01-08-2007 | Experimental

Evaluation of the influence of acquisition and reconstruction parameters for 16-row multidetector CT on coronary calcium scoring using a stationary and dynamic cardiac phantom

Authors: Philipp G. C. Begemann, Udo van Stevendaal, Ralph Koester, Andreas H. Mahnken, Andreas Koops, Gerhard Adam, Michael Grass, Claus Nolte-Ernsting

Published in: European Radiology | Issue 8/2007

Login to get access

Abstract

A calcium-scoring phantom with hydroxyapatite-filled cylindrical holes (0.5 to 4 mm) was used. High-resolution scans were performed for an accuracy baseline. The phantom was mounted to a moving heart phantom. Non-moving data with the implementation of an ECG-signal were acquired for different pitches (0.2/0.3), heart rates (60/80/95 bpm) and collimations (16 × 0.75/16 × 1.5 mm). Images were reconstructed with a cone-beam multi-cycle algorithm at a standard thickness/increment of 3 mm/1.5 mm and the thinnest possible thickness (0.8/0.4 and 2/1). Subsequently, ECG-gated moving calcium-scoring phantom data were acquired. The calcium volume and Agatston score were measured. The temporal resolution and reconstruction cycles were calculated. High-resolution scans determine the calcium volume with a high accuracy (mean overestimation, 0.8%). In the non-moving measurements, the volume underestimation ranged from about 6% (16 × 0.75 mm; 0.8/0.4 mm) to nearly 25% (16 × 1.5 mm; 3/1.5 mm). Moving scans showed increased measurement errors depending on the reconstructed RR interval, collimation, pitch, heart rate and gantry rotation time. Also, a correlation with the temporal resolution could be found. The reliability of calcium-scoring results can be improved with the use of a narrower collimation, a lower pitch and the reconstruction of thinner images, resulting in higher patient doses. The choice of the correct cardiac phase within the RR interval is essential to minimize measurement errors.
Literature
1.
go back to reference Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M Jr, Detrano R (1990) Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 15:827–832PubMedCrossRef Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M Jr, Detrano R (1990) Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol 15:827–832PubMedCrossRef
2.
go back to reference Rumberger JA, Brundage BH, Rader DJ, Kondos G (1999) Electron beam computed tomographic coronary calcium scanning: a review and guidelines for use in asymptomatic persons. Mayo Clin Proc 74:243–252PubMedCrossRef Rumberger JA, Brundage BH, Rader DJ, Kondos G (1999) Electron beam computed tomographic coronary calcium scanning: a review and guidelines for use in asymptomatic persons. Mayo Clin Proc 74:243–252PubMedCrossRef
3.
go back to reference Becker CR (2005) Estimation of cardiac event risk by MDCT. Eur Radiol 15(Suppl 2):B17–B22PubMed Becker CR (2005) Estimation of cardiac event risk by MDCT. Eur Radiol 15(Suppl 2):B17–B22PubMed
4.
go back to reference Becker CR, Majeed A, Crispin A, Knez A, Schoepf UJ, Boekstegers P, Steinbeck G, Reiser MF (2005) CT measurement of coronary calcium mass: impact on global cardiac risk assessment. Eur Radiol 15:96–101PubMedCrossRef Becker CR, Majeed A, Crispin A, Knez A, Schoepf UJ, Boekstegers P, Steinbeck G, Reiser MF (2005) CT measurement of coronary calcium mass: impact on global cardiac risk assessment. Eur Radiol 15:96–101PubMedCrossRef
5.
go back to reference Weber C, Begemann P, Wedegartner U, Meinertz T, Adam G (2005) Calcium scoring and coronary angiography performed with multislice spiral CT-clinical experience. Fortschr Geb Rontgenstr 177:50–59CrossRef Weber C, Begemann P, Wedegartner U, Meinertz T, Adam G (2005) Calcium scoring and coronary angiography performed with multislice spiral CT-clinical experience. Fortschr Geb Rontgenstr 177:50–59CrossRef
6.
go back to reference Horiguchi J, Yamamoto H, Akiyama Y, Marukawa K, Hirai N, Ito K (2004) Coronary artery calcium scoring using 16-MDCT and a retrospective ECG-gating reconstruction algorithm. Am J Roentgenol 183:103–108 Horiguchi J, Yamamoto H, Akiyama Y, Marukawa K, Hirai N, Ito K (2004) Coronary artery calcium scoring using 16-MDCT and a retrospective ECG-gating reconstruction algorithm. Am J Roentgenol 183:103–108
7.
go back to reference Ohnesorge B, Flohr T, Fischbach R, Kopp AF, Knez A, Schroder S, Schopf UJ, Crispin A, Klotz E, Reiser MF, Becker CR (2002) Reproducibility of coronary calcium quantification in repeat examinations with retrospectively ECG-gated multisection spiral CT. Eur Radiol 12:1532–1540PubMedCrossRef Ohnesorge B, Flohr T, Fischbach R, Kopp AF, Knez A, Schroder S, Schopf UJ, Crispin A, Klotz E, Reiser MF, Becker CR (2002) Reproducibility of coronary calcium quantification in repeat examinations with retrospectively ECG-gated multisection spiral CT. Eur Radiol 12:1532–1540PubMedCrossRef
8.
go back to reference Horiguchi J, Yamamoto H, Akiyama Y, Hirai N, Marukawa K, Fukuda H, Ito K (2005) Variability of repeated coronary artery calcium measurements by 16-MDCT with retrospective reconstruction. Am J Roentgenol 184:1917–1923 Horiguchi J, Yamamoto H, Akiyama Y, Hirai N, Marukawa K, Fukuda H, Ito K (2005) Variability of repeated coronary artery calcium measurements by 16-MDCT with retrospective reconstruction. Am J Roentgenol 184:1917–1923
9.
go back to reference Kopp AF, Ohnesorge B, Becker C, Schroder S, Heuschmid M, Kuttner A, Kuzo R, Claussen CD (2002) Reproducibility and accuracy of coronary calcium measurements with multi-detector row versus electron-beam CT. Radiology 225:113–119PubMedCrossRef Kopp AF, Ohnesorge B, Becker C, Schroder S, Heuschmid M, Kuttner A, Kuzo R, Claussen CD (2002) Reproducibility and accuracy of coronary calcium measurements with multi-detector row versus electron-beam CT. Radiology 225:113–119PubMedCrossRef
10.
go back to reference Carr JJ, Nelson JC, Wong ND, McNitt-Gray M, Arad Y, Jacobs DR Jr, Sidney S, Bild DE, Williams OD, Detrano RC (2005) Calcified coronary artery plaque measurement with cardiac CT in population-based studies: standardized protocol of Multi-Ethnic Study of Atherosclerosis (MESA) and Coronary Artery Risk Development in Young Adults (CARDIA) study. Radiology 234:35–43PubMedCrossRef Carr JJ, Nelson JC, Wong ND, McNitt-Gray M, Arad Y, Jacobs DR Jr, Sidney S, Bild DE, Williams OD, Detrano RC (2005) Calcified coronary artery plaque measurement with cardiac CT in population-based studies: standardized protocol of Multi-Ethnic Study of Atherosclerosis (MESA) and Coronary Artery Risk Development in Young Adults (CARDIA) study. Radiology 234:35–43PubMedCrossRef
11.
go back to reference Detrano RC, Anderson M, Nelson J, Wong ND, Carr JJ, McNitt-Gray M, Bild DE (2005) Coronary calcium measurements: effect of CT scanner type and calcium measure on rescan reproducibility-MESA study. Radiology 236:477–484PubMedCrossRef Detrano RC, Anderson M, Nelson J, Wong ND, Carr JJ, McNitt-Gray M, Bild DE (2005) Coronary calcium measurements: effect of CT scanner type and calcium measure on rescan reproducibility-MESA study. Radiology 236:477–484PubMedCrossRef
12.
go back to reference Ulzheimer S, Kalender WA (2003) Assessment of calcium scoring performance in cardiac computed tomography. Eur Radiol 13:484–497PubMed Ulzheimer S, Kalender WA (2003) Assessment of calcium scoring performance in cardiac computed tomography. Eur Radiol 13:484–497PubMed
13.
go back to reference Mahnken AH, Wildberger JE, Sinha AM, Flohr T, Truong HT, Krombach GA, Gunther RW (2002) Variation of the coronary calcium score depending on image reconstruction interval and scoring algorithm. Invest Radiol 37:496–502PubMedCrossRef Mahnken AH, Wildberger JE, Sinha AM, Flohr T, Truong HT, Krombach GA, Gunther RW (2002) Variation of the coronary calcium score depending on image reconstruction interval and scoring algorithm. Invest Radiol 37:496–502PubMedCrossRef
14.
go back to reference Schlosser T, Hunold P, Schmermund A, Kuhl H, Waltering KU, Debatin JF, Barkhausen J (2004) Coronary artery calcium score: influence of reconstruction interval at 16-detector row CT with retrospective electrocardiographic gating. Radiology 233:586–589PubMedCrossRef Schlosser T, Hunold P, Schmermund A, Kuhl H, Waltering KU, Debatin JF, Barkhausen J (2004) Coronary artery calcium score: influence of reconstruction interval at 16-detector row CT with retrospective electrocardiographic gating. Radiology 233:586–589PubMedCrossRef
15.
go back to reference Horiguchi J, Nakanishi T, Tamura A, Ito K (2002) Coronary artery calcium scoring using multicardiac computed tomography. J Comput Assist Tomogr 26:880–885PubMedCrossRef Horiguchi J, Nakanishi T, Tamura A, Ito K (2002) Coronary artery calcium scoring using multicardiac computed tomography. J Comput Assist Tomogr 26:880–885PubMedCrossRef
16.
go back to reference Hong C, Pilgram TK, Zhu F, Joe BN, Towler DA, Bae KT (2003) Improving mass measurement of coronary artery calcification using threshold correction and thin collimation in multi-detector row computed tomography: in vitro experiment. Acad Radiol 10:969–977PubMedCrossRef Hong C, Pilgram TK, Zhu F, Joe BN, Towler DA, Bae KT (2003) Improving mass measurement of coronary artery calcification using threshold correction and thin collimation in multi-detector row computed tomography: in vitro experiment. Acad Radiol 10:969–977PubMedCrossRef
17.
go back to reference Ohnesorge B, Becker CR, Flohr T, Dorgelo J, Oudkerk M (2004) Multi-slice computed tomography-technical principles, clinical application and future perspectives in book. In: Oudkerk M (ed) Coronary radiology. Springer, Berlin Heidelberg New York, pp 87–115 Ohnesorge B, Becker CR, Flohr T, Dorgelo J, Oudkerk M (2004) Multi-slice computed tomography-technical principles, clinical application and future perspectives in book. In: Oudkerk M (ed) Coronary radiology. Springer, Berlin Heidelberg New York, pp 87–115
18.
go back to reference Muhlenbruch G, Thomas C, Wildberger JE, Koos R, Das M, Hohl C, Katoh M, Gunther RW, Mahnken AH (2005) Effect of varying slice thickness on coronary calcium scoring with multislice computed tomography in vitro and in vivo. Invest Radiol 40:695–699PubMedCrossRef Muhlenbruch G, Thomas C, Wildberger JE, Koos R, Das M, Hohl C, Katoh M, Gunther RW, Mahnken AH (2005) Effect of varying slice thickness on coronary calcium scoring with multislice computed tomography in vitro and in vivo. Invest Radiol 40:695–699PubMedCrossRef
19.
go back to reference Mahnken AH, Muhlenbruch G, Koos R, Das M, Pohl S, Stanzel S, Gunther RW, Wildberger JE (2006) Influence of a small field-of-view size on the detection of coronary artery calcifications with MSCT: in vitro and in vivo study. Eur Radiol 16:358–364PubMedCrossRef Mahnken AH, Muhlenbruch G, Koos R, Das M, Pohl S, Stanzel S, Gunther RW, Wildberger JE (2006) Influence of a small field-of-view size on the detection of coronary artery calcifications with MSCT: in vitro and in vivo study. Eur Radiol 16:358–364PubMedCrossRef
20.
go back to reference Hong C, Bae KT, Pilgram TK (2003) Coronary artery calcium: accuracy and reproducibility of measurements with multi-detector row CT-assessment of effects of different thresholds and quantification methods. Radiology 227:795–801PubMedCrossRef Hong C, Bae KT, Pilgram TK (2003) Coronary artery calcium: accuracy and reproducibility of measurements with multi-detector row CT-assessment of effects of different thresholds and quantification methods. Radiology 227:795–801PubMedCrossRef
21.
go back to reference Hong C, Bae KT, Pilgram TK, Zhu F (2003) Coronary artery calcium quantification at multi-detector row CT: influence of heart rate and measurement methods on interacquisition variability initial experience. Radiology 228:95–100PubMedCrossRef Hong C, Bae KT, Pilgram TK, Zhu F (2003) Coronary artery calcium quantification at multi-detector row CT: influence of heart rate and measurement methods on interacquisition variability initial experience. Radiology 228:95–100PubMedCrossRef
22.
go back to reference van Ooijen PM, Vliegenthart R, Witteman JC, Oudkerk M (2005) Influence of scoring parameter settings on Agatston and volume scores for coronary calcification. Eur Radiol 15:102–110PubMedCrossRef van Ooijen PM, Vliegenthart R, Witteman JC, Oudkerk M (2005) Influence of scoring parameter settings on Agatston and volume scores for coronary calcification. Eur Radiol 15:102–110PubMedCrossRef
23.
go back to reference Hong C, Pilgram TK, Zhu F, Bae KT (2004) Is coronary artery calcium mass related to Agatston score? Acad Radiol 11:286–292PubMedCrossRef Hong C, Pilgram TK, Zhu F, Bae KT (2004) Is coronary artery calcium mass related to Agatston score? Acad Radiol 11:286–292PubMedCrossRef
24.
go back to reference Mahnken AH, Wildberger JE, Simon J, Koos R, Flohr TG, Schaller S, Gunther RW (2003) Detection of coronary calcifications: feasibility of dose reduction with a body weight-adapted examination protocol. Am J Roentgenol 181:533–538 Mahnken AH, Wildberger JE, Simon J, Koos R, Flohr TG, Schaller S, Gunther RW (2003) Detection of coronary calcifications: feasibility of dose reduction with a body weight-adapted examination protocol. Am J Roentgenol 181:533–538
25.
go back to reference Hong C, Bae KT, Pilgram TK, Suh J, Bradley D (2002) Coronary artery calcium measurement with multi-detector row CT: in vitro assessment of effect of radiation dose. Radiology 225:901–906PubMedCrossRef Hong C, Bae KT, Pilgram TK, Suh J, Bradley D (2002) Coronary artery calcium measurement with multi-detector row CT: in vitro assessment of effect of radiation dose. Radiology 225:901–906PubMedCrossRef
26.
go back to reference Jakobs TF, Wintersperger BJ, Herzog P, Flohr T, Suess C, Knez A, Reiser MF, Becker CR (2003) Ultra-low-dose coronary artery calcium screening using multislice CT with retrospective ECG gating. Eur Radiol 13:1923–1930PubMedCrossRef Jakobs TF, Wintersperger BJ, Herzog P, Flohr T, Suess C, Knez A, Reiser MF, Becker CR (2003) Ultra-low-dose coronary artery calcium screening using multislice CT with retrospective ECG gating. Eur Radiol 13:1923–1930PubMedCrossRef
27.
go back to reference Sediono W, Dossel O (2002) Elastomechanics of the ventricles: development of a phantom and results of simulation. Biomed Tech (Berl) 47(Suppl 1 Pt 1):243–245CrossRef Sediono W, Dossel O (2002) Elastomechanics of the ventricles: development of a phantom and results of simulation. Biomed Tech (Berl) 47(Suppl 1 Pt 1):243–245CrossRef
28.
go back to reference Timinger H, Krueger S, Borgert J et al (2004) Motion compensation for interventional navigation on 3D static roadmaps based on an affine model and gating. Phys Med Biol 49:719–732PubMedCrossRef Timinger H, Krueger S, Borgert J et al (2004) Motion compensation for interventional navigation on 3D static roadmaps based on an affine model and gating. Phys Med Biol 49:719–732PubMedCrossRef
29.
go back to reference Begemann PG, van Stevendaal U, Manzke R, Stork A, Weiss F, Nolte-Ernsting C, Grass M, Adam G (2005) Evaluation of spatial and temporal resolution for ECG-gated 16-row multidetector CT using a dynamic cardiac phantom. Eur Radiol 15:1015–1026PubMedCrossRef Begemann PG, van Stevendaal U, Manzke R, Stork A, Weiss F, Nolte-Ernsting C, Grass M, Adam G (2005) Evaluation of spatial and temporal resolution for ECG-gated 16-row multidetector CT using a dynamic cardiac phantom. Eur Radiol 15:1015–1026PubMedCrossRef
30.
go back to reference Grass M, Manzke R, Nielsen T, Koken P, Proksa R, Natanzon M, Shechter G (2003) Helical cardiac cone beam reconstruction using retrospective ECG gating. Phys Med Biol 48:3069–3084PubMedCrossRef Grass M, Manzke R, Nielsen T, Koken P, Proksa R, Natanzon M, Shechter G (2003) Helical cardiac cone beam reconstruction using retrospective ECG gating. Phys Med Biol 48:3069–3084PubMedCrossRef
31.
go back to reference Manzke R, Kohler T, Nielsen T, Hawkes D, Grass M (2004) Automatic phase determination for retrospectively gated cardiac CT. Med Phys 31:3345–3362PubMedCrossRef Manzke R, Kohler T, Nielsen T, Hawkes D, Grass M (2004) Automatic phase determination for retrospectively gated cardiac CT. Med Phys 31:3345–3362PubMedCrossRef
32.
go back to reference Manzke R, Grass M, Nielsen T, Shechter G, Hawkes D (2003) Adaptive temporal resolution optimization in helical cardiac cone beam CT reconstruction. Med Phys 30:3072–3080PubMedCrossRef Manzke R, Grass M, Nielsen T, Shechter G, Hawkes D (2003) Adaptive temporal resolution optimization in helical cardiac cone beam CT reconstruction. Med Phys 30:3072–3080PubMedCrossRef
33.
go back to reference Shechter G, Naveh G, Altmann A et al (2003) Cardiac image reconstruction on a 16-slice CT scanner using a retrospectively gated, multi-cycle 3D back projection algorithm. Proceedings of SPIE Medical Imaging Conference 5032:1820–1828 Shechter G, Naveh G, Altmann A et al (2003) Cardiac image reconstruction on a 16-slice CT scanner using a retrospectively gated, multi-cycle 3D back projection algorithm. Proceedings of SPIE Medical Imaging Conference 5032:1820–1828
34.
go back to reference Horiguchi J, Shen Y, Akiyama Y, Hirai N, Sasaki K, Ishifuro M, Nakanishi T, Ito K (2005) Electron Beam CT Versus 16-MDCT on the variability of repeated coronary artery calcium measurements in a variable heart rate phantom. Am J Roentgenol 185:995–1000CrossRef Horiguchi J, Shen Y, Akiyama Y, Hirai N, Sasaki K, Ishifuro M, Nakanishi T, Ito K (2005) Electron Beam CT Versus 16-MDCT on the variability of repeated coronary artery calcium measurements in a variable heart rate phantom. Am J Roentgenol 185:995–1000CrossRef
35.
go back to reference Horiguchi J, Shen Y, Akiyama Y, Hirai N, Sasaki K, Ishifuro M, Ito K (2006) Electron beam CT versus 16-slice spiral CT: how accurately can we measure coronary artery calcium volume? Eur Radiol 16:374–380PubMedCrossRef Horiguchi J, Shen Y, Akiyama Y, Hirai N, Sasaki K, Ishifuro M, Ito K (2006) Electron beam CT versus 16-slice spiral CT: how accurately can we measure coronary artery calcium volume? Eur Radiol 16:374–380PubMedCrossRef
36.
go back to reference Hoffmann MH, Lessick J, Manzke R, Schmid FT, Gershin E, Boll DT, Rispler S, Aschoff AJ, Grass M (2006) Automatic determination of minimal cardiac motion phases for computed tomography imaging: initial experience. Eur Radiol 16:365–373PubMedCrossRef Hoffmann MH, Lessick J, Manzke R, Schmid FT, Gershin E, Boll DT, Rispler S, Aschoff AJ, Grass M (2006) Automatic determination of minimal cardiac motion phases for computed tomography imaging: initial experience. Eur Radiol 16:365–373PubMedCrossRef
37.
go back to reference Horiguchi J, Nakanishi T, Ito K (2001) Quantification of coronary artery calcium using multidetector CT and a retrospective ECG-gating reconstruction algorithm. Am J Roentgenol 177:1429–1435 Horiguchi J, Nakanishi T, Ito K (2001) Quantification of coronary artery calcium using multidetector CT and a retrospective ECG-gating reconstruction algorithm. Am J Roentgenol 177:1429–1435
Metadata
Title
Evaluation of the influence of acquisition and reconstruction parameters for 16-row multidetector CT on coronary calcium scoring using a stationary and dynamic cardiac phantom
Authors
Philipp G. C. Begemann
Udo van Stevendaal
Ralph Koester
Andreas H. Mahnken
Andreas Koops
Gerhard Adam
Michael Grass
Claus Nolte-Ernsting
Publication date
01-08-2007
Publisher
Springer-Verlag
Published in
European Radiology / Issue 8/2007
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-006-0564-z

Other articles of this Issue 8/2007

European Radiology 8/2007 Go to the issue

Hepatobiliary-Pancreas

Imaging of Budd-Chiari syndrome