Three-dimensional computer-assisted navigation for the placement of cannulated hip screws. A pilot study

Authors: Marcus Christian Müller, Peter Belei, Peter H. Pennekamp, Koroush Kabir, Dieter C. Wirtz, Christof Burger, Oliver Weber

Published in: International Orthopaedics | Issue 7/2012

Login to get access

Abstract

Purpose

Medial femoral neck fractures are common, and closed reduction and internal fixation by three cannulated screws is an accepted method for the surgical treatment. Computer navigation for screw placement may reduce fluoroscopy time, the number of guidewire passes and optimise screw placement.

Methods

In the context of a sawbone study, a computer-assisted planning and navigation system based on 3D-imaging for guidewire placement in the femoral neck was tested to improve screw placement. Three screws were inserted into 12, intact, femoral sawbones using the conventional technique and into 12, intact, femoral sawbones guided by the computer-based navigation system. Guidewire and subsequent screw placement in the femoral neck were evaluated.

Results

Use of the navigation system resulted in a significant reduction of the number of drilling attempts (p ≤ 0.05) and achieved optimised accuracy of implant placement by attaining significantly better screw parallelism (p ≤ 0.05) and significantly enlarged neck-width coverage by the three screws (p ≤ 0.0001). Computer assistance significantly increased the number of fluoroscopic images (p ≤ 0.001) and the operation time (p ≤ 0.0001).

Conclusions

Three-dimensional computer-assisted navigation improves accuracy of cannulated screw placement in femoral neck while increasing the number of fluoroscopic images and operation time. Additional studies including fractured sawbones and cadaver models with the goal of reducing operation time are indispensable before introduction of this navigation system into clinical practice.
Literature
1.
go back to reference Kannus P, Niemi S, Parkkari J, Palvanen M, Vuori I, Järvinen M (1999) Hip fractures in Finland between 1970 and 1977 and predictions for the future. Lancet 353:802–805PubMedCrossRef Kannus P, Niemi S, Parkkari J, Palvanen M, Vuori I, Järvinen M (1999) Hip fractures in Finland between 1970 and 1977 and predictions for the future. Lancet 353:802–805PubMedCrossRef
2.
go back to reference Selvan VT, Oakley MJ, Rangan A, Al-Lami MK (2004) Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study. Injury 35:136–141PubMedCrossRef Selvan VT, Oakley MJ, Rangan A, Al-Lami MK (2004) Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study. Injury 35:136–141PubMedCrossRef
3.
go back to reference Kloen P, Rubel IF, Lyden JP, Helfet DL (2003) Subtrochanteric fracture after cannulated screw fixation of femoral neck fractures: a case report of four cases. J Orthop Trauma 17:225–229PubMedCrossRef Kloen P, Rubel IF, Lyden JP, Helfet DL (2003) Subtrochanteric fracture after cannulated screw fixation of femoral neck fractures: a case report of four cases. J Orthop Trauma 17:225–229PubMedCrossRef
4.
go back to reference Lindequist S, Tornkvist H (1995) Quality of reduction and cortical screw support in femoral neck fractures. An analysis of 72 fractures with a new computerized measuring method. J Orthop Trauma 9:215–221PubMedCrossRef Lindequist S, Tornkvist H (1995) Quality of reduction and cortical screw support in femoral neck fractures. An analysis of 72 fractures with a new computerized measuring method. J Orthop Trauma 9:215–221PubMedCrossRef
5.
go back to reference Szita J, Cserhati P, Bosch U, Manninger J, Bodzay T, Fekete K (2002) Intracapsular femoral neck fractures: the importance of early reduction and stable osteosynthesis. Injury 33:41–46CrossRef Szita J, Cserhati P, Bosch U, Manninger J, Bodzay T, Fekete K (2002) Intracapsular femoral neck fractures: the importance of early reduction and stable osteosynthesis. Injury 33:41–46CrossRef
6.
go back to reference Mehlman CT, DiPasquale TG (1997) Radiation exposure to the orthopaedic surgical team during fluoroscopy: “how far away is far enough?”. J Orthop Trauma 11:392–398PubMedCrossRef Mehlman CT, DiPasquale TG (1997) Radiation exposure to the orthopaedic surgical team during fluoroscopy: “how far away is far enough?”. J Orthop Trauma 11:392–398PubMedCrossRef
7.
go back to reference Sanders R, Koval KJ, Di Pasquale T, Schmelling G, Stenzler S, Ross E (1993) Exposure of the orthopaedic surgeon to radiation. J Bone Joint Surg [Am] 75:326–330 Sanders R, Koval KJ, Di Pasquale T, Schmelling G, Stenzler S, Ross E (1993) Exposure of the orthopaedic surgeon to radiation. J Bone Joint Surg [Am] 75:326–330
8.
go back to reference Kendoff D, Citak M, Gardner MJ, Kfuri M Jr, Thumes B, Krettek C, Hüfner T (2007) Three-dimensional fluoroscopy for evaluation of articular reduction and screw placement in calcaneal fractures. Foot Ankle Int 28:1165–1171PubMedCrossRef Kendoff D, Citak M, Gardner MJ, Kfuri M Jr, Thumes B, Krettek C, Hüfner T (2007) Three-dimensional fluoroscopy for evaluation of articular reduction and screw placement in calcaneal fractures. Foot Ankle Int 28:1165–1171PubMedCrossRef
9.
go back to reference Kotsianos D, Rock C, Euler E, Wirth S, Linsenmaier U, Brandl R, Mutschler W, Pfeifer KJ (2001) 3-D imaging with a mobile surgical image enhancement equipment (ISO-C-3D). Initial examples of fracture diagnosis of peripheral joints in comparison with spiral CT and conventional radiography. Unfallchirurg 104:834–838PubMedCrossRef Kotsianos D, Rock C, Euler E, Wirth S, Linsenmaier U, Brandl R, Mutschler W, Pfeifer KJ (2001) 3-D imaging with a mobile surgical image enhancement equipment (ISO-C-3D). Initial examples of fracture diagnosis of peripheral joints in comparison with spiral CT and conventional radiography. Unfallchirurg 104:834–838PubMedCrossRef
10.
go back to reference Wendl K, von Recum J, Wentzensen A, Grützner PA (2003) Iso-C 3D assisted navigated implantation of pedicle screws in thoracic lumbar vertebrae. Unfallchirurg 106:907–913PubMed Wendl K, von Recum J, Wentzensen A, Grützner PA (2003) Iso-C 3D assisted navigated implantation of pedicle screws in thoracic lumbar vertebrae. Unfallchirurg 106:907–913PubMed
11.
go back to reference Stöckle U, König B, Schaser K, Melcher I, Haas NP (2003) CT and fluoroscopy based navigation in pelvic surgery. Unfallchirurg 106:914–920PubMed Stöckle U, König B, Schaser K, Melcher I, Haas NP (2003) CT and fluoroscopy based navigation in pelvic surgery. Unfallchirurg 106:914–920PubMed
12.
go back to reference Stöckle U, Schaser K, König B (2007) Image guidance in pelvic and acetabular surgery—expectations, success and limitations. Injury 38:450–462PubMedCrossRef Stöckle U, Schaser K, König B (2007) Image guidance in pelvic and acetabular surgery—expectations, success and limitations. Injury 38:450–462PubMedCrossRef
13.
go back to reference Hamelinck HKM, Haagmans M, Snoeren MM (2006) Safety of computer-assisted surgery for cannulated hip screws. Clin Orthop Rel Res 455:241–245 Hamelinck HKM, Haagmans M, Snoeren MM (2006) Safety of computer-assisted surgery for cannulated hip screws. Clin Orthop Rel Res 455:241–245
14.
go back to reference Mayman D, Vasarhelyi EM, Long W, Ellis RE, Rudan J, Pichora DR (2005) Computer-assisted guidewire insertion for hip fracture fixation. J Orthop Trauma 19:610–615PubMedCrossRef Mayman D, Vasarhelyi EM, Long W, Ellis RE, Rudan J, Pichora DR (2005) Computer-assisted guidewire insertion for hip fracture fixation. J Orthop Trauma 19:610–615PubMedCrossRef
15.
go back to reference Suhm N, Beck T, Zuna I, Marazzi C, Staedele H, Jacob AL, Messmer P, Regazzoni P (2004) Fluoroscopy-based surgical navigation versus fluoroscopic guidance to control guide wire insertion for osteosynthesis of femoral neck fractures. Eur J Trauma 1:23–28CrossRef Suhm N, Beck T, Zuna I, Marazzi C, Staedele H, Jacob AL, Messmer P, Regazzoni P (2004) Fluoroscopy-based surgical navigation versus fluoroscopic guidance to control guide wire insertion for osteosynthesis of femoral neck fractures. Eur J Trauma 1:23–28CrossRef
16.
go back to reference Mueller MC, Belei P, de la Fuente M, Strake M, Weber O, Burger C, Radermacher K, Wirtz DC (2011) Evaluation of a fluoroscopy-based navigation system enabling a virtual radiation-free preview of X-ray images for placement of cannulated hip screws. A cadaver study. Comp Aided Surg 16:22–31CrossRef Mueller MC, Belei P, de la Fuente M, Strake M, Weber O, Burger C, Radermacher K, Wirtz DC (2011) Evaluation of a fluoroscopy-based navigation system enabling a virtual radiation-free preview of X-ray images for placement of cannulated hip screws. A cadaver study. Comp Aided Surg 16:22–31CrossRef
17.
go back to reference Gurusamy K, Parker MJ, Rowlands TK (2005) The complications of displaced intracapsular fractures of the hip: the effect of screw positioning and angulation on fracture healing. J Bone Joint Surg Br 87:632–634PubMedCrossRef Gurusamy K, Parker MJ, Rowlands TK (2005) The complications of displaced intracapsular fractures of the hip: the effect of screw positioning and angulation on fracture healing. J Bone Joint Surg Br 87:632–634PubMedCrossRef
18.
go back to reference Liebergall M, Ben-David D, Weil Y, Peyser A, Mosheiff R (2006) Computerized navigation for the internal fixation of femoral neck fractures. J Bone Joint Surg [Am] 88:1748–1754CrossRef Liebergall M, Ben-David D, Weil Y, Peyser A, Mosheiff R (2006) Computerized navigation for the internal fixation of femoral neck fractures. J Bone Joint Surg [Am] 88:1748–1754CrossRef
19.
go back to reference Kendoff D, Hüfner T, Citak M, Maier C, Wesemeier F, Pearle A, Krettek C (2006) A new parallel drill guide for navigating femoral neck screw placement. Development and evaluation. Unfallchirurg 109:875–880PubMedCrossRef Kendoff D, Hüfner T, Citak M, Maier C, Wesemeier F, Pearle A, Krettek C (2006) A new parallel drill guide for navigating femoral neck screw placement. Development and evaluation. Unfallchirurg 109:875–880PubMedCrossRef
20.
go back to reference Rock C, Kotsianos D, Linsenmaier U, Fischer T, Brandl R, Vill F, Wirth S, Kaltschmidt R, Euler E, Pfeifer KJ, Reiser M (2002) Studies on image quality, high contrast resolution and dose for the axial skeleton and limbs with a new, dedicated CT system (ISO-D-3D). Rofo 174:170–176PubMedCrossRef Rock C, Kotsianos D, Linsenmaier U, Fischer T, Brandl R, Vill F, Wirth S, Kaltschmidt R, Euler E, Pfeifer KJ, Reiser M (2002) Studies on image quality, high contrast resolution and dose for the axial skeleton and limbs with a new, dedicated CT system (ISO-D-3D). Rofo 174:170–176PubMedCrossRef
21.
go back to reference Kwong L, Johanson PH, Zinar DM, Lenihan MR, Herman MW (1990) Shielding of the patient’s gonads during intramedullary interlocking femoral nailing. J Bone Joint Surg [Am] 72:1523–1526 Kwong L, Johanson PH, Zinar DM, Lenihan MR, Herman MW (1990) Shielding of the patient’s gonads during intramedullary interlocking femoral nailing. J Bone Joint Surg [Am] 72:1523–1526
22.
go back to reference Kendoff D, Bogojevic A, Citak M, Citak M, Maier C, Maier G, Krettek C, Hüfner T (2007) Experimental validation of non-invasive referencing in navigated procedures on long bones. J Orthop Res 25:201–207PubMedCrossRef Kendoff D, Bogojevic A, Citak M, Citak M, Maier C, Maier G, Krettek C, Hüfner T (2007) Experimental validation of non-invasive referencing in navigated procedures on long bones. J Orthop Res 25:201–207PubMedCrossRef
Metadata
Title
Three-dimensional computer-assisted navigation for the placement of cannulated hip screws. A pilot study
Authors
Marcus Christian Müller
Peter Belei
Peter H. Pennekamp
Koroush Kabir
Dieter C. Wirtz
Christof Burger
Oliver Weber
Publication date
01-07-2012
Publisher
Springer-Verlag
Published in
International Orthopaedics / Issue 7/2012
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-012-1496-7

Other articles of this Issue 7/2012

International Orthopaedics 7/2012 Go to the issue