Published in: Archives of Orthopaedic and Trauma Surgery 3/2006

01-04-2006 | Case Report

Serious metallosis of a metal head due to fragmented ceramic screws in a cemented THA

Authors: Hidenobu Miki, Nobuhiko Sugano, Mitsuyoshi Yamamura, Nobuo Nakamura, Takashi Nishii, Hideki Yoshikawa

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 3/2006

Login to get access

Abstract

We reported a rare revision case necessitated by massive metallosis of a metal head due to a fragmented ceramic screw which had been extraarticularily used in the primary THA here and reviewed previous literature about metallosis of metal heads associated with ceramic materials. The ceramic screw had been used for the fixation of a bone graft and reattachment of the greater trochanter in the primary THA. Scanning electron microscope analysis revealed large alumina ceramic particles embedded on the bearing surface of the polyethylene. It has previously been pointed out that metallosis of an inner head occurred in bipolar case due to a fracture of the ceramic screws in contact with a proximally migrated outer head. However, this case showed that if a ceramic screw fragmented even though it was used outside of the hip joint, the fragments could migrate into bearing surface and cause serious metallosis of the metal head. Thus, we recommend that surgeons should pay special attention to the radiographic signs of ceramic failure or metallosis such as a change of the metal head contour, if an aseptic loosening or severe osteolysis is observed in a THA in which ceramic materials were used near the hip joint. And, when a surgical revision is planned, the surgeon should be prepared to exchange with a new ceramic head.
Literature
1.
go back to reference Allain J, Roudot-Thoraval F, Delecrin J, Anract P, Migaud H, Goutallier D (2003) Revision total hip arthroplasty performed after fracture of a ceramic femoral head. A multicenter survivorship study. J Bone Joint Surg Am 85:825PubMed Allain J, Roudot-Thoraval F, Delecrin J, Anract P, Migaud H, Goutallier D (2003) Revision total hip arthroplasty performed after fracture of a ceramic femoral head. A multicenter survivorship study. J Bone Joint Surg Am 85:825PubMed
2.
go back to reference Besong AA, Hailey JL, Ingham E, Stone M, Wroblewski BM, Fisher J (1997) A study of the combined effects of shelf ageing following irradiation in air and counterface roughness on the wear of UHMWPE. Biomed Mater Eng 7:59PubMed Besong AA, Hailey JL, Ingham E, Stone M, Wroblewski BM, Fisher J (1997) A study of the combined effects of shelf ageing following irradiation in air and counterface roughness on the wear of UHMWPE. Biomed Mater Eng 7:59PubMed
3.
go back to reference Bohler M, Mochida Y, Bauer TW, Plenk H Jr, Salzer M (2000) Wear debris from two different alumina-on-alumina total hip arthroplasties. J Bone Joint Surg Br 82:901CrossRefPubMed Bohler M, Mochida Y, Bauer TW, Plenk H Jr, Salzer M (2000) Wear debris from two different alumina-on-alumina total hip arthroplasties. J Bone Joint Surg Br 82:901CrossRefPubMed
4.
go back to reference Davidson JA, Poggie RA, Mishra AK (1994) Abrasive wear of ceramic, metal, and UHMWPE bearing surfaces from third-body bone, PMMA bone cement, and titanium debris. Biomed Mater Eng 4:213PubMed Davidson JA, Poggie RA, Mishra AK (1994) Abrasive wear of ceramic, metal, and UHMWPE bearing surfaces from third-body bone, PMMA bone cement, and titanium debris. Biomed Mater Eng 4:213PubMed
5.
go back to reference Iizuka T, Fujimoto H, Ono T (1987) A new material (single crystal sapphire screw) for internal fixation of the mandibular ramus. J Craniomaxillofac Surg 15:24PubMed Iizuka T, Fujimoto H, Ono T (1987) A new material (single crystal sapphire screw) for internal fixation of the mandibular ramus. J Craniomaxillofac Surg 15:24PubMed
6.
go back to reference Ishizuki M, Furuya K (1991) Clinical application of sapphire pins as an internal fixation device for the upper extremity. J Hand Surg [Am] 16:922 Ishizuki M, Furuya K (1991) Clinical application of sapphire pins as an internal fixation device for the upper extremity. J Hand Surg [Am] 16:922
7.
go back to reference Kempf I, Semlitsch M (1994) Massive wear of a steel ball head by ceramic fragments in the polyethylene acetabular cup after revision of a total hip prosthesis with fractured ceramic ball. Med Prog Technol 20:185PubMed Kempf I, Semlitsch M (1994) Massive wear of a steel ball head by ceramic fragments in the polyethylene acetabular cup after revision of a total hip prosthesis with fractured ceramic ball. Med Prog Technol 20:185PubMed
8.
go back to reference Matsuda Y, Yamamuro T (1990) Metallosis due to abnormal abrasion of the femoral head in bipolar hip prosthesis. Implant retrieval and analysis in six cases. Arch Orthop Trauma Surg 109:284CrossRefPubMed Matsuda Y, Yamamuro T (1990) Metallosis due to abnormal abrasion of the femoral head in bipolar hip prosthesis. Implant retrieval and analysis in six cases. Arch Orthop Trauma Surg 109:284CrossRefPubMed
9.
go back to reference Matziolis G, Perka C, Disch A (2003) Massive metallosis after revision of a fractured ceramic head onto a metal head. Arch Orthop Trauma Surg 123:48PubMed Matziolis G, Perka C, Disch A (2003) Massive metallosis after revision of a fractured ceramic head onto a metal head. Arch Orthop Trauma Surg 123:48PubMed
10.
go back to reference McNie CM, Barton DC, Fisher J, Stone MH (2000) Modeling of damage to articulating surfaces by third body particles in total joint replacements. J Mater Sci: Mater Med 11:569CrossRef McNie CM, Barton DC, Fisher J, Stone MH (2000) Modeling of damage to articulating surfaces by third body particles in total joint replacements. J Mater Sci: Mater Med 11:569CrossRef
11.
go back to reference Ooi Y, Suka T, Satoh Y, Shibuya K, Ohki I, Masubuchi M, Mikanagi K (1982) Clinical experience of Al2O3 ceramics as a surgical implant. Int Orthop 6:93CrossRefPubMed Ooi Y, Suka T, Satoh Y, Shibuya K, Ohki I, Masubuchi M, Mikanagi K (1982) Clinical experience of Al2O3 ceramics as a surgical implant. Int Orthop 6:93CrossRefPubMed
12.
go back to reference Shikita T, Inoue A, Hamaguchi T, Oonishi H (1980) Experience of peripheral osteosynthesis using single crystal alumina ceramics screws. Z Orthop Ihre Grenzgeb 118:975PubMedCrossRef Shikita T, Inoue A, Hamaguchi T, Oonishi H (1980) Experience of peripheral osteosynthesis using single crystal alumina ceramics screws. Z Orthop Ihre Grenzgeb 118:975PubMedCrossRef
13.
go back to reference Su EP, Callander PW, Salvati EA (2003) The bubble sign: a new radiographic sign in total hip arthroplasty. J Arthroplasty 18:110CrossRefPubMed Su EP, Callander PW, Salvati EA (2003) The bubble sign: a new radiographic sign in total hip arthroplasty. J Arthroplasty 18:110CrossRefPubMed
14.
go back to reference Tipper JL, Hatton A, Nevelos JE, Ingham E, Doyle C, Streicher R, Nevelos AB, Fisher J (2002) Alumina–alumina artificial hip joints. Part II; characterisation of the wear debris from in vitro hip joint simulations. Biomaterials 23:3441CrossRefPubMed Tipper JL, Hatton A, Nevelos JE, Ingham E, Doyle C, Streicher R, Nevelos AB, Fisher J (2002) Alumina–alumina artificial hip joints. Part II; characterisation of the wear debris from in vitro hip joint simulations. Biomaterials 23:3441CrossRefPubMed
15.
go back to reference Watanabe M, Okumura H, Kihara Y, Shibata T (1993) Metallosis caused by alumina ceramic screw fixation of grafted bone in a bipolar hip endoprosthesis. Arch Orthop Trauma Surg 112:296CrossRefPubMed Watanabe M, Okumura H, Kihara Y, Shibata T (1993) Metallosis caused by alumina ceramic screw fixation of grafted bone in a bipolar hip endoprosthesis. Arch Orthop Trauma Surg 112:296CrossRefPubMed
Metadata
Title
Serious metallosis of a metal head due to fragmented ceramic screws in a cemented THA
Authors
Hidenobu Miki
Nobuhiko Sugano
Mitsuyoshi Yamamura
Nobuo Nakamura
Takashi Nishii
Hideki Yoshikawa
Publication date
01-04-2006
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 3/2006
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-006-0112-y

Other articles of this Issue 3/2006

Archives of Orthopaedic and Trauma Surgery 3/2006 Go to the issue

Original Article

The male neck of femur