Purpose: To describe our early clinical experience using the Prostar Perclose device for arterial closure after placement of a stent-graft.
Methods: Sixty-three patients had stent-graft devices placed for iliac (n = 7), abdominal aortic (n = 43) or thoracic applications (n = 13). The devices were introduced through 18 Fr (22 Fr O.D.) (n = 3), 16 Fr (20 Fr O.D.) (n = 15), 14 Fr (18 Fr O.D.) (n = 44) or 12 Fr (14 Fr O.D.) (n = 38) sheaths. After Perclose deployment, the arteriotomies were dilated to the appropriate sizes for the deployment sheaths. All patients were anticoagulated during the procedure. After stent-graft deployment, hemostasis was achieved by the Perclose sutures that were in place. All the femoral sites were followed with physical examination and CT angiography (CTA).
Results:The follow-up periods have ranged from 2 to 18 months (mean 7.5 months). All 100 femoral sites were closed without the need for surgical closure. In 4% (4/100) of the femoral sites, the Perclose sutures were pulled out while securing the knots and these patients required manual compression. In 2% (2/100) of the sites, additional manual compression was required after successful deployment of the Perclose devices. In 1% (1/100) of the femoral sites, there was acute femoral popliteal thrombosis perhaps related to common femoral artery narrowing at the site of Perclose deployment. In 5% (5/100) of the femoral sites, pseudoaneurysms were seen on follow-up CTA examinations. In 1% (1/100) of patients, there was infection at the puncture site requiring surgical bypass. In 1% (1/100) of patients, there was a stable, asymptomatic intimal dissection at the puncture site. By CTA criteria, none of the patients have developed stenotic lesions at the sites of Perclose deployment with follow-up periods limited to 6 months. Fifty-one of 63 patients (81%) patients ambulated within 4–6 hr.
Conclusion: The Perclose device, when placed prior to arterial dilatation, can be used to achieve hemostasis for stent-graft procedures. The device used in this manner is associated with complications, the most common being pseudoaneurysm formation. The long-term effect on the femoral arteries is uncertain.
Literature
1.
go back to reference Shawl, FA, Quyyumi, AA, Bajaj, S, Hoff, SB, Dougherty, KG 1996Percutaneous cardiopulmonary bypass-supported coronary angioplasty in patients with unstable angina pectoris or myocardial infarction and a left ventricular ejection fraction <25%Am J Cardiol771419CrossRefPubMed Shawl, FA, Quyyumi, AA, Bajaj, S, Hoff, SB, Dougherty, KG 1996Percutaneous cardiopulmonary bypass-supported coronary angioplasty in patients with unstable angina pectoris or myocardial infarction and a left ventricular ejection fraction <25%Am J Cardiol771419CrossRefPubMed
2.
go back to reference Gerckens, U, Cattelaens, N, Lampe, EG, Grube, E 1999Management of arterial puncture site after catheterization procedures: Evaluating a suture-mediated closure mediated deviceAm J Cardiol8316581663CrossRefPubMed Gerckens, U, Cattelaens, N, Lampe, EG, Grube, E 1999Management of arterial puncture site after catheterization procedures: Evaluating a suture-mediated closure mediated deviceAm J Cardiol8316581663CrossRefPubMed
3.
go back to reference Chamberlin, JR, Lardi, AB, McKeever, LS, et al. 1999; discussion 148Use of vascular sealing devices (VasoSeal and Perclose) versus assisted manual compression (Femostop) in transcatheter coronary interventions requiring abciximab (ReoPro)Catheter Cardiovasc Interv47143147 Chamberlin, JR, Lardi, AB, McKeever, LS,  et al. 1999; discussion 148Use of vascular sealing devices (VasoSeal and Perclose) versus assisted manual compression (Femostop) in transcatheter coronary interventions requiring abciximab (ReoPro)Catheter Cardiovasc Interv47143147
4.
go back to reference Duda, SH, Wiskirchen, J, Erb, M, et al. 1999Suture-mediated percutaneous closure of antegrade femoral arterial access sites in patients who have received full anticoagulation therapyRadiology2104752PubMed Duda, SH, Wiskirchen, J, Erb, M,  et al. 1999Suture-mediated percutaneous closure of antegrade femoral arterial access sites in patients who have received full anticoagulation therapyRadiology2104752PubMed
5.
go back to reference Mann, T, Cowper, PA, Peterson, ED, et al. 2000Transradial coronary stenting: Comparison with femoral access closed with an arterial suture device [see comments]Catheter Cardiovasc Interv49150156CrossRefPubMed Mann, T, Cowper, PA, Peterson, ED,  et al. 2000Transradial coronary stenting: Comparison with femoral access closed with an arterial suture device [see comments]Catheter Cardiovasc Interv49150156CrossRefPubMed
6.
go back to reference Hass, PC, Krajcer, Z, Diethrich, EB 1999Closure of larger percutaneous access sites using the Prostar XL percutaneous vascular surgery deviceJ Endovasc Surg6168170CrossRefPubMed Hass, PC, Krajcer, Z, Diethrich, EB 1999Closure of larger percutaneous access sites using the Prostar XL percutaneous vascular surgery deviceJ Endovasc Surg6168170CrossRefPubMed
7.
go back to reference Traul, DK, Clair, DG, Gray, B, O’Hara, PJ, Ouriel, K 2000Percutaneous endovascular repair of infrarenal abdominal aortic aneurysms: A feasibility studyJ Vasc Surg32770776CrossRefPubMed Traul, DK, Clair, DG, Gray, B, O’Hara, PJ, Ouriel, K 2000Percutaneous endovascular repair of infrarenal abdominal aortic aneurysms: A feasibility studyJ Vasc Surg32770776CrossRefPubMed
8.
go back to reference Teh, LG, Sieunarine, K, van Schie, G, et al. 2001Use of the percutaneous vascular surgery device for closure of femoral access sites during endovascular aneurysm repair: Lessons from our experienceEur J Vasc Endovasc Surg22418423CrossRefPubMed Teh, LG, Sieunarine, K, van Schie, G,  et al. 2001Use of the percutaneous vascular surgery device for closure of femoral access sites during endovascular aneurysm repair: Lessons from our experienceEur J Vasc Endovasc Surg22418423CrossRefPubMed
9.
go back to reference Howell, M, Villareal, R, Krajer, Z (2001) “Percutaneous access and closure of femoral artery access sites associated with endoluminal repair of abdominal aortic aneurysms.” J. Endovasc. Ther. 8: 68–74 Howell, M, Villareal, R, Krajer, Z (2001) “Percutaneous access and closure of femoral artery access sites associated with endoluminal repair of abdominal aortic aneurysms.” J. Endovasc. Ther. 8: 68–74
10.
go back to reference Katzen B, Becker G, Mathison M, Benenati J, Zemel G, Powell A (2001) Percutaneous or cut-down for contralateral femoral artery access in endovascular abdominal aortic repair: Comparison of complications. In: SCVIR. San Antonio, TX, 2001. Katzen B, Becker G, Mathison M, Benenati J, Zemel G, Powell A (2001) Percutaneous or cut-down for contralateral femoral artery access in endovascular abdominal aortic repair: Comparison of complications. In: SCVIR. San Antonio, TX, 2001.
11.
go back to reference Blum, U, Voshage, G, Lammer, J, et al. 1997Endoluminal stent-grafts for infrarenal abdominal aortic aneurysmsN Engl J Med3361320CrossRefPubMed Blum, U, Voshage, G, Lammer, J,  et al. 1997Endoluminal stent-grafts for infrarenal abdominal aortic aneurysmsN Engl J Med3361320CrossRefPubMed
12.
go back to reference Chuter, TA, Gordon, RL, Reilly, LM, et al. 1999Abdominal aortic aneurysm in high-risk patients: Short- to intermediate-term results of endovascular repairRadiology210361365PubMed Chuter, TA, Gordon, RL, Reilly, LM,  et al. 1999Abdominal aortic aneurysm in high-risk patients: Short- to intermediate-term results of endovascular repairRadiology210361365PubMed
13.
go back to reference Torsello, GB, Kasprzak, B, Klenk, E, Tessarek, J, Osada, N, Torsello, GF 2003Endovascular suture versus cutdown for endovascular aneurysm repair: A prospective randomized pilot studyJ Vasc Surg387882CrossRefPubMed Torsello, GB, Kasprzak, B, Klenk, E, Tessarek, J, Osada, N, Torsello, GF 2003Endovascular suture versus cutdown for endovascular aneurysm repair: A prospective randomized pilot studyJ Vasc Surg387882CrossRefPubMed
14.
go back to reference Geary, K, Landers, JT, Fiore, W, Riggs, P 2002Management of infected femoral closure devicesCardiovasc Surg10161163CrossRefPubMed Geary, K, Landers, JT, Fiore, W, Riggs, P 2002Management of infected femoral closure devicesCardiovasc Surg10161163CrossRefPubMed
15.
go back to reference Nehler, MR, Lawrence, WA, Whitehill, TA, Charette, SD, Jones, DN, Krupski, WC 2001Iatrogenic vascular injuries from percutaneous vascular suturing devicesJ Vasc Surg33943947CrossRefPubMed Nehler, MR, Lawrence, WA, Whitehill, TA, Charette, SD, Jones, DN, Krupski, WC 2001Iatrogenic vascular injuries from percutaneous vascular suturing devicesJ Vasc Surg33943947CrossRefPubMed
16.
go back to reference Smith, TP, Cruz, CP, Moursi, MM, Eidt, JF 2001Infectious complications resulting from use of hemostatic puncture closure devicesAm J Surg182658662CrossRefPubMed Smith, TP, Cruz, CP, Moursi, MM, Eidt, JF 2001Infectious complications resulting from use of hemostatic puncture closure devicesAm J Surg182658662CrossRefPubMed
17.
go back to reference Hofmann, LV, Sood, S, Liddell, RP, et al. 2003Arteriographic and pathologic evaluation of two suture-mediated arterial closure devices in a porcine modelJ Vasc Interv Radiol14755761PubMed Hofmann, LV, Sood, S, Liddell, RP,  et al. 2003Arteriographic and pathologic evaluation of two suture-mediated arterial closure devices in a porcine modelJ Vasc Interv Radiol14755761PubMed
18.
go back to reference Brown, DB, Crawford, ST, Norton, PL, Hovsepian, DM 2002Angiographic follow-up after suture-mediated femoral artery closureJ Vasc Interv Radiol13677680PubMed Brown, DB, Crawford, ST, Norton, PL, Hovsepian, DM 2002Angiographic follow-up after suture-mediated femoral artery closureJ Vasc Interv Radiol13677680PubMed
Metadata
Title
Percutaneous Femoral Closure Following Stent-Graft Placement: Use of the Perclose Device
Authors
Stephen F. Quinn
Jinho Kim
Publication date
01-09-2004
Publisher
Springer-Verlag
Published in
CardioVascular and Interventional Radiology / Issue 3/2004
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-003-2713-y

Other articles of this Issue 3/2004

CardioVascular and Interventional Radiology 3/2004 Go to the issue