Whole-body MRI at high field: technical limits and clinical potential

Author: Fritz Schick

Published in: European Radiology | Issue 5/2005

Login to get access

Abstract

This review seeks to clarify the most important implications of higher magnetic field strength for clinical examinations of the whole body. An overview is provided on the resulting advantages and disadvantages for anatomical, functional and biochemical magnetic resonance examinations in different regions of the body. It is demonstrated that susceptibility-dependent imaging, chemical shift selective (e.g., fat-suppressed) imaging, and spectroscopic techniques clearly gain from higher field strength. Problems due to shorter wavelength and higher radio frequency energy deposition at higher field strength are reported, especially in examinations of the body trunk. Thorax examinations provided sufficient homogeneity of the radio frequency field for common examination techniques in most cases, whereas abdominal and pelvic imaging was often hampered by undesired dielectric effects. Currently available and potential future strategies to overcome related limitations are discussed. Whole-body MRI at higher field strength currently leads to clearly improved image quality using a variety of established sequence types and for examination of many body regions. But some major problems at higher field strength have to be solved before high-field magnetic resonance systems can really replace the well-established and technically developed magnetic resonance systems operating at 1.5 T for each clinical application.
Literature
1.
go back to reference Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P (1999) SENSE: sensitivity encoding for fast MRI. Magn Reson Med 42:952–962CrossRefPubMed Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P (1999) SENSE: sensitivity encoding for fast MRI. Magn Reson Med 42:952–962CrossRefPubMed
2.
go back to reference Wang Y (2000) Description of parallel imaging in MRI using multiple coils. Magn Reson Med 44:495–499 Wang Y (2000) Description of parallel imaging in MRI using multiple coils. Magn Reson Med 44:495–499
3.
go back to reference Paschal CB, Morris HD (2004) K-space in the clinic. J Magn Reson Imaging 19:145–159CrossRef Paschal CB, Morris HD (2004) K-space in the clinic. J Magn Reson Imaging 19:145–159CrossRef
4.
go back to reference De Zwart JA, Ledden PJ, van Gelderen P, Bodurka J, Chu R, Duyn JH (2004) Signal-to-noise ratio and parallel imaging performance of a 16-channel receive-only brain coil array at 3.0 Tesla. Magn Reson Med 51:22–26CrossRef De Zwart JA, Ledden PJ, van Gelderen P, Bodurka J, Chu R, Duyn JH (2004) Signal-to-noise ratio and parallel imaging performance of a 16-channel receive-only brain coil array at 3.0 Tesla. Magn Reson Med 51:22–26CrossRef
5.
go back to reference Marzola P, Osculati F, Sbarbati A (2003) High field MRI in preclinical research. Eur J Radiol 48:165–170CrossRef Marzola P, Osculati F, Sbarbati A (2003) High field MRI in preclinical research. Eur J Radiol 48:165–170CrossRef
6.
go back to reference Chen W, Ugurbil K (1999) High spatial resolution functional magnetic resonance imaging at very-high-magnetic field. Top Magn Reson Imaging 10:63–78 Chen W, Ugurbil K (1999) High spatial resolution functional magnetic resonance imaging at very-high-magnetic field. Top Magn Reson Imaging 10:63–78
7.
go back to reference Silvennoinen MJ, Clingman CS, Golay X, Kauppinen RA, van Zijl PC (2003) Comparison of the dependence of blood R2 and R2* on oxygen saturation at 1.5 and 4.7 Tesla. Magn Reson Med 49:47–60CrossRef Silvennoinen MJ, Clingman CS, Golay X, Kauppinen RA, van Zijl PC (2003) Comparison of the dependence of blood R2 and R2* on oxygen saturation at 1.5 and 4.7 Tesla. Magn Reson Med 49:47–60CrossRef
8.
go back to reference Tkac I, Andersen P, Adriany G, Merkle H, Ugurbil K, Gruetter R (2001) In vivo 1H NMR spectroscopy of the human brain at 7 T. Magn Reson Med 46:451–456CrossRef Tkac I, Andersen P, Adriany G, Merkle H, Ugurbil K, Gruetter R (2001) In vivo 1H NMR spectroscopy of the human brain at 7 T. Magn Reson Med 46:451–456CrossRef
9.
go back to reference Ugurbil K, Adriany G, Andersen P, Chen W, Garwood M, Gruetter R, Henry PG, Kim SG, Lieu H, Tkac I, Vaughan T, Van De Moortele PF, Yacoub E, Zhu XH (2003) Ultrahigh field magnetic resonance imaging and spectroscopy. Magn Reson Imaging 21:1263–1281CrossRef Ugurbil K, Adriany G, Andersen P, Chen W, Garwood M, Gruetter R, Henry PG, Kim SG, Lieu H, Tkac I, Vaughan T, Van De Moortele PF, Yacoub E, Zhu XH (2003) Ultrahigh field magnetic resonance imaging and spectroscopy. Magn Reson Imaging 21:1263–1281CrossRef
10.
go back to reference Rotter M, Berg A, Langenberger H, Grampp S, Imhof H, Moser E (2001) Autocorrelation analysis of bone structure. J Magn Reson Imaging 14(1):87–93CrossRef Rotter M, Berg A, Langenberger H, Grampp S, Imhof H, Moser E (2001) Autocorrelation analysis of bone structure. J Magn Reson Imaging 14(1):87–93CrossRef
11.
go back to reference Uematsu H, Dougherty L, Takahashi M, Ohno Y, Nakatsu M, Song HK, Ferrari VA, Gefter WB, Schnall MD, Hatabu H (2001) Pulmonary MR angiography with contrast agent at 4 Tesla: a preliminary result. Magn Reson Med 46:1028–1030 Uematsu H, Dougherty L, Takahashi M, Ohno Y, Nakatsu M, Song HK, Ferrari VA, Gefter WB, Schnall MD, Hatabu H (2001) Pulmonary MR angiography with contrast agent at 4 Tesla: a preliminary result. Magn Reson Med 46:1028–1030
12.
go back to reference Dougherty L, Connick TJ, Mizsei G (2001) Cardiac imaging at 4 Tesla. Magn Reson Med 45:176–178CrossRef Dougherty L, Connick TJ, Mizsei G (2001) Cardiac imaging at 4 Tesla. Magn Reson Med 45:176–178CrossRef
13.
go back to reference Klarhofer M, Csapo B, Balassy C, Szeles JC, Moser E (2001) High-resolution blood flow velocity measurements in the human finger. Magn Reson Med 45:716–719CrossRef Klarhofer M, Csapo B, Balassy C, Szeles JC, Moser E (2001) High-resolution blood flow velocity measurements in the human finger. Magn Reson Med 45:716–719CrossRef
14.
go back to reference Szeles JC, Csapo B, Klarhofer M, Balassy C, Hoda R, Berg A, Roden M, Polterauer P, Waldhausl W, Moser E (2001) In vivo magnetic resonance micro-imaging of the human toe at 3 Tesla. Magn Reson Imaging 19:1235–1238CrossRef Szeles JC, Csapo B, Klarhofer M, Balassy C, Hoda R, Berg A, Roden M, Polterauer P, Waldhausl W, Moser E (2001) In vivo magnetic resonance micro-imaging of the human toe at 3 Tesla. Magn Reson Imaging 19:1235–1238CrossRef
15.
go back to reference Csapo B, Szeles J, Helbich TH, Klarhofer M, Balassy C, Pammer J, Obermair A (2002) Histopathologic correlation of high-resolution magnetic resonance imaging of human cervical tissue samples at 3 Tesla: validation of a technique. Invest Radiol 37:381–385CrossRef Csapo B, Szeles J, Helbich TH, Klarhofer M, Balassy C, Pammer J, Obermair A (2002) Histopathologic correlation of high-resolution magnetic resonance imaging of human cervical tissue samples at 3 Tesla: validation of a technique. Invest Radiol 37:381–385CrossRef
16.
go back to reference Lin W, An H, Chen Y, Nicholas P, Zhai G, Gerig G, Gilmore J, Bullitt E (2003) Practical consideration for 3T imaging. Magn Reson Imaging Clin N Am 11:615–639 Lin W, An H, Chen Y, Nicholas P, Zhai G, Gerig G, Gilmore J, Bullitt E (2003) Practical consideration for 3T imaging. Magn Reson Imaging Clin N Am 11:615–639
17.
go back to reference Uematsu H, Dougherty L, Takahashi M, Ohno Y, Nakatsu M, Schnall MD, Hatabu H (2003) Abdominal imaging at 4 T MR system: a preliminary result. Eur J Radiol 47:161–163CrossRef Uematsu H, Dougherty L, Takahashi M, Ohno Y, Nakatsu M, Schnall MD, Hatabu H (2003) Abdominal imaging at 4 T MR system: a preliminary result. Eur J Radiol 47:161–163CrossRef
18.
go back to reference Price SJ, Burnet NG, Donovan T, Green HA, Pena A, Antoun NM, Pickard JD, Carpenter TA, Gillard JH (2003) Diffusion tensor imaging of brain tumours at 3T: a potential tool for assessing white matter tract invasion? Clin Radiol 58:455–462CrossRefPubMed Price SJ, Burnet NG, Donovan T, Green HA, Pena A, Antoun NM, Pickard JD, Carpenter TA, Gillard JH (2003) Diffusion tensor imaging of brain tumours at 3T: a potential tool for assessing white matter tract invasion? Clin Radiol 58:455–462CrossRefPubMed
20.
go back to reference Hormann M, Traxler H, Ba-Ssalamah A, Mlynarik V, Shodaj-Baghini M, Kubiena H, Trattnig S (2003) Correlative high-resolution MR-anatomic study of sciatic, ulnar, and proper palmar digital nerve. Magn Reson Imaging 21:879–885CrossRef Hormann M, Traxler H, Ba-Ssalamah A, Mlynarik V, Shodaj-Baghini M, Kubiena H, Trattnig S (2003) Correlative high-resolution MR-anatomic study of sciatic, ulnar, and proper palmar digital nerve. Magn Reson Imaging 21:879–885CrossRef
21.
go back to reference Hinton DP, Wald LL, Pitts J, Schmitt F (2003) Comparison of cardiac MRI on 1.5 and 3.0 Tesla clinical whole body systems. Invest Radiol 38:436–442CrossRef Hinton DP, Wald LL, Pitts J, Schmitt F (2003) Comparison of cardiac MRI on 1.5 and 3.0 Tesla clinical whole body systems. Invest Radiol 38:436–442CrossRef
22.
go back to reference Sosna J, Pedrosa I, Dewolf WC, Mahallati H, Lenkinski RE, Rofsky NM (2004) MR imaging of the prostate at 3 Tesla: comparison of an external phased-array coil to imaging with an endorectal coil at 1.5 Tesla. Acad Radiol 11:857–862CrossRef Sosna J, Pedrosa I, Dewolf WC, Mahallati H, Lenkinski RE, Rofsky NM (2004) MR imaging of the prostate at 3 Tesla: comparison of an external phased-array coil to imaging with an endorectal coil at 1.5 Tesla. Acad Radiol 11:857–862CrossRef
23.
go back to reference Uematsu H, Takahashi M, Dougherty L, Hatabu H (2004) High field body MR imaging—preliminary experiences. J Clin Imaging 28:159–162CrossRef Uematsu H, Takahashi M, Dougherty L, Hatabu H (2004) High field body MR imaging—preliminary experiences. J Clin Imaging 28:159–162CrossRef
24.
go back to reference Karlberg M, Annertz M, Magnusson M (2004) Acute vestibular neuritis visualized by 3-T magnetic resonance imaging with high-dose gadolinium. Arch Otolaryngol Head Neck Surg 130:229–232CrossRef Karlberg M, Annertz M, Magnusson M (2004) Acute vestibular neuritis visualized by 3-T magnetic resonance imaging with high-dose gadolinium. Arch Otolaryngol Head Neck Surg 130:229–232CrossRef
25.
go back to reference Huber ME, Kozerke S, Pruessmann KP, Smink J, Boesiger P (2004) Sensitivity-encoded coronary MRA at 3T. Magn Reson Med 52:221–227CrossRef Huber ME, Kozerke S, Pruessmann KP, Smink J, Boesiger P (2004) Sensitivity-encoded coronary MRA at 3T. Magn Reson Med 52:221–227CrossRef
26.
go back to reference Nayak KS, Cunningham CH, Santos JM, Pauly JM (2004) Real-time cardiac MRI at 3 Tesla. Magn Reson Med 51:655–660CrossRef Nayak KS, Cunningham CH, Santos JM, Pauly JM (2004) Real-time cardiac MRI at 3 Tesla. Magn Reson Med 51:655–660CrossRef
27.
go back to reference Fuetterer JJ, Scheenen TW, Huisman HJ, Klomp DW, van Dorsten FA, Hulsbergen-van de Kaa CA, Witjes JA, Heerschap A, Barentsz JO (2004) Initial experience of 3 Tesla endorectal coil magnetic resonance imaging and 1H-spectroscopic imaging of the prostate. Invest Radiol 39:671–680CrossRef Fuetterer JJ, Scheenen TW, Huisman HJ, Klomp DW, van Dorsten FA, Hulsbergen-van de Kaa CA, Witjes JA, Heerschap A, Barentsz JO (2004) Initial experience of 3 Tesla endorectal coil magnetic resonance imaging and 1H-spectroscopic imaging of the prostate. Invest Radiol 39:671–680CrossRef
28.
go back to reference Graf H, Schick F, Claussen CD, Seemann MD (2004) MR visualization of the inner ear structures: comparison of 1.5 Tesla and 3 Tesla images. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 176:17–20CrossRef Graf H, Schick F, Claussen CD, Seemann MD (2004) MR visualization of the inner ear structures: comparison of 1.5 Tesla and 3 Tesla images. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 176:17–20CrossRef
29.
go back to reference Baudendistel KT, Heverhagen JT, Knopp MV (2004) Clinical MR at 3 Tesla: current status. Radiologe 44:11–18CrossRefPubMed Baudendistel KT, Heverhagen JT, Knopp MV (2004) Clinical MR at 3 Tesla: current status. Radiologe 44:11–18CrossRefPubMed
30.
go back to reference Lenk S, Ludescher B, Martirosan P, Schick F, Claussen CD, Schlemmer HP (2004) 3.0 T high-resolution MR imaging of carpal ligaments and TFCC. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 176:664–667CrossRef Lenk S, Ludescher B, Martirosan P, Schick F, Claussen CD, Schlemmer HP (2004) 3.0 T high-resolution MR imaging of carpal ligaments and TFCC. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 176:664–667CrossRef
31.
go back to reference Bloch BN, Rofsky NM, Baroni RH, Marquis RP, Pedrosa I, Lenkinski RE (2004) 3 Tesla magnetic resonance imaging of the prostate with combined pelvic phased-array and endorectal coils; initial experience (1). Acad Radiol 11:863–877CrossRef Bloch BN, Rofsky NM, Baroni RH, Marquis RP, Pedrosa I, Lenkinski RE (2004) 3 Tesla magnetic resonance imaging of the prostate with combined pelvic phased-array and endorectal coils; initial experience (1). Acad Radiol 11:863–877CrossRef
32.
go back to reference Lutterbey G, Gieseke J, von Falkenhausen M, Morakkabati N, Schild H (2005) Lung MRI at 3.0T: a comparison of helical CT and high-field MRI in the detection of diffuse lung disease. Eur Radiol. DOI 10.1007/s00330-004-2548-1 Lutterbey G, Gieseke J, von Falkenhausen M, Morakkabati N, Schild H (2005) Lung MRI at 3.0T: a comparison of helical CT and high-field MRI in the detection of diffuse lung disease. Eur Radiol. DOI 10.​1007/​s00330-004-2548-1
33.
go back to reference Hoult DI (2000) Sensitivity and power deposition in a high-field imaging experiment. J Magn Reson Imag 12:46–67CrossRef Hoult DI (2000) Sensitivity and power deposition in a high-field imaging experiment. J Magn Reson Imag 12:46–67CrossRef
34.
go back to reference Michaeli S, Garwood M, Zhu XH, DelaBarre L, Andersen P, Adriany G, Merkle H, Ugurbil K, Chen W (2002) Proton T2 relaxation study of water, N-acetylaspartate, and creatine in human brain using Hahn and Carr–Purcell spin echoes at 4T and 7T. Magn Reson Med 47:629–633CrossRef Michaeli S, Garwood M, Zhu XH, DelaBarre L, Andersen P, Adriany G, Merkle H, Ugurbil K, Chen W (2002) Proton T2 relaxation study of water, N-acetylaspartate, and creatine in human brain using Hahn and Carr–Purcell spin echoes at 4T and 7T. Magn Reson Med 47:629–633CrossRef
35.
go back to reference Ethofer T, Mader I, Seeger U, Helms G, Erb M, Grodd W, Ludolph A, Klose U (2003) Magn Reson Med 50:1296–1301CrossRef Ethofer T, Mader I, Seeger U, Helms G, Erb M, Grodd W, Ludolph A, Klose U (2003) Magn Reson Med 50:1296–1301CrossRef
36.
go back to reference Bezelaire CMJ, Duhamel GD, Rofsky NM, Alsop DC (2004) MR imaging relaxation times of abdominal and pelvic tissues measured in vivo at 3.0 T: preliminary results. Radiology 230:652–659 Bezelaire CMJ, Duhamel GD, Rofsky NM, Alsop DC (2004) MR imaging relaxation times of abdominal and pelvic tissues measured in vivo at 3.0 T: preliminary results. Radiology 230:652–659
37.
go back to reference Hennig J, Scheffler K (2001) Hyperechoes. Magn Reson Med 46:6–12 Hennig J, Scheffler K (2001) Hyperechoes. Magn Reson Med 46:6–12
38.
go back to reference Carr JC, Simonetti O, Bundy J, Li D, Pereles S, Finn JP (2001) Cine MR angiography of the heart with segmented true fast imaging with steady-state precession. Radiology 219:828–834PubMed Carr JC, Simonetti O, Bundy J, Li D, Pereles S, Finn JP (2001) Cine MR angiography of the heart with segmented true fast imaging with steady-state precession. Radiology 219:828–834PubMed
39.
go back to reference Posse S, Kemna LJ, Elghahwagi B, Wiese S, Kiselev VG (2001) Effect of graded hypo- and hypercapnia on fMRI contrast in visual cortex: quantification of T2* changes by multiecho EPI. Magn Reson Med 46:264–271CrossRef Posse S, Kemna LJ, Elghahwagi B, Wiese S, Kiselev VG (2001) Effect of graded hypo- and hypercapnia on fMRI contrast in visual cortex: quantification of T2* changes by multiecho EPI. Magn Reson Med 46:264–271CrossRef
40.
go back to reference Reimer P, Muller M, Marx C, Wiedermann D, Muller R, Rummeny EJ, Ebert W, Shamsi K, Peters PE (1998) T1 effects of a bolus-injectable superparamagnetic iron oxide, SH U 555 A: dependence on field strength and plasma concentration—preliminary clinical experience with dynamic T1-weighted MR imaging. Radiology 209:831–836PubMed Reimer P, Muller M, Marx C, Wiedermann D, Muller R, Rummeny EJ, Ebert W, Shamsi K, Peters PE (1998) T1 effects of a bolus-injectable superparamagnetic iron oxide, SH U 555 A: dependence on field strength and plasma concentration—preliminary clinical experience with dynamic T1-weighted MR imaging. Radiology 209:831–836PubMed
41.
go back to reference Huppertz A, Rohrer M (2004) Gadobutrol, a highly concentrated MR imaging contrast agent: its physicochemical characteristics and the basis for its use in contrast enhanced MR angiography and perfusion imaging. Eur Radiol Suppl 14(Suppl 5):M12–M18CrossRef Huppertz A, Rohrer M (2004) Gadobutrol, a highly concentrated MR imaging contrast agent: its physicochemical characteristics and the basis for its use in contrast enhanced MR angiography and perfusion imaging. Eur Radiol Suppl 14(Suppl 5):M12–M18CrossRef
42.
go back to reference Kim SG (1995) Quantification of relative cerebral blood flow change by flow-sensitive alternating inversion recovery (FAIR) technique: application to functional mapping. Magn Reson Med 37:293–301 Kim SG (1995) Quantification of relative cerebral blood flow change by flow-sensitive alternating inversion recovery (FAIR) technique: application to functional mapping. Magn Reson Med 37:293–301
43.
go back to reference Calamante F, Thomas DL, Pell GS, Wiersma J, Turner R (1999) Measuring cerebral blood flow using magnetic resonance imaging techniques. J Cereb Blood Flow Metab 19:701–735CrossRef Calamante F, Thomas DL, Pell GS, Wiersma J, Turner R (1999) Measuring cerebral blood flow using magnetic resonance imaging techniques. J Cereb Blood Flow Metab 19:701–735CrossRef
44.
go back to reference Martirosian P, Klose U, Mader I, Schick F (2004) FAIR True-FISP perfusion imaging of the kidneys. Magn Reson Med 51:353–361CrossRef Martirosian P, Klose U, Mader I, Schick F (2004) FAIR True-FISP perfusion imaging of the kidneys. Magn Reson Med 51:353–361CrossRef
45.
go back to reference Kurhanewicz J, Swanson MG, Nelson SJ, Vigneron DB (2002) Combined magnetic resonance imaging and spectroscopic imaging approach to molecular imaging of prostate cancer. J Magn Reson Imaging 16:451–463CrossRef Kurhanewicz J, Swanson MG, Nelson SJ, Vigneron DB (2002) Combined magnetic resonance imaging and spectroscopic imaging approach to molecular imaging of prostate cancer. J Magn Reson Imaging 16:451–463CrossRef
46.
go back to reference Hasumi M, Suzuki K, Taketomi A, Matsui H, Yamamoto T, Ito K, Kurokawa K, Aoki J, Endo K, Yamanaka H (2003) The combination of multi-voxel MR spectroscopy with MR imaging improve the diagnostic accuracy for localization of prostate cancer. Anticancer Res 23:4223–4227 Hasumi M, Suzuki K, Taketomi A, Matsui H, Yamamoto T, Ito K, Kurokawa K, Aoki J, Endo K, Yamanaka H (2003) The combination of multi-voxel MR spectroscopy with MR imaging improve the diagnostic accuracy for localization of prostate cancer. Anticancer Res 23:4223–4227
47.
go back to reference Schmitt M, Feiweier T, Horger W, Krueger G, Schoen L, Lazar R, Kiefer B (2004) Improved uniformity of RF-distribution in clinical whole body—imaging at 3T by means of dielectric pads. In: Proceedings of the 12th annual meeting of ISMRM, p. 197 Schmitt M, Feiweier T, Horger W, Krueger G, Schoen L, Lazar R, Kiefer B (2004) Improved uniformity of RF-distribution in clinical whole body—imaging at 3T by means of dielectric pads. In: Proceedings of the 12th annual meeting of ISMRM, p. 197
48.
go back to reference Vaughan JT, Adriany G, Snyder CJ, Tian J, Thiel T, Bolinger L, Liu H, DelaBarre L, Ugurbil K (2004) Efficient high-frequency body coil for high-field MRI. Magn Reson Med 52:851–859CrossRef Vaughan JT, Adriany G, Snyder CJ, Tian J, Thiel T, Bolinger L, Liu H, DelaBarre L, Ugurbil K (2004) Efficient high-frequency body coil for high-field MRI. Magn Reson Med 52:851–859CrossRef
49.
go back to reference Nitz WR, Oppelt A, Renz W, Manke C, Lenhart M, Link J (2001) On the heating of linear conductive structures as guide wires and catheters in interventional MRI. J Magn Reson Imaging 13:105–114 Nitz WR, Oppelt A, Renz W, Manke C, Lenhart M, Link J (2001) On the heating of linear conductive structures as guide wires and catheters in interventional MRI. J Magn Reson Imaging 13:105–114
50.
go back to reference Kangarlu A, Shellock FG, Chakeres DW (2003) 8.0-Tesla human MR system: temperature changes associated with radiofrequency-induced heating of a head phantom. J Magn Reson Imaging 17:220–226CrossRef Kangarlu A, Shellock FG, Chakeres DW (2003) 8.0-Tesla human MR system: temperature changes associated with radiofrequency-induced heating of a head phantom. J Magn Reson Imaging 17:220–226CrossRef
51.
go back to reference Graf H, Lauer UA, Berger A, Schick F (2005) RF interactions with metallic instruments and implants get more prominent at 3 Tesla—an in-vitro study. Magn Reson Imaging (accepted for publication) Graf H, Lauer UA, Berger A, Schick F (2005) RF interactions with metallic instruments and implants get more prominent at 3 Tesla—an in-vitro study. Magn Reson Imaging (accepted for publication)
52.
go back to reference Yacoub E, Shmuel A, Pfeuffer J, Van De Moortele PF, Adriany G, Andersen P, Vaughan JT, Merkle H, Ugurbil K, Hu X (2001) Imaging brain function in humans at 7 Tesla. Magn Reson Med 45:588–594CrossRef Yacoub E, Shmuel A, Pfeuffer J, Van De Moortele PF, Adriany G, Andersen P, Vaughan JT, Merkle H, Ugurbil K, Hu X (2001) Imaging brain function in humans at 7 Tesla. Magn Reson Med 45:588–594CrossRef
53.
go back to reference Hattori N, Abe K, Sakoda S, Sawada T (2001) Proton MR spectroscopic study at 3 Tesla glutamate/glutamine in Alzheimer’s disease. NeuroReport 13:183–186 Hattori N, Abe K, Sakoda S, Sawada T (2001) Proton MR spectroscopic study at 3 Tesla glutamate/glutamine in Alzheimer’s disease. NeuroReport 13:183–186
54.
go back to reference Barker PB, Hearshen DO, Boska MD (2001) Single-voxel proton MRS of the human brain at 1.5T and 3.0T. Magn Reson Med 45:765–769CrossRef Barker PB, Hearshen DO, Boska MD (2001) Single-voxel proton MRS of the human brain at 1.5T and 3.0T. Magn Reson Med 45:765–769CrossRef
55.
go back to reference Schubert F, Gallinat J, Seifert F, Rinneberg H (2004) Glutamate concentrations in human brain using single voxel proton magnetic resonance spectroscopy at 3 Tesla. NeuroImage 21:1762–1771CrossRef Schubert F, Gallinat J, Seifert F, Rinneberg H (2004) Glutamate concentrations in human brain using single voxel proton magnetic resonance spectroscopy at 3 Tesla. NeuroImage 21:1762–1771CrossRef
56.
go back to reference Kim DS, Garwood M (2003) High-field magnetic resonance techniques for brain research. Curr Opin Neurobiol 13:612–619CrossRef Kim DS, Garwood M (2003) High-field magnetic resonance techniques for brain research. Curr Opin Neurobiol 13:612–619CrossRef
57.
go back to reference Hennig J, Speck O, Koch MA, Weiller C (2003) Functional magnetic resonance imaging: a review of methodological aspects and clinical applications. J Magn Reson Imaging 18:1–15CrossRef Hennig J, Speck O, Koch MA, Weiller C (2003) Functional magnetic resonance imaging: a review of methodological aspects and clinical applications. J Magn Reson Imaging 18:1–15CrossRef
58.
go back to reference Frayne R, Goodyear BG, Dickhoff P, Lauzon ML, Sevick RJ (2003) Magnetic resonance imaging at 3.0 Tesla: challenges and advantages in clinical neurological imaging. Invest Radiol 38:385–402CrossRef Frayne R, Goodyear BG, Dickhoff P, Lauzon ML, Sevick RJ (2003) Magnetic resonance imaging at 3.0 Tesla: challenges and advantages in clinical neurological imaging. Invest Radiol 38:385–402CrossRef
59.
go back to reference Schwindt W, Kugel H, Bachmann R, Kloska S, Allkemper T, Maintz D, Pfleiderer B, Tombach B, Heindel W (2003) Magnetic resonance imaging protocols for examination of the neurocranium at 3T. Eur Radiol 13:2170–2179CrossRefPubMed Schwindt W, Kugel H, Bachmann R, Kloska S, Allkemper T, Maintz D, Pfleiderer B, Tombach B, Heindel W (2003) Magnetic resonance imaging protocols for examination of the neurocranium at 3T. Eur Radiol 13:2170–2179CrossRefPubMed
60.
go back to reference Allkemper T, Schwindt W, Maintz D, Heindel W, Tombach B (2004) Sensitivity of T2-weighted FSE sequences towards physiological iron depositions in normal brains at 1.5 and 3.0 T. Eur Radiol 14:1000–1004CrossRefPubMed Allkemper T, Schwindt W, Maintz D, Heindel W, Tombach B (2004) Sensitivity of T2-weighted FSE sequences towards physiological iron depositions in normal brains at 1.5 and 3.0 T. Eur Radiol 14:1000–1004CrossRefPubMed
Metadata
Title
Whole-body MRI at high field: technical limits and clinical potential
Author
Fritz Schick
Publication date
01-05-2005
Publisher
Springer-Verlag
Published in
European Radiology / Issue 5/2005
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-005-2678-0

Other articles of this Issue 5/2005

European Radiology 5/2005 Go to the issue

Letter to the Editor

Letter to the editor