Published in: European Radiology 1/2006

01-01-2006 | Physics

Survey of computed tomography techniques and absorbed dose in Italian hospitals: a comparison between two methods to estimate the dose–length product and the effective dose and to verify fulfilment of the diagnostic reference levels

Authors: Daniela Origgi, Sabrina Vigorito, Gaetano Villa, Massimo Bellomi, Giampiero Tosi

Published in: European Radiology | Issue 1/2006

Login to get access

Abstract

The aim of this study was the production of the first Italian survey of radiation dose in computed tomography (CT) prior to the widespread adoption of multislice CT, in order to have a reference point to facilitate later investigation of dose exposure changes brought by this new CT modality. The collected dose data were compared with diagnostic reference levels (DRLs). The agreement between experimental dose evaluation and Monte Carlo (MC) simulations was investigated. The survey was carried out in 29 Italian hospitals, covered 48 CT scanners and 232 examinations. The dose–length product (DLP) and effective dose (E) values were estimated based on MC simulations for seven clinical protocols using the CT-Dose program. Statistical analysis showed a significant difference (p<0.01) in the DLP between the two methods, with MC values being greater than the experimental ones. For E, the MC values were greater in routine head (8.2%), cervical spine (2.7%) and lumbar spine (2.9%) studies. The weighted CT dose index, the DLP and E were always below the DRLs set by the European Community. This dose survey gives a good but incomplete picture of the Italian CT dose situation and may be useful as a reference baseline for defining clinical multislice protocols in the near future.
Appendix
Available only for authorised users
Literature
1.
go back to reference Shrimpton PC, Wall BF (1995) The increasing importance of X ray computed tomography as a source of medical exposure. Radiat Prot Dosim 57:413–415 Shrimpton PC, Wall BF (1995) The increasing importance of X ray computed tomography as a source of medical exposure. Radiat Prot Dosim 57:413–415
2.
go back to reference Jurik AG, Jessen KA, Hansen J (1997) Image quality and dose in computed tomography. Eur Radiol 7:77–81 Jurik AG, Jessen KA, Hansen J (1997) Image quality and dose in computed tomography. Eur Radiol 7:77–81
3.
go back to reference Smith A, Shah GA, Kron T (1998) Variation of patient dose in head CT. Br J Radiol 71:1296–1301 Smith A, Shah GA, Kron T (1998) Variation of patient dose in head CT. Br J Radiol 71:1296–1301
4.
go back to reference Kalender WA (1998) Computed tomography: influence of exposure parameters and the establishment of reference dose values. Radiat Prot Dosim 80:163–166 Kalender WA (1998) Computed tomography: influence of exposure parameters and the establishment of reference dose values. Radiat Prot Dosim 80:163–166
5.
go back to reference Calzado A, Rodriguez R, Munoz A (2000) Quality criteria implementation for brain and lumbar spine CT examinations. Br J Radiol 73:384–395 Calzado A, Rodriguez R, Munoz A (2000) Quality criteria implementation for brain and lumbar spine CT examinations. Br J Radiol 73:384–395
6.
go back to reference European guidelines on quality criteria for computed tomography. Report EUR 16262. European Community, Brussels, Belgium, 1998 European guidelines on quality criteria for computed tomography. Report EUR 16262. European Community, Brussels, Belgium, 1998
7.
go back to reference Clarke J, Cranley K, Robinson J, Smith PH, Workman A (2000) Application of draft European Commission reference levels to a regional CT dose survey. Br J Radiol 73:43–50 Clarke J, Cranley K, Robinson J, Smith PH, Workman A (2000) Application of draft European Commission reference levels to a regional CT dose survey. Br J Radiol 73:43–50
8.
go back to reference Crawley MT, Booth A, Wainwright A (2001) A practical approach to the first iteration in the optimization of radiation dose and image quality in CT: estimates of the collective dose savings achieved. Br J Radiol 74:607–614 Crawley MT, Booth A, Wainwright A (2001) A practical approach to the first iteration in the optimization of radiation dose and image quality in CT: estimates of the collective dose savings achieved. Br J Radiol 74:607–614
9.
go back to reference Dixon AK, Goldstone KE (2002) Abdominal CT and the Euratom directive. Eur Radiol 12:1567–1570 Dixon AK, Goldstone KE (2002) Abdominal CT and the Euratom directive. Eur Radiol 12:1567–1570
10.
go back to reference Goddard CC, al-Farsi A (1999) Radiation doses from CT in the sultanate of Oman. Br J Radiol 72:1073–1077 Goddard CC, al-Farsi A (1999) Radiation doses from CT in the sultanate of Oman. Br J Radiol 72:1073–1077
11.
go back to reference Jessen KA, Shrimpton PC, Geleijns J, Panzer W, Tosi G (1999) Dosimetry for optimisation of patient protection in computed tomography. Appl Radiat Isotopes 50:165–172 Jessen KA, Shrimpton PC, Geleijns J, Panzer W, Tosi G (1999) Dosimetry for optimisation of patient protection in computed tomography. Appl Radiat Isotopes 50:165–172
12.
go back to reference Shrimpton PC, Jessen KA, Geleijns J, Panzer W, Tosi G (1998) Reference doses in computed tomography. Radiat Prot Dosim 80:55–59 Shrimpton PC, Jessen KA, Geleijns J, Panzer W, Tosi G (1998) Reference doses in computed tomography. Radiat Prot Dosim 80:55–59
13.
go back to reference IEC. Medical diagnostic X-ray equipment-radiation conditions for use in the determination of characteristics, IEC 61267. IEC, Geneva, 1994 IEC. Medical diagnostic X-ray equipment-radiation conditions for use in the determination of characteristics, IEC 61267. IEC, Geneva, 1994
14.
go back to reference Baadegaard N, Jensen LC (1999) A CT dose calculation software “CT-Dose”. National Board of Health, Aarhus University Hospital, Denmark Baadegaard N, Jensen LC (1999) A CT dose calculation software “CT-Dose”. National Board of Health, Aarhus University Hospital, Denmark
15.
go back to reference IEC. Medical electrical equipment. Part 2: particular requirements for the safety of X-ray equipment for computed tomography, IEC 60601-2-44. IEC, Geneva, 1999 IEC. Medical electrical equipment. Part 2: particular requirements for the safety of X-ray equipment for computed tomography, IEC 60601-2-44. IEC, Geneva, 1999
16.
go back to reference Jones DG, Shrimpton PC (1993) Normalized organ doses for X-ray computed tomography calculated using Monte Carlo techniques NRPB-SR250 Jones DG, Shrimpton PC (1993) Normalized organ doses for X-ray computed tomography calculated using Monte Carlo techniques NRPB-SR250
17.
go back to reference Kalender WA, Schmidt B, Zankl M, Schmidt M (1999) A PC program for estimating organ dose and effective dose values in computed tomography. Eur Radiol 9:555–562CrossRefPubMed Kalender WA, Schmidt B, Zankl M, Schmidt M (1999) A PC program for estimating organ dose and effective dose values in computed tomography. Eur Radiol 9:555–562CrossRefPubMed
18.
go back to reference Brix G et al (2004) Assessment of a theoretical formalism for dose estimation in CT: an anthropomorphic phantom study. Eur Radiol 14(7):1275–1284 Brix G et al (2004) Assessment of a theoretical formalism for dose estimation in CT: an anthropomorphic phantom study. Eur Radiol 14(7):1275–1284
19.
go back to reference Hidajat N, Maurer J, Schroder RJ, Nunnemann A, Wolf M, Pauli K, Felix R (1999) Relationships between physical dose quantities and patient dose in CT. Br J Radiol 72:556–561 Hidajat N, Maurer J, Schroder RJ, Nunnemann A, Wolf M, Pauli K, Felix R (1999) Relationships between physical dose quantities and patient dose in CT. Br J Radiol 72:556–561
20.
go back to reference Jarry G, De Marco JJ, Beifuss U, Cagnon CH, McNitt-Gray MF (2003) A Monte Carlo-based method to estimate radiation dose from spiral CT: from phantom testing to patient-specific models. Phys Med Biol 48:2645–2663 Jarry G, De Marco JJ, Beifuss U, Cagnon CH, McNitt-Gray MF (2003) A Monte Carlo-based method to estimate radiation dose from spiral CT: from phantom testing to patient-specific models. Phys Med Biol 48:2645–2663
21.
go back to reference Hiles P, Brennen S, Scott S, Davies J (2001) A survey of patient dose and image quality for computed tomography scanners in Wales. J Radiol Prot 21:345–354 Hiles P, Brennen S, Scott S, Davies J (2001) A survey of patient dose and image quality for computed tomography scanners in Wales. J Radiol Prot 21:345–354
22.
go back to reference Tsapaki V, Kottou S, Papadimitriou D (2001) Application of European Commission reference dose levels in CT examinations in Crete Greece. Br J Radiol 74:836–840 Tsapaki V, Kottou S, Papadimitriou D (2001) Application of European Commission reference dose levels in CT examinations in Crete Greece. Br J Radiol 74:836–840
23.
go back to reference Hidajat N, Wolf M, Nunnemann A, Liersch P, Gebauer B, Teichgraber U, Schroder RJ, Felix R (2001) Survey of conventional and spiral CT doses. Radiology 218:395–401 Hidajat N, Wolf M, Nunnemann A, Liersch P, Gebauer B, Teichgraber U, Schroder RJ, Felix R (2001) Survey of conventional and spiral CT doses. Radiology 218:395–401
24.
go back to reference Papadimitriou D, Perris A, Manetou A, Molfetas M, Panagiotakis N, Lyra-Georgosopoulou M, Hourdakis K, Kottou S, Tosi G, Origgi D, Vigorito S (2003) A survey of 14 computed tomography scanners in Greece and 32 scanners in Italy. Examination frequencies, dose reference values, effective doses and doses to organs. Radiat Prot Dosim 104(1):47–53 Papadimitriou D, Perris A, Manetou A, Molfetas M, Panagiotakis N, Lyra-Georgosopoulou M, Hourdakis K, Kottou S, Tosi G, Origgi D, Vigorito S (2003) A survey of 14 computed tomography scanners in Greece and 32 scanners in Italy. Examination frequencies, dose reference values, effective doses and doses to organs. Radiat Prot Dosim 104(1):47–53
Metadata
Title
Survey of computed tomography techniques and absorbed dose in Italian hospitals: a comparison between two methods to estimate the dose–length product and the effective dose and to verify fulfilment of the diagnostic reference levels
Authors
Daniela Origgi
Sabrina Vigorito
Gaetano Villa
Massimo Bellomi
Giampiero Tosi
Publication date
01-01-2006
Publisher
Springer-Verlag
Published in
European Radiology / Issue 1/2006
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-005-2682-4

Other articles of this Issue 1/2006

European Radiology 1/2006 Go to the issue