Published in: Child's Nervous System 1/2013

01-01-2013 | Technical Note

Smartphone-assisted guide for the placement of ventricular catheters

Published in: Child's Nervous System | Issue 1/2013

Login to get access

Abstract

Objective

Freehand placement of ventricular catheters (VC) is reported to be inaccurate in 10–40 %. Endoscopy, ultrasound, or neuronavigation are used in selected cases with significant technical and time-consuming efforts. We suggest a smartphone-assisted guiding tool for the placement of VC.

Methods

Measurements of relevant parameters in 3D-MRI datasets in a patient cohort with narrow ventricles for a frontal precoronal VC placement were performed. In this context, a guiding tool was developed to apply the respective measures for VC placement. The guiding tool was tested in a phantom followed by CT imaging to quantify placement precision. A smartphone application was designed to assist the relevant measurements. The guide was applied in 35 patients for VC placement.

Results

MRI measurements revealed the rectangular approach in the sagittal plane and the individual angle towards the tangent in the coronal section as relevant parameter for a frontal approach. The latter angle ranged from medial (91.96° ± 2.75°) to lateral margins (99.56° ± 4.14°) of the ventricle, which was similar in laterally shifted (±5 mm) entry points. The subsequently developed guiding tool revealed precision measurements in an agarose model with 1.1° ± 0.7° angle deviation. Using the smartphone-assisted guide in patients with narrow ventricles (frontal occipital horn ratio, 0.38 ± 0.05), a primary puncture of the ventricles was possible in all cases. No VC failure was observed during follow-up (9.1 ± 5.3 months).

Conclusions

VC placement in narrow ventricles requires accurate placement with simple means in an every-case routine. The suggested smartphone-assisted guide meets these criteria. Further data are planned to be collected in a prospective randomized study.
Literature
1.
go back to reference Abdoh MG, Bekaert O, Hodel J, Diarra SM, Le Guerinel C, Nseir R, Bastuji-Garin S, Decq P (2012) Accuracy of external ventricular drainage catheter placement. Acta Neurochir (Wien) 154(1):153–159CrossRef Abdoh MG, Bekaert O, Hodel J, Diarra SM, Le Guerinel C, Nseir R, Bastuji-Garin S, Decq P (2012) Accuracy of external ventricular drainage catheter placement. Acta Neurochir (Wien) 154(1):153–159CrossRef
2.
go back to reference Abu-Serieh B, Ghassempour K, Duprez T, Raftopoulos C (2007) Stereotactic ventriculoperitoneal shunting for refractory idiopathic intracranial hypertension. Neurosurgery 60(6):1039–1043PubMedCrossRef Abu-Serieh B, Ghassempour K, Duprez T, Raftopoulos C (2007) Stereotactic ventriculoperitoneal shunting for refractory idiopathic intracranial hypertension. Neurosurgery 60(6):1039–1043PubMedCrossRef
3.
go back to reference Azeem SS, Origitano TC (2007) Ventricular catheter placement with a frameless neuronavigational system: a 1-year experience. Neurosurgery 60(4 Suppl 2):243–247PubMed Azeem SS, Origitano TC (2007) Ventricular catheter placement with a frameless neuronavigational system: a 1-year experience. Neurosurgery 60(4 Suppl 2):243–247PubMed
4.
go back to reference Di Rocco C, Marchese E, Velardi F (1994) A survey of the first complication of newly implanted CSF shunt devices for the treatment of nontumoral hydrocephalus. Cooperative survey of the 1991–1992 Education Committee of the ISPN. Childs Nerv Syst 10(5):321–327PubMedCrossRef Di Rocco C, Marchese E, Velardi F (1994) A survey of the first complication of newly implanted CSF shunt devices for the treatment of nontumoral hydrocephalus. Cooperative survey of the 1991–1992 Education Committee of the ISPN. Childs Nerv Syst 10(5):321–327PubMedCrossRef
5.
go back to reference Ghajar JB (1985) A guide for ventricular catheter placement. Technical note. J Neurosurg 63(6):985–986PubMedCrossRef Ghajar JB (1985) A guide for ventricular catheter placement. Technical note. J Neurosurg 63(6):985–986PubMedCrossRef
6.
go back to reference Giese H, Hoffmann KT, Winkelmann A, Stockhammer F, Jallo GI, Thomale UW (2010) Precision of navigated stereotactic probe implantation into the brainstem. J Neurosurg Pediatr 5(4):350–359PubMedCrossRef Giese H, Hoffmann KT, Winkelmann A, Stockhammer F, Jallo GI, Thomale UW (2010) Precision of navigated stereotactic probe implantation into the brainstem. J Neurosurg Pediatr 5(4):350–359PubMedCrossRef
7.
go back to reference Harris CA, McAllister JP 2nd (2012) What we should know about the cellular and tissue response causing catheter obstruction in the treatment of hydrocephalus. Neurosurgery Jun 70(6):1589–1601 Harris CA, McAllister JP 2nd (2012) What we should know about the cellular and tissue response causing catheter obstruction in the treatment of hydrocephalus. Neurosurgery Jun 70(6):1589–1601
8.
go back to reference Hayhurst C, Beems T, Jenkinson MD, Byrne P, Clark S, Kandasamy J, Goodden J, Nandoe Tewarie RD, Mallucci CL (2010) Effect of electromagnetic-navigated shunt placement on failure rates: a prospective multicenter study. J Neurosurg Dec 113(6):1273–1278CrossRef Hayhurst C, Beems T, Jenkinson MD, Byrne P, Clark S, Kandasamy J, Goodden J, Nandoe Tewarie RD, Mallucci CL (2010) Effect of electromagnetic-navigated shunt placement on failure rates: a prospective multicenter study. J Neurosurg Dec 113(6):1273–1278CrossRef
9.
go back to reference Hermann EJ, Capelle HH, Tschan CA, Krauss JK (2012) Electromagnetic-guided neuronavigation for safe placement of intraventricular catheters in pediatric neurosurgery. J Neurosurg Pediatr. 10:327–333PubMedCrossRef Hermann EJ, Capelle HH, Tschan CA, Krauss JK (2012) Electromagnetic-guided neuronavigation for safe placement of intraventricular catheters in pediatric neurosurgery. J Neurosurg Pediatr. 10:327–333PubMedCrossRef
10.
go back to reference Hsieh CT, Chen GJ, Ma HI, Chang CF, Cheng CM, Su YH, Ju DT, Hsia CC, Chen YH, Wu HY, Liu MY (2011) The misplacement of external ventricular drain by freehand method in emergent neurosurgery. Acta Neurol Belg 111(1):22–28PubMed Hsieh CT, Chen GJ, Ma HI, Chang CF, Cheng CM, Su YH, Ju DT, Hsia CC, Chen YH, Wu HY, Liu MY (2011) The misplacement of external ventricular drain by freehand method in emergent neurosurgery. Acta Neurol Belg 111(1):22–28PubMed
11.
go back to reference Huyette DR, Turnbow BJ, Kaufman C, Vaslow DF, Whiting BB, Oh MY (2008) Accuracy of the freehand pass technique for ventriculostomy catheter placement: retrospective assessment using computed tomography scans. J Neurosurg 108(1):88–91PubMedCrossRef Huyette DR, Turnbow BJ, Kaufman C, Vaslow DF, Whiting BB, Oh MY (2008) Accuracy of the freehand pass technique for ventriculostomy catheter placement: retrospective assessment using computed tomography scans. J Neurosurg 108(1):88–91PubMedCrossRef
12.
go back to reference Kakarla UK, Kim LJ, Chang SW, Theodore N, Spetzler RF (2008) Safety and accuracy of bedside external ventricular drain placement. Neurosurgery 63(1 Suppl 1):ONS162-6 Kakarla UK, Kim LJ, Chang SW, Theodore N, Spetzler RF (2008) Safety and accuracy of bedside external ventricular drain placement. Neurosurgery 63(1 Suppl 1):ONS162-6
13.
go back to reference Kandasamy J, Hayhurst C, Clark S, Jenkinson MD, Byrne P, Karabatsou K, Mallucci CL (2011) Electromagnetic stereotactic ventriculoperitoneal csf shunting for idiopathic intracranial hypertension: a successful step forward? World Neurosurg 75(1):155–160PubMedCrossRef Kandasamy J, Hayhurst C, Clark S, Jenkinson MD, Byrne P, Karabatsou K, Mallucci CL (2011) Electromagnetic stereotactic ventriculoperitoneal csf shunting for idiopathic intracranial hypertension: a successful step forward? World Neurosurg 75(1):155–160PubMedCrossRef
14.
go back to reference Kestle JR, Drake JM, Cochrane DD, Milner R, Walker ML, Abbott R 3rd, Boop FA (2003) Endoscopic Shunt Insertion Trial participants. Lack of benefit of endoscopic ventriculoperitoneal shunt insertion: a multicenter randomized trial. J Neurosurg 98(2):284–290PubMedCrossRef Kestle JR, Drake JM, Cochrane DD, Milner R, Walker ML, Abbott R 3rd, Boop FA (2003) Endoscopic Shunt Insertion Trial participants. Lack of benefit of endoscopic ventriculoperitoneal shunt insertion: a multicenter randomized trial. J Neurosurg 98(2):284–290PubMedCrossRef
15.
go back to reference Knaus H, Abbushi A, Hoffmann KT, Schwarz K, Haberl H, Thomale UW (2009) Measurements of burr-hole localization for endoscopic procedures in the third ventricle in children. Childs Nerv Syst 25(3):293–299PubMedCrossRef Knaus H, Abbushi A, Hoffmann KT, Schwarz K, Haberl H, Thomale UW (2009) Measurements of burr-hole localization for endoscopic procedures in the third ventricle in children. Childs Nerv Syst 25(3):293–299PubMedCrossRef
16.
go back to reference Leonardo J, Hanel RA, Grand W (2009) Endoscopic tracking of a ventricular catheter for entry into the lateral ventricle: technical note. Minim Invasive Neurosurg. 52(5–6):287–289PubMedCrossRef Leonardo J, Hanel RA, Grand W (2009) Endoscopic tracking of a ventricular catheter for entry into the lateral ventricle: technical note. Minim Invasive Neurosurg. 52(5–6):287–289PubMedCrossRef
17.
go back to reference Lind CR, Tsai AM, Lind CJ, Law AJ (2009) Ventricular catheter placement accuracy in non-stereotactic shunt surgery for hydrocephalus. J Clin Neurosci 16(7):918–920PubMedCrossRef Lind CR, Tsai AM, Lind CJ, Law AJ (2009) Ventricular catheter placement accuracy in non-stereotactic shunt surgery for hydrocephalus. J Clin Neurosci 16(7):918–920PubMedCrossRef
18.
go back to reference Lind CR, Correia JA, Law AJ, Kejriwal R (2008) A survey of surgical techniques for catheterising the cerebral lateral ventricles. J Clin Neurosci 15(8):886–890PubMedCrossRef Lind CR, Correia JA, Law AJ, Kejriwal R (2008) A survey of surgical techniques for catheterising the cerebral lateral ventricles. J Clin Neurosci 15(8):886–890PubMedCrossRef
19.
go back to reference Lollis SS, Roberts DW (2008) Robotic catheter ventriculostomy: feasibility, efficacy, and implications. J Neurosurg 108(2):269–274PubMedCrossRef Lollis SS, Roberts DW (2008) Robotic catheter ventriculostomy: feasibility, efficacy, and implications. J Neurosurg 108(2):269–274PubMedCrossRef
20.
go back to reference O'Neill BR, Velez DA, Braxton EE, Whiting D, Oh MY (2008) A survey of ventriculostomy and intracranial pressure monitor placement practices. Surg Neurol 70(3):268–273PubMedCrossRef O'Neill BR, Velez DA, Braxton EE, Whiting D, Oh MY (2008) A survey of ventriculostomy and intracranial pressure monitor placement practices. Surg Neurol 70(3):268–273PubMedCrossRef
21.
go back to reference Phillips SB, Gates M, Krishnamurthy S (2012) Strategic placement of bedside ventriculostomies using ultrasound image guidance: report of three cases. Neurocrit Care 17(2):255–259PubMedCrossRef Phillips SB, Gates M, Krishnamurthy S (2012) Strategic placement of bedside ventriculostomies using ultrasound image guidance: report of three cases. Neurocrit Care 17(2):255–259PubMedCrossRef
22.
go back to reference Rehman T, Rehman AU, Rehman A, Bashir HH, Ali R, Bhimani SA, Khan S (2012) A US-based survey on ventriculostomy practices. Clin Neurol Neurosurg 114(6):651–654PubMedCrossRef Rehman T, Rehman AU, Rehman A, Bashir HH, Ali R, Bhimani SA, Khan S (2012) A US-based survey on ventriculostomy practices. Clin Neurol Neurosurg 114(6):651–654PubMedCrossRef
23.
go back to reference Reig AS, Stevenson CB, Tulipan NB (2010) CT-based, fiducial-free frameless stereotaxy for difficult ventriculoperitoneal shunt insertion: experience in 26 consecutive patients. Stereotact Funct Neurosurg 88(2):75–80PubMedCrossRef Reig AS, Stevenson CB, Tulipan NB (2010) CT-based, fiducial-free frameless stereotaxy for difficult ventriculoperitoneal shunt insertion: experience in 26 consecutive patients. Stereotact Funct Neurosurg 88(2):75–80PubMedCrossRef
24.
go back to reference Roth J, Constantini S (2012) Selective use of intra-catheter endoscopic-assisted ventricular catheter placement: indications and outcome. Childs Nerv Syst 28(8):1163–1169PubMedCrossRef Roth J, Constantini S (2012) Selective use of intra-catheter endoscopic-assisted ventricular catheter placement: indications and outcome. Childs Nerv Syst 28(8):1163–1169PubMedCrossRef
25.
go back to reference Sainte-Rose C, Piatt JH, Renier D, Pierre-Kahn A, Hirsch JF, Hoffman HJ, Humphreys RP, Hendrick EB (1991) Mechanical complications in shunts. Pediatr Neurosurg 17(1):2–9PubMedCrossRef Sainte-Rose C, Piatt JH, Renier D, Pierre-Kahn A, Hirsch JF, Hoffman HJ, Humphreys RP, Hendrick EB (1991) Mechanical complications in shunts. Pediatr Neurosurg 17(1):2–9PubMedCrossRef
26.
go back to reference Saladino A, White JB, Wijdicks EF, Lanzino G (2009) Malplacement of ventricular catheters by neurosurgeons: a single institution experience. Neurocrit Care 10(2):248–252PubMedCrossRef Saladino A, White JB, Wijdicks EF, Lanzino G (2009) Malplacement of ventricular catheters by neurosurgeons: a single institution experience. Neurocrit Care 10(2):248–252PubMedCrossRef
27.
go back to reference Sampath R, Wadhwa R, Tawfik T, Nanda A, Guthikonda B (2012) Stereotactic placement of ventricular catheters: does it affect proximal malfunction rates? Stereotact Funct Neurosurg 90(2):97–103PubMedCrossRef Sampath R, Wadhwa R, Tawfik T, Nanda A, Guthikonda B (2012) Stereotactic placement of ventricular catheters: does it affect proximal malfunction rates? Stereotact Funct Neurosurg 90(2):97–103PubMedCrossRef
28.
go back to reference Schulz M, Bohner G, Knaus H, Haberl H, Thomale UW (2010) Navigated endoscopic surgery for multiloculated hydrocephalus in children. J Neurosurg Pediatr 5(5):434–442PubMedCrossRef Schulz M, Bohner G, Knaus H, Haberl H, Thomale UW (2010) Navigated endoscopic surgery for multiloculated hydrocephalus in children. J Neurosurg Pediatr 5(5):434–442PubMedCrossRef
29.
go back to reference Sekula RF, Cohen DB, Patek PM, Jannetta PJ, Oh MY (2008) Epidemiology of ventriculostomy in the United States from 1997 to 2001. Br J Neurosurg 22(2):213–218PubMedCrossRef Sekula RF, Cohen DB, Patek PM, Jannetta PJ, Oh MY (2008) Epidemiology of ventriculostomy in the United States from 1997 to 2001. Br J Neurosurg 22(2):213–218PubMedCrossRef
30.
go back to reference Stieglitz LH, Giordano M, Samii M, Luedemann WO (2010) A new tool for frameless stereotactic placement of ventricular catheters. Neurosurgery 67(3 Suppl Operative):ons131–ons135PubMedCrossRef Stieglitz LH, Giordano M, Samii M, Luedemann WO (2010) A new tool for frameless stereotactic placement of ventricular catheters. Neurosurgery 67(3 Suppl Operative):ons131–ons135PubMedCrossRef
31.
go back to reference Strowitzki M, Komenda Y, Eymann R, Steudel WI (2008) Accuracy of ultrasound-guided puncture of the ventricular system. Childs Nerv Syst 24(1):65–69PubMedCrossRef Strowitzki M, Komenda Y, Eymann R, Steudel WI (2008) Accuracy of ultrasound-guided puncture of the ventricular system. Childs Nerv Syst 24(1):65–69PubMedCrossRef
32.
go back to reference Thomale UW, Hosch H, Koch A, Schulz M, Stoltenburg G, Haberl EJ, Sprung C (2010) Perforation holes in ventricular catheters—is less more? Childs Nerv Syst 26(6):781–789PubMedCrossRef Thomale UW, Hosch H, Koch A, Schulz M, Stoltenburg G, Haberl EJ, Sprung C (2010) Perforation holes in ventricular catheters—is less more? Childs Nerv Syst 26(6):781–789PubMedCrossRef
34.
go back to reference Toma AK, Camp S, Watkins LD, Grieve J, Kitchen ND (2009) External ventricular drain insertion accuracy: is there a need for change in practice? Neurosurgery 65(6):1197–1200PubMedCrossRef Toma AK, Camp S, Watkins LD, Grieve J, Kitchen ND (2009) External ventricular drain insertion accuracy: is there a need for change in practice? Neurosurgery 65(6):1197–1200PubMedCrossRef
35.
go back to reference Wan KR, Toy JA, Wolfe R, Danks A (2011) Factors affecting the accuracy of ventricular catheter placement. J Clin Neurosci 18(4):485–488PubMedCrossRef Wan KR, Toy JA, Wolfe R, Danks A (2011) Factors affecting the accuracy of ventricular catheter placement. J Clin Neurosci 18(4):485–488PubMedCrossRef
36.
go back to reference Whitehead WE, Jea A, Vachhrajani S, Kulkarni AV, Drake JM (2007) Accurate placement of cerebrospinal fluid shunt ventricular catheters with real-time ultrasound guidance in older children without patent fontanelles. J Neurosurg 107(5 Suppl):406–410PubMed Whitehead WE, Jea A, Vachhrajani S, Kulkarni AV, Drake JM (2007) Accurate placement of cerebrospinal fluid shunt ventricular catheters with real-time ultrasound guidance in older children without patent fontanelles. J Neurosurg 107(5 Suppl):406–410PubMed
37.
go back to reference Villavicencio AT, Leveque JC, McGirt MJ, Hopkins JS, Fuchs HE, George TM (2003) Comparison of revision rates following endoscopically versus nonendoscopically placed ventricular shunt catheters. Surg Neurol 59(5):375–379PubMedCrossRef Villavicencio AT, Leveque JC, McGirt MJ, Hopkins JS, Fuchs HE, George TM (2003) Comparison of revision rates following endoscopically versus nonendoscopically placed ventricular shunt catheters. Surg Neurol 59(5):375–379PubMedCrossRef
38.
go back to reference Vinchon M, Rekate HL, Kulkarni AV (2012) Pediatric hydrocephalus outcomes: a review. Fluids Barriers CNS 9(1):18PubMedCrossRef Vinchon M, Rekate HL, Kulkarni AV (2012) Pediatric hydrocephalus outcomes: a review. Fluids Barriers CNS 9(1):18PubMedCrossRef
39.
go back to reference Yamada SM, Yamada S, Goto Y, Nakaguchi H, Murakami M, Hoya K, Matsuno A (2012) A simple and consistent technique for ventricular catheter insertion using a tripod. Clin Neurol Neurosurg 114(6):622–626PubMedCrossRef Yamada SM, Yamada S, Goto Y, Nakaguchi H, Murakami M, Hoya K, Matsuno A (2012) A simple and consistent technique for ventricular catheter insertion using a tripod. Clin Neurol Neurosurg 114(6):622–626PubMedCrossRef
Metadata
Title
Smartphone-assisted guide for the placement of ventricular catheters
Publication date
01-01-2013
Published in
Child's Nervous System / Issue 1/2013
Print ISSN: 0256-7040
Electronic ISSN: 1433-0350
DOI
https://doi.org/10.1007/s00381-012-1943-1

Other articles of this Issue 1/2013

Child's Nervous System 1/2013 Go to the issue