01-09-2007 | Review / State of the Art

Long-Term Retrievability of IVC Filters: Should We Abandon Permanent Devices?

Authors: V. Berczi, J. R. Bottomley, S. M. Thomas, S. Taneja, P. A. Gaines, T. J. Cleveland

Published in: CardioVascular and Interventional Radiology | Issue 5/2007

Login to get access

Abstract

Thromboembolic disease produces a considerable disease burden, with death from pulmonary embolism in the UK alone estimated at 30,000–40,000 per year. Whilst it is unproven whether filters actually improve longevity, the morbidity and mortality associated with thromboembolic disease in the presence of contraindications to anticoagulation is high. Thus complications associated with filter insertion, and whilst they remain in situ, must be balanced against the alternatives. Permanent filters remain in situ for the remainder of the patient’s life and any complications from the filters are of significant concern. Filters that are not permanent are therefore attractive in these circumstances. Retrievable filters, to avoid or decrease long-term filter complications, appear to be a significant advance in the prevention of pulmonary embolism. In this review, we discuss the safety and effectiveness of both permanent and retrievable filters as well as the retrievability of retrievable inferior vena cava (IVC) filters, to explore whether the use of permanent IVC filters can be abandoned in favor of retrievable filters. Currently four types of retrievable filters are available: the Recovery filter (Bard Peripheral Vascular, Tempe, AZ, USA), the Günther Tulip filter (Cook, Bloomington, IN, USA), the OptEase Filter (Cordis, Roden, The Netherlands), and the ALN filter (ALN Implants Chirurgicaux, Ghisonaccia, France). Efficacy and safety data for retrievable filters are as yet based on small series, with a total number of fewer than 1,000 insertions, and follow-up is mostly short term. Current long-term data are poor and insufficient to warrant the long-term implantation of these devices into humans. The case of fractured wire from a Recovery filter that migrated to the heart causing pericardial tamponade requiring open heart surgery is a reminder that any new endovascular device remaining in situ in the long term may produce unexpected problems. We should also bear in mind that the data on permanent filters are much more robust, with reports on over 9,500 cases with follow-up of up to 8 years. The original implantation time of 10–14 days has been extended to more than 100 days as the mean implantation time with some of the filter types. Follow-up (preferably prospective) is necessary for all retrievable filters, whether or not they are retrieved. Until these data become available we should restrict ourselves to the present indications of permanent and retrievable filters. If long-term follow-up data on larger numbers of cases confirm the initial data that retrievable filters are as safe and effective as permanent filters, the use of the retrievable filters is likely to expand.
Literature
1.
go back to reference Sing RF, Rogers FB, Novitsky YW, Heniford BT (2005) Optional vena cava filters for patients with high thromboembolic risk: Questions to be answered. Surg Innov 12:195–202PubMedCrossRef Sing RF, Rogers FB, Novitsky YW, Heniford BT (2005) Optional vena cava filters for patients with high thromboembolic risk: Questions to be answered. Surg Innov 12:195–202PubMedCrossRef
2.
go back to reference Hann CL, Streiff MB (2005) The role of vena caval filters in the management of venous thromboembolism. Blood Rev 19:179–202PubMedCrossRef Hann CL, Streiff MB (2005) The role of vena caval filters in the management of venous thromboembolism. Blood Rev 19:179–202PubMedCrossRef
3.
go back to reference Ku GH, Billett HH (2005) Long lives, short indications. The case for removable inferior cava filters. Thromb Haemost 93:17–22PubMed Ku GH, Billett HH (2005) Long lives, short indications. The case for removable inferior cava filters. Thromb Haemost 93:17–22PubMed
4.
go back to reference Robertson F, Platts A (2004) IVC filters, venous thrombolysis and thrombectomy. In: Wyatt MG, Watkinson AF (ed) Endovascular intervention: Current controversies. TFM Publishing, pp 285–298 Robertson F, Platts A (2004) IVC filters, venous thrombolysis and thrombectomy. In: Wyatt MG, Watkinson AF (ed) Endovascular intervention: Current controversies. TFM Publishing, pp 285–298
5.
go back to reference Chiou AC (2005) IVC filter retrieval: Long-term data on today’s filter are needed. Endovascular Today 4:81–84 Chiou AC (2005) IVC filter retrieval: Long-term data on today’s filter are needed. Endovascular Today 4:81–84
6.
go back to reference Imberti D, Bianchi M, Farina A, Siragusa S, Silingardi M, Ageno W (2005) Clinical experience with retrievable vena cava filters: Results of a prospective observational multicenter study. J Thromb Haemost 3:1370–1375PubMedCrossRef Imberti D, Bianchi M, Farina A, Siragusa S, Silingardi M, Ageno W (2005) Clinical experience with retrievable vena cava filters: Results of a prospective observational multicenter study. J Thromb Haemost 3:1370–1375PubMedCrossRef
7.
go back to reference Brountzos EN, Kaufman JA, Venbrux AC, Brown PR, Harry J, Kinst TF, Kleshinski S, Ravenscroft AC (2003) A new optional vena cava filter: Retrieval at 12 weeks in an animal model. J Vasc Interv Radiol 14:763–772PubMed Brountzos EN, Kaufman JA, Venbrux AC, Brown PR, Harry J, Kinst TF, Kleshinski S, Ravenscroft AC (2003) A new optional vena cava filter: Retrieval at 12 weeks in an animal model. J Vasc Interv Radiol 14:763–772PubMed
8.
go back to reference Decousus H, Leizorovicz A, Parent F, Page Y, Tardy B, Girard P, Laporte S, Faivre R, Charbonnier B, Barral FG, Huet Y, Simonneau G (1998) A clinical trial of vena caval filters in the prevention of PE in patients with proximal deep-vein thrombosis. Prevention du Risque d’Embolie Pulmonaire par Interruption Cave Study Group. N Engl J Med 338:409–415PubMedCrossRef Decousus H, Leizorovicz A, Parent F, Page Y, Tardy B, Girard P, Laporte S, Faivre R, Charbonnier B, Barral FG, Huet Y, Simonneau G (1998) A clinical trial of vena caval filters in the prevention of PE in patients with proximal deep-vein thrombosis. Prevention du Risque d’Embolie Pulmonaire par Interruption Cave Study Group. N Engl J Med 338:409–415PubMedCrossRef
9.
go back to reference PREPIC Study Group (2005) Eight-year follow-up of patients with permanent vena cava filters in the prevention of PE: The PREPIC (Prevention du Risque d’Embolie Pulmonaire par Interruption Cave) randomized study. Circulation 112:416–422CrossRef PREPIC Study Group (2005) Eight-year follow-up of patients with permanent vena cava filters in the prevention of PE: The PREPIC (Prevention du Risque d’Embolie Pulmonaire par Interruption Cave) randomized study. Circulation 112:416–422CrossRef
10.
go back to reference Lorch H, Welger D, Wagner V, Hillner B, Strecker EP, Herrmann H, Voshage G, Zur C, Schwarzbach C, Schroder J, Gullotta U, Pleissner J, Huttner S, Siering U, Marcklin C, Chavan A, Glaser F, Apitzsch DE, Moubayed K, Leonhardi J, Schuchard UM, Weiss HD, Zwaan M (2000) Current practice of temporary vena cava filter insertion: A multicenter registry. J Vasc Interv Radiol. 11:83–88PubMedCrossRef Lorch H, Welger D, Wagner V, Hillner B, Strecker EP, Herrmann H, Voshage G, Zur C, Schwarzbach C, Schroder J, Gullotta U, Pleissner J, Huttner S, Siering U, Marcklin C, Chavan A, Glaser F, Apitzsch DE, Moubayed K, Leonhardi J, Schuchard UM, Weiss HD, Zwaan M (2000) Current practice of temporary vena cava filter insertion: A multicenter registry. J Vasc Interv Radiol. 11:83–88PubMedCrossRef
11.
go back to reference Poletti PA, Becker CD, Prina L, Ruijs P, Bounameaux H, Didier D, Schneider PA, Terrier F (1998) Long-term results of the Simon nitinol inferior vena cava filter. Eur Radiol 8:289–294PubMedCrossRef Poletti PA, Becker CD, Prina L, Ruijs P, Bounameaux H, Didier D, Schneider PA, Terrier F (1998) Long-term results of the Simon nitinol inferior vena cava filter. Eur Radiol 8:289–294PubMedCrossRef
12.
go back to reference Asch MR (2002) Initial experience in humans with a new retrievable inferior vena cava filter. Radiology 225:835–844PubMedCrossRef Asch MR (2002) Initial experience in humans with a new retrievable inferior vena cava filter. Radiology 225:835–844PubMedCrossRef
13.
go back to reference Grande WJ, Trerotola SO, Reilly PM, Clark TW, Soulen MC, Patel A, Shlansky-Goldberg RD, Tuite CM, Solomon JA, Mondschein JI, Fitzpatrick MK, Stavropoulos SW (2005) Experience with the recovery filter as a retrievable inferior vena cava filter. J Vasc Interv Radiol 16:1189–1193PubMed Grande WJ, Trerotola SO, Reilly PM, Clark TW, Soulen MC, Patel A, Shlansky-Goldberg RD, Tuite CM, Solomon JA, Mondschein JI, Fitzpatrick MK, Stavropoulos SW (2005) Experience with the recovery filter as a retrievable inferior vena cava filter. J Vasc Interv Radiol 16:1189–1193PubMed
14.
go back to reference Bottomley J, Bérczi V, Cleveland TJ, Gaines PA, Thomas SM (2006) Initial clinical experience at a single tertiary centre with the “recovery” vena caval filter. Cardiovasc Intervent Radiol 29:497 (abstract) Bottomley J, Bérczi V, Cleveland TJ, Gaines PA, Thomas SM (2006) Initial clinical experience at a single tertiary centre with the “recovery” vena caval filter. Cardiovasc Intervent Radiol 29:497 (abstract)
15.
go back to reference Kalva SP, Athanasoulis CA, Fan CM, Curvelo M, Geller SC, Greenfield AJ, Waltman AC, Wicky S (2006) “Recovery” vena cava filter: Experience in 96 patients. Cardiovasc Intervent Radiol 29:559–564PubMedCrossRef Kalva SP, Athanasoulis CA, Fan CM, Curvelo M, Geller SC, Greenfield AJ, Waltman AC, Wicky S (2006) “Recovery” vena cava filter: Experience in 96 patients. Cardiovasc Intervent Radiol 29:559–564PubMedCrossRef
16.
go back to reference Saeed I, Garcia M, McNicholas K (2006) Right ventricular migration of a recovery IVC filter’s fractured wire with subsequent pericardial tamponade. Cardiovasc Intervent Radiol 29:685–686PubMedCrossRef Saeed I, Garcia M, McNicholas K (2006) Right ventricular migration of a recovery IVC filter’s fractured wire with subsequent pericardial tamponade. Cardiovasc Intervent Radiol 29:685–686PubMedCrossRef
17.
go back to reference Hoppe H, Nutting CW, Smouse HR, Vesely TM, Pohl C, Bettmann MA, Kaufman JA (2006) Günther Tulip filter retrievability multicenter study including CT follow-up: Final report. J Vasc Interv Radiol 17:1017–1023PubMedCrossRef Hoppe H, Nutting CW, Smouse HR, Vesely TM, Pohl C, Bettmann MA, Kaufman JA (2006) Günther Tulip filter retrievability multicenter study including CT follow-up: Final report. J Vasc Interv Radiol 17:1017–1023PubMedCrossRef
18.
go back to reference De Gregorio MA, Gamboa P, Bonilla DL, Sanchez M, Higuera MT, Medrano J, Mainar A, Lostale F, Laborda A (2006) Retrieval of Günther Tulip optional vena cava filters 30 days after implantation: A prospective clinical study. J Vasc Interv Radiol 17:1781–1789PubMed De Gregorio MA, Gamboa P, Bonilla DL, Sanchez M, Higuera MT, Medrano J, Mainar A, Lostale F, Laborda A (2006) Retrieval of Günther Tulip optional vena cava filters 30 days after implantation: A prospective clinical study. J Vasc Interv Radiol 17:1781–1789PubMed
19.
go back to reference Looby S, Given MF, Geoghegan T, McErlean A, Lee MJ (2007) Günther tulip retrievable inferior vena caval filters: Indications, efficacy, retrieval, and complications. Cardiovasc Intervent Radiol 30:59–65PubMedCrossRef Looby S, Given MF, Geoghegan T, McErlean A, Lee MJ (2007) Günther tulip retrievable inferior vena caval filters: Indications, efficacy, retrieval, and complications. Cardiovasc Intervent Radiol 30:59–65PubMedCrossRef
20.
go back to reference Oliva VL, Szatmari F, Giroux MF, Flemming BK, Cohen SA, Soulez G (2006) The Jonas study: Evaluation of the retrievability of the Cordis OptEase inferior vena cava filter. J Vasc Interv Radiol. 16:1439–1445 Oliva VL, Szatmari F, Giroux MF, Flemming BK, Cohen SA, Soulez G (2006) The Jonas study: Evaluation of the retrievability of the Cordis OptEase inferior vena cava filter. J Vasc Interv Radiol. 16:1439–1445
21.
go back to reference Rosenthal D, Wellons ED, Lai KM, Bikk A, Henderson VJ (2006) Retrievable inferior vena cava filters: Initial clinical results. Ann Vasc Surg 20:157–165PubMedCrossRef Rosenthal D, Wellons ED, Lai KM, Bikk A, Henderson VJ (2006) Retrievable inferior vena cava filters: Initial clinical results. Ann Vasc Surg 20:157–165PubMedCrossRef
22.
go back to reference Lam RC, Bush RL, Lin PH, Lumsden AB (2004) Early technical and clinical results with retrievable inferior vena caval filters. Vascular 12:233–237PubMedCrossRef Lam RC, Bush RL, Lin PH, Lumsden AB (2004) Early technical and clinical results with retrievable inferior vena caval filters. Vascular 12:233–237PubMedCrossRef
23.
go back to reference Rajan DK, Sniderman KW, Rubin BB (2004) Retrieval of the Bard recovery filter from the superior vena cava. J Vasc Interv Radiol 15:1169–1171PubMed Rajan DK, Sniderman KW, Rubin BB (2004) Retrieval of the Bard recovery filter from the superior vena cava. J Vasc Interv Radiol 15:1169–1171PubMed
24.
go back to reference Millward SF, Oliva VL, Bell SD, Valenti DA, Rasuli P, Asch M, Hadziomerovic A, Kachura JR (2001) Günther Tulip Retrievable Vena Cava Filter: Results from the Registry of the Canadian Interventional Radiology Association. J Vasc Interv Radiol 12:1053–1058PubMedCrossRef Millward SF, Oliva VL, Bell SD, Valenti DA, Rasuli P, Asch M, Hadziomerovic A, Kachura JR (2001) Günther Tulip Retrievable Vena Cava Filter: Results from the Registry of the Canadian Interventional Radiology Association. J Vasc Interv Radiol 12:1053–1058PubMedCrossRef
25.
go back to reference Terhaar OA, Lyon SM, Given MF, Foster AE, McGrath F, Lee MJ (2004) Extended interval for retrieval of Günther Tulip filters. J Vasc Interv Radiol 15:1257–1262PubMed Terhaar OA, Lyon SM, Given MF, Foster AE, McGrath F, Lee MJ (2004) Extended interval for retrieval of Günther Tulip filters. J Vasc Interv Radiol 15:1257–1262PubMed
26.
go back to reference Binkert CA, Bansal A, Gates JD (2005) Inferior vena cava filter removal after 317-day implantation. J Vasc Interv Radiol. 16:395–398PubMed Binkert CA, Bansal A, Gates JD (2005) Inferior vena cava filter removal after 317-day implantation. J Vasc Interv Radiol. 16:395–398PubMed
27.
go back to reference Nadkarni S, Macdonald S, Cleveland TJ, Gaines PA (2002) Placement of a retrievable Günther Tulip filter in the superior vena cava for upper extremity deep venous thrombosis. Cardiovasc Intervent Radiol 25:524–526PubMedCrossRef Nadkarni S, Macdonald S, Cleveland TJ, Gaines PA (2002) Placement of a retrievable Günther Tulip filter in the superior vena cava for upper extremity deep venous thrombosis. Cardiovasc Intervent Radiol 25:524–526PubMedCrossRef
28.
go back to reference Smouse B (2005) IVC filter retrieval: The argument for long-term retrievability. Endovascular Today 4:77–80 Smouse B (2005) IVC filter retrieval: The argument for long-term retrievability. Endovascular Today 4:77–80
29.
go back to reference Kumar BC, Chakraverty S, Zealley I (2006) Removal of a permanent IVC filter. Cardiovasc Intervent Radiol 29:124–125PubMedCrossRef Kumar BC, Chakraverty S, Zealley I (2006) Removal of a permanent IVC filter. Cardiovasc Intervent Radiol 29:124–125PubMedCrossRef
30.
go back to reference Kaufman JA, Kinney TB, Streiff MB, et al. (2006) Guidelines for the use of retrievable and convertible vena cava filters: Report from the society of interventional radiology multidisciplinary consensus conference. J Vasc Interv Radiol 17:449–459PubMed Kaufman JA, Kinney TB, Streiff MB, et al. (2006) Guidelines for the use of retrievable and convertible vena cava filters: Report from the society of interventional radiology multidisciplinary consensus conference. J Vasc Interv Radiol 17:449–459PubMed
Metadata
Title
Long-Term Retrievability of IVC Filters: Should We Abandon Permanent Devices?
Authors
V. Berczi
J. R. Bottomley
S. M. Thomas
S. Taneja
P. A. Gaines
T. J. Cleveland
Publication date
01-09-2007
Publisher
Springer-Verlag
Published in
CardioVascular and Interventional Radiology / Issue 5/2007
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-007-9153-z

Other articles of this Issue 5/2007

CardioVascular and Interventional Radiology 5/2007 Go to the issue