based surgical planning software improves the accuracy of total hip replacement preoperative planning. Med Eng Phys 25(5): 371–377PubMedCrossRef
Metadata
Title
Preoperative templating of femoral components on plain X-rays
Rotational evaluation with synthetic X-rays on ORTHODOC
Authors
Takehito Hananouchi
Nobuhiko Sugano
Nobuo Nakamura
Takashi Nishii
Hidenobu Miki
Mitsuyoshi Yamamura
Hideki Yoshikawa
Publication date
01-07-2007
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 5/2007
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-007-0349-0

Other articles of this Issue 5/2007

Archives of Orthopaedic and Trauma Surgery 5/2007 Go to the issue