Published in: Aesthetic Plastic Surgery 5/2010

01-10-2010 | Original Article

High-Intensity Focused Ultrasound Effectively Reduces Waist Circumference by Ablating Adipose Tissue from the Abdomen and Flanks: A Retrospective Case Series

Authors: Afschin Fatemi, Michael A. C. Kane

Published in: Aesthetic Plastic Surgery | Issue 5/2010

Login to get access

Abstract

Background

Currently available technologies for performing aesthetic body sculpting are either noninvasive but require multiple treatments to achieve relatively superficial effects or very effective but also invasive and sometimes associated with serious complications. A new, noninvasive alternative is to ablate adipose tissue using high-intensity focused ultrasound (HIFU). When focused within subcutaneous adipose tissue, HIFU quickly raises the local temperature, resulting in instantaneous cell death via coagulative necrosis within the targeted area but no damage to the surrounding tissue.

Methods

A new HIFU device, the LipoSonix system (Medicis Technologies Corporation, Bothell, WA, USA), was used by our clinic staff to reduce waist circumference via removal of excess adipose tissue from the anterior abdomen and flank areas. This report describes the results of a retrospective chart review of patients at one clinic who underwent HIFU treatment of the anterior abdomen and flank areas.

Results

A total of 85 men and women with a mean age of 43.8 years underwent a single HIFU treatment session. The time required to complete treatment was approximately 1 to 1.5 h. Using a mean energy level of 134.8 J/cm2 and a focal depth of 1.1 to 1.6 cm, the waist circumference was decreased by an average of 4.6 cm after 3 months. Of the 85 patients, 10 (11.8%) reported adverse events including prolonged tenderness (n = 3), ecchymosis (n = 3), hard lumps (n = 2), edema (n = 1), and pain (n = 1), which resolved spontaneously.

Conclusions

The authors conclude from their experience that HIFU represents a safe and effective means for performing noninvasive body sculpting.
Footnotes
1
The highest and lowest energy doses were compared using the median energy dose as the cutoff.
 
2
Comparisons were made using a two-sample t test assuming equal variances.
 
Literature
1.
go back to reference van der Lugt C, Romero C, Ancona D, Al-Zarouni M, Perera J, Trelles MA (2009) A multicenter study of cellulite treatment with a variable emission radio frequency system. Dermatol Ther 22:74–84CrossRefPubMed van der Lugt C, Romero C, Ancona D, Al-Zarouni M, Perera J, Trelles MA (2009) A multicenter study of cellulite treatment with a variable emission radio frequency system. Dermatol Ther 22:74–84CrossRefPubMed
2.
go back to reference Nootheti PK, Magpantay A, Yosowitz G, Calderon S, Goldman MP (2006) A single-center, randomized, comparative, prospective clinical study to determine the efficacy of the VelaSmooth system versus the Triactive system for the treatment of cellulite. Lasers Surg Med 38:908–912CrossRefPubMed Nootheti PK, Magpantay A, Yosowitz G, Calderon S, Goldman MP (2006) A single-center, randomized, comparative, prospective clinical study to determine the efficacy of the VelaSmooth system versus the Triactive system for the treatment of cellulite. Lasers Surg Med 38:908–912CrossRefPubMed
3.
go back to reference Kulick M (2005) Evaluation of a combined laser-radio frequency device (Polaris WR) for the nonablative treatment of facial wrinkles. J Cosmet Laser Ther 7:87–92CrossRefPubMed Kulick M (2005) Evaluation of a combined laser-radio frequency device (Polaris WR) for the nonablative treatment of facial wrinkles. J Cosmet Laser Ther 7:87–92CrossRefPubMed
4.
go back to reference Sadick N, Magro C (2007) A study evaluating the safety and efficacy of the VelaSmooth system in the treatment of cellulite. J Cosmet Laser Ther 9:15–20CrossRefPubMed Sadick N, Magro C (2007) A study evaluating the safety and efficacy of the VelaSmooth system in the treatment of cellulite. J Cosmet Laser Ther 9:15–20CrossRefPubMed
5.
go back to reference Romero C, Caballero N, Herrero M, Ruíz R, Sadick NS, Trelles MA (2008) Effects of cellulite treatment with RF, IR light, mechanical massage and suction treating one buttock with the contralateral as a control. J Cosmet Laser Ther 10:193–201CrossRefPubMed Romero C, Caballero N, Herrero M, Ruíz R, Sadick NS, Trelles MA (2008) Effects of cellulite treatment with RF, IR light, mechanical massage and suction treating one buttock with the contralateral as a control. J Cosmet Laser Ther 10:193–201CrossRefPubMed
6.
go back to reference Lehnhardt M, Homann HH, Daigeler A, Hauser J, Palka P, Steinau HU (2008) Major and lethal complications of liposuction: a review of 72 cases in Germany between 1998 and 2002. Plast Reconstr Surg 121:396–403eCrossRef Lehnhardt M, Homann HH, Daigeler A, Hauser J, Palka P, Steinau HU (2008) Major and lethal complications of liposuction: a review of 72 cases in Germany between 1998 and 2002. Plast Reconstr Surg 121:396–403eCrossRef
7.
go back to reference Gasperoni C, Gasperoni P (2006) Subdermal liposuction: long-term experience. Clin Plast Surg 33:63–73CrossRefPubMed Gasperoni C, Gasperoni P (2006) Subdermal liposuction: long-term experience. Clin Plast Surg 33:63–73CrossRefPubMed
9.
go back to reference Katz B, McBean J (2008) Laser-assisted lipolysis: a report on complications. J Cosmet Laser Ther 10:2331–2333CrossRef Katz B, McBean J (2008) Laser-assisted lipolysis: a report on complications. J Cosmet Laser Ther 10:2331–2333CrossRef
10.
go back to reference Ter Haar G, Coussios C (2007) High-intensity focused ultrasound: physical principles and devices. Int J Hyperthermia 23:89–104CrossRef Ter Haar G, Coussios C (2007) High-intensity focused ultrasound: physical principles and devices. Int J Hyperthermia 23:89–104CrossRef
11.
go back to reference LipoSonix™ System User Manual. Medicis Technologies Corporation, Scottsdale, AZ LipoSonix System User Manual. Medicis Technologies Corporation, Scottsdale, AZ
12.
go back to reference Fodor PB, Smoller BR, Stecco KA, Desilets CS (2006) Biochemical changes in adipocytes and lipid metabolism secondary to the use of high-intensity focused ultrasound for noninvasive body sculpting. Paper presented at the American Society of Aesthetic Plastic Surgery Annual Meeting, April 2006, Orlando, FL Fodor PB, Smoller BR, Stecco KA, Desilets CS (2006) Biochemical changes in adipocytes and lipid metabolism secondary to the use of high-intensity focused ultrasound for noninvasive body sculpting. Paper presented at the American Society of Aesthetic Plastic Surgery Annual Meeting, April 2006, Orlando, FL
13.
go back to reference Fodor PB, Smoller BR, Stecco KA, Desilets CS (2006) A noninvasive method for body sculpting: the use of high-intensity focused ultrasound. Paper presented at the American Society of Aesthetic Plastic Surgery Annual Meeting, April 2006, Orlando, FL Fodor PB, Smoller BR, Stecco KA, Desilets CS (2006) A noninvasive method for body sculpting: the use of high-intensity focused ultrasound. Paper presented at the American Society of Aesthetic Plastic Surgery Annual Meeting, April 2006, Orlando, FL
Metadata
Title
High-Intensity Focused Ultrasound Effectively Reduces Waist Circumference by Ablating Adipose Tissue from the Abdomen and Flanks: A Retrospective Case Series
Authors
Afschin Fatemi
Michael A. C. Kane
Publication date
01-10-2010
Publisher
Springer-Verlag
Published in
Aesthetic Plastic Surgery / Issue 5/2010
Print ISSN: 0364-216X
Electronic ISSN: 1432-5241
DOI
https://doi.org/10.1007/s00266-010-9503-0

Other articles of this Issue 5/2010

Aesthetic Plastic Surgery 5/2010 Go to the issue

Announcements

Announcements