Published in: European Radiology 5/2005

01-05-2005 | Physics

Study on motion artifacts in coronary arteries with an anthropomorphic moving heart phantom on an ECG-gated multidetector computed tomography unit

Authors: Marcel J. W. Greuter, Joost Dorgelo, Wim G. J. Tukker, Matthijs Oudkerk

Published in: European Radiology | Issue 5/2005

Login to get access

Abstract

Acquisition time plays a key role in the quality of cardiac multidetector computed tomography (MDCT) and is directly related to the rotation time of the scanner. The purpose of this study is to examine the influence of heart rate and a multisector reconstruction algorithm on the image quality of coronary arteries of an anthropomorphic adjustable moving heart phantom on an ECG-gated MDCT unit. The heart phantom and a coronary artery phantom were used on a MDCT unit with a rotation time of 500 ms. The movement of the heart was determined by analysis of the images taken at different phases. The results indicate that the movement of the coronary arteries on the heart phantom is comparable to that in a clinical setting. The influence of the heart rate on image quality and artifacts was determined by analysis of several heart rates between 40 and 80 bpm where the movement of the heart was synchronized using a retrospective ECG-gated acquisition protocol. The resulting reformatted volume rendering images of the moving heart and the coronary arteries were qualitatively compared as a result of the heart rate. The evaluation was performed on three independent series by two independent radiologists for the image quality of the coronary arteries and the presence of artifacts. The evaluation shows that at heart rates above 50 bpm the influence of motion artifacts in the coronary arteries becomes apparent. In addition the influence of a dedicated multisector reconstruction technique on image quality was determined. The results show that the image quality of the coronary arteries is not only related to the heart rate and that the influence of the multisector reconstruction technique becomes significant above 70 bpm. Therefore, this study proves that from the actual acquisition time per heart cycle one cannot determine an actual acquisition time, but only a mathematical acquisition time.
Literature
1.
go back to reference Giesler T, Baum U, Ropers D et al (2002) Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT. Am J Roentgenol 179:911–916 Giesler T, Baum U, Ropers D et al (2002) Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT. Am J Roentgenol 179:911–916
2.
go back to reference Hong C, Becker CR, Huber A et al (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220:712–717PubMed Hong C, Becker CR, Huber A et al (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220:712–717PubMed
3.
go back to reference Kopp AF, Schroeder S, Kuettner A et al (2001) Coronary arteries: retrospectively ECG-gated multi-detector row CT angiography with selective optimization of the image reconstruction window. Radiology 221:683–688PubMed Kopp AF, Schroeder S, Kuettner A et al (2001) Coronary arteries: retrospectively ECG-gated multi-detector row CT angiography with selective optimization of the image reconstruction window. Radiology 221:683–688PubMed
4.
go back to reference Rumberger J, Kaufman L (2003) A Rosetta stone for coronary calcium risk stratification. Am J Roentgenol 181:743–748 Rumberger J, Kaufman L (2003) A Rosetta stone for coronary calcium risk stratification. Am J Roentgenol 181:743–748
5.
go back to reference Brown SJ, Hayball MP, Coulden RAR (2000) Impact of motion artefact on the measurement of coronary calcium score. Br J Radiol 73:956–962 Brown SJ, Hayball MP, Coulden RAR (2000) Impact of motion artefact on the measurement of coronary calcium score. Br J Radiol 73:956–962
6.
go back to reference Ulzheimer S, Kalender WA (2003) Assessment of calcium scoring performance in cardiac computed tomography. Eur Radiol 13:484–497PubMed Ulzheimer S, Kalender WA (2003) Assessment of calcium scoring performance in cardiac computed tomography. Eur Radiol 13:484–497PubMed
7.
go back to reference Hopper KD, Strollo DC, Mauger DT (2002) Comparison of electron-beam and ungated helical CT in detecting coronary arterial calcification by using a working heart phantom and artificial coronary arteries. Radiology 222:474–482 Hopper KD, Strollo DC, Mauger DT (2002) Comparison of electron-beam and ungated helical CT in detecting coronary arterial calcification by using a working heart phantom and artificial coronary arteries. Radiology 222:474–482
8.
go back to reference McCollough CH, Bruesewitz MR, Daly TR, Zink FE (2000) Motion artifacts in subsecond conventional CT and electron-beam CT: pictorial demonstration of temporal resolution. Radiographics 20:1675–1681 McCollough CH, Bruesewitz MR, Daly TR, Zink FE (2000) Motion artifacts in subsecond conventional CT and electron-beam CT: pictorial demonstration of temporal resolution. Radiographics 20:1675–1681
9.
go back to reference Wicky S, Rosol M, Hamberg LM et al (2002) Evaluation of retrospective multisector and half scan ECG-gated multidetector cardiac CT protocols with moving phantoms. J Comput Assist Tomogr 26(5):768–776 Wicky S, Rosol M, Hamberg LM et al (2002) Evaluation of retrospective multisector and half scan ECG-gated multidetector cardiac CT protocols with moving phantoms. J Comput Assist Tomogr 26(5):768–776
10.
go back to reference Wicky S, Rosol M, Hoffmann U, Graziano M, Yucel KE, Brady TJ (2003) Comparative study with a moving heart phantom of the impact of temporal resolution on image quality with two multidetector electrocardiography-gated computed tomography units. J Comput Assist Tomogr 27(3):392–398 Wicky S, Rosol M, Hoffmann U, Graziano M, Yucel KE, Brady TJ (2003) Comparative study with a moving heart phantom of the impact of temporal resolution on image quality with two multidetector electrocardiography-gated computed tomography units. J Comput Assist Tomogr 27(3):392–398
11.
go back to reference Parker DL (1982) Optimal short scan convolution reconstruction for fanbeam CT. Med Phys 9(2):254–257 Parker DL (1982) Optimal short scan convolution reconstruction for fanbeam CT. Med Phys 9(2):254–257
12.
go back to reference Flohr T, Ohnesorge B (2001) Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart. J Comput Assist Tomogr 25(6):907–923CrossRefPubMed Flohr T, Ohnesorge B (2001) Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart. J Comput Assist Tomogr 25(6):907–923CrossRefPubMed
13.
go back to reference Halliburton SS, Stillman AE, Flohr T et al (2003) Do segmented reconstruction algorithms for cardiac multi-slice computed tomography improve image quality? Herz 28:20–31 Halliburton SS, Stillman AE, Flohr T et al (2003) Do segmented reconstruction algorithms for cardiac multi-slice computed tomography improve image quality? Herz 28:20–31
14.
go back to reference Bruder H, Schaller S, Ohnesorge B, Mertelmeier T (1999) High temporal resolution volume heart imaging with multirow computed tomography. SPIE Med Imaging 3661:420–432 Bruder H, Schaller S, Ohnesorge B, Mertelmeier T (1999) High temporal resolution volume heart imaging with multirow computed tomography. SPIE Med Imaging 3661:420–432
15.
go back to reference Flohr T, Bruder H, Stierstorfer K, Simon J, Schaller B, Ohnesorge B (2002) New technical developments in multislice CT. Part 2. Sub-millimeter 16-slice scanning and increased gantry rotation speed for cadiac imaging. RoFo Fortschr Geb Rontgenstrahlen Neuen Bildgeb Verfahr 174(8):1022–1027CrossRef Flohr T, Bruder H, Stierstorfer K, Simon J, Schaller B, Ohnesorge B (2002) New technical developments in multislice CT. Part 2. Sub-millimeter 16-slice scanning and increased gantry rotation speed for cadiac imaging. RoFo Fortschr Geb Rontgenstrahlen Neuen Bildgeb Verfahr 174(8):1022–1027CrossRef
16.
go back to reference Mao S, Lu B, Takasu J, Oudiz RJ, Budoff MJ (2003) Measurement of the RT interval on ECG records during electron-beam CT. Acad Radiol 10:638–643 Mao S, Lu B, Takasu J, Oudiz RJ, Budoff MJ (2003) Measurement of the RT interval on ECG records during electron-beam CT. Acad Radiol 10:638–643
17.
go back to reference Herzog C, Abolmaali N, Balzer JO et al (2002) Heart-rate-adapted image reconstruction in multidetector-row cardiac CT: influence of physiological and technical prerequisite on image quality. Eur Radiol 12(11):2670–2678PubMed Herzog C, Abolmaali N, Balzer JO et al (2002) Heart-rate-adapted image reconstruction in multidetector-row cardiac CT: influence of physiological and technical prerequisite on image quality. Eur Radiol 12(11):2670–2678PubMed
18.
go back to reference Mao S, Budoff MJ, Bin L, Liu SCK (2001) Optimal ECG trigger point in electron-beam CT studies. Acad Radiol 8:1107–1115 Mao S, Budoff MJ, Bin L, Liu SCK (2001) Optimal ECG trigger point in electron-beam CT studies. Acad Radiol 8:1107–1115
19.
go back to reference Ohnesorge B, Flohr T, Becker C et al (2000) Cardiac imaging by means of electrocardiographically gated multisection spiral CT: initial experience. Radiology 217:564–571PubMed Ohnesorge B, Flohr T, Becker C et al (2000) Cardiac imaging by means of electrocardiographically gated multisection spiral CT: initial experience. Radiology 217:564–571PubMed
20.
go back to reference Achenbach S, Ropers D, Holle J, Muschiol G, Daniel WG, Moshage W (2000) In-plane coronary arterial motion velocity: measurement with electron-beam CT. Radiology 216:457–463PubMed Achenbach S, Ropers D, Holle J, Muschiol G, Daniel WG, Moshage W (2000) In-plane coronary arterial motion velocity: measurement with electron-beam CT. Radiology 216:457–463PubMed
21.
go back to reference Lu B, Mao SS, Zhuang N et al (2001) Coronary artery motion during the cardiac cycle and optimal ECG triggering for coronary artery imaging. Invest Radiol 36(5):250–256 Lu B, Mao SS, Zhuang N et al (2001) Coronary artery motion during the cardiac cycle and optimal ECG triggering for coronary artery imaging. Invest Radiol 36(5):250–256
22.
go back to reference Hofman MBM, Wickline SA, Lorenz CH (1998) Quantification of in-plane motion of the coronary arteries during the cardiac cycle: implications for acquisition window duration for MR flow quantification. J Magn Reson Imaging 8:568–576 Hofman MBM, Wickline SA, Lorenz CH (1998) Quantification of in-plane motion of the coronary arteries during the cardiac cycle: implications for acquisition window duration for MR flow quantification. J Magn Reson Imaging 8:568–576
23.
go back to reference Dodge JT, Brown BG, Bolson EL, Dodge HT (1992) Lumen diameter of normal human coronary arteries. Influence of age, sex, anatomic variation, and left ventricular hypertrophy or dilation. Circulation 86:232–246 Dodge JT, Brown BG, Bolson EL, Dodge HT (1992) Lumen diameter of normal human coronary arteries. Influence of age, sex, anatomic variation, and left ventricular hypertrophy or dilation. Circulation 86:232–246
24.
go back to reference Schroeder S, Kopp AF, Kuettner A et al (2002) Influence of heart rate on vessel visibility in noninvasive coronary angiography using new multislice computed tomography; experience in 94 patients. Clin Imaging 26:106–111CrossRefPubMed Schroeder S, Kopp AF, Kuettner A et al (2002) Influence of heart rate on vessel visibility in noninvasive coronary angiography using new multislice computed tomography; experience in 94 patients. Clin Imaging 26:106–111CrossRefPubMed
25.
go back to reference Jakobs TF, Wintersperger BJ, Herzog P, Flohr T, Suess C, Knez A, Reiser MF, Becker CR (2003) Ultra-low-dose coronary artery calcium screening using multislice CT with retrospective ECG gating. Eur Radiol 13:1923–1930 Jakobs TF, Wintersperger BJ, Herzog P, Flohr T, Suess C, Knez A, Reiser MF, Becker CR (2003) Ultra-low-dose coronary artery calcium screening using multislice CT with retrospective ECG gating. Eur Radiol 13:1923–1930
26.
go back to reference Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PMT, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054 Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PMT, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054
27.
go back to reference Herzog C, Britten M, Balzer JO, Mack MG, Zangos S, Ackermann H, Schaechinger V, Schaller S, Flohr T, Vogl TJ (2004) Multidetector-row cardiac CT: diagnostic value of calcium scoring and CT coronary angiography in patients with symptomatic, but atypical, chest pain. Eur Radiol 14:169–177CrossRefPubMed Herzog C, Britten M, Balzer JO, Mack MG, Zangos S, Ackermann H, Schaechinger V, Schaller S, Flohr T, Vogl TJ (2004) Multidetector-row cardiac CT: diagnostic value of calcium scoring and CT coronary angiography in patients with symptomatic, but atypical, chest pain. Eur Radiol 14:169–177CrossRefPubMed
Metadata
Title
Study on motion artifacts in coronary arteries with an anthropomorphic moving heart phantom on an ECG-gated multidetector computed tomography unit
Authors
Marcel J. W. Greuter
Joost Dorgelo
Wim G. J. Tukker
Matthijs Oudkerk
Publication date
01-05-2005
Publisher
Springer-Verlag
Published in
European Radiology / Issue 5/2005
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-004-2602-z

Other articles of this Issue 5/2005

European Radiology 5/2005 Go to the issue