Published in: Child's Nervous System 1/2014

01-01-2014 | Original Paper

Computational fluid dynamics of ventricular catheters used for the treatment of hydrocephalus: a 3D analysis

Authors: Marcelo Galarza, Ángel Giménez, José Valero, Olga Porcar Pellicer, José María Amigó

Published in: Child's Nervous System | Issue 1/2014

Login to get access

Abstract

Introduction

The most common treatment for hydrocephalus remains the ventriculoperitoneal shunt. Yet, the most frequent complication is ventricular catheter obstruction, which may account for 50–80 % of newly inserted shunts. Although many factors contribute to this, the main one is related to flow characteristics of the catheter within the hydrocephalic brain. A landmark study by Lin et al. addressed the problem of fluid characteristics in ventricular catheters using a two-dimensional simulation program of computational fluid dynamics (CFD).

Methods

The authors have studied five current commercially available ventricular catheter designs using CFD in three-dimensional automated designs. The general procedure for the development of a CFD model involves incorporating the physical dimensions of the system to be studied into a virtual wire-frame model. The shape and features of the actual physical model are transformed into coordinates for the virtual space of the computer and a CFD computational grid (mesh) is generated. The fluid properties and motion are calculated at each of these grid points. After grid generation, flow field boundary conditions are applied, and the fluid’s thermodynamic and transport properties are included. At the end, a system of strongly coupled, nonlinear, partial differential conservation equations governing the motion of the flow field are numerically solved. This numerical solution describes the fluid motion and properties.

Results

The authors calculated that most of the total fluid mass flows into the catheter’s most proximal holes. Fifty to 75 % flows into the two most proximal sets of inlets of current commercially available 12–32-hole catheters. Some flow uniformity was disclosed in Rivulet-type catheter.

Conclusions

Most commercially available ventricular catheters have an abnormally increase flow distribution pattern. New catheter designs with variable hole diameters along the catheter tip will allow the fluid to enter the catheter more uniformly along its length, thereby reducing the probability of its becoming occluded.
Literature
1.
go back to reference Payr E (1908) Drainage der Hirnventrikel Mittelst frei Transplantirter Blutgefasse; Bemerkungen ueber Hydrocephalus. Arch Klin Chir 87:801–885 Payr E (1908) Drainage der Hirnventrikel Mittelst frei Transplantirter Blutgefasse; Bemerkungen ueber Hydrocephalus. Arch Klin Chir 87:801–885
2.
go back to reference Bergsneider M, Egnor MR, Johnston M, Kranz D, Madsen JR, McAllister JP 2nd, Stewart C, Walker ML, Williams MA (2006) What we don’t (but should) know about hydrocephalus. J Neurosurg 104:157–159PubMedCrossRef Bergsneider M, Egnor MR, Johnston M, Kranz D, Madsen JR, McAllister JP 2nd, Stewart C, Walker ML, Williams MA (2006) What we don’t (but should) know about hydrocephalus. J Neurosurg 104:157–159PubMedCrossRef
3.
go back to reference Drake JM, Sainte-Rose C (1995) The shunt book. Blackwell, Cambridge Drake JM, Sainte-Rose C (1995) The shunt book. Blackwell, Cambridge
4.
go back to reference Drake JM, Kestle JR, Tuli S (2000) CSF shunts 50 years on—past, present and future. Childs Nerv Syst 16:800–804PubMedCrossRef Drake JM, Kestle JR, Tuli S (2000) CSF shunts 50 years on—past, present and future. Childs Nerv Syst 16:800–804PubMedCrossRef
5.
go back to reference Sainte-Rose C, Piatt JH, Renier D, Pierre-Kahn A, Hirsch JF, Hoffman HJ, Humphreys RP, Hendrick EB (1991) Mechanical complications in shunts. Pediatr Neurosurg 17:2–9PubMedCrossRef Sainte-Rose C, Piatt JH, Renier D, Pierre-Kahn A, Hirsch JF, Hoffman HJ, Humphreys RP, Hendrick EB (1991) Mechanical complications in shunts. Pediatr Neurosurg 17:2–9PubMedCrossRef
6.
go back to reference Tuli S, Drake J, Lawless J, Wigg M, Math M, Lamberti-Pasculli M (2000) Risk factors for repeated cerebrospinal shunt failures in pediatric patients with hydrocephalus. J Neurosurg 92:31–38PubMedCrossRef Tuli S, Drake J, Lawless J, Wigg M, Math M, Lamberti-Pasculli M (2000) Risk factors for repeated cerebrospinal shunt failures in pediatric patients with hydrocephalus. J Neurosurg 92:31–38PubMedCrossRef
7.
go back to reference Drake J, Kestle JR, Milner R, Cinalli G, Boop F, Piatt J Jr, Haines S, Schiff SJ, Cochrane DD, Steinbok P, MacNeil N (1998) Randomized trial of cerebrospinal fluid shunt valve design in pediatric hydrocephalus. Neurosurgery 43:294–305PubMedCrossRef Drake J, Kestle JR, Milner R, Cinalli G, Boop F, Piatt J Jr, Haines S, Schiff SJ, Cochrane DD, Steinbok P, MacNeil N (1998) Randomized trial of cerebrospinal fluid shunt valve design in pediatric hydrocephalus. Neurosurgery 43:294–305PubMedCrossRef
8.
go back to reference Harris CA, McAllister JP 2nd (2012) What we should know about the cellular and tissue response causing catheter obstruction in the treatment of hydrocephalus. Neurosurgery 70:1589–1601PubMedCrossRef Harris CA, McAllister JP 2nd (2012) What we should know about the cellular and tissue response causing catheter obstruction in the treatment of hydrocephalus. Neurosurgery 70:1589–1601PubMedCrossRef
9.
go back to reference Ginsberg HJ, Sum A, Drake JM (2000) Ventriculoperitoneal shunt flow dependency on the number of patent holes in a ventricular catheter. Pediatr Neurosurg 33:7–11PubMedCrossRef Ginsberg HJ, Sum A, Drake JM (2000) Ventriculoperitoneal shunt flow dependency on the number of patent holes in a ventricular catheter. Pediatr Neurosurg 33:7–11PubMedCrossRef
10.
go back to reference Lin J, Morris M, Olivero W, Boop F, Sanford RA (2003) Computational and experimental study of proximal flow in ventricular catheters. Technical note. J Neurosurg 99:426–431PubMedCrossRef Lin J, Morris M, Olivero W, Boop F, Sanford RA (2003) Computational and experimental study of proximal flow in ventricular catheters. Technical note. J Neurosurg 99:426–431PubMedCrossRef
11.
go back to reference Clark TW, Van Canneyt K, Verdonck P (2012) Computational flow dynamics and preclinical assessment of a novel hemodialysis catheter. Semin Dial 25:574–581PubMedCrossRef Clark TW, Van Canneyt K, Verdonck P (2012) Computational flow dynamics and preclinical assessment of a novel hemodialysis catheter. Semin Dial 25:574–581PubMedCrossRef
12.
go back to reference Frawley P, Geron M (2009) Combination of CFD and DOE to analyze and improve the mass flow rate in urinary catheters. J Biomech Eng 131(8):084501PubMedCrossRef Frawley P, Geron M (2009) Combination of CFD and DOE to analyze and improve the mass flow rate in urinary catheters. J Biomech Eng 131(8):084501PubMedCrossRef
13.
go back to reference Harris CA, McAllister JP 2nd (2011) Does drainage hole size influence adhesion on ventricular catheters? Childs Nerv Syst 27:1221–1232PubMedCrossRef Harris CA, McAllister JP 2nd (2011) Does drainage hole size influence adhesion on ventricular catheters? Childs Nerv Syst 27:1221–1232PubMedCrossRef
14.
go back to reference Hakim S (1969) Observations on the physiopathology of the CSF pulse and prevention of ventricular catheter obstruction in valve shunts. Dev Med Child Neurol Suppl 20:42–48PubMed Hakim S (1969) Observations on the physiopathology of the CSF pulse and prevention of ventricular catheter obstruction in valve shunts. Dev Med Child Neurol Suppl 20:42–48PubMed
15.
go back to reference Sood S, Kim S, Ham SD, Canady AI, Greninger N (1993) Useful components of the shunt tap test for evaluation of shunt malfunction. Childs Nerv Syst 9:157–161PubMed Sood S, Kim S, Ham SD, Canady AI, Greninger N (1993) Useful components of the shunt tap test for evaluation of shunt malfunction. Childs Nerv Syst 9:157–161PubMed
16.
go back to reference Schley D, Billingham J, Marchbanks RJ (2004) A model of in-vivo hydrocephalus shunt dynamics for blockage and performance diagnostics. Math Med Biol 21:347–368PubMedCrossRef Schley D, Billingham J, Marchbanks RJ (2004) A model of in-vivo hydrocephalus shunt dynamics for blockage and performance diagnostics. Math Med Biol 21:347–368PubMedCrossRef
17.
go back to reference Penn RD, Basati S, Sweetman B, Guo X, Linninger A (2011) Ventricle wall movements and cerebrospinal fluid flow in hydrocephalus. J Neurosurg 115:159–164PubMedCrossRef Penn RD, Basati S, Sweetman B, Guo X, Linninger A (2011) Ventricle wall movements and cerebrospinal fluid flow in hydrocephalus. J Neurosurg 115:159–164PubMedCrossRef
18.
go back to reference Sood S, Lokuketagoda J, Ham SD (2005) Periventricular rigidity in long-term shunt-treated hydrocephalus. J Neurosurg 102:146–149PubMedCrossRef Sood S, Lokuketagoda J, Ham SD (2005) Periventricular rigidity in long-term shunt-treated hydrocephalus. J Neurosurg 102:146–149PubMedCrossRef
19.
go back to reference Sood S, Kumar CR, Jamous M, Schuhmann MU, Ham SD, Canady AI (2004) Pathophysiological changes in cerebrovascular distensibility in patients undergoing chronic shunt therapy. J Neurosurg Pediatr 100:447–453CrossRef Sood S, Kumar CR, Jamous M, Schuhmann MU, Ham SD, Canady AI (2004) Pathophysiological changes in cerebrovascular distensibility in patients undergoing chronic shunt therapy. J Neurosurg Pediatr 100:447–453CrossRef
20.
go back to reference Penn RD, Lee MC, Linninger AA, Miesel K, Lu SN, Stylos L (2005) Pressure gradients in the brain in an experimental model of hydrocephalus. J Neurosurg 102:1069–1075PubMedCrossRef Penn RD, Lee MC, Linninger AA, Miesel K, Lu SN, Stylos L (2005) Pressure gradients in the brain in an experimental model of hydrocephalus. J Neurosurg 102:1069–1075PubMedCrossRef
21.
go back to reference Stein SC, Guo W (2008) Have we made progress in preventing shunt failure? A critical analysis. J Neurosurg Pediatr 1:40–47PubMedCrossRef Stein SC, Guo W (2008) Have we made progress in preventing shunt failure? A critical analysis. J Neurosurg Pediatr 1:40–47PubMedCrossRef
22.
go back to reference Portnoy HD (1971) New ventricular catheter for hydrocephalic shunt. Technical note. J Neurosurg 34:702–703PubMedCrossRef Portnoy HD (1971) New ventricular catheter for hydrocephalic shunt. Technical note. J Neurosurg 34:702–703PubMedCrossRef
23.
go back to reference Haase J, Weeth R (1976) Multiflanged ventricular Portnoy catheter for hydrocephalus shunts. Acta Neurochir (Wien) 33:213–218CrossRef Haase J, Weeth R (1976) Multiflanged ventricular Portnoy catheter for hydrocephalus shunts. Acta Neurochir (Wien) 33:213–218CrossRef
24.
go back to reference Harris CA, Resau JH, Hudson EA, West RA, Moon C, Black AD, McAllister JP 2nd (2011) Reduction of protein adsorption and macrophage and astrocyte adhesion on ventricular catheters by polyethylene glycol and N-acetyl-L-cysteine. J Biomed Mater Res A 98:425–433PubMedCrossRef Harris CA, Resau JH, Hudson EA, West RA, Moon C, Black AD, McAllister JP 2nd (2011) Reduction of protein adsorption and macrophage and astrocyte adhesion on ventricular catheters by polyethylene glycol and N-acetyl-L-cysteine. J Biomed Mater Res A 98:425–433PubMedCrossRef
25.
go back to reference Harris CA, Resau JH, Hudson EA, West RA, Moon C, McAllister JP 2nd (2010) Mechanical contributions to astrocyte adhesion using a novel in vitro model of catheter obstruction. Exp Neurol 222:204–210PubMedCrossRef Harris CA, Resau JH, Hudson EA, West RA, Moon C, McAllister JP 2nd (2010) Mechanical contributions to astrocyte adhesion using a novel in vitro model of catheter obstruction. Exp Neurol 222:204–210PubMedCrossRef
26.
go back to reference Prasad A, Madan VS, Buxi TB, Renjen PN, Vohra R (1991) The role of the perforated segment of the ventricular catheter in cerebrospinal fluid leakage into the brain. Br J Neurosurg 5:299–302PubMedCrossRef Prasad A, Madan VS, Buxi TB, Renjen PN, Vohra R (1991) The role of the perforated segment of the ventricular catheter in cerebrospinal fluid leakage into the brain. Br J Neurosurg 5:299–302PubMedCrossRef
27.
go back to reference Thomale UW, Hosch H, Koch A, Schulz M, Stoltenburg G, Haberl EJ, Sprung C (2010) Perforation holes in ventricular catheters—is less more? Childs Nerv Syst 26:781–789PubMedCrossRef Thomale UW, Hosch H, Koch A, Schulz M, Stoltenburg G, Haberl EJ, Sprung C (2010) Perforation holes in ventricular catheters—is less more? Childs Nerv Syst 26:781–789PubMedCrossRef
28.
go back to reference Cheatle JT, Bowder AN, Agrawal SK, Sather MD, Hellbusch LC (2012) Flow characteristics of cerebrospinal fluid shunt tubing. J Neurosurg Pediatr 9:191–197PubMedCrossRef Cheatle JT, Bowder AN, Agrawal SK, Sather MD, Hellbusch LC (2012) Flow characteristics of cerebrospinal fluid shunt tubing. J Neurosurg Pediatr 9:191–197PubMedCrossRef
29.
go back to reference Mukerji N, Cahill J, Rodrigues D, Prakash S, Strachan R (2009) Flow dynamics in lumboperitoneal shunts and their implications in vivo. J Neurosurg 111:632–637PubMedCrossRef Mukerji N, Cahill J, Rodrigues D, Prakash S, Strachan R (2009) Flow dynamics in lumboperitoneal shunts and their implications in vivo. J Neurosurg 111:632–637PubMedCrossRef
Metadata
Title
Computational fluid dynamics of ventricular catheters used for the treatment of hydrocephalus: a 3D analysis
Authors
Marcelo Galarza
Ángel Giménez
José Valero
Olga Porcar Pellicer
José María Amigó
Publication date
01-01-2014
Publisher
Springer Berlin Heidelberg
Published in
Child's Nervous System / Issue 1/2014
Print ISSN: 0256-7040
Electronic ISSN: 1433-0350
DOI
https://doi.org/10.1007/s00381-013-2226-1

Other articles of this Issue 1/2014

Child's Nervous System 1/2014 Go to the issue